CREATININE

K083809 · Abbott Laboratories · CGX · Apr 7, 2009 · Clinical Chemistry

Device Facts

Record IDK083809
Device NameCREATININE
ApplicantAbbott Laboratories
Product CodeCGX · Clinical Chemistry
Decision DateApr 7, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1225
Device ClassClass 2

Indications for Use

The Creatinine assay is used for the quantitation of creatinine in human serum, plasma, or urine.

Device Story

Creatinine assay; in vitro diagnostic test for serum, plasma, or urine; performed on Abbott ARCHITECT c Systems. Modification involves calibration update for urine application; transition from Use-factor Calibration to independent Linear Calibration using NIST SRM 914 reference material and urine-specific calibrator values. Device measures creatinine levels to assist clinicians in diagnosing and managing renal disease. Operates in clinical laboratory settings; processed by laboratory technicians. Output provides quantitative creatinine concentration; informs clinical assessment of kidney function.

Clinical Evidence

Bench testing only. Verification and validation activities performed per design control procedures to confirm performance following calibration modification. Results met predetermined acceptance criteria.

Technological Characteristics

In vitro diagnostic reagent; modified Jaffe (alkaline picrate) methodology. Measures absorbance at 500 nm. Calibrated using NIST SRM 914. Designed for use on automated clinical chemistry analyzers (AEROSET, ARCHITECT cSystems).

Indications for Use

Indicated for the quantitative determination of creatinine in human serum, plasma, or urine to aid in the diagnosis and treatment of renal diseases, in patients requiring monitoring of renal function.

Regulatory Classification

Identification

A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} SPECIAL 510(k): Device Modification ODE Review Memorandum (Decision Making Document is Attached) To: THE FILE RE: DOCUMENT NUMBER k083809 This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary): 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.) Creatinine, k061193 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. This change was for modification to the calibration for the urine application of the Creatinine assay. Abbott has implemented NIST SRM 914 as the reference material for standardization of the urine application of the Creatinine assay, which uses the Multiconstituent Calibrator. The urine assay parameters have changed from the Use-factor Calibration to an independent Linear Calibration, utilizing urine application-specific calibrator values. 4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics and performance characteristics. 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied c) A declaration of conformity with design controls. The declaration of conformity should include: i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review. 6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices). The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device. Comments revised:8/1/03
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...