K191360 · Abbott Point of Care, Inc. · JFP · Feb 14, 2020 · Clinical Chemistry
Device Facts
Record ID
K191360
Device Name
i-STAT CHEM8+ cartridge with the i-STAT 1 System
Applicant
Abbott Point of Care, Inc.
Product Code
JFP · Clinical Chemistry
Decision Date
Feb 14, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1145
Device Class
Class 2
Intended Use
The i-STAT CHEM8+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of ionized calcium in arterial or venous whole blood in point of care or clinical laboratory settings. Ionized calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany.
Device Story
The i-STAT CHEM8+ cartridge is a single-use, disposable device for the in vitro quantification of ionized calcium (iCa) in whole blood. Used with the handheld i-STAT 1 Analyzer, the system processes 95 µL of arterial or venous whole blood. The analyzer moves the sample across sensors within the cartridge to perform an ion-selective electrode measurement, providing quantitative results in approximately 2 minutes. The system is intended for use by trained medical professionals in point-of-care or clinical laboratory settings. The analyzer performs automated internal quality checks on every test; if specifications are not met, it suppresses results and displays a Quality Check Code (QCC) to guide the operator. Results assist clinicians in diagnosing and managing parathyroid, bone, and chronic renal diseases, as well as tetany.
Clinical Evidence
No clinical trials were performed. Evidence consists of analytical bench testing, including precision/reproducibility studies (n=90 per level for POC precision), linearity (R²=0.9992), and method comparison against the predicate. Method comparison used 250 specimens (136 venous, 114 arterial), showing strong correlation (r=0.99-1.00 for combined samples). Interference testing evaluated exogenous and endogenous substances; dose-response analysis established limits for substances like leflunomide, teriflunomide, and sodium thiosulfate.
Technological Characteristics
Microfabricated thin-film ion-selective electrodes on a silicon chip. Potentiometric measurement principle. Single-use disposable cartridge format. Requires i-STAT 1 Analyzer for fluid movement and signal processing. Traceable to NIST SRM 956. Measurement range: 0.25–2.50 mmol/L. Sample type: lithium heparinized venous or arterial whole blood.
Indications for Use
Indicated for patients requiring ionized calcium quantification in arterial or venous whole blood for the diagnosis and treatment of parathyroid disease, bone diseases, chronic renal disease, and tetany.
Regulatory Classification
Identification
A calcium test system is a device intended to measure the total calcium level in serum. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).
Predicate Devices
Ionized Calcium test with the Epocal EPOC Blood Analysis System (K061597)
Related Devices
K230300 — i-STAT CG8+ cartridge with the i-STAT 1 System · Abbott Point of Care, Inc. · Oct 27, 2023
K210958 — i-STAT CHEM8+ cartridge with the i-STAT 1 System · Abbott Point of Care, Inc. · Aug 13, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The logos are placed side by side.
February 14, 2020
Abbott Point of Care, Inc. Susan Tibedo Director, Regulatory Affairs 400 College Road East Princeton, NJ 08540
Re: K191360
Trade/Device Name: i-STAT CHEM8+ cartridge with the i-STAT 1 System Regulation Number: 21 CFR 862.1145 Regulation Name: Calcium Test System Regulatory Class: Class II Product Code: JFP Dated: January 15, 2020 Received: January 16, 2020
Dear Susan Tibedo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Marianela Perez-Torres, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K191360
Device Name
i-STAT CHEM8+ cartridge with the i-STAT 1 System
Indications for Use (Describe)
The i-STAT CHEM8+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of ionized calcium in arterial or venous whole blood in point of care or clinical laboratory settings.
Ionized calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany.
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;"> <svg height="12" width="12"> <rect fill="none" height="12" stroke="black" width="12" x="0" y="0"></rect> <path d="M2,2 L10,10 M2,10 L10,2" stroke="black"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;"> <svg height="12" width="12"> <rect fill="none" height="12" stroke="black" width="12" x="0" y="0"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
The information in this 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
| 1. | Submitter Information | | |
|----|-----------------------|-----------------------------|--|
| | Owner | Abbott Point of Care Inc. | |
| | | 400 College Road East | |
| | | Princeton, NJ 08540 | |
| | Contact | Primary: Susan Tibedo | |
| | | Director Regulatory Affairs | |
| | | susan.tibedo@abbott.com | |
| | | Phone: 609-454-9360 | |
| | | Secondary: Maria Figueroa | |
| | | Manager Regulatory Affairs | |
| | | maria.l.figueroa@abbott.com | |
| | | Phone: 609-454-9271 | |
| | | | |
| | Date Prepared | February 14, 2020 | |
| | 510(k) Number | K191360 | |
### 2. Device Information
Proprietary Name i-STAT CHEM8+ cartridge with i-STAT 1 System
Common Name Chemistry test, analyzer, handheld
| Product<br>code | Device Classification<br>name | Regulation<br>Number | Class | Panel |
|-----------------|-------------------------------------|----------------------|-------|--------------------|
| JFP | Electrode, Ion Specific,<br>Calcium | 862.1145 | II | Clinical Chemistry |
## 3. Predicate Device
Proprietary Name Ionized Calcium test with the Epocal EPOC Blood Analysis System
510(k) Number K061597
| Product<br>code | Device Classification<br>name | Regulation<br>Number | Class | Panel |
|-----------------|-------------------------------------|----------------------|-------|--------------------|
| JFP | Electrode, Ion Specific,<br>Calcium | 862.1145 | II | Clinical Chemistry |
{4}------------------------------------------------
# 4. Device Description
The i-STAT CHEM8+ test cartridge contains test reagents to analyze whole blood at the point of care or in the clinical laboratory for ionized calcium (iCa). The test is contained in a single-use, disposable cartridge. Cartridges require two to three drops of whole blood which are typically applied to the cartridge using a transfer device.
The i-STAT 1 Analyzer is a handheld, in vitro diagnostic analytical device designed to run only i-STAT test cartridges. The instrument interacts with the cartridge to move fluid across the sensors and generate a quantitative result (within approximately 2 minutes).
The i-STAT 1 System is comprised of the i-STAT 1 analyzer, the i-STAT test cartridges and accessories (i-STAT 1 Downloader/Recharger, electronic simulator and portable printer). The system is designed for use by trained medical professionals at the patient point of care or in the clinical laboratory and is for prescription use only.
# 5. Intended Use Statement
The i-STAT CHEM8+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of ionized calcium in arterial or venous whole blood in point of care or clinical laboratory settings.
Ionized calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany.
| Feature or<br>Characteristic | Predicate<br>lonized Calcium Test with<br>EPOC Blood Analysis System<br>(K061597) | Candidate<br>lonized Calcium Test with i-STAT 1<br>System |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Measurement of Ionized Calcium<br>is used in diagnosis and<br>treatment of parathyroid disease,<br>a variety of bone diseases,<br>chronic renal disease and<br>tetany. | The i-STAT CHEM8+ cartridge with<br>the i-STAT 1 System is intended for<br>use in the in vitro quantification of<br>ionized calcium in arterial or venous<br>whole blood in point of care or clinical<br>laboratory settings. |
| | | lonized calcium measurements are<br>used in the diagnosis and treatment<br>of parathyroid disease, a variety of<br>bone diseases, chronic renal disease |
# 6. Summary Comparison of Technological Characteristics
{5}------------------------------------------------
| Feature or<br>Characteristic | Predicate<br>lonized Calcium Test with<br>EPOC Blood Analysis System<br>(K061597) | Candidate<br>lonized Calcium Test with i-STAT 1<br>System |
|------------------------------|-----------------------------------------------------------------------------------|-----------------------------------------------------------|
| | | and tetany. |
| Reportable | 0.25 - 4.00 mmol/L | 0.25 - 2.50 mmol/L (mEq/L) |
| Range | | |
| Sample Type | Arterial or venous whole blood | Arterial or venous whole blood |
| Sample Volume | At least 92 µL | 95 µL |
| Sample | Ready to Use | Ready to Use |
| Preparation | | |
| Traceability | NIST SRM 956 | NIST SRM 956 |
| Calibration | On board the instrument initiated | 1-point on-board contained within the |
| | once test card is inserted | cartridge |
| Time to Test | Immediately after drawing | Heparinized samples: within 10 |
| (Sample | sample | minutes of collection |
| Stability) | | |
| Principle of | lon selective electrode | lon selective electrode |
| Measurement | | |
| Reagent Format | Test Card | Cartridge |
| Reagent Storage | 15 to 30°C (59 to 86°F) | 2°C to 8°C (35-46°F) |
| and Stability | | |
| Analyzer Type | Handheld | Handheld |
#### 7. Performance Characteristics
### Analytical Performance
#### a. Precision
## Precision 20 days (Aqueous Materials)
The precision of the i-STAT iCa test on the i-STAT 1 Wireless Analyzer was evaluated using 5 levels of aqueous materials. This 20-day multi-day precision testing was based on CLSI document EP05-A3: Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline-Third Edition. The study was conducted using multiple instruments and one test cartridge lot over 20 days at one site. Total precision ('within-laboratory', ST), within-run, (Sr), between-run, (Sr) and between-day, (Sad) were estimated for each level. The results of the 20-day precision study are shown in Table 1.
| i-STAT<br>Test | Calibration<br>Verification<br>Level | N | Mean /<br>Median | Total | | Within-run | | Between-run | | Between-day | |
|-----------------|--------------------------------------|----|------------------|--------|------------|------------|------------|-------------|-------------|-------------|-------------|
| | | | | ST | CVT<br>(%) | Sr | CVr<br>(%) | Srr | CVrr<br>(%) | Sdd | CVdd<br>(%) |
| iCa<br>(mmol/L) | CV L1 | 81 | 2.328 | 0.0132 | 0.6 | 0.0121 | 0.5 | 0.0037 | 0.2 | 0.0038 | 0.2 |
| iCa<br>(mmol/L) | CV L2 /<br>Control L1*,a | 81 | 1.484 | 0.0103 | 0.7 | 0.0096 | 0.6 | 0.0025 | 0.2 | 0.0026 | 0.2 |
{6}------------------------------------------------
| i-STAT<br>Test | Calibration<br>Verification<br>Level | N | Mean /<br>Median | Total | | Within-run | | Between-run | | Between-day | |
|----------------|--------------------------------------|----|------------------|--------|------------|------------|------------|-------------|-------------|-------------|-------------|
| | | | | ST | CVT<br>(%) | Sr | CVr<br>(%) | Srr | CVrr<br>(%) | Sdd | CVdd<br>(%) |
| | CV L3 /<br>Control L2* | 81 | 1.299 | 0.0067 | 0.5 | 0.0062 | 0.5 | 0.0020 | 0.2 | 0.0018 | 0.1 |
| | CV L4 /<br>Control L3*,a | 80 | 0.724 | 0.0038 | 0.5 | 0.0036 | 0.5 | 0.0009 | 0.1 | 0.0008 | 0.1 |
| | CV L5 | 80 | 0.262 | 0.0040 | 1.5 | 0.0035 | 1.3 | 0.0015 | 0.6 | 0.0010 | 0.4 |
* The aqueous control materials (Control L1, L2, L3) are also used as the middle levels of the calibration verification set (CV L2, CV L3, CV L4). The aqueous fluids are the same.
a Non-normal distribution (p-value <0.010), the median value was used to calculate %CV.
### Precision (Whole Blood)
A whole blood repeatability analysis was conducted using the data collected across three point of care sites. Two hundred and forty-one samples (132 venous and 109 arterial) were measured in duplicate. The mean values for each sample were divided into four subintervals for each sample type taking into consideration the lower and upper limits of the reference interval, 1.12 mmol/L and 1.32 mmol/L.
The results are provided in Table 2 and Table 3 below:
| Table 2: Venous whole blood | | | | | | |
|-----------------------------|-----|------------------|--------|--------|--|--|
| Sample Range<br>(mmol/L) | N | Mean<br>(mmol/L) | SD | CV (%) | | |
| 0.25 - 0.75 | 10 | 0.438 | 0.0097 | 2.2 | | |
| 0.75 – 1.2 | ರಿತ | 1.094 | 0.0173 | 1.6 | | |
| 1.2 — 1.5 | 22 | 1.278 | 0.0134 | 1.0 | | |
| 1.5 — 2.5 | 7 | 2.109 | 0.0183 | 0.9 | | |
| Table 3: Arterial whole blood | | | | |
|-------------------------------|----|------------------|--------|--------|
| Sample Range<br>(mmol/L) | N | Mean<br>(mmol/L) | SD | CV (%) |
| 0.25 - 0.75 | 3 | 0.445 | 0.0041 | 0.9 |
| 0.75 – 1.2 | 73 | 1.110 | 0.0329 | 3.0 |
| 1.2 – 1.5 | 27 | 1.244 | 0.0105 | 0.8 |
| 1.5 – 2.5 | 6 | 1.725 | 0.0091 | 0.5 |
#### b. Linearity
The study was designed based on CLSI EP06-A: Evaluation of the linearity of quantitative measurement procedures.
The linearity of the i-STAT iCa test on the i-STAT 1 Analyzer was evaluated by preparing whole blood samples of varying analyte levels that spanned the reportable range of the test. The best fitting regression model was a second order model. The absolute degree of nonlinearity results met the acceptance criteria for each of the levels tested. Therefore, the i-STAT iCa demonstrated linearity over the reportable range
{7}------------------------------------------------
0.25 - 2.50 mmol/L. Regression summary of the iCa response versus the concentration of the whole blood samples of varying analyte levels is provided in Table 4.
| Table 4: Regression Summary for the i-STAT iCa test on the i-STAT 1 Analyzer | | | | | | |
|------------------------------------------------------------------------------|------------------------------|--------------------------|--------|-----------|-------|--|
| i-STAT Test<br>(mmol/L) | Reportable<br>Range (mmol/L) | Range Tested<br>(mmol/L) | Slope | Intercept | R2 | |
| iCa | 0.25 - 2.50 | 0.22 - 2.81 | 0.9738 | 0.360 | 0.999 | |
# c. Limit of Quantitation (LoO)
The study was based on the CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline-Second Edition.
The LoQ of the i-STAT iCa test was evaluated on the i-STAT 1 Analyzer using whole blood that was altered to low iCa (< 0.25 mmol/L). The study was conducted over four (4) days using two (2) cartridge lots. The LoQ for the i-STAT iCa test was determined to be 0.21 mmol/L, which is below the lower limit of the i-STAT iCa test reportable range (0.25 - 2.50 mmol/L).
# d. Limit of Blank and Detection (LoB/LoD)
The study was based on CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline–Second Edition.
The LoB/LoD of the i-STAT iCa test was evaluated on the i-STAT 1 Analyzer using whole blood that was altered to a "blank" ionized calcium concentration for LoB testing and two "low" ionized calcium concentrations for LoD testing. The LoB and LoD were determined based on the maximal LoB or LoD value obtained for each cartridge lot tested. The LoB for i-STAT iCa test was determined to be 0.14 mmol/L and the LoD was determined to be 0.15 mmol/L.
# e. Interference
The interference performance of the i-STAT iCa test on the i-STAT 1 Analyzer was evaluated using whole blood samples based on CLSI EP07 ED3: Interference Testing in Clinical Chemistry, Third Edition. The effect of each substance was evaluated by comparing the performance of a control sample, spiked with blank solvent solution, with the test results from a sample spiked with the potentially interfering substance at the toxic/pathological concentration based on CLSI EP37 ED1: Supplemental Tables for Interference Testing in Clinical Chemistry, First Edition, as applicable. A substance was identified as an interferent if the difference between the control and test samples was outside of the allowed error (Ea) for the i-STAT test.
Table 5 contains the list of potentially interfering substances tested for the i-STAT Ionized Calcium test and the interference results.
{8}------------------------------------------------
| test | | Test Concentration | Interference | | |
|-----------------------|-----------------|--------------------|--------------|-------------------------------------|--|
| Substance | mmol/L<br>mg/dL | | (Yes/No) | Interference Results | |
| Acetaminophen | 1.03 | 15.6 | No | | |
| Bilirubin | 0.684 | 40 | No | | |
| Cholesterol | 10.3 | 400 | No | | |
| Hemoglobin | 10 g/L | 1000 | No | | |
| β-Hydroxybutyric Acid | 6.0* | 62.46 | No | | |
| L-Ascorbic Acid | 0.298 | 5.25 | No | | |
| Leflunomide | 0.722* | 19.5 | Yes | Decreased results<br>≥ 0.4 mmol/L | |
| Lithium Bromide | 37.5* | 325.69 | No | | |
| Lithium Lactate | 10 | 90 | Yes | Decreased results<br>≥ 6.3 mmol/L | |
| Lithium Salicylate | 0.207 | 2.86 | No | | |
| Lithium Thiocyanate | 0.898 | 5.22 | Yes | Decreased results<br>≥ 0.874 mmol/L | |
| Magnesium Chloride | 4.1 | 10 | No | | |
| N-Acetyl-L-Cysteine | 0.92 | 15.0 | No | | |
| Potassium Chloride | 8 | 59.6 | No | | |
| Sodium Chloride | 170 | 993.48 | No | | |
| Sodium lodide | 2.99* | 44.82 | No | | |
| Sodium Thiosulfate | 16.7* | 264.04 | Yes | Decreased results<br>≥ 5.5 mmol/L | |
| Teriflunomide | 0.722* | 19.5 | Yes | Decreased results<br>≥ 0.1 mmol/L | |
| Triglyceride | 16.94 | 1500 | No | | |
Table E. Cummary
* No CLSI EP37-ED1 test concentration available. The molecular weight of the substance tested was used to convert the test concentration from mmol/L to mg/dL. The molecular weight of each substance could vary depending on the form chosen.
## Comparison Study
## f. Method Comparison with Predicate Device
Method comparison was demonstrated in a study comparing the i-STAT iCa test performance on the i-STAT 1 to the EPOC Blood Analysis System. The study was based on CLSI guideline EP09c-ED3. Venous and arterial blood specimens were evaluated and analyzed on the i-STAT 1 against venous and arterial blood specimens on the EPOC Blood Analysis System. A Passing-Bablok linear regression analysis was performed using the first replicate result from the i-STAT 1 versus the singlicate result of the comparative method.
The i-STAT System automatically runs a comprehensive set of quality checks of both the analyzer and cartridge performance each time a sample is tested. This internal quality system will suppress results by generating a Quality Check Code (QCC) if the analyzer, cartridge or sample does not meet certain internal specifications. When a QCC occurs, a single code number, the type of problem and
{9}------------------------------------------------
the next step to be taken will be displayed on the i-STAT Analyzer. The failure rate for a single cartridge due to QCCs may be as high as 4%. The rate of failure for two consecutive cartridges due to QCCs may be as high as 1.7%.
| Table 6: Method Comparison Results | | | | |
|------------------------------------|-----|-------|-----------|------|
| i-STAT Test | N | Slope | Intercept | r |
| iCa | 250 | 1.00 | -0.02 | 0.99 |
# 8. Conclusion
The results of these studies demonstrate that performance of the i-STAT CHEM8+ Ionized Calcium test with the i-STAT 1 System are substantially equivalent to the comparative method.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.