The Roche cobas c 111 analyzer is an in-vitro diagnostic analyzer capable of performing clinical chemistry, specific protein and electrolyte tests in the professional setting and small laboratories, specialized testing and CLIA-licensed doctor's offices. Analytes are measured photometrically or turbidimetrically; the analyzer also has an optional ISE module for measuring sodium, potassium and chloride.
Device Story
Cobas c 111 analyzer performs clinical chemistry, specific protein, and electrolyte testing. Inputs: human serum, plasma, or urine samples. Operation: photometric/turbidimetric assays or optional ion-selective electrode (ISE) module. Output: quantitative analyte concentrations. Used in professional settings, small labs, and CLIA-licensed doctor's offices by trained technicians. Healthcare providers use results for diagnosis and monitoring of liver/heart disease, diabetes, tissue injury, and electrolyte imbalances. System provides automated analysis to support clinical decision-making at point-of-care.
Clinical Evidence
Bench testing only. Performance evaluated via method comparison studies (n=280-333) against the Integra 400 analyzer, demonstrating high correlation (Passing Bablok regression). Precision studies (within-run and total CV%) conducted on human sera and controls for all analytes (AST, Glucose, CRP, ISE-Cl/K/Na). Interference studies confirmed no significant impact from common drugs, bilirubin, hemolysis, or lipemia within specified limits.
Technological Characteristics
Photometric/turbidimetric analyzer with optional ISE module. Reagents include enzymes (MDH, hexokinase, G6PDH), latex particles, and buffers. ISE module uses neutral carrier (Na, K) and ion exchanger (Cl) electrodes. Standardized against IFCC, ID/MS, or gravimetric primary calibrators. Complies with CLSI EP5-A for precision.
Indications for Use
Indicated for professional use in small labs, specialized testing, and CLIA-licensed doctor's offices for quantitative determination of AST, glucose, CRP, sodium, potassium, and chloride in human serum, plasma, and urine (electrolytes only). Used for diagnosis/treatment of liver/heart disease, carbohydrate metabolism disorders, tissue injury, and electrolyte/metabolic disorders.
Regulatory Classification
Identification
An aspartate amino transferase (AST/SGOT) test system is a device intended to measure the activity of the enzyme aspartate amino transferase (AST) (also known as a serum glutamic oxaloacetic transferase or SGOT) in serum and plasma. Aspartate amino transferase measurements are used in the diagnosis and treatment of certain types of liver and heart disease.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
{0}------------------------------------------------
K071211
## JUL 3 0 2007
# 510(k) Summary
| Introduction | According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence. | | | |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|----------------------------------------------------------|---------------------|
| Submitter name, address, contact | Roche Diagnostics Operations<br>9115 Hague Road<br>Indianapolis, IN 46250<br>317-521-3831<br><br>Contact Person: Corina Harper<br><br>Date Prepared: April 2007 | | | |
| Device name | Proprietary name: cobas c 111 Analyzer<br>Common name: Analyzer, Chemistry (Photometric, Discrete), for clinical use<br>Classification name: Discrete photometric chemistry analyzer for clinical use | | | |
| Establishment information | The establishment registration number for Roche Diagnostics Ltd. is 3003795116.<br><br>The establishment registration number for Roche Diagnostics GmbH is 9610126.<br><br>The establishment registration number for Roche Diagnostics Corporation is 1823260. | | | |
| Classification | The FDA has classified a discrete photometric analyzer in Class I. | | | |
| | Panel | Classification Number | Classification Name | Regulation Citation |
| | 75 Clinical Chemistry | 1 | Discrete photometric chemistry analyzer for clinical use | 21 CFR 862.2160 |
{1}------------------------------------------------
| Performance<br>standards | To date, no performance standards that affect this device have been finalized<br>under Section 514 of the Act. |
|-------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submission<br>history | The cobas c111 analyzer is a member of the COBAS INTEGRA® family of<br>analyzers cleared in the Total Bilirubin reagent submission as K063744. The<br>reagents applied to this analyzer are cleared reagents for the COBAS INTEGRA<br>family of analyzers and were successfully applied to the cobas c 111 analyzer<br>using the Replacement Reagent and Instrument Family Policy - Dec 11, 2003, to<br>demonstrate equivalence to the original reagent/instrument performance. |
| Device<br>modification -<br>Additional<br>indication for<br>cobas c 111<br>analyzer | The table below compares the additional intended use for the device, cobas c<br>111 analyzer to the current device, cobas c 111 for professional use. |
| Topic | cobas c 111 Analyzer<br>professional use<br>(K063744) | cobas c 111 Analyzer<br>with extended intended<br>use |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Analyzer Description | | |
| Intended<br>Use/Indications of Use | The Roche cobas c 111 analyzer is an <i>in-vitro</i> diagnostic analyzer capable of performing clinical chemistry, specific protein and electrolyte tests. Analytes are measured photometrically or turbidimetrically; the analyzer also has an optional ISE module for measuring sodium, potassium and chloride. | The Roche cobas c 111 analyzer is an <i>in-vitro</i> diagnostic analyzer capable of performing clinical chemistry, specific protein and electrolyte tests in the professional setting and small laboratories, specialized testing and CLIA-licensed doctor's offices. Analytes are measured photometrically or turbidimetrically; the analyzer also has an optional ISE module for measuring sodium, potassium and chloride |
:
{2}------------------------------------------------
Device modifications reagents comparison
The following table compares the panel of six assays performance characteristics on cobas c 111 analyzer for professional use with the point-ofcare use.
| Feature | Reagent performance<br>characteristics on cobas c 111<br>analyzer for professional use | Reagent<br>performance<br>characteristics<br>for point-of-<br>care use |
|------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| Intended Use for<br>representative<br>reagents | AST:<br>In vitro test for the quantitative determination<br>of aspartate aminotransferase (AST) in human<br>serum and plasma on the cobas c 111 system.<br>Glucose:<br>In vitro test for the quantitative determination<br>of glucose in human serum and plasma on the<br>cobas c 111 system.<br>CRP:<br>In vitro test for the quantitative determination<br>of C-reactive protein in human serum and<br>plasma on the cobas c 111 system.<br>ISE-C1:<br>The chloride electrode for the cobas c 111<br>system is intended for the quantitative<br>determination of chloride in diluted serum,<br>plasma and urine.<br>ISE-K:<br>The potassium electrode for the cobas c 111<br>system is intended for the quantitative<br>determination of chioride in diluted serum,<br>plasma and urine.<br>ISE-Na:<br>The sodium electrode for the cobas c 111<br>system is intended for the quantitative<br>determination of chloride in diluted serum,<br>plasma and urine. | Same |
| Instrument | cobas c 111 analyzer | Same |
| Operator | Professional setting | Point-of-care<br>setting |
| Sample type | AST: serum and plasma<br>Glucose: serum and plasma<br>CRP: serum and plasma<br>ISE-Cl: serum, plasma and urine<br>ISE-K: serum, plasma and urine<br>ISE-Na: serum, plasma and urine | Same |
| Feature | Reagent performance<br>characteristics on cobas c 111<br>analyzer for professional use | Reagent<br>performance<br>characteristics<br>for point-of-care<br>use |
| Traceability/<br>Standardization | AST: standardized against the original<br>IFCC formulation using calibrated pipettes<br>together with manual photometer providing<br>absolute values and the substrate-<br>specific absorptivity, $ε$<br>Glucose: standardized against ID/MS<br>CRP: standardized against the reference<br>preparation of the IRMM - BCR470/<br>CRM470 (RPPHS - Reference Preparation<br>for Proteins in Human Serum)<br>ISE-Cl/K/Na: standardized against primary<br>calibrators prepared gravimetrically from<br>purified salts | Same |
| Measuring range | AST: 2-700 U/L<br>Glucose: 0.11-40 mmol/L<br>CRP: 1-200 mg/L<br>ISE-Cl: 20-250 mmol/L<br>ISE-K: 1-100 mmol/L<br>ISE-Na: 20-250 mmol/L | Same |
| Analytical<br>sensitivity<br>(LDL) | AST: 2 U/L<br>Glucose: 0.11 mmol/L<br>CRP: 1.0 mg/L<br>ISE-Cl: slope range -25 to -56 mV/dec<br>ISE-K: slope range 45 to 63 mV/dec<br>ISE-Na: slope range 45 to 63 mV/dec | Same |
{3}------------------------------------------------
Device modifications – reagents comparison (continued)
Continued on next page
{4}------------------------------------------------
| Feature | Reagent performance characteristics on cobas c 111 analyzer for professional use | Reagent performance characteristics for point-of-care use |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Precision<br>Within run<br>(CV%/SD) | AST<br>Human Sera :<br>SD 0.37 at 26.2 U/L<br>0.5% at 221 U/L<br>Controls:<br>1.1% at 39.7 U/L<br>0.4% at 123 U/L | Human sera:<br>SD 0.78 at 18.98 U/L<br>2.07% at 46.92 U/L<br>Controls:<br>1.99% at 41.05 U/L<br>0.63% at 137.93 U/L |
| | Glucose<br>Human Sera:<br>SD 0.5 at 40.9 mg/dL<br>0.8% at 180 mg/dL<br>Controls:<br>1.0% at 90.6 mg/dL<br>0.5% at 252 mg/dL | Human sera:<br>0.56% at 89.14 mg/L<br>0.63% at 168.29 mg/L<br>Controls:<br>0.63% at 96.11 mg/L<br>0.66% at 255.47 mg/L |
| | CRP<br>Human Sera:<br>SD 0.01 at 0.42 mg/dL<br>1.3% at 23.4 mg/dL<br>Controls:<br>0.8% at 1.83 mg/dL<br>0.6% at 3.77 mg/dL | Human sera:<br>SD 0.072 at 0.356 mg/dL<br>1.15%-- 0t 0.554 mg/dL<br>1.23% at 2.385 mg/dL<br>Controls:<br>0.68% at 0.842 mg/dL<br>0.41% at 4.792 mg/dL |
| | ISE – Cl<br>Human Sera:<br>0.30% at 97 mmol/L<br>0.24% at 1.27 mmol/L<br>Plasma:<br>0.29% at 93 mmol/L | Human sera:<br>0.72% at 109.03 mmol/L<br>0.72% at 98.55 mmol/L<br>Controls:<br>SD 0.43 at 86.22 mmol/L |
Device modifications – reagents comparison (continued)
{5}------------------------------------------------
| Feature | Reagent performance<br>characteristics on cobas c<br>111 analyzer for<br>professional use | Reagent performance<br>characteristics for point-<br>of-care use |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Precision<br>Within run<br>(CV%) | ISE - K<br>Human Sera:<br>0.49% at 2.1 mmol/L<br>0.40% at 5.7 mmol/L<br>Plasma:<br>0.41% at 2.0 mmol/L<br>0.32% at 60. mmol/L | Human sera:<br>0.79% at 4.65 mmol/L<br>0.71% at 4.69 mmol/L<br>Controls:<br>0.47% at 3.465 mmol/L<br>SD 0.040 at 6.621 mmol/L |
| | ISE - Na<br>Human Sera:<br>0.32% at 130 mmol/L<br>0.31% at 156 mmol/L<br>Plasma<br>0.31% at 127 mmol/L<br>0.30% at 151 mmol/L | Human sera:<br>0.59% at 144.69 mmol/L<br>0.61% at 133.70 mmol/L<br>Controls:<br>0.39% at 126.66 mmol/L<br>0.41% at 150.50 mmol/L |
| Precision<br>Total<br>(CV%/SD unit) | AST<br>Human Sera:<br>SD 0.64 at 19.5 U/L<br>1.0% at 306 U/L<br>Controls:<br>2.4% at 38.6 U/L<br>0.9% at 126 U/L | Human sera:<br>SD 1.2 at 16.4 U/L<br>3.7% at 48.7 U/L<br>Controls:<br>3.3% at 40.24 U/L<br>2.2% at 137.04 U/L |
| | Glucose<br>Human Sera:<br>SD 0.2 at 45.4 mg/dL<br>0.6% at 178 mg/dL<br>Controls:<br>0.7% at 92.3 mg/dL<br>0.5% at 254 mg/dL | Human sera:<br>2.6% at 97.5 mg/L<br>2.8% at 130.7 mg/L<br>Controls:<br>2.5% at 93.03 mg/L<br>2.5% at 247.08 mg/L |
Device modifications - reagents comparison (continued)
{6}------------------------------------------------
| Feature | Reagent performance characteristics on cobas c<br>111 analyzer for<br>professional use | Reagent performance characteristics for<br>point-of-care use |
|--------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Precision<br>Total<br>(CV%/SD unit) | CRP<br>Human Sera:<br>SD 0.01 at 0.47 mg/dL<br>2.6% at 2.33 mg/dL<br>Controls:<br>2.1% at 1.86 mg/dL<br>1.6% at 3.84 mg/dL | Human sera:<br>3.7% at 4.044 mg/dL<br>3.2% at 4.670 mg/L<br>Controls:<br>2.6% at 0.835 mg/dL<br>2.9% at 4.764 mg/dL |
| | ISE – Cl (Between-run)<br>Human sera:<br>0.44% at 97 mmol/L<br>0.48% at 128 mmol/L<br>Plasma (Between-run):<br>0.51% at 93 mmol/L<br>0.67% at 125 mmol/L | Human sera:<br>1.4% at 104.7 mmol/L<br>1.6% at 104.6 mmol/L<br>Controls:<br>SD 2.0 at 87.16 mmol/L<br>2.0% at 119.97 mmol/L |
| | ISE – K (Between –run)<br>Human Sera:<br>0.45% at 2.1 mmol/L<br>0.30% at 5.7 mmol/L<br>Plasma (Between-run)<br>0.78% at 2.0 mmol/L<br>0.72% at 6.0 mmol/L | Human sera:<br>1.4% at 4.465 mmol/L<br>1.6% at 4.540 mmol/L<br>Controls:<br>1.8% at 3.466 mmol/L<br>SD 1.8 at 6.646 mmol/L |
| | ISE – Na (Between-run)<br>Human Sera:<br>0.42% at 130 mmol/L<br>0.48% at 155 mmol/L<br>Plasma:<br>0.80% at 128 mmol/L<br>0.75% at 151 mmol/L | Human sera:<br>1.6% at 139.9 mmol/L<br>1.6% at 131.6 mmol/L<br>Controls:<br>1.8% at 126.48 mmol/L<br>1.7% at 151.0 mmol/L |
| Feature | Reagent performance<br>characteristics on cobas c<br>111 analyzer for professional<br>use | Reagent performance<br>characteristics for<br>point-of-care use |
| Limitations -<br>interferences | AST:<br>Bilirubin- no interference up to I<br>index of 67 (conjugated) and I<br>index of 65 (unconjugated)<br>Hemolysis: elevated results with<br>contamination of erythrocytes<br>Lipemia: interferences<br>Anticoagulants: citrate and<br>fluoride interfere<br>Glucose:<br>Bilirubin– no interference up to I<br>index of 67 (conjugated or<br>unconjugated)<br>Hemolysis: no interference up to H<br>index of 1046<br>Lipemia: no interference up to L<br>index of 2126<br>CRP:<br>Bilirubin– no interference up to I<br>index of 60 (conjugated or<br>unconjugated)<br>Hemolysis: no interference up to H<br>index of 700<br>Lipemia: no interference up to L<br>index of 700<br>Rheumatoid factors up to 1200<br>IU/mL do not interfere<br>No high dose hook-effect is seen up<br>to a CRP concentration of 3100<br>mg/L | Same |
| Feature | Reagent performance<br>characteristics on cobas c<br>111 analyzer for professional<br>use | Reagent performance<br>characteristics for<br>point-of-care use |
| Limitations -<br>interferences | ISE-Cl:<br>Serum/Plasma:<br>Hemolysis: Avoid hemolyzed<br>samples. No significant<br>interferences up to a<br>hemoglobin level of 10 g/L<br>Bilirubin/Lipemia: No<br>significant interferences | Same |
| | ISE-K:<br>Serum/Plasma:<br>Hemolysis: Avoid hemolyzed<br>samples. No significant<br>interferences up to a<br>hemoglobin level of 1 g/L.<br>Potassium centration in<br>erythrocytes is 25 times higher<br>than in normal plasma. The<br>level of interference may be<br>variable depending on the<br>excat content of erythrocytes.<br>Bilirubin/Lipemia: No<br>significant interferences | |
| | ISE-Na:<br>Serum/Plasma:<br>Hemolysis: Avoid hemolyzed<br>samples. No significant<br>interferences up to a<br>hemoglobin level of 10 g/L<br>Bilirubin/Lipemia: No<br>significant interferences | |
| Feature | Reagent performance<br>characteristics on cobas c<br>111 analyzer for professional<br>use | Reagent performance<br>characteristics for<br>point-of-care use |
| Endogenous<br>interferences | AST:<br>No interferences found using<br>common drug panel, except:<br>Doxycycline HCl causes artificially<br>low results<br>Glucose:<br>No interferences found using<br>common drug panel.<br>CRP:<br>No interferences found using<br>common drug panel.<br>ISE-Cl:<br>Serum/Plasma:<br>Panel of drugs tested causes no<br>significant interferences<br>Salicylic acid in concentration of 5<br>mmol/L causes artificially elevated<br>results<br>Urine:<br>Panel of drugs tested causes no<br>significant interferences, except:<br>Salicylic acid, Ca-dobesilate and Na-<br>cefoxitin which causes artificially<br>elevated Cl concentrations<br>ISE-K:<br>Serum/Plasma: | Same |
| | Panel of drugs tested causes no<br>significant interferences up to the<br>indicated concentration.<br>Urine:<br>Panel of drugs tested causes no<br>significant interferences up to the<br>indicated concentration. | |
| Feature | Reagent performance<br>characteristics on cobas c<br>111 analyzer for professional<br>use | Reagent performance<br>characteristics for<br>point-of-care use |
| Endogenous<br>interferences | ISE-Na:<br>Serum/Plasma:<br>Panel of drugs tested causes no<br>significant interferences up to<br>the indicated concentration.<br>Urine:<br>Panel of drugs tested causes no<br>significant interferences up to<br>the indicated concentration.<br>pH – Acidified urines can give<br>false results. | Same |
| Calibration<br>frequency | AST/Glucose/CRP:<br>Each lot and as required<br>following quality control<br>procedures<br><br>ISE-CI/K/Na:<br>24 hours (main calibration)<br>After ISE cleaning and<br>maintenance, changing reagent<br>bottles, replacing electrodes | Same |
| Feature | Reagent performance<br>characteristics on<br>cobas c 111 analyzer<br>for professional use | Reagent performance<br>characteristics for<br>point-of-care use |
| Method Comparison<br>Professional use<br>y = cobas c 111<br>x = Integra 400 | AST:<br>Passing Bablok:<br>y = 0.989x + 1.869 U/L<br>$\tau$ = 0.981<br>n = 82 | AST:<br>Passing Bablok:<br>y = 0.989x + 1.276 U/L<br>$\tau$ = 0.8316<br>n = 333 |
| Point-of-care use:<br>y = cobas c 111<br>x = Integra 400 | Glucose:<br>y = 1.02x - 0.009<br>mmol/L<br>$\tau$ = 0.983<br>n = 80 | Glucose:<br>y = 0.997x + 2.069 mg/dL<br>$\tau$ = 0.9217<br>n = 333 |
| | CRP:<br>y = 0.995x + 1.334 mg/L<br>$\tau$ = 0.970<br>n = 63 | CRP:<br>y = 1.058x + 0.022 mg/L<br>$\tau$ = 0.9789<br>n = 326 |
| | ISE-Cl:<br>y = 1.014x - 3.236<br>mmol/L<br>r = 0.982<br>n = 51 | ISE-Cl:<br>y = 1.011x - 0.51 mmol/L<br>$\tau$ = 0.7532<br>n = 280 |
| | ISE-K:<br>y = 0.984x - 0.003<br>mmol/L<br>r = 1.000<br>n = 51 | y = 0.943x + 0.189 mmol/L<br>$\tau$ = 0.8835<br>n = 280 |
| | ISE-Na:<br>y = 0.986x - 0.364<br>mmol/L<br>$\tau$ = 0.983<br>n = 51 | ISE-Na:<br>y = 1.064x - 9.818 mmol/L<br>$\tau$ = 0.6920<br>n = 280 |
| Feature | Reagent performance characteristics on cobas c 111 analyzer for professional use | Reagent performance characteristics for point-of-care use |
| Expected values<br>(from reference)<br>(also please reference method<br>sheet) | AST:<br>Female: up to 32 U/L<br>Males: up to 38 U/L<br><br>Glucose:<br>Plasma fasting: 3.88-6.38 mmol/L<br>Serum/Plasma Adults: 4.11-5.89 mmol/L<br><br>CRP:<br>Adults: less than 5 mg/L<br><br>ISE-C1:<br>Serum/Plasma (adults): 98-107 mmol/L<br>Urine (24h, adults): 110-250 mmol/L<br><br>ISE-K:<br>Serum (adults): 3.5-5.1 mmol/L<br>Plasma (adults): 3.4-4.5 mmol/L<br>Urine (24h, adults) 25-125 mmol/L<br><br>ISE-Na:<br>Serum (adults): 136-145 mmol/L<br>Plasma (adults): 136-145 mmol/L<br>Urine (24h, adults): 40-220 mmol/L | Same |
Device modifications – reagents comparison (continued)
{7}------------------------------------------------
Device modifications – reagents comparison (continued)
Continued on next page
.
{8}------------------------------------------------
.
Device modifications - reagents comparison (continued)
{9}------------------------------------------------
Device modifications - reagents comparison (continued)
Continued on next page
.
{10}------------------------------------------------
Device modifications – reagents comparison (continued)
{11}------------------------------------------------
#### Device modifications -- reagents comparison (continued)
{12}------------------------------------------------
Device modifications - reagents comparison (continued)
{13}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/13/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is positioned inside a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written around the circle's perimeter. The text is in all capital letters.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Roche Diagnostics Corporation c/o Corina Harper, Regulatory Affairs Consultant 9115 Hague Road Indianapolis, IN 46250
JUL 3 0 2007
Re: k071211 Trade/Device Name: cobas c 111 analyzer and applied reagents Regulation Number: 21 CFR 862.1100 Regulation Name: Aspartate amino transferase (AST/SGO) test system Regulatory Class: Class II Product Code: CIT, CFR, DCN, CEM, CGZ, JGS, JJE Dated: April 30, 2007 Received: May 1, 2007
Dear Ms. Harper:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{14}------------------------------------------------
Page 2 --
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device. or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Jean M. Cooper, M.S., D.V.M.
Yean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{15}------------------------------------------------
#### Indications for Use
510(k) Number (if known):
Device Name: cobas c 111 analyzer and applied reagents
Indications For Use:
cobas c 111 analyzer:
The Roche cobas c 111 analyzer is an in-vitro diagnostic analyzer capable of performing clinical chemistry, specific protein and electrolyte tests for professional settings and small laboratories, specialized testing and CLIA-licensed doctor's offices.
Analytes are measured photometrically or turbidimetrically; the analyzer also has an optional ISE module for measuring sodium, potassium and chloride.
#### Reagents:
Aspartate aminotransferase (ASTL/ASTPL) In vitro test for the quantitative determination of AST in human serum and plasma on the cobas c1 11 system. Aspartate amino transferase measurements are used in the diagnosis and treatment of certain types of liver and heart disease.
Prescription Use XXXX (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
CarolC. Benam
Sign-O
്ffice of In Vitro Diagnostic Device Evaluation and Safety
K071211
{16}------------------------------------------------
#### Indications for Use
510(k) Number (if known): _ K071211
Device Name: cobas c 111 analyzer and reagents
Indications For Use:
C-Reactive Protein Latex (CRPLX)
In vitro test for the quantitative immunological determination of human C-reactive protein in human serum and plasma on the cobas c111 system. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues.
Glucose HK (GLUC2)
In vitro test for the quantitative determination of glucose concentration in human serum and plasma on the cobas c111 system. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus and idiopathic hypoglycemia.
#### ISE Chloride Electrode
The chloride electrode for the cobas c111 system is intended for the quantitative determination of chloride in diluted serum, plasma, and urine. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
Prescription Use XXXX (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Carol C. Benson
on Sign-Off
Page 2 of 3
ice of In Vitro Diagnostic Device Juation and Safety
K071211
{17}------------------------------------------------
#### Indications for Use
510(k) Number (if known):
Device Name: cobas c 111 analyzer and reagents
Indications For Use:
ISE Potassium Electrode
The potassium electrode for the cobas c111 system is intended for the quantitative determination of potassium in diluted serum, plasma, and urine. Measurements of potassium are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.
ISE Sodium Electrode
The sodium electrode for the cobas c111 system is intended for the quantitative determination of sodium in diluted serum, plasma, and urine. Measurements of sodium are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insibidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.
Prescription Use XXXX (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Page 3 of 3
Sign-Off
be of In Vitro Diagnostic Device uation and Safety
4071211
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.