ELITECH CLINICAL SYSTEMS CREATINNE PAP SL MODEL CRSL-0250, CLITECH CLINICAL SYSTEMS ELICAL 2 MODEL CALI-0580, ELITECH CL
Applicant
Seppim S.A.S.
Product Code
JFY · Clinical Chemistry
Decision Date
Nov 18, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1225
Device Class
Class 2
Indications for Use
ELITech Clinical Systems Creatinine PAP SL reagent is for the quantitative in vitro diagnostic determination of creatinine in human serum and plasma on the Vital Scientific Selectra/Flexor Analyzers. It is not intended for use in Point of Care settings. Creatinine measurements are used in the diagnosis and treatment of renal diseases, and in monitoring renal dialysis. ELITech Clinical Systems ELICAL 2 is a multi-parametric calibrator for in vitro diagnostic use in the calibration of quantitative ELITech Clinical Systems methods on Vital Scientific Selectra Junior Analyzer and the Vital Scientific Flexor Junior Analyzers. ELITech Clinical Systems ELITROL I is a multi-parametric control serum for in vitro diagnostic use in accuracy control of quantitative ELITech Clinical Systems methods on the Vital Scientific Selectra Junior Analyzer and the Vital Scientific Flexor Junior Analyzer. ELITech Clinical Systems ELITROL II is a multi-parametric control serum for in vitro diagnostic use in accuracy control of quantitative ELITech Clinical Systems methods on the Vital Scientific Selectra Junior analyzer and the Vital Scientific Flexor Junior Analyzers.
Device Story
The ELITech Clinical Systems CREATININE PAP SL is an enzymatic colorimetric assay used for the quantitative determination of creatinine in human serum and plasma. The device consists of two liquid, ready-to-use reagents (R1 and R2) containing enzymes (creatinase, sarcosine oxidase, ascorbate oxidase, creatininase, peroxidase) and buffers. It is designed for use on Vital Scientific Selectra/Flexor analyzers in clinical laboratory settings. The assay operates by converting creatinine into a measurable colorimetric signal proportional to the creatinine concentration in the sample. Healthcare providers use the resulting creatinine measurements to diagnose and treat renal diseases and monitor patients undergoing renal dialysis. The device is intended for professional use only and is not for point-of-care settings.
Clinical Evidence
Bench testing only. Precision studies (n=80 and n=40) showed total CVs 1.6-3.2%. Linearity confirmed 0.28-22.30 mg/dL (r2=0.9999). Method comparison with predicate (n=91) yielded slope 1.045, intercept -0.01. Matrix comparison (serum vs lithium heparin) showed slope 1.013, r=1.000. Interference testing performed per CLSI EP7-A2; positive bias noted with creatine >= 5 mg/dL; negative interference with calcium dobesilate, methyl-dopa, and L-dopa.
Technological Characteristics
Enzymatic colorimetric assay. Reagents: R1 (MOPS buffer, EHSPT, creatinase, sarcosine oxidase, ascorbate oxidase); R2 (MOPS buffer, 4-AAP, creatininase, peroxidase, sodium azide). Liquid, ready-to-use form. Designed for Vital Scientific Selectra/Flexor analyzers. Storage: 2-8 °C, protected from light.
Indications for Use
Indicated for the quantitative determination of creatinine in human serum and plasma for the diagnosis and treatment of renal diseases and monitoring of renal dialysis. Not for Point of Care use.
Regulatory Classification
Identification
A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.
Predicate Devices
Roche cobas c 111 Creatinine plus ver. 2 (k024098)
Roche Diagnostics Calibrator for Automated Systems (C.f.a.s) (k033501)
Roche Diagnostics Precinorm Universal and Precipath Universal Control Sera (k041227)
Related Devices
K132399 — ELITECH CLINICAL SYSTEMS CREATININE PAP SL, ELICAL 2, ELITROL I AND ELITROL II, URINE CONTROL BI-LEVEL · Elitechgroup · Jan 3, 2014
K193649 — Yumizen C1200 Creatinine PAP · HORIBA ABX SAS · May 10, 2021
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k103376
B. Purpose for Submission:
New device
C. Measurand:
Creatinine
D. Type of Test:
Enzymatic, colorimetric, quantitative
E. Applicant:
Seppim S.A.S.
F. Proprietary and Established Names:
ELITech Clinical Systems Creatinine PAP SL
ELITech Clinical Systems Elical 2
ELITech Clinical Systems Elitrol I and II
G. Regulatory Information:
| Regulation | Classification | Product Code | Panel |
| --- | --- | --- | --- |
| 21 CFR 862.1255
Creatinine Test system | Class II | JFY | Chemistry (75) |
| 21 CFR 862.1150
Calibrator | Class II | JIX | Chemistry (75) |
| 21 CFR 862.1660
Quality Control material
(assayed and unassayed) | Class I, reserved | JJY | Chemistry (75) |
{1}
H. Intended Use:
1. Intended use(s):
See indications for use below.
2. Indication(s) for use:
ELITech Clinical Systems Creatinine PAP SL reagent is for the quantitative in vitro diagnostic determination of creatinine in human serum and plasma on the Vital Scientific Selectra/Flexor Analyzers. It is not intended for use in Point of Care settings.
Creatinine measurements are used in the diagnosis and treatment of renal diseases, and in monitoring renal dialysis.
ELITech Clinical Systems ELICAL 2 is a multi-parametric calibrator for in vitro diagnostic use in the calibration of quantitative ELITech Clinical Systems methods on Vital Scientific Selectra Junior Analyzer and the Vital Scientific Flexor Junior Analyzers.
ELITech Clinical Systems ELITROL I is a multi-parametric control serum for in vitro diagnostic use in accuracy control of quantitative ELITech Clinical Systems methods on the Vital Scientific Selectra Junior Analyzer and the Vital Scientific Flexor Junior Analyzer.
ELITech Clinical Systems ELITROL II is a multi-parametric control serum for in vitro diagnostic use in accuracy control of quantitative ELITech Clinical Systems methods on the Vital Scientific Selectra Junior analyzer and the Vital Scientific Flexor Junior Analyzers.
3. Special conditions for use statement(s):
The test system is not for use in Point of Care settings
4. Special instrument requirements:
Performance was provided for the Vital Scientific Selectra Junior Analyzer which is also trademarked as the Vital Scientific Flexor Junior Analyzer.
I. Device Description:
The ELITech Clinical Systems Creatinine PAP SL reagent kit consists of two liquid ready to use reagents, "R1" and "R2". Reagent R1 contains: MOPS buffer (pH 7.50), EHSPT (N-Ethyl-N-2(-Hydroxy-3-Sulfopropyl)-m-Toluidine), Creatinase (microorganism), Sarcosine oxidase (microorganism), Ascorbate oxidase (vegetal).
2
{2}
Reagent R2 contains: MOPS buffer (pH 7.50), Amino-4-antipyrine (4-AAP), Creatininase (bacterial), Peroxidase (horseradish), and Sodium azide.
ELITech Clinical Systems ELICAL 2 is a lyophilized calibrator based on human serum containing added constituents.
ELITech Clinical Systems ELITROL I and II is a two level quality control product consisting of lyophilized human serum containing constituents at desired levels.
ELICAL2, ELITROL I and II are prepared exclusively from the blood of donors tested individually and found to be negative for HbsAg and to antibodies to HCV and HIV according to FDA-approved methods or methods in compliance with the European Directive 98/79/EC, Annex II, List A.
## J. Substantial Equivalence Information:
Predicate device name(s) and K number(s):
Roche cobas c 111 Creatinine plus ver. 2 (k024098)
Roche Diagnostics Calibrator for Automated Systems (C.f.a.s) (k033501)
Roche Diagnostics Precinorm Universal and Precipath Universal Control Sera (k041227)
## 3. Comparison with predicate:
| Similarities and Differences ELITECH Clinical Systems Creatinine PAP SL | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Indications for use | Same | Creatinine measurements are used in the diagnosis and treatment of renal diseases, and in monitoring renal dialysis |
| Type of test | Same | Enzymatic colorimetric |
| Specimen | Serum and lithium heparin plasma | Serum, lithium heparin and EDTA plasma, urine |
| Standardization | Same | Traceable to ID-MS |
| Reportable range | 0.28-22.30 mg/dL | 0.06-30.5 mg/dL |
| Similarities and Differences ELITech Clinical Systems ELICAL 2 | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended use | Same | For in vitro diagnostic use |
{3}
| Similarities and Differences ELITech Clinical Systems ELICAL 2 | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | | in the calibration of quantitative methods |
| Matrix | Same | Lyophilized calibrator based on human serum with constituents added as required to obtain desired component levels |
| Level | Same | Single level |
| Storage and stability | Same | Lyophilized: store at 2-8°C and protect from light until the expiry date. After reconstitution: 8 hrs between 15-25°C, 2 days between 2-8°C, 4 weeks between -25° to -15°C (freeze only once). |
| Similarities and Differences ELITech Clinical Systems ELITROL I and II | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended use | Same | For in vitro diagnostic use in accuracy control of quantitative methods |
| Matrix | Same | Lyophilized calibrator based on human serum with constituents added as required to obtain desired component levels |
| Levels | Same | Two levels |
| Stability | Same | Lyophilized: store at 2-8°C and protect from light until the expiry date. After reconstitution: 12 hrs between 15-25°C, 5 days between 2-8°C, 4 weeks between -25° to -15°C (when frozen once). |
{4}
5
K. Standard/Guidance Document Referenced (if applicable):
Evaluation of precision performance of quantitative measurement methods; Approved guideline (CLSI EP5-A2)
Evaluation of the linearity of the measurement of quantitative procedures: a statistical approach, Approved guideline (CLSI EP6-A)
Interference Testing in Clinical Chemistry, Approved guideline (CLSI EP7-A2)
Method comparison and bias estimate using patient samples-Second Edition (CLSI EP9-A2)
Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline (CLSI EP17-A)
L. Test Principle:
Creatininase hydrolyzes creatinine in the sample to creatine. Creatine is hydrolyzed by creatinase to sarcosine and urea. Sarcosine is then oxidized by sarcosine oxidase to produce hydrogen peroxide (H₂O₂). H₂O₂ reacts with 4-amino-antipyrine (4-AAP) and EHSPT (N-Ethyl-N-2(-Hydroxy-3-Sulfopropyl)-m-Toluidine) under the catalytic action of peroxidase to form a colored quinoneimine. The absorbance of the quinoneimine at 546 nm is proportional to the concentration of creatinine in the sample.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
A precision study was performed by assaying three serum pools at varying concentrations over 20 operating days, on one Selectra Junior analyzer and with one lot of reagent. Assay runs were performed twice per day and each sample was tested in duplicate. Calibration was performed for each run. The results are summarized below.
| Samples | n | Mean (mg/dL) | Within-run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SD | CV(%) | SD | CV(%) |
| Serum pool 1 | 80 | 0.59 | 0.01 | 1.3% | 0.02 | 3.2% |
| Serum pool 2 | 80 | 1.62 | 0.01 | 0.8% | 0.03 | 1.6% |
| Serum pool 3 | 80 | 6.93 | 0.09 | 1.4% | 0.19 | 2.8% |
{5}
Another precision study was performed by assaying three serum pools and one control at varying concentrations over 10 operating days, on two Selectra Junior analyzers and with two lots of reagents. Assay runs were performed once per day and calibrated according to the device labeling. Each sample was tested in duplicate. The results are summarized below.
| Samples | n | Mean (mg/dL) | Within-run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SD | CV(%) | SD | CV(%) |
| Serum pool 1 | 40 | 0.58 | 0.01 | 2.3% | 0.02 | 2.9 % |
| Serum pool 2 | 40 | 1.57 | 0.02 | 1.5 % | 0.03 | 2.0 % |
| Serum pool 3 | 40 | 6.70 | 0.05 | 0.7 % | 0.12 | 1.7 % |
| Control | 40 | 3.74 | 0.02 | 0.7 % | 0.06 | 1.6% |
b. Linearity/assay reportable range:
A linearity study across the claimed assay range was performed by preparing samples from patient serum pools at a low creatinine concentration (0.28 mg/dL) and a high concentration (22.32 mg/dL). The low pool contained serum diluted with saline. The low and high pools were mixed in different proportions to create 9 dilutions. Eleven samples in all were assayed in triplicate
Data was analyzed using 1st, 2nd, and 3rd order least square regressions according to CLSI Protocol EP6-A. Based on the analysis the sponsor determined that first order regression was the best fit. The analysis of the first order regression yielded the following equation:
$$
1 ^ {\mathrm {s t}} \text {order} \mathrm {y} = 2. 2 \mathrm {x} - 1. 8 4
$$
Additionally, the expected values were plotted against the observed values and a linear regression line was fitted with the following regression equation:
$$
y = 0. 9 9 3 9 x + 0. 1 4, r ^ {2} = 0. 9 9 9 9
$$
Based on these results, the sponsor claims that the assay is linear from 0.28-22.30 mg/dL.
c. Traceability, Stability, Expected values (controls, calibrators, or methods): The calibrator material is purchased from a commercially available source. The creatinine in ELICAL 2 is traceable to ID-MS (Isotope dilution-Mass Spectrometry). Elical 2 calibrator is value assigned from an internal master lot of calibrator using three bottles of reconstituted Elical 2 calibrator evaluated in two runs on two Vital Scientific Flexor Junior Analyzers. The target value is the mean of the observed range of values and is listed on value sheet accompanying each lot of Elical 2. The sponsor claims the following
{6}
stability for Elical 2 after reconstitution: up to 8 hours when stored at 15-25° C, 2 days when stored at 2-8° C, and 4 weeks when stored at -25° to -15° C (freeze only one). The lyophilized material (unopened) is stable until the expiration date printed on the vial, 11 months when stored at 2-8° C.
Elitrol I and II control solutions are value assigned using two Vital Scientific Flexor Junior Analyzers. Each sample is tested in triplicate over several days. The target value of Level I and II are the median of the observed values range. After validation of the target value, a confidence range (high and low values) is calculated. The confidence range and target values are listed on the value sheet accompanying each lot of Elitrol I and II. The sponsor claims the following stability after reconstitution: up to 12 hrs between 15-25° C, 5 days between 2-8° C, 4 weeks between -25° to -15° C (when frozen once). The lyophilized material (unopened) is stable until the expiration date printed on the vial, 30 months when stored at 2-8° C.
d. Detection limit:
The sponsor states that the Limits of Blank (LoB), Detection (LoD) and Quantitation (LoQ) studies were conducted following CLSI EP17-A. The LoB study was performed by assaying 60 replicates of a blank sample in one assay run. The LoD was determined by assaying 15 replicates of four diluted patient pools with very low creatinine concentrations in one assay run. The LoQ was determined by assaying 15 replicates of four diluted patient pools with low creatinine concentrations (approximately 0.28 mg/dL) in one assay run. Each study was performed using one Selectra Junior analyzer and one lot of creatinine reagent.
The sponsor states that the limit of blank is 0.001 mg/dL and the limit of detection 0.0065 mg/dL, however the analyzer does not report creatinine values below the claimed limit of quantitation. The limit of quantitation based on the study described above is stated to be 0.28 mg/dL. Imprecision of the samples evaluated in the LoQ study ranged from 1.4-4.8% (CV%).
e. Analytical specificity:
Testing for interfering substances was based on CLSI EP-7A. Testing was performed on a minimum of five concentrations for each interfering substance. Two different clinically relevant concentrations of creatinine (1.58 and 6.67 mg/dL) were used for evaluation. Samples with increasing amounts of potential interferents were tested in triplicate and compared to the same sample without the interferent.
The sponsor defined non-significant interference as the highest level tested that does not cause >10% change between the tested samples and the control sample.
7
{7}
8
| | Highest concentration tested that showing no significant interference (mg/dL) (≤10%) |
| --- | --- |
| Triglyceride | 3,198 |
| Unconjugated Bilirubin | 30 |
| Conjugated Bilirubin | 14.8 |
| Hemoglobin | 500 |
| Glucose | 550 |
| Ascorbic Acid | 20 |
| Uric acid | 24 |
A positive bias was observed with creatine ≥ 5 mg/dL.
Calcium dobesilate, Methyl-dopa and L-dopa cause falsely low results at therapeutic concentrations.
A summary of these results and interferences is found in the labeling.
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
The performance of the Creatinine PAP SL reagent on the Selectra Junior analyzer was compared with the Roche Creatinine plus v2 reagent on the cobas c111 analyzer. Study performed according to CLSI protocol EP9-A2 using 91 serum samples covering the measuring range with values from 0.3 to 20.30 mg/dL. The linear regression results are summarized below:
| | Regression parameters | Confidence Interval 95% | |
| --- | --- | --- | --- |
| | | Lower 95% | Upper 95% |
| Slope (b) | 1.045 | 1.039 | 1.052 |
| Intercept (a) | -0.01 | -0.05 | 0.04 |
b. Matrix comparison:
To demonstrate comparable performance between serum and lithium heparin, the sponsor evaluated 36 matched serum and plasma sets ranging from 0.33 to
{8}
21.46 mg/dL on the Selectra Junior analyzer. The linear regression analysis resulted in a slope of 1.013 (95% CI 1.008-1.018), an intercept of 0.00 (95% CI -0.03 to 0.03), correlation coefficient of r=1.000
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
In the labeling, the sponsor recommends that each laboratory should establish and maintain its own reference values. The data given are for information only.
REFERENCE VALUES
| | Men | Women |
| --- | --- | --- |
| Serum : | 0.72 - 1.18 | 0.55 - 1.02 mg/dL |
| | 64 - 104 | 49 - 90 μmol/L |
Ceriotti, F., Reference Intervals for Serum Creatinine Concentrations: Assessment of Available Data for Global Application. Clin. Chem., (2008), 54, 559.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.