K040976 · Genchem, Inc. · CGX · Dec 27, 2004 · Clinical Chemistry
Device Facts
Record ID
K040976
Device Name
GENCHEM CREATININE REAGENT
Applicant
Genchem, Inc.
Product Code
CGX · Clinical Chemistry
Decision Date
Dec 27, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1225
Device Class
Class 2
Intended Use
The GenChem Creatinine Test Reagent System is intended for the quantitative determination of creatinine in serum, plasma and urine on the Beckman SYNCHRON CX3® System and as an aid in the diagnosis of renal impairment and diseases such as chronic glomerulonephritis, diabetic nephropathy, chronic interstitial nephritis and as an indicator of glomerular filtration rate.
Device Story
Reagent kit containing alkaline picrate for quantitative creatinine determination. Operates on Beckman SYNCHRON CX3 System. Principle: Jaffe rate method; creatinine reacts with alkaline picrate to form red complex; rate of absorbance increase at 520 nm measured photometrically. Output: creatinine concentration in serum, plasma, or urine. Used in clinical laboratory settings by trained personnel. Results assist clinicians in diagnosing renal impairment and monitoring glomerular filtration rate. Reagent replaces predicate for use on existing CX3 hardware.
Clinical Evidence
Bench testing only. Precision/reproducibility evaluated over 30 days (n=60 per serum sample, n=59-60 per urine sample) per NCCLS EP3-T; total %CV ranged from 0.5% to 9.8%. Linearity assessed per NCCLS EP6-A (r=1.00). Analytical specificity tested for hemoglobin, bilirubin, and lipemia per NCCLS EP7-A with no adverse effects. Clinical comparison study (n=80 serum, n=80 plasma, n=79 urine) against the predicate device demonstrated high correlation (slope 0.991-1.000, R2 0.997-1.000).
Technological Characteristics
Alkaline picrate reagent for colorimetric creatinine determination. Designed for use on the Beckman SYNCHRON CX3® System. Analytical method follows standard colorimetric principles. Compatible with heparin and EDTA anticoagulants. No specific materials of construction or software algorithms described; device is a chemical reagent system.
Indications for Use
Indicated for quantitative determination of creatinine in serum, plasma, and urine to aid in diagnosis of renal impairment and diseases (e.g., chronic glomerulonephritis, diabetic nephropathy, chronic intestinal nephritis) as an indicator of glomerular filtration rate. For prescription use on Beckman SYNCHRON CX3 System.
Regulatory Classification
Identification
A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.
Predicate Devices
Beckman Creatinine Reagent for the CX3 (k915077)
Related Devices
K033058 — CREATININE REAGENT KIT FOR BECKMAN SYNCHRON CX & CX DELTA SYSTEMS · Diamond Diagnostics, Inc. · Jan 30, 2004
K983937 — DUAL VIAL CREATININE REAGENT · Intersect Systems, Inc. · Jan 6, 1999
K971908 — CREA OR CREATININE · Carolina Liquid Chemistries Corp. · Jun 17, 1997
K193649 — Yumizen C1200 Creatinine PAP · HORIBA ABX SAS · May 10, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
DEC 27 2004
# 510(k) SUMMARY K040976
### 1.0 Submitted Bv:
C.C. Allain, Ph.D. Chief Scientific Officer GenChem. Inc. 471 W. Lambert Road, Ste 107 Brea, CA 92821 Telephone: (714) 529-7125 FAX: (714) 529-3339
#### 2.0 Date of Preparation: June 1, 2004
### 3.0 Regulatory Information:
- 3.1 Regulation section:
- 3.2 21 CFR § 862.1225, Creatinine, Alkaline Picrate, Colorimetry Reagent for Beckman Synchron CX3® System
- 3.3 Clasification : Class II
- 3.4 Product Code: CGX
- 3.5 Panel: Clinical Chemistry (75)
### 4.0 Device Description:
The Device is a Reagent containing alkaline picrate for the determination of creatinine for optimum system operation on the Beckman Synchron CX3® System.
### 5.0 Substantial Equivalence Information:
- a. Predicate Device Name: Beckman Creatinine Reagent for the CX3
- b. Predicate K Number: K915077
- c. Comparison with Predicate: Both Reagents are similar in design, function and chemical principle as well as ingredient composition and concentration.
### Performance Characteristics: All studies were performed on the Beckman CX3 6.0 Synchron Analyzer
### 6.1 Precision/Reproducibility:
Control sera and urine pools were each assayed twice per day in triplicate on a SYNCHRON CX3® System. Data were collected on ten different days over a thirty day
{1}------------------------------------------------
period. Estimates of within run and total imprecision were calculated as described in NCCLS publication EP3-T.
| Precision of Creatinine Recoveries (mg/dL) | | | | | |
|--------------------------------------------|----|---------------------------|-------------|-----|--|
| | | | | | |
| Sample | n | Within Run<br>mean SD %CV | Total<br>SD | %CV | |
| Serum 1 | 60 | 0.5 0.05 9.8 | 0.05 | 9.8 | |
| Serum 2 | 60 | 4.0 0.02 0.5 | 0.02 | 0.5 | |
| Serum 3 | 60 | 7.4 0.03 0.5 | 0.05 | 0.7 | |
| Urine 1 | 59 | 40.3 0.41 1.0 | 0.56 | 1.4 | |
| Urine 2 | 60 | 222.9 2.29 1.0 | 2.79 | 1.3 | |
## Precision of Creatinine Recoveries (mg/dl )
#### 6.2 Linearity/assay reportable range:
Linearity was performed according to NCCLS Guideline EP6-A. Commercially available linearity standards ranging from 0 to 25.7 mg/dl were analyzed in triplicate on the Beckman CX3® and the results analyzed by the Least Squares method. The results gave a slope of 0.994 with an intercept of -0.05, a standard error of estimate of 0.11 and r = 1.00 and is shown below. Specimens exceeding these limits should be diluted with normal saline and reanalyzed. Multiply the result by the appropriate dilution factor.
| Specimens | Conventional Units | SI Units |
|--------------|--------------------|---------------------|
| Serum/Plasma | 0.2 to 25 mg/dL | 0.2 - 2210 mmol/L |
| Urine | 10 - 400 mg/dL | 0.88 - 35.36 mmol/L |
#### SENSITIVITY: 6.3
The sensitivity of this method is 0.2 mg/dL and is documented through the repetitive assay of a diluted serum control. The observed sensitivity limit, calculated as three standard deviations of a 21 replicate within run precision study, is 0.01 mg/dL and is below the claimed limit of 0.2 mg/dL.
### 6.4 Analytical Specificity:
Determined according to NCCLS EP7-A. Hemoglobin levels up to 500 mg/dL, Bilirubin levels up to 20 mg/dL, and Lipemia levels up to 1800 mg/dL were tested and did not show any adverse effect on a stock sample with a creatinne level of 1.1 mg/dL. Stock solutions of the substance to be tested were prepared at 20x concentrations and 0.5 ml of this stock was placed in a 10 ml volumetric flask and made up to volume with the base pool. The control stock was prepared similarly but with water as the diluent.
{2}------------------------------------------------
Heparin, Lithium Heparin Ammonium Heparin, and EDTA are acceptable anticoagulants.
### 7.0 Patient Comparison
Serum and plasma, ranging from 0.4 to 30.4 mg/dl and urine specimens ranging from 12.1to 400 mg/dL were collected from adult patients and assayed for creatinine on a SYNCHRON CX3® System using GenChem and Beckman creatinine reagents. Results were compared by least squares linear regression and the following statistics were obtained:
| VALUE | SERUM | PLASMA | URINE |
|-----------|------------|------------|------------------|
| Intercept | 0.0 | 0.05 | -0.3 |
| Slope | 0.991 | 0.998 | 1.000 |
| R2 Value | 0.997 | 0.998 | 1.000 |
| N | 80 | 80 | 79 |
| Range | 0.4 - 30.4 | 0.4 - 30.4 | 12.1 – 400 mg/dl |
### 8.0 Expected Values
The expected values for creatinine are listed below. Use these ranges only as guides. Each laboratory should establish its own normal ranges.
Normal Ranges
| Specimens | Conventional Units | SI Units |
|--------------|--------------------|----------------------|
| Serum/Plasma | 0.6 - 1.3 mg/dL | 53 - 115 mmol/L |
| Urine | 11 - 26 mg/day/kg | 97 - 230 mmol/day/kg |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services-USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol of medicine.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
DEC 27 2004
C.C. Allain, Ph.D. Chief Scientistic Officer GenChem, Inc. 471 W. Lambert Road, Suite 107 Brea, CA 92821
Re: k040976
Trade/Device Name: GenChem Creatinine Test Reagent Regulation Number: 21 CFR 862.1225 Regulation Name: Creatinine test system Regulatory Class: Class II Product Code: CGX Dated: October 15, 2004 Received: October 15, 2004
Dear Dr. Allain:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely vours.
Sincerely yours,
Cornelia B. Parks
Cornelia B. Rooks, MA Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE
510(k) Number: K040976
Device Name: GenChem Creatinine Test Reagent
Indications For Use:
The GenChem Creatinine Test Reagent System is intended for the quantitative determination of creatinine in serum, plasma and urine on the Beckman SYNCHRON CX3® System and as an aid in the diagnosis of renal impairment and diseases such as chronic glomerulonephritis, diabetic nephropathy, chronic interstitial nephritis and as an indicator of glomerular filtration rate.
Division Sign-Off
Office of In Vitro Diag Device Evaluation an
5100k k040976
Prescription Use X
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE
AND/OR
IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.