PICCOLO TRGLYCERIDES-CAPILLARY TEST SYSTEM

K120664 · Abaxis, Inc. · JGY · Apr 24, 2012 · Clinical Chemistry

Device Facts

Record IDK120664
Device NamePICCOLO TRGLYCERIDES-CAPILLARY TEST SYSTEM
ApplicantAbaxis, Inc.
Product CodeJGY · Clinical Chemistry
Decision DateApr 24, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1705
Device ClassClass 1

Intended Use

The Piccolo Triglycerides – Capillary Test System used with the Piccolo xpress Chemistry Analyzer is intended for the in vitro quantitative determination of triglycerides in capillary (fingerstick) heparinized whole blood in a clinical laboratory setting or point-of-care location. Triglyceride measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, and other diseases involving lipid metabolism, or various endocrine disorders.

Device Story

System uses Piccolo Lipid Panel Reagent Disc with Piccolo xpress Chemistry Analyzer; processes heparinized capillary whole blood samples. Disc separates plasma from blood cells; meters plasma and diluent; mixes with reagent beads. Analyzer monitors enzymatic endpoint chemical reactions at 37°C. Used in clinical labs or point-of-care settings by professional operators. Output is quantitative triglyceride concentration (mg/dL). Results assist clinicians in diagnosing and managing lipid metabolism, endocrine, and metabolic diseases.

Clinical Evidence

Method comparison study performed at three point-of-care sites using 588 capillary fingerstick samples. Results compared to Roche Cobas Triglycerides Test (k893973) via Deming regression. Combined data showed slope of 0.96 (95% CI: 0.96-0.97), intercept of 3.5 (95% CI: 2.1-4.9), and R² of 0.992. Precision study using 5 whole blood samples across 20 analyzers showed %CV ranging from 0.8% to 3.1%.

Technological Characteristics

Single-use disc constructed of polymethylmethacrylate and ABS plastic. Employs enzymatic colorimetric assay. Sensing via optical system measuring color intensity at 37°C. Reagents provided as lyophilized beads with liquid diluent. Dimensions: 8 cm diameter. Standalone operation on Piccolo xpress Chemistry Analyzer. Calibration via disc-specific bar code.

Indications for Use

Indicated for the quantitative determination of triglycerides in capillary heparinized whole blood to aid in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, lipid metabolism disorders, or endocrine disorders.

Regulatory Classification

Identification

A triglyceride test system is a device intended to measure triglyceride (neutral fat) in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Piccolo® Triglycerides - Capillary Test System 510(k) Premarket Notification K120664 APR 24 2012 APPENDIX D # 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS 3240 Whipple Road, Union City, CA 94587 Phone 510 . 675-6500 Fax 510 . 441-6150 Image /page/0/Picture/20 description: The image shows the word "ABAXIS" in a bold, sans-serif font. A stylized oval shape encircles the "X" in the word, adding a dynamic element to the design. The overall impression is clean and modern, suggesting a professional and innovative company. This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is: ## 1. Applicant Information: | Date Prepared:<br>Name:<br>Address: | March 2, 2012<br>Abaxis, Inc.<br>3240 Whipple Road<br>Union City, CA 94587 | | |-------------------------------------|----------------------------------------------------------------------------|--| | Contact Person:<br>Dhana Mundau | Dennis M. Bleile, PhD<br>ICHOS COLE | | | Contact Person: | Dennis M. Bielle, PhD | |-----------------|-----------------------| | Phone Number: | (510) 675-6515 | | Fax Number: | (510) 441-6150 | #### 2. Device Information: | Classification | Class I | |----------------|------------------------------------------------| | Trade Name: | Piccolo® Triglycerides - Capillary Test System | Classification Name: Triglycerides Test system 862.1705 #### 3. Identification of Legally Marketed Devices to which the Submitter Claims Equivalence: The following table identifies the legally marketed device to which Abaxis claims equivalence: | Predicate Devices | | | | |---------------------------------------|---------------------------------------|------------------|-----------------------------| | Predicate Device | Manufacturer | 510(k)<br>Number | Date of SE<br>Determination | | Piccolo® Triglycerides<br>Test System | Abaxis<br>Union City, CA | K023639 | 1/24/2003 | | Cobas® Triglycerides<br>Test | Roche Diagnostics<br>Indianapolis, IN | K893973 | 08/11/1989 | {1}------------------------------------------------ ## Summary of Safety and Effectiveness (continued) #### 4. Description of the Device: The Piccolo® Lipid Panel - Capillary Reagent Disc (which contains the Piccolo® Triglycerides - Capillary Test System) is designed to separate a heparinized whole blood sample into plasma and blood cells. The disc meters the required quantity of plasma and diluent, mixes the plasma with diluent, and delivers the mixture to the reaction cuvettes along the disc perimeter. The diluted plasma mixes with the reagent beads, initiating the chemical reactions that are then monitored by the analyzer. #### Statement of Intended Use: 5. The Piccolo Triglycerides - Capillary Test System used with the Piccolo xpress Chemistry Analyzer is intended for the in vitro quantitative determination of triglycerides in capillary (fingerstick) heparinized whole blood in a clinical laboratory setting or point-of-care location. #### Summary of the Technological Characteristics of the New Device in 6. Comparison to those of the Predicate Device: Table 1 outlines the technological characteristics of the Piccolo® Triglycerides -Capillary Test System in comparison to the legally marketed predicate device. {2}------------------------------------------------ ಿ Triglycerides - Capillary Test System Picc Summary of Safety and Effectiveness (continued) . Table 1: Specification Comparison for Piccolo Triglycerides – Capillary Test System, Piccolo Triglycerides Test System and the Roche Triglycerides Test on the Cobas 6000 Ana | | Piccolo Triglycerides –<br>Capillary Test on Abaxis<br>Chemistry Analyzer | Piccolo Triglycerides Test on<br>Abaxis Chemistry Analyzer | Roche Triglycerides Test on<br>the Cobas 6000 Analyzer | |------------------------------|------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------| | Intended Use | Quantitative analysis of<br>Triglycerides | Quantitative analysis of<br>Triglycerides | Quantitative analysis of<br>Triglycerides | | Methodology | Enzymatic endpoint reaction | Enzymatic endpoint reaction | Enzymatic endpoint reaction | | Sample Type | Lithium heparinized capillary whole<br>blood | Lithium heparinized whole blood,<br>heparinized plasma, and<br>serum | Lithium heparin and potassium<br>EDTA plasma and serum | | Dynamic Range<br>Lower Limit | 20 mg/dL | 20 mg/dL | 3.86 mg/dL | | Reagents | Dry test-specific reagent beads<br>and liquid diluent; reconstitution<br>performed by analyzer | Dry test-specific reagent beads<br>and liquid diluent; reconstitution<br>performed by analyzer | Liquid reagent | | Temperature of<br>Reaction | 37°C | 37°C | 37°C | | Calibration | Bar code with factory<br>calibrated lot specific data | Bar code with factory<br>calibrated lot specific data | Calibrated periodically using<br>calibrators supplied by vendor | | Assay Range | 20-500 mg/dL | 20-500 mg/dL | 8.85-885 mg/dL | | Testing Environment | Professional use | Professional use | Professional use | | Sample Size | Approx 100 µL | Approx 100 μL | 2 µL | baxis Confidentia ・ . . {3}------------------------------------------------ # Summary of Safety and Effectiveness (continued) #### 7. Brief Discussion of the Clinical and Nonclinical Tests Relied on for a Determination of Substantial Equivalence. The following information summarizes the results of clinical and non-clinical tests performed using the Piccolo® Triglycerides Test System. ## Linearity: ### Table 2: Summary of Linearity | | Triglycerides | |-----------------------------|---------------| | Slope | 1.000 | | Intercept | 3.0 | | Correlation Coefficient (r) | 1.000 | ### Precision: Precision studies were designed to evaluate within-run and total precision of triglycerides included on the Piccolo® Lipid Panel Reagent Disc when run on the Abaxis analyzer. # Table 3: Within-Run and Total Precision of Triglycerides Assayed on the Abaxis Analyzer | | Within-Run<br>(n =160) | Total<br>(n =160) | |-----------------------|------------------------|-------------------| | Triglycerides (mg/dL) | | | | Serum 1 | | | | Mean | 206.8 | 206.8 | | SD | 4.7 | 5.5 | | %CV | 2.3 | 2.6 | | Serum 2 | | | | Mean | 163.7 | 163.7 | | SD | 1.8 | 2.4 | | %CV | 1.1 | 1.5 | {4}------------------------------------------------ # Summary of Safety and Effectiveness (continued) # Method Comparison: # Table 4: Method Comparison Data for Triglycerides Assayed by the Abaxis Piccolo Triglycerides - Capillary Test System and the Roche Triglycerides Test | Parameters | Statistics | |---------------------------------------------------------|------------| | Piccolo Triglycerides - Capillary: Singlicate Values, N | 588 | | Roche Triglycerides Assay: Average of Duplicates, N | 588 | | Piccolo Triglycerides - Capillary: Mean | 155.4 | | Roche Triglycerides Assay: Mean | 157.5 | | Piccolo Triglycerides - Capillary: Std. Dev. | 88.3 | | Roche Triglycerides Assay: Std. Dev | 91.6 | | Piccolo Triglycerides - Capillary: Range of Samples | 36 - 496 | | Roche Triglycerides Assay: Range of Samples | 33.5 - 528 | | Parameters (Roche on X Axis) | Linear Regression | Deming Regression | |----------------------------------|---------------------|---------------------| | N | 588 | 588 | | Slope (95% CI) | 0.96 (0.95 to 0.97) | 0.96 (0.95 to 0.97) | | Intercept (95% CI) | 4.1 (2.8 - 5.4) | 3.5 (2.1 to 4.9) | | Correlation Coefficient (R²) | 0.992 | 0.992 | | Std. Error of the Estimate (SEE) | 7.9 | 7.9 | #### 8. Conclusions The clinical and non-clinical tests performed using the Piccolo® Triglycerides = Capillary Test System, when run on the Piccolo xpress Chemistry Analyzer, Supillary - Foot - System is as safe, effective and performs as well as the legally marketed devices identified above. {5}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of a human figure, represented by flowing lines, suggesting a sense of movement and connection. ### Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 Abaxis, Inc. c/o Dennis M. Bleile, Ph.D., RAC Director of Assay Performance and Compliance 3240 Whipple Road Union City, CA 94587 APR 2 4 2012 ### Re: k120664 Trade/Device Name: Piccolo Triglycerides - Capillary Test System Regulation Number: 21 CFR 862.1705 Regulation Name: Triglyceride Test System Regulatory Class: Class I, meets limitations of exemptions per 21 CFR 862.9 (c)(4) Product Code: JGY Dated: March 2, 2012 Received: March 5, 2012 Dear Dr. Bleile: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. {6}------------------------------------------------ Page 2 If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance ... You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm Sincerely yours, \$\mathcal{N}\$ Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ Piccolo® Triglycerides - Capillary Test System #### INDICATIONS FOR USE STATEMENT 12.0 # Indications for Use 510(k) Number (if known): #### Piccolo® Triglycerides Test System Device Name: ### Intended Use: The Piccolo Triglycerides - Capillary Test System used with the Piccolo xpress Chemistry Analyzer is intended for the in vitro quantitative determination of triglycerides in capillary (fingerstick) heparinized whole blood in a clinical laboratory setting or point-of-care location. # Indications for Use: Triglyceride measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, and other diseases involving lipid metabolism, or various endocrine disorders. Prescription Use × (Part 21 CFR 801.109 Subpart D) AND/OR Over- The Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K12066 Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...