NOVA STATSENSOR CREATININE HOSPITAL METER, CONTROL SOLUTIONS,LINEARITY SOLUTIONS, AND TEST STRIPS

K070068 · Nova Biomedical Corp. · CGL · May 7, 2007 · Clinical Chemistry

Device Facts

Record IDK070068
Device NameNOVA STATSENSOR CREATININE HOSPITAL METER, CONTROL SOLUTIONS,LINEARITY SOLUTIONS, AND TEST STRIPS
ApplicantNova Biomedical Corp.
Product CodeCGL · Clinical Chemistry
Decision DateMay 7, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1225
Device ClassClass 2
AttributesPediatric

Intended Use

The Nova StatSensor Creatinine Hospital Meter is intended for in vitro diagnostic use by health care professionals and for Point-Of-Care usage for the quantitative measurement of creatinine in capillary, venous, and arterial whole blood. Creatinine measurements are used in the diagnosis and treatment of renal diseases and in monitoring renal dialysis. Not for use in neonates. Nova StatSensor Creatinine Test Strips are intended for use only with the StatSensor Creatinine Hospital Meter for quantitative tests. The Nova StatSensor Creatinine Hospital Meter is intended to quantitatively measure creatinine in whole blood. The Creatinine Meter is factory calibrated to provide plasma equivalent results to laboratory methods. Nova StatSensor Creatinine Test Strips are for testing outside the body (in vitro diagnostic use only). Nova StatSensor Creatinine Control Solutions are intended for use with the Nova StatSensor Creatinine Hospital Meter and Nova StatSensor Creatinine Test Strips as a quality control check to verify the accuracy of blood creatinine test results. There are three levels of controls, (Level 1, Level 2, and Level 3). These solutions will be offered for sale separately from the meter. Nova StatSensor Creatinine Linearity Kit solutions are used to check the linearity of the Nova StatSensor Creatinine Hospital Meter System. There are five levels of Linearity Solutions, (Level 1, Level 2, Level 3, Level 4, and Level 5).

Device Story

Point-of-care diagnostic system for quantitative creatinine measurement in whole blood; utilizes electrochemical test strips; operated by healthcare professionals in clinical settings. Device processes blood samples to provide plasma-equivalent creatinine results; aids clinicians in diagnosing renal disease and monitoring dialysis. System includes meter, test strips, quality control solutions, and linearity kits for accuracy verification. Provides rapid results to inform immediate clinical decision-making regarding renal function.

Clinical Evidence

No clinical data provided in the document; substantial equivalence determination based on 510(k) regulatory review.

Technological Characteristics

Electrochemical biosensor; uses test strips containing creatinine amidohydrolase, creatine amidinohydrolase, and sarcosine oxidase. Handheld form factor; 3.7V rechargeable lithium battery. Factory calibrated; no calibration code required. Connectivity via charging/data upload station. Operating range: 15°C-40°C, 20%-80% humidity. Measuring range: 0.3-12.5 mg/dL.

Indications for Use

Indicated for quantitative measurement of creatinine in capillary, venous, and arterial whole blood by healthcare professionals in point-of-care settings to aid in diagnosis/treatment of renal disease and monitoring dialysis. Contraindicated for use in neonates.

Regulatory Classification

Identification

A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and three human figures below it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Nova Biomedical Corp. c/o Dr. Paul W. MacDonald Chief Quality and Regulatory Affairs Officer 200 Prospect Street Waltham, MA 02454-9141 MAY - 7 2007 Re: k070068 > Trade/Device Name: Nova Statsensor Creatinine Hospital Meter, Nova Statsenor Creatinine Quality Control Levels 1,2,3, Nova Statsensor Creatinine Linearity Solutions Levels 1,2,3,4.5 Regulation Number: 21 CFR 862.1225 Regulation Name: Creatinine test system. Regulatory Class: Class II Product Code: CGL, JJX Dated: March 19, 2007 Received: March 20, 2007 Dear Dr. MacDonald: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {1}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Jean M. Cooper, M.S., D.V.M. Yean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known): k070068 Device Name:_____Nova StatSensor Creatinine Hospital Meter Indications For Use: The Nova StatSensor Creatinine Hospital Meter is intended for in vitro diagnostic use by health care professionals and for Point-Of-Care usage for the quantitative measurement of creatinine in capillary, venous, and arterial whole blood. Creatinine measurements are used in the diagnosis and treatment of renal diseases and in monitoring renal dialysis. Not for use in neonates. Nova StatSensor Creatinine Test Strips are intended for use only with the StatSensor Creatinine Hospital Meter for quantitative tests. The Nova StatSensor Creatinine Hospital Meter is intended to quantitatively measure creatinine in whole blood. The Creatinine Meter is calibrated to provide plasma equivalent results to laboratory methods. Nova StatSensor Creatinine Test Strips are for testing outside the body (in vitro diagnostic use only). Nova StatSensor Creatinine Control Solutions are intended for use with the Nova StatSensor Creatinine Hospital Meter and Nova StatSensor Creatinine Test Strips as a quality control check to verify the accuracy of blood creatinine test results. There are three levels of controls, (Level 1, Level 2, and Level 3). These solutions will be offered for sale separately from the meter. Nova StatSensor Creatinine Linearity Kit solutions are used to check the linearity of the Nova StatSensor Creatinine Hospital Meter System. There are five levels of Linearity Solutions, (Level 1, Level 2, Level 3, Level 4, and Level 5). Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Page 1 of Office of In Vitro Diagnostic Device Evaluation and Safety
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...