ELITECH CLINICAL SYSTEMS BILIRUBIN TOTAL 4+1, ELITECH CLINICAL SYSTEMS ELICAL 2, ELITECH CLINICAL SYSTMES ELITROL I & II
Applicant
Elitechgroup
Product Code
CIG · Clinical Chemistry
Decision Date
Oct 7, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1110
Device Class
Class 2
Intended Use
ELITech Clinical Systems BILIRUBIN TOTAL 4+1 is intended for the quantitative in vitro diagnostic determination of total bilirubin in human serum and plasma on ELITech Clinical Systems Selectra analyzers. It is not intended for use in Point of Care settings. Measurements of the levels of bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, are used in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder blockage. ELITech Clinical Systems BILIRUBIN DIRECT 4+1 is intended for the quantitative in vitro diagnostic determination of direct bilirubin in human serum and plasma on ELITech Clinical Systems Selectra analyzers. It is not intended for use in Point of Care settings. Measurements of the levels of bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, are used in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder blockage. ELITech Clinical Systems ELICAL 2 is a multi-parametric calibrator for in vitro diagnostic use in the calibration of quantitative ELITech Clinical Systems methods on ELITech Clinical Systems Selectra analyzers. ELITech Clinical Systems ELITROL I & ELITROL II are multi-parametric control sera for in vitro diagnostic use in the quality control of quantitative ELITech Clinical Systems methods on ELITech Clinical Systems Selectra analyzers.
Device Story
ELITech Clinical Systems Bilirubin Total 4+1 and Bilirubin Direct 4+1 are in vitro diagnostic reagent kits used on Selectra analyzers. The system utilizes a modified Malloy-Evelyn method (diazo colorimetry) to measure bilirubin levels in human serum or plasma. Reagents (R1 and R2) are mixed with the patient sample; the resulting colorimetric change is measured photometrically by the analyzer. The output is a quantitative bilirubin concentration (mg/dL). These results assist clinicians in diagnosing and monitoring liver, hemolytic, and metabolic conditions. The system includes ELICAL 2 (lyophilized human serum calibrator) and ELITROL I/II (lyophilized human serum quality controls) to ensure measurement accuracy and precision. The device is intended for professional use in clinical laboratory settings, not for point-of-care.
Clinical Evidence
Bench testing only. Performance validated via precision (Total CV < 4.7%), linearity (Total: 0.28-20.22 mg/dL; Direct: 0.18-6.82 mg/dL), and interference studies (CLSI EP07-A2). Method comparison against predicate (n=100) showed high correlation (Total R²=0.998; Direct R²=0.974). Matrix comparison confirmed suitability of lithium heparin plasma.
Technological Characteristics
Diazo colorimetry (Malloy-Evelyn modified method). Reagents are liquid, ready-to-use. Calibrator and controls are lyophilized human serum-based. Analyzers: Selectra ProM. Connectivity: Standalone clinical chemistry analyzer. Storage: 2-8°C, protected from light.
Indications for Use
Indicated for quantitative in vitro diagnostic determination of total and direct bilirubin in human serum and plasma to aid in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders (e.g., hepatitis, gall bladder blockage). For prescription use only; not for Point of Care settings.
Regulatory Classification
Identification
A bilirubin (total or direct) test system is a device intended to measure the levels of bilirubin (total or direct) in plasma or serum. Measurements of the levels of bilirubin, an organic compound formed during the normal and abnormal distruction of red blood cells, if used in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block.
Predicate Devices
ABX PENTRA BILIRUBIN, TOTAL CP (k060325)
ABX PENTRA BILIRUBIN, DIRECT CP (k060325)
Roche Diagnostics Calibrator for Automated Systems (C.f.a.s) (k033501)
Roche Diagnostics Precinorm U and Precipath U (k041227)
Related Devices
K171401 — ELITech Clinical Systems BILIRUBIN TOTAL 4+1, ELITech Clinical Systems BILIRUBIN DIRECT 4+1 · Elitechgroup · Jul 11, 2017
K040631 — VITALAB DIRECT BILIRUBIN REAGENT · Clinical Data, Inc. · Mar 18, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
ﻟﺴﻴﺮ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ
・
-
OCT -
7 2011
## SECTION 5 - 510(k) Summary
| Introduction | According to the requirements of 21 CFR 807.92, the following information<br>provides sufficient detail to understand the basis for a determination of<br>substantial equivalence. | |
|---------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The assigned 510(k) number is: K110780 | |
| Submitter<br>Address<br>Phone number<br>Fax number | SEPPIM S.A.S.<br>Zone Industrielle, 61500 SEES, FRANCE<br>+ 33 (0)2 33 81 21 00<br>+ 33 (0)2 33 28 77 51 | |
| Contact | Valérie GOURDON (Email: v.gourdon@elitechgroup.com)<br>Debra K. Hutson (email: d.hutson@elitechgroup.com) | |
| Date of Preparation | Sept 9, 2011 | |
| Device names | | |
| REAGENT:<br>Trade/proprietary Name:<br>Common or Usual Name:<br>Device Class<br>Classification name<br>Product code | ELITech Clinical Systems BILIRUBIN TOTAL 4+1<br>Bilirubin total, "BILIRUBIN TOTAL 4+1"<br>Class II<br>Bilirubin (Total or Direct) test System (Sec.862.1110)<br>CIG - Diazo Colorimetry, Bilirubin | |
| Predicate device | ABX PENTRA BILIRUBIN, TOTAL CP (K060325) | |
| Device description | The device for this submission is available as kit only. It consists of 2<br>reagents, "R1" and "R2".<br>Reagent R1 contains sulfanilic acid, Hydrochloric acid and cetrimide.<br>Reagent R2 contains sodium nitrite. | |
| Intended Use | ELITech Clinical Systems BILIRUBIN TOTAL 4+1 is intended for the<br>quantitative <i>in vitro</i> diagnostic determination of total bilirubin in human serum<br>and plasma on ELITech Clinical Systems Selectra analyzers.<br>It is not intended for use in Point of Care settings. | |
| Indication for use | Measurements of the levels of bilirubin, an organic compound formed during<br>the normal and abnormal destruction of red blood cells, are used in the<br>diagnosis and treatment of liver, hemolytic hematological, and metabolic<br>disorders, including hepatitis and gall bladder block | |
| | ELITech Clinical Systems Device<br>BILIRUBIN TOTAL 4+1 | Predicate device<br>(ABX PENTRA BILIRUBIN, TOTAL CP<br>K060325) |
| Intended use | Intended for the quantitative in vitro<br>diagnostic determination of total<br>bilirubin in human serum and<br>plasma on ELITech Clinical<br>Systems Selectra analyzers.<br>It is not intended for use in Point of<br>Care settings. | Intended for use with associated<br>calibrators and controls for quantitative<br>in vitro diagnostic determination of total<br>bilirubin in human serum and plasma<br>on ABX PENTRA 400 Clinical<br>Chemistry Analyzer. |
| Indication for Use | Measurements of the levels of<br>bilirubin, an organic compound<br>formed during the normal and<br>abnormal destruction of red blood<br>cells, are used in the diagnosis and<br>treatment of liver, hemolytic<br>hematological, and metabolic<br>disorders, including hepatitis and<br>gall bladder block. | Measurements of the levels of bilirubin,<br>an organic compound formed during<br>the normal and abnormal destruction of<br>red blood cells, are used in the<br>diagnosis and treatment of liver,<br>hemolytic hematological, and metabolic<br>disorders, including hepatitis and gall<br>bladder block. |
| Assay protocol | Malloy-Evelyn modified method | Photometric test using 2,4-<br>dichloroaniline (DCA) and a specific<br>mixture of detergents |
| Composition | Reagent 1 :<br>Sulfanilic acid 29 mmol/L<br>Hydrochloric acid 67 mmol/L<br>Cetrimide 37 mmol/L | Reagent 1 :<br>Phosphate buffer 50 mmol/L<br>NaCl 150 mmol/L<br>Detergents , Stabilizers |
| | Reagent 2 :<br>Sodium nitrite 5.8 mmol/L | Reagent 2 :<br>2,4-Dichlorophenyl -<br>diazonium salt 5 mmol/L<br>HCI 130 mmol/L<br>Detergent |
| Appearance of reagent | Liquid form, ready to use | Liquid form, ready to use |
| Sample type | Serum<br>Lithium heparin plasma | Serum<br>Plasma in lithium heparin |
| Reagent storage | Store at 2-8 °C and protect from<br>light. The reagent is stable until the<br>expiry date stated on the label. | Reagents, in unopened cassette, are<br>stable up to expiry date on the label if<br>stored at 2-8°C and contamination is<br>avoided. |
| Expected values | Adults: 0.3 - 1.2 mg/dL | Adults : 0.1 - 1.2 mg/dL |
| Instrument | Selectra ProM | ABX PENTRA 400 |
| Measuring range | 0.28 to 20.22 mg/dL | 0.2 to 26.3 mg/dL |
| Limit of detection (LoD) | 0.06 mg/dL | 0.09 mg/dL |
| Limit of quantification (LoQ) | 0.17 mg/dL | 0.14 mg/dL |
| Precision | Within run<br>Level 1.04 mg/dL CV=2.7%<br>Level 3.67 mg/dL CV=0.8%<br>Level 14.90 mg/dLCV=0.5% | Within run<br>Level 0.97 mg/dL CV=2.14%<br>Level 5.13 mg/dL CV=0.99%<br>Level 0.61 mg/dL CV=3.09%<br>Level 0.85 mg/dL CV=2.23%<br>Level 2.20 mg/dL CV=1.33% |
| | ELITech Clinical Systems Device<br>BILIRUBIN TOTAL 4+1 | Predicate device<br>(ABX PENTRA BILIRUBIN, TOTAL CP<br>K060325) |
| | Total<br>Level 1.04 mg/dL CV=4.0%<br>Level 3.67 mg/dL CV=2.0%<br>Level 14.90 mg/dL CV=1.8% | Total<br>Level 1.0 mg/dL CV=4.04%<br>Level 5.5 mg/dL CV=1.70%<br>Level 0.8 mg/dL CV=5.97%<br>Level 2.9 mg/dL CV=2.78%<br>Level 9.1 mg/dL CV=2.20% |
| Method comparison | y=0.924x + 0.02 mg/dL<br>$r^2$ = 0.998<br>range: 0.25 to 22.00 mg/dL | y=1.03x - 0.14 mg/dL<br>$r^2$ = 0.9965<br>range: 0.3 to 25.8 mg/dL |
| Limitations | Triglycerides: No significant<br>interference up to 2779 mg/dL<br>Hemoglobin:<br>No significant interference up to<br>500 mg/dL.<br>Acetaminophen: No significant<br>interference up to 30 mg/dL.<br>Ascorbic acid:<br>Concentrations greater than 2.0<br>mg/dL will interfere and cause<br>erroneous results.<br>Acetylsalicylic acid:<br>No significant interference up to<br>200 mg/dL | Hemoglobin: No significant influence<br>is observed up to 500 mg/dL.<br>Triglycerides: No significant influence<br>is observed up to 612.5 mg/dL (as<br>Intralipid®, representative of lipemia). |
| Calibration Frequency | 28 days | 10 days |
| On board stability | refrigerated area : 28 days | refrigerated area: 25 days |
| Calibrator | Recommended calibration material<br>(not included):<br>ELITech Clinical Systems ELICAL 2 | Recommended calibration material (not<br>included):<br>ABX Pentra Multical |
| Controls | Recommended quality control<br>material (not included):<br>ELITech Clinical Systems ELITROL<br>I (Normal control)<br>ELITech Clinical Systems ELITROL<br>II (Pathologic control) | Recommended quality control material<br>(not included):<br>ABX Pentra N Control<br>(Normal control)<br>ABX Pentra P Control (Pathologic<br>control) |
:: ..
{1}------------------------------------------------
{2}------------------------------------------------
{3}------------------------------------------------
| Device names | |
|---------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| REAGENT:<br>Trade/proprietary Name:<br>Common or Usual Name:<br>Device Class<br>Classification name<br>Product code | ELITech Clinical Systems BILIRUBIN DIRECT 4+1<br>Bilirubin direct, "BILIRUBIN DIRECT 4+1"<br>Class II<br>Bilirubin (Total or Direct) test System (Sec.862.1110)<br>CIG - Diazo Colorimetry, Bilirubin |
| Predicate device | ABX PENTRA BILIRUBIN, DIRECT CP (K060325) |
| Device description | The device for this submission is available as kit only. It consists of 2<br>reagents, "R1" and "R2".<br>Reagent R1 contains sulfanilic acid and Hydrochloric acid.<br>Reagent R2 contains sodium nitrite. |
| Intended Use | ELITech Clinical Systems BILIRUBIN DIRECT 4+1 is intended the<br>quantitative in vitro diagnostic determination of direct bilirubin in human<br>serum and plasma on ELITech Clinical Systems Selectra analyzers.<br>It is not intended for use in Point of Care settings. |
| Indication for use | Measurements of the levels of bilirubin, an organic compound formed during<br>the normal and abnormal destruction of red blood cells, are used in the<br>diagnosis and treatment of liver, hemolytic hematological, and metabolic<br>disorders, including hepatitis and gall bladder block |
| | ELITech Clinical Systems Device<br>BILIRUBIN DIRECT 4+1 | Predicate device<br>(ABX PENTRA BILIRUBIN, DIRECT CP<br>K060325) |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | Intended for the quantitative in<br>vitro diagnostic determination of<br>direct bilirubin in human serum and<br>plasma on ELITech Clinical<br>Systems Selectra analyzers.<br>It is not intended for use in Point of<br>Care settings. | Intended for use with associated calibrators<br>and controls for quantitative in vitro diagnostic<br>determination of direct bilirubin in human<br>serum and plasma on ABX PENTRA 400<br>Clinical Chemistry Analyzer. |
| Indication for Use | Measurements of the levels of<br>bilirubin, an organic compound<br>formed during the normal and<br>abnormal destruction of red blood<br>cells, are used in the diagnosis<br>and treatment of liver, hemolytic<br>hematological, and<br>metabolic<br>disorders, including hepatitis and<br>gall bladder block | Measurements of the levels of bilirubin, an<br>organic compound formed during the normal<br>and abnormal destruction of red blood cells,<br>are used in the diagnosis and treatment of<br>liver, hemolytic hematological, and metabolic<br>disorders, including hepatitis and gall bladder<br>block |
| Assay protocol | Malloy-Evelyn modified method | Photometric test using<br>2,4-dichloroaniline<br>(DCA) |
| Composition | Reagent 1:<br>Sulfanilic acid<br>$29$ mmol/L<br>Hydrochloric acid<br>$67$ mmol/L | Reagent 1:<br>Sulfamic acid<br>$100$ mmol/L<br>EDTA-Na2<br>$0.1$ mmol/L<br>NaCl<br>$150$ mmol/L |
| | ELITech Clinical Systems Device<br>BILIRUBIN DIRECT 4+1 | Predicate device<br>(ABX PENTRA BILIRUBIN, DIRECT CP<br>K060325) |
| | Reagent 2 :<br>Sodium nitrite 5.8 mmol/L | Reagent 2 :<br>2,4-Dichlorophenyl -<br>diazonium salt 0.5 mmol/L<br>HCI 900 mmol/L<br>EDTA-Na2 0.13 mmol/L |
| Appearance of reagent | Liquid form, ready to use | Liquid form, ready to use |
| Sample type | Serum<br>Lithium heparin plasma | Serum<br>Plasma in lithium heparin |
| Reagent storage | Store at 2-8 °C and protect from<br>light. The reagent is stable until the<br>expiry date stated on the label. | Reagents, in unopened cassette, are stable up<br>to expiry date on the label if stored at 2-8°C<br>and contamination is avoided. |
| Expected values | < 0.2 mg/dL | ≤ 0.2 mg/dL |
| Instrument | Selectra ProM | ABX PENTRA 400 |
| Measuring range | 0.18 to 6.82 mg/dL | 0.09 to 6.71 mg/dL |
| Limit of detection (LoD) | 0.06 mg/dL | 0.04 mg/dL |
| Limit of quantification (LoQ) | 0.17 mg/dL | 0.04 mg/dL |
| Precision | Within run<br>Level 0.49 mg/dL CV=2.0%<br>Level 1.89 mg/dL CV=0.6%<br>Level 4.96 mg/dL CV=0.5% | Within run<br>Level 0.90 mg/dL CV=0.67%<br>Level 1.85 mg/dL CV=0.44%<br>Level 0.23 mg/dL CV=3.23%<br>Level 1.52 mg/dL CV=0.59%<br>Level 7.88 mg/dL CV=2.69% |
| | Total<br>Level 0.49 mg/dL CV=4.7%<br>Level 1.89 mg/dL CV=3.3%<br>Level 4.96 mg/dL CV=3.2% | Total<br>Level 0.94 mg/dL CV=4.26%<br>Level 2.02 mg/dL CV=4.22%<br>Level 0.69 mg/dL CV=3.27%<br>Level 3.83 mg/dL CV=2.98% |
| Method comparison | y=0.988x + 0.07 mg/dL<br>r²= 0.974<br>range: 0.10 to 6.23 mg/dL | y=1.06x + 0.04 mg/dL<br>r²= 0.9928<br>range: 0.09 to 6.71 mg/dL |
| Limitations | Triglycerides<br>No significant interference up to<br>2106 mg/dL<br>Hemoglobin:<br>No significant interference up to<br>125 mg/dL .<br>Acetaminophen: No significant<br>interference up to 30 mg/dL.<br>Ascorbic acid:<br>Concentrations greater than 0.3<br>mg/dL will interfere and cause<br>erroneous results. | Hemoglobin: do not use hemolyzed samples<br>Triglycerides: No significant influence is<br>observed up to 612.5 mg/dL (As Intralipid®,<br>representative of lipemia) |
| | ELITech Clinical Systems Device<br>BILIRUBIN DIRECT 4+1 | Predicate device<br>(ABX PENTRA BILIRUBIN, DIRECT CP<br>K060325) |
| | No significant interference up to<br>200 mg/dL | |
| Calibration Frequency | 28 days | 10 days |
| On board stability | refrigerated area : 28 days | refrigerated area: 30 days |
| Calibrator | Recommended calibration material<br>(not included):<br>ELITech Clinical Systems ELICAL<br>2 | Recommended calibration material (not<br>included):<br>ABX Pentra Multical |
| Controls | Recommended quality control<br>material (not included):<br>ELITech Clinical Systems<br>ELITROL I (Normal control)<br>ELITech Clinical Systems<br>ELITROL II (Pathologic control) | Recommended quality control material (not<br>included):<br>ABX Pentra N Control<br>(Normal control)<br>ABX Pentra P Control (Pathologic control) |
{4}------------------------------------------------
{5}------------------------------------------------
#### Device Names
| CALIBRATOR: | |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade/proprietary Name: | ELITech Clinical Systems ELICAL 2 |
| Common or Usual Name: | Calibrator, multi-analyte mixture, "ELICAL 2" |
| Device Class | Class II |
| Classification name | Calibrator (21 CFR 862.1150) |
| Product code | JIX- Calibrator, multi-analyte mixture |
| Predicate device | Roche Diagnostics Calibrator for Automated Systems (C.f.a.s) (K033501) |
| Device description | ELITech Clinical Systems ELICAL 2 is a lyophilized calibrator based on human serum containing constituents to ensure optimal calibration.<br>ELICAL 2 is prepared exclusively from the blood of donors tested individually and found to be negative for HbsAg and to antibodies to HCV and HIV according to FDA-approved methods or methods in compliance with the European Directive 98/79/EC, Annex II, List A. |
| Intended Use | ELITech Clinical Systems ELICAL 2 is a multi-parametric calibrator for <i>in vitro</i> diagnostic use in the calibration of quantitative ELITech Clinical Systems methods on ELITech Clinical Systems Selectra analyzers. |
{6}------------------------------------------------
| | ELITech Clinical Systems Device<br>(ELICAL 2) | Predicate device<br>(Roche Calibrator f.a.s. K033501) |
|--------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | ELITech Clinical Systems ELICAL 2<br>is a multi-parametric calibrator for <i>in vitro</i> diagnostic use in the calibration<br>of quantitative ELITech Clinical<br>Systems methods on the ELITech<br>Clinical Systems Selectra analyzers. | For <i>in vitro</i> diagnostic use in the<br>calibration of quantitative Roche<br>methods on Roche clinical chemistry<br>analysers as specified in the value<br>sheets. |
| Format | Lyophilized calibrator based on human<br>serum with constituents added as<br>required to obtain desired components<br>levels | Lyophilized calibrator based on human<br>serum with constituents added as<br>required to obtain desired components<br>levels |
| Level | Single level | Single level |
| Handling | Carefully open the vial, avoiding the<br>loss of lyophilizate, and pipette in<br>exactly 3 mL of distilled/deionized<br>water. Carefully close the vial and<br>dissolve the contents completely by<br>occasional gentle swirling within 30<br>minutes avoiding the formation of<br>foam. | Carefully open one bottle, avoiding the<br>loss of lyophilizate, and pipette in<br>exactly 3 mL of distilled/deionized<br>water. Carefully close the bottle and<br>dissolve the contents completely by<br>occasional gentle swirling within 30<br>minutes. Avoid the formation of foam. |
| Traceability | Traceability information is given in the<br>value sheet included in the box. | Traceability of the target value is given<br>in the respective instruction for use of<br>the system reagents. |
| Stability | Lyophilized:<br>To store at 2-8℃ and protected from<br>light until the expiry date<br><br>After reconstitution, the stabilities are :<br>- Stability of direct bilirubin (when<br>stored protected from light): | Lyophilized:<br>Stable at 2-8℃ up to expiration date.<br><br>After reconstitution, the stabilities* are :<br>- Stability of direct bilirubin (when<br>stored protected from light): |
| | Between 15-25 ℃:<br>3 hours | Between 15-25 ℃:<br>3 hours |
| | Between 2-8 ℃:<br>8 hours<br>Between -25 and -15 ℃: 2 weeks<br>(when frozen once) | Between 2-8 ℃<br>8 hours<br>Between (-25)-(-15) ℃: 2 weeks (when<br>frozen once) |
| | - Stability of total bilirubin (when stored<br>protected from light): | - Stability of total bilirubin (when stored<br>protected from light): |
| | Between 15-25 ℃:<br>6 hours | Between 15-25 ℃:<br>6 hours |
| | Between 2-8 ℃:<br>1 day<br>Between -25 and -15 ℃: 2 weeks<br>(when frozen once) | Between 2-8 ℃:<br>1 day<br>Between (-25)-(-15) ℃: 2 weeks<br>(when frozen once) |
{7}------------------------------------------------
#### Device names
·
| CONTROLS:<br>Trade/proprietary Name:<br>Common or Usual Name:<br>Device Class<br>Classification name<br>Product code | ELITech Clinical Systems ELITROL I and ELITROL II<br>Multi-analyte controls - all kinds, "ELITROL I"- "ELITROL II"<br>Class I<br>Quality control material (assayed and unassayed).<br>(21 CFR 862.1660)<br>JJY- Multi-analyte controls – all kinds |
|----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate device | Roche Diagnostics Precinorm U (K041227)<br>Roche Diagnostics Precipath U (K041227) |
| Device description | ELITech Clinical Systems ELITROL I and ELITROL II are two level quality<br>control products consisting of lyophilized human serum containing constituents<br>at desired levels.<br>Elitrol I and Elitrol II are prepared exclusively from the blood of donors tested<br>individually and found to be negative for HbsAg and to antibodies to HCV and<br>HIV according to FDA-approved methods or methods in compliance with the<br>European Directive 98/79/EC, Annex II, List A. |
| Intended Use | ELITech Clinical Systems ELITROL I & ELITROL II are multi parametric |
ELITech Clinical Systems ELITROL I & ELITROL II are multi-parametric intended Use control sera for in vitro diagnostic use in quality control of ELITech Clinical Systems methods on ELITech Clinical Systems Selectra analyzers.
| Intended use | ELITech Clinical Systems Device<br>ELITROL I/ELITROL II | Predicate Device<br>Roche Precinorm U/ Precipath U<br>K041227 |
|--------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Format | Lyophilized human sera with constituents<br>added as required to obtain desired<br>components levels | Lyophilized human sera with<br>constituents added as required to<br>obtain desired components levels |
| Levels | Two levels | Two levels |
| Handling | Carefully open the vial, avoiding the loss<br>of lyophilizate, and pipette in exactly 5<br>mL of distilled/deionized water. Carefully<br>close the vial and dissolve the contents<br>completely by occasional gentle swirling<br>within 30 minutes avoiding the formation<br>of foam. | For in vitro diagnostic use in quality<br>control by monitoring accuracy and<br>precision for the quantitative methods<br>as specified in the value sheet. |
{8}------------------------------------------------
| Stability | Lyophilized:<br>To store at 2-8°C and protected from<br>light until the expiry date<br>After reconstitution, the stabilities are : | Lyophilized:<br>Stable at 2-8°C up to expiration date.<br>- Stability of direct bilirubin (when<br>stored protected from light): |
|-----------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | - Stability of direct bilirubin (when stored<br>protected from light):<br>Between 15-25 °C: 4 hours<br>Between 2-8 °C: 8 hours<br>Between -25 and -15 °C: 2 weeks<br>(when frozen once) | Between 15-25 °C: 4 hours<br>Between 2-8 °C: 8 hours<br>Between (-25)-(-15) °C: 2 weeks<br>(when frozen once)<br>- Stability of total bilirubin (when stored<br>protected from light): |
| | - Stability of total bilirubin (when stored<br>protected from light):<br>Between 15-25 °C: 8 hours<br>Between 2-8 °C: 24 hours<br>Between -25 and -15 °C: 2 weeks<br>(when frozen once) | Between 15-25 °C: 8 hours<br>Between 2-8 °C: 24 hours<br>Between (-25)-(-15) °C: 2 weeks<br>(when frozen once) |
#### Conclusion
The performance data and other information demonstrate that the safety and effectiveness of these devices versus the predicate devices are not compromised, and that it met all acceptance criteria, demonstrating that the device is substantially equivalent to its respective predicate device.
.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a circular design with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract emblem resembling an eagle or bird in flight, composed of several curved lines. The overall design is simple and monochromatic.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
ELITech Group c/o Debra Hutson 21720 23td Dr. SE. Suite 150 Bothell. Washington 98021
Food and Druq Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
Re: k110780 Trade Name: ELITech Clinical Systems Bilirubin Total 4+1, ELITech Clinical Systems Bilirubin Direct 4+1. ELITech Clinical Systems ELICAL 2, ELITech Clinical Systems ELITROL. I and ELITROL H Regulation Number: 21 CFR §862.1110 Regulation Name: Bilirubin Test System Regulatory Class: Class H Product Codes: CIG. JIX and JJY Dated: September 12, 2011 Received: September 14, 2011
OCT - 7 2011
Dear Ms. Hutson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21. Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{10}------------------------------------------------
Page 2 -
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesSorYou/Industry/default.htm.
Sincerely yours,
signature
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
510(k) Number (if known): K110780
Device Name: _________________________________________________________________________________________________________________________________________________________________ ELITech Clinical Systems BILIRUBIN TOTAL 4+1
Indications for Use:
ELITech Clinical Systems BILIRUBIN TOTAL 4+1 is intended for the quantitative in vitro diagnostic determination of total bilirubin in human serum and plasma on ELITech Clinical Systems Selectra analyzers.
It is not intended for use in Point of Care settings.
Measurements of the levels of bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, are used in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder blockage.
Prescription Use x (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Rate Chen
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) 110780
{12}------------------------------------------------
510(k) Number (if known):
ELITech Clinical Systems BILIRUBIN DIRECT 4+1 Device Name:
Indications for Use:
ELITech Clinical Systems BILIRUBIN DIRECT 4+1 is intended for the quantitative in vitro diagnostic determination of direct bilirubin in human serum and plasma on ELITech Clinical Systems Selectra analyzers.
It is not intended for use in Point of Care settings.
Measurements of the levels of bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, are used in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder blockage.
Prescription Use x (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OVD)
Ruta Cime
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
11 2780 510(k)
Page Zof 4
{13}------------------------------------------------
510(k) Number (if known): _ _ K | 10780
ELITech Clinical Systems ELICAL 2 Device Name:
Indications for Use:
ELITech Clinical Systems ELICAL 2 is a multi-parametric calibrator for in vitro diagnostic use in the calibration of quantitative ELITech Clinical Systems methods on ELITech Clinical Systems Selectra analyzers.
Prescription Use × AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Rutul Chosler
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
いいつでない 510(k)
{14}------------------------------------------------
510(k) Number (if known): _ K 110780
ELITech Clinical Systems ELITROL I & ELITROL II Device Name:
Indications for Use:
ELITech Clinical Systems ELITROL I & ELITROL II are multi-parametric control sera for in vitro diagnostic use in the quality control of quantitative ELITech Clinical Systems methods on ELITech Clinical Systems Selectra analyzers.
Prescription Use × (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
AND/OR
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Rute chere
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
11 07 8 0 510(k)
Page of of 4
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.