CREATININE

K061193 · Abbott Laboratories, Inc. · CGX · Jun 23, 2006 · Clinical Chemistry

Device Facts

Record IDK061193
Device NameCREATININE
ApplicantAbbott Laboratories, Inc.
Product CodeCGX · Clinical Chemistry
Decision DateJun 23, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1225
Device ClassClass 2

Intended Use

A creatinine test system is a device intended to measure creatinine levels in serum. plasma, and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes

Device Story

Liquid, ready-to-use, two-reagent kit (Sodium Hydroxide, Picric Acid) for quantitative measurement of creatinine in human serum, plasma, or urine. Used on Abbott Aeroset and Architect c8000 Systems. Principle: alkaline pH reaction of creatinine with picrate to form creatinine-picrate complex; rate of absorbance increase at 500 nm proportional to creatinine concentration. Operated by laboratory professionals in clinical settings. Output provides quantitative creatinine levels to assist clinicians in diagnosing/treating renal disease and monitoring dialysis.

Clinical Evidence

Bench testing only. Performance established via precision studies (CLSI EP5-A, EP10-A), linearity, limit of detection/quantitation (CLSI EP17-A), and interference testing. Method comparison against predicate showed high correlation (r=0.995-0.999) for serum and urine across both Aeroset and Architect systems.

Technological Characteristics

Quantitative colorimetric assay; Modified Jaffe method. Reagents: Sodium Hydroxide (0.8 mol/L) and Picric Acid (25.0 mmol/L). Energy source: photometric detection at 500 nm. Form factor: liquid reagent kit for automated clinical chemistry analyzers. Connectivity: integrated with Abbott Aeroset and Architect c8000 systems.

Indications for Use

Indicated for patients requiring creatinine measurement in serum, plasma, or urine for diagnosis and treatment of renal diseases, monitoring renal dialysis, and as a calculation basis for other urine analytes. For professional use.

Regulatory Classification

Identification

A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k061193 B. Purpose for Submission: New device C. Measurand: Creatinine D. Type of Test: Quantitative, Colorimetric (absorbance change at 500 nm) E. Applicant: Abbott Laboratories F. Proprietary and Established Names: Abbott Laboratories Clinical Chemistry Creatinine G. Regulatory Information: 1. Regulation section: 21 CFR § 862.1225 2. Classification: Class II 3. Product code: CGX 4. Panel: 75 (Clinical Chemistry) {1} H. Intended Use: 1. Intended use(s): A creatinine test system is a device intended to measure creatinine levels in serum, plasma and urine. 2. Indication(s) for use: A creatinine test system is a device intended to measure creatinine levels in serum, plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. 3. Special conditions for use statement(s): For professional use. 4. Special instrument requirements: Abbott Laboratories Aeroset and Architect c8000 Systems I. Device Description: The Aeroset and Architect c8000 creatinine assay is a liquid, ready-to-use, two-reagent kit. R1 contains Sodium Hydroxide at a concentration of 0.8 mol/L and R2 contains Picric Acid at a concentration of 25.0 mmol/L. The same reagent kit may be used on either the Aeroset or Architect c8000 Systems. J. Substantial Equivalence Information: 1. Predicate device name(s): Roche creatinine assay 2. Predicate 510(k) number(s): k941837 3. Comparison with predicate: {2} | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Assay | Same | Automated Quantitative Measurement of Creatinine | | Methodology | Same | Modified Jaffe | | Matrices | Same | Serum, Urine, Plasma | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Reportable Range | Serum: 0.20 – 37.00 mg/dL | Serum: 0.20 – 25.00 mg/dL | | | Plasma: 0.20 – 37.00 mg/dL | Plasma: 0.20 – 25.00 mg/dL | | | Urine: 5.0 – 740.0 mg/dL | Urine: 0.2 – 650 mg/dL | # K. Standard/Guidance Document Referenced (if applicable): CLSI (formerly NCCLS) Document EP9-A2: Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline – Second Edition CLSI (formerly NCCLS) Document EP5-A: Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline CLSI (formerly NCCLS) Document EP17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline CLSI (formerly NCCLS) Document EP10-A: Preliminary Evaluation of Quantitative Clinical Laboratory Methods; Approved Guideline # L. Test Principle: Abbott Clinical Chemistry Creatinine is an in vitro diagnostic assay for the quantitation of creatinine in human serum, plasma, or urine. At an alkaline pH, creatinine in the sample reacts with picrate to form a creatinine-picrate complex. The rate of increase in absorbance at 500 nm due to the formation of this complex is directly proportional to the concentration of creatinine in the sample. # M. Performance Characteristics (if/when applicable): # 1. Analytical performance: a. Precision/Reproducibility: Serum Application: The total precision as well as the precision for each component of variation (between-day, between-run, and within-run) was estimated for creatinine. Two control levels (Level 1 and Level 2) at normal and abnormal analyte concentrations were tested. These controls were evaluated over 20 days, two runs per day, and two replicates per run. Precision {3} was reported as the total percent CV. This study was conducted in accordance with CLSI Document EP5-A. Summary results were as follows: ## AEROSET Serum Precision | Control | | Level 1 | Level 2 | | --- | --- | --- | --- | | N | | 80 | 80 | | Mean (mg/dL) | | 1.20 | 4.66 | | Within Run | SD | 0.01 | 0.03 | | | %CV | 0.76 | 0.57 | | Between Run | SD | 0.01 | 0.04 | | | %CV | 0.81 | 0.80 | | Between Day | SD | 0.06 | 0.14 | | | %CV | 4.83 | 3.02 | | Total | SD | 0.06 | 0.15 | | | %CV | 4.95 | 3.18 | ## ARCHITECT Serum Precision | Control | | Level 1 | Level 2 | | --- | --- | --- | --- | | N | | 80 | 80 | | Mean (mg/dL) | | 1.27 | 4.81 | | Within Run | SD | 0.02 | 0.02 | | | %CV | 1.74 | 0.44 | | Between Run | SD | 0.01 | 0.04 | | | %CV | 1.01 | 0.92 | | Between Day | SD | 0.03 | 0.06 | | | %CV | 2.36 | 1.15 | | Total | SD | 0.04 | 0.07 | | | %CV | 3.10 | 1.54 | Urine Application: Five day precision studies were conducted on the AEROSET and ARCHITECT c8000 Systems in accordance with CLSI Document EP10-A. This study was intended to supplement data obtained from the twenty-day serum precision study and provides a limited assessment of the performance of the assay with the urine matrix. Two control levels (Level 1 and Level 2) at normal and abnormal analyte concentrations were tested for the urine application. These controls were evaluated over five days, two runs per day, and five replicates per run. Precision was reported as the total percent CV. {4} 5 # AEROSET Urine Precision | Control | | Level 1 | Level 2 | | --- | --- | --- | --- | | N | | 50 | 50 | | Mean (mg/dL) | | 42.28 | 92.75 | | Within Run | SD | 0.96 | 1.80 | | | %CV | 2.26 | 1.94 | | Between Run | SD | 0.14 | 1.01 | | | %CV | 0.32 | 1.09 | | Between Day | SD | 0.32 | 0.85 | | | %CV | 0.77 | 0.92 | | Total | SD | 1.02 | 2.23 | | | %CV | 2.41 | 2.41 | # ARCHITECT Urine Precision | Control | | Level 1 | Level 2 | | --- | --- | --- | --- | | N | | 50 | 50 | | Mean (mg/dL) | | 42.59 | 93.91 | | Within Run | SD | 0.37 | 0.73 | | | %CV | 0.87 | 0.78 | | Between Run | SD | 0.11 | 0.56 | | | %CV | 0.25 | 0.60 | | Between Day | SD | 0.12 | 0.14 | | | %CV | 0.28 | 0.15 | | Total | SD | 0.40 | 0.93 | | | %CV | 0.94 | 0.99 | ## b. Linearity/assay reportable range: The sponsor’s reportable range for serum and plasma is 0.20 – 37.00 mg/dL and 5.0 – 740.0 mg/dL for urine. Linearity of the assay across the measuring range was established by measuring serum and urine samples at concentrations spanning the measuring range of the assay in four replicates. At least one level was included which exceeded the desired linear range. To measure the highest concentration sample, the analyzer’s “linear high” and “>” error codes were suppressed. The {5} percent recovery for each sample was determined by dividing the mean observed result by the expected value. The sponsor also provided data demonstrating equivalence between the analyzer's auto-dilution feature and manual dilutions. c. Traceability, Stability, Expected values (controls, calibrators, or methods): This assay utilizes the Abbott Multiconstituent Calibrator which was cleared under k981706. Shelf life, calibration, and on-board stability study protocols and acceptance criteria were described and found to be acceptable. d. Detection limit: The functional sensitivity of the Creatinine assay was determined based on the Limit of Quantitation (LOQ). To determine the LOQ, test levels near the linear low for the Creatinine assay were run in replicates of 10, on three instruments, two runs per instrument. The limit of quantitation was defined as the lowest concentration of analyte which has imprecision less than or equal to 20% CV. An internal verification study supported an LOQ of 0.10 mg/dL for the serum application and 2.0 mg/dL for the urine application. The (Limit of Detection (LOD) testing for Creatinine was performed using a study design based on CLSI EP17-A. An internal verification study supported an LOD of 0.05 mg/dL for the serum application, and 1.0 mg/dL for the urine application. The proportions of false positives (α) and false negatives (β) were less than 5% and the limit of blank (LOB) was 0.0224 mg/dL for serum and 0.3732 mg/dL for urine. Conclusions: The LOQ for Creatinine serum is 0.10 mg/dL and the LOD is 0.05 mg/dL. The LOQ for Creatinine urine is 2.0 mg/dL and the LOD is 1.0 mg/dL. Limit of Detection (Serum) | | Instrument | N | Limit of Blank (mg/dL) | SD | Limit of Detection (mg/dL) | | --- | --- | --- | --- | --- | --- | | AEROSET | 1 | 20 | 0.0153 | 0.0064 | 0.0258 | | | 2 | 20 | 0.0197 | 0.0100 | 0.0362 | | | 3 | 20 | 0.0081 | 0.0067 | 0.0191 | | ARCHITE | 1 | 20 | 0.0108 | 0.0071 | 0.0224 | | | 2 | 20 | 0.0224 | 0.0046 | 0.0300 | | | 3 | 20 | 0.0224 | 0.0051 | 0.0307 | {6} 7 Limit of Quantitation (Serum) | | Instrument | N | Mean (mg/dL) | SD | % CV | | --- | --- | --- | --- | --- | --- | | AEROSET | 1 | 20 | 0.0721 | 0.0093 | 12.8771 | | | 2 | 20 | 0.0713 | 0.0127 | 17.8165 | | | 3 | 20 | 0.0684 | 0.0073 | 10.6171 | | ARCHITECT | 1 | 20 | 0.0368 | 0.0072 | 19.6566 | | | 2 | 20 | 0.0457 | 0.0058 | 12.7092 | | | 3 | 20 | 0.0506 | 0.0057 | 11.2252 | Limit of Detection (Urine) | | Instrument | N | Limit of Blank (mg/dL) | SD | Limit of Detection | | --- | --- | --- | --- | --- | --- | | AEROSET | 1 | 20 | 0.2557 | 0.1770 | 0.5508 | | | 2 | 20 | 0.3732 | 0.1673 | 0.6484 | | | 3 | 20 | 0.2833 | 0.1301 | 0.4973 | | ARCHITECT | 1 | 20 | 0.3024 | 0.2009 | 0.6329 | | | 2 | 20 | 0.2856 | 0.2204 | 0.6482 | | | 3 | 20 | 0.2509 | 0.2466 | 0.6565 | Limit of Quantitation (Urine) | | Instrument | N | Mean (mg/dL) | SD | % CV | | --- | --- | --- | --- | --- | --- | | AEROSET | 1 | 20 | 1.8140 | 0.1262 | 6.9579 | | | 2 | 20 | 1.7137 | 0.1741 | 10.1572 | | | 3 | 20 | 0.7340 | 0.1413 | 19.2540 | | ARCHITECT | 1 | 20 | 1.8112 | 0.1465 | 8.0874 | | | 2 | 20 | 2.0207 | 0.2300 | 11.3821 | | | 3 | 20 | 1.8596 | 0.2018 | 10.8516 | {7} e. Analytical specificity: Interferents, which may falsely elevate or reduce the concentration of an analyte, were tested. Human serum samples at the medical decision level of the analyte and urine samples were spiked with various levels of interferents. Four replicates of each interferent level and four replicates of an unspiked reference sample were run. The percent recovery was determined by dividing the mean result of each interferent sample by the mean result of the reference sample. Testing was performed using the AEROSET System only. The percent interference was within ± 10% difference for serum samples containing 30 mg/dL bilirubin; 2,000 mg/dL hemoglobin; 1,000 mg/dL Intralipid; 3.0 mg/dL ascorbate; 300 mg/dL glucose; and 10.6 g/dL protein at Medical Decision Level 1. The percent interference was within ± 10% difference for serum samples containing 30 mg/dL bilirubin; 2,000 mg/dL hemoglobin; 1,000 mg/dL Intralipid; 3.0 mg/dL ascorbate; 600 mg/dL glucose; and 14.7 g/dL protein at Medical Decision Level 2. Interfering Substances for Serum - Level 1 | Interfering Substance | Interfering Substance Concentration | Target (mg/dL) | Observed (% of Target) | | --- | --- | --- | --- | | Bilirubin | 30 mg/dL | 1.549 | 98.11 | | | 60 mg/dL | 1.549 | 71.87 | | Hemoglobin | 1,000 mg/dL | 1.403 | 104.79 | | | 2,000 mg/dL | 1.403 | 108.86 | | Intralipid | 750 mg/dL | 1.425 | 99.44 | | | 1,000 mg/dL | 1.425 | 98.39 | | Ascorbate | 1.5 mg/dL | 1.522 | 99.49 | | | 3.0 mg/dL | 1.522 | 98.83 | | Glucose | 300 mg/dL | 1.522 | 107.08 | | | 600 mg/dL | 1.522 | 115.92 | | Protein | 10.6 g/dL | 1.537 | 107.97 | | | 14.3 g/dL | 1.537 | 115.37 | {8} Interfering Substances for Serum - Level 2 | Interfering Substance | Interfering Substance Concentration | Target (mg/dL) | Observed (% of Target) | | --- | --- | --- | --- | | Bilirubin | 30 mg/dL | 5.334 | 94.91 | | | 60 mg/dL | 5.334 | 74.71 | | Hemoglobin | 1,000 mg/dL | 4.701 | 102.07 | | | 2,000 mg/dL | 4.701 | 103.29 | | Intralipid | 750 mg/dL | 4.619 | 98.99 | | | 1,000 mg/dL | 4.619 | 99.10 | | Ascorbate | 1.5 mg/dL | 5.230 | 99.61 | | | 3.0 mg/dL | 5.230 | 99.71 | | Glucose | 300 mg/dL | 4.999 | 101.46 | | | 600 mg/dL | 4.999 | 103.16 | | Protein | 10.8 g/dL | 5.571 | 99.13 | | | 14.7 g/dL | 5.571 | 99.08 | The percent difference was within $\pm 10\%$ for urine samples containing 6.25 mL/dL acetic acid (8.5 N), 200 mg/dL ascorbate, 250 mg/dL boric acid, 1,000 mg/dL glucose, 2.5 mL/dL hydrochloric acid (6 N), 5.0 mL/dL nitric acid (6 N), 50 mg/dL protein, 1.25 g/dL sodium carbonate, 400 mg/dL sodium fluoride, and 60 mg/dL sodium oxalate. Interfering Substances for Urine | Interfering Substance | Interfering Substance Concentration | Target (mg/dL) | Observed (% of Target) | | --- | --- | --- | --- | | Acetic Acid (8.5 N) | 6.25 ml/dL | 95.543 | 100.10 | | Ascorbate | 200 mg/dL | 90.533 | 99.75 | | Boric Acid | 250 ml/dL | 95.035 | 100.14 | | Glucose | 1,000 mg/dL | 94.128 | 100.39 | | Hydrochloric Acid (6 N) | 2.5 ml/dL | 95.043 | 100.75 | | Nitric Acid (6 N) | 5.0 ml/dL | 95.338 | 99.82 | | Protein | 50 mg/dL | 95.470 | 101.93 | | Sodium Carbonate | 1.25 g/dL | 95.805 | 99.96 | | Sodium Fluoride | 400 mg/dL | 96.150 | 99.78 | | Sodium Oxalate | 60 mg/dL | 95.678 | 100.50 | {9} f. Assay cut-off: Not Applicable. 2. Comparison studies: a. Method comparison with predicate device: AEROSET: The sponsor compared 117 serum samples ranging from 0.69 to 38.36 mg/dL (as measured by the predicate method). Samples with concentrations greater than 25 mg/dL were diluted X2 on the predicate method but run undiluted on the Aeroset. Linear regression produced a slope of 0.98, a y-intercept of -0.18, and a correlation coefficient of 0.995. The sponsor also compared 50 urine samples ranging from 7.14 to 401.72 mg/dL (as measured by the predicate method). Linear regression produced a slope of 0.94, a y-intercept of -3.66, and a correlation coefficient of 0.999. ARCHITECT: The sponsor compared 117 serum samples ranging from 0.69 to 38.36 mg/dL (as measured by the predicate method). Samples with concentrations greater than 25 mg/dL were diluted X2 on the predicate method but run undiluted on the Architect. Linear regression produced a slope of 0.96, a y-intercept of -0.24, and a correlation coefficient of 0.999 The sponsor also compared 50 urine samples ranging from 7.14 to 401.72 mg/dL (as measured by the predicate method). Linear regression produced a slope of 0.93, a y-intercept of -4.16, and a correlation coefficient of 0.999. b. Matrix comparison: Serum, lithium heparin plasma, and sodium heparin plasma are recommended as acceptable matrices in the sponsor’s labeling. To demonstrate comparable performance, the sponsor compared serum from a plain glass tube (baseline) with serum from a gel tube, lithium heparin plasma, lithium heparin plasma from a gel tube, and sodium heparin plasma. All of the recommended matrices showed acceptable performance. 3. Clinical studies: a. Clinical Sensitivity: Not Applicable. b. Clinical specificity: Not Applicable. {10} c. Other clinical supportive data (when a. and b. are not applicable): 4. Clinical cut-off: Not Applicable. 5. Expected values/Reference range: The sponsor cites reference ranges from Tietz Fundamentals of Clinical Chemistry, 5th ed, 2001:975. | SERUM/PLASMA | Range (mg/dL) | Range (μmol/L) | | --- | --- | --- | | Cord | 0.6 to 1.2 | 53 to 106 | | Newborn, 1 to 4 days | 0.3 to 1.0 | 27 to 88 | | Infant | 0.2 to 0.4 | 18 to 35 | | Child | 0.3 to 0.7 | 27 to 62 | | Adolescent | 0.5 to 1.0 | 44 to 88 | | Adult, Male | 0.7 to 1.3 | 62 to 115 | | Adult, Female | 0.6 to 1.1 | 53 to 97 | | URINE | Range (mg/kg/day) | Range (μmol/kg/day) | | --- | --- | --- | | Infant | 8 to 20 | 71 to 177 | | Child | 8 to 22 | 71 to 194 | | Adolescent | 8 to 30 | 71 to 265 | | Adult, Male | 14 to 26 | 124 to 230 | | Adult, Female | 11 to 20 | 97 to 177 | (Declines with age to 10 mg/kg/day at age 90) The sponsor recommends that each laboratory determine its own reference ranges based upon its particular locale and patient characteristics. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...