ELTECH CHINICAL SYSTEM AMYLASE SL ELITECH CLINICAL SYSTEMS ELICAL2 ELTECH CLINICAL SYSTEMS ELITROL I AND ELITROL II
Device Facts
| Record ID | K122858 |
|---|---|
| Device Name | ELTECH CHINICAL SYSTEM AMYLASE SL ELITECH CLINICAL SYSTEMS ELICAL2 ELTECH CLINICAL SYSTEMS ELITROL I AND ELITROL II |
| Applicant | Elitechgroup |
| Product Code | JFJ · Clinical Chemistry |
| Decision Date | Oct 3, 2012 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1070 |
| Device Class | Class 2 |
Intended Use
ELITech Clinical Systems AMYLASE SL is intended for the quantitative in vitro determination of amylase in human serum and plasma on ELITech Clinical Systems Selectra analyzers. It is not intended for use in Point of Care settings. Measurements of amylase are used primarily for the diagnosis and treatment of pancreatitis (inflammation of the pancreas).
Device Story
ELITech Clinical Systems AMYLASE SL is an in vitro diagnostic reagent kit for measuring amylase activity in human serum and plasma. The device uses an enzymatic method where the substrate CNP-G3 is hydrolyzed by amylase to produce CNP. The rate of increase in absorbance at 405 nm is measured on ELITech Clinical Systems Selectra analyzers; this rate is directly proportional to amylase activity. The system includes ELICAL 2 (calibrator) and ELITROL I/II (control sera). The device is used in clinical laboratory settings by trained personnel. Results assist clinicians in diagnosing and monitoring pancreatitis. The system provides quantitative data to support clinical decision-making regarding pancreatic inflammation.
Clinical Evidence
Bench testing only. Performance validated per CLSI guidelines (EP5-A2, EP6-A, EP17-A, EP7-A2, EP9-A2). Precision: Total CV% 2.2-2.7%. Linearity: 20-1500 U/L. LoD: 6 U/L; LoQ: 13 U/L. Method comparison (n=100) vs. Roche AMYL2 yielded y = 0.976x - 1, r=0.999. Plasma comparison (n=48) yielded y = 0.907x + 1, r=1.000. No significant interference from bilirubin, triglycerides, hemoglobin, acetaminophen, ascorbic acid, or acetylsalicylic acid at tested concentrations.
Technological Characteristics
Enzymatic assay using CNP-G3 substrate. Reagent composition: MES buffer (pH 6.15), NaCl, CaCl2, potassium thiocyanate, CNP-G3, sodium azide. Lyophilized human serum-based calibrators and controls. Photometric detection at 405 nm. Designed for use on Selectra ProM analyzers. No software algorithm class specified; standard clinical chemistry photometric analysis.
Indications for Use
Indicated for the quantitative in vitro determination of amylase in human serum and plasma to aid in the diagnosis and treatment of pancreatitis. Not for Point of Care use. Prescription use only.
Regulatory Classification
Identification
An amylase test system is a device intended to measure the activity of the enzyme amylase in serum and urine. Amylase measurements are used primarily for the diagnosis and treatment of pancreatitis (inflammation of the pancreas).
Related Devices
- K040534 — VITALAB AMYLASE REAGENT · Clinical Data, Inc. · Mar 11, 2004
- K974125 — IL TEST AMYLASE · Instrumentation Laboratory CO · Dec 17, 1997
- K013101 — WIENER LAB. AMILASA 405 CINETICA AA, WIENER LAB.AMILASA 405 CINETICA UNITEST,WIENER LAB.AMILASA CINETICA AA · Wiener Laboratories Saic · Nov 16, 2001
- K040920 — AMYLASE LIQUID REAGENT, MODEL A532-200 · Teco Diagnostics · May 10, 2004