DEKA Loop

K234055 · Deka Research and Development · QJI · Mar 13, 2024 · Clinical Chemistry

Device Facts

Record IDK234055
Device NameDEKA Loop
ApplicantDeka Research and Development
Product CodeQJI · Clinical Chemistry
Decision DateMar 13, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1356
Device ClassClass 2
AttributesSoftware as a Medical Device, Therapeutic, Pediatric

Intended Use

DEKA Loop is intended for use with compatible integrated continuous glucose monitors (iCGM) and the DEKA alternate controller enabled (ACE) insulin infusion pump to automatically increase, decrease, and suspend delivery of basal insulin based on iCGM readings and predicted glucose values. It can also recommend, and with the user's confirmation, control the delivery of correction boluses when glucose values are predicted to exceed user configurable thresholds. DEKA Loop is intended for the management of Type 1 diabetes mellitus in persons six years of age and greater. DEKA Loop is intended for single patient use and requires a prescription.

Device Story

DEKA Loop is an interoperable Automated Glycemic Controller (iAGC) that manages Type 1 diabetes. It receives input from an integrated continuous glucose monitor (iCGM) and uses a predictive algorithm to adjust basal insulin delivery via an Alternate Controller Enabled (ACE) insulin pump. The system predicts glucose levels up to 6 hours into the future based on iCGM readings, insulin history, and user inputs (carbohydrate intake, exercise). It automatically increases, decreases, or suspends basal insulin and recommends correction boluses, which the user must confirm. The device is operated by the patient in home environments via an iOS app. It benefits patients by automating insulin delivery to maintain glucose within user-configurable thresholds, reducing the burden of manual adjustments.

Clinical Evidence

No new clinical trials were conducted. Substantial equivalence is supported by leveraging clinical validation data from the predicate (Tidepool Loop) and extensive in-silico simulation testing. In-silico testing included 24 million simulated-use hours across 1 million random scenarios, demonstrating 99.76% clinical equivalence. Discrepancies in 0.24% of scenarios were attributed to floating-point calculation differences between hardware platforms and determined not to impact clinical safety. Human factors validation confirmed the device is safe for its intended use.

Technological Characteristics

Software-only iAGC embedded in DEKA ACE pump. Interoperable via Bluetooth LE with iCGMs (e.g., Dexcom G6). Operates on user-defined therapy settings (basal rates, insulin sensitivity, carb ratios). Complies with ANSI/AAMI/UL 2800-1:2019 for interoperability. Software is a ported version of Tidepool Loop algorithm.

Indications for Use

Indicated for management of Type 1 diabetes mellitus in persons 6 years of age and older. Requires prescription. For use with compatible iCGM and ACE insulin infusion pumps.

Regulatory Classification

Identification

An interoperable automated glycemic controller is a device intended to automatically calculate drug doses based on inputs such as glucose and other relevant physiological parameters, and to command the delivery of such drug doses from a connected infusion pump. Interoperable automated glycemic controllers are designed to reliably and securely communicate with digitally connected devices to allow drug delivery commands to be sent, received, executed, and confirmed. Interoperable automated glycemic controllers are intended to be used in conjunction with digitally connected devices for the purpose of maintaining glycemic control.

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include: (i) An appropriate, as determined by FDA, clinical implementation strategy, including data demonstrating appropriate, as determined by FDA, clinical performance of the device for its intended use, including all of its indications for use. (A) The clinical data must be representative of the performance of the device in the intended use population and in clinically relevant use scenarios and sufficient to demonstrate appropriate, as determined by FDA, clinical performance of the device for its intended use, including all of its indications for use. (B) For devices indicated for use with multiple therapeutic agents for the same therapeutic effect ( *e.g.,* more than one type of insulin), data demonstrating performance with each product or, alternatively, an appropriate, as determined by FDA, clinical justification for why such data are not needed.(C) When determined to be necessary by FDA, the strategy must include postmarket data collection to confirm safe real-world use and monitor for rare adverse events. (ii) Results obtained through a human factors study that demonstrates that an intended user can safely use the device for its intended use. (iii) A detailed and appropriate, as determined by FDA, strategy to ensure secure and reliable means of data transmission with other intended connected devices. (iv) Specifications that are appropriate, as determined by FDA, for connected devices that shall be eligible to provide input to ( *e.g.,* specification of glucose sensor performance) or accept commands from (*e.g.,* specifications for drug infusion pump performance) the controller, and a detailed strategy for ensuring that connected devices meet these specifications.(v) Specifications for devices responsible for hosting the controller, and a detailed and appropriate, as determined by FDA, strategy for ensuring that the specifications are met by the hosting devices. (vi) Documentation demonstrating that appropriate, as determined by FDA, measures are in place ( *e.g.,* validated device design features) to ensure that safe therapy is maintained when communication with digitally connected devices is interrupted, lost, or re-established after an interruption. Validation testing results must demonstrate that critical events that occur during a loss of communications (*e.g.,* commands, device malfunctions, occlusions, etc.) are handled and logged appropriately during and after the interruption to maintain patient safety.(vii) A detailed plan and procedure for assigning postmarket responsibilities including adverse event reporting, complaint handling, and investigations with the manufacturers of devices that are digitally connected to the controller. (2) Design verification and validation documentation must include appropriate design inputs and design outputs that are essential for the proper functioning of the device that have been documented and include the following: (i) Risk control measures to address device system hazards; (ii) Design decisions related to how the risk control measures impact essential performance; and (iii) A traceability analysis demonstrating that all hazards are adequately controlled and that all controls have been validated in the final device design. (3) The device shall include appropriate, as determined by FDA, and validated interface specifications for digitally connected devices. These interface specifications shall, at a minimum, provide for the following: (i) Secure authentication (pairing) to connected devices; (ii) Secure, accurate, and reliable means of data transmission between the controller and connected devices; (iii) Sharing of necessary state information between the controller and any connected devices ( *e.g.,* battery level, reservoir level, sensor use life, pump status, error conditions);(iv) Ensuring that the controller continues to operate safely when data is received in a manner outside the bounds of the parameters specified; (v) A detailed process and procedures for sharing the controller's interface specification with connected devices and for validating the correct implementation of that protocol; and (vi) A mechanism for updating the controller software, including any software that is required for operation of the controller in a manner that ensures its safety and performance. (4) The device design must ensure that a record of critical events is stored and accessible for an adequate period to allow for auditing of communications between digitally connected devices, and to facilitate the sharing of pertinent information with the responsible parties for those connected devices. Critical events to be stored by the controller must, at a minimum, include: (i) Commands issued by the controller, and associated confirmations the controller receives from digitally connected devices; (ii) Malfunctions of the controller and malfunctions reported to the controller by digitally connected devices ( *e.g.,* infusion pump occlusion, glucose sensor shut down);(iii) Alarms and alerts and associated acknowledgements from the controller as well as those reported to the controller by digitally connected devices; and (iv) Connectivity events ( *e.g.,* establishment or loss of communications).(5) The device must only receive glucose input from devices cleared under § 862.1355 (integrated continuous glucose monitoring system), unless FDA determines an alternate type of glucose input device is designed appropriately to allow the controller to meet the special controls contained within this section. (6) The device must only command drug delivery from devices cleared under § 880.5730 of this chapter (alternate controller enabled infusion pump), unless FDA determines an alternate type of drug infusion pump device is designed appropriately to allow the controller to meet the special controls contained within this section. (7) An appropriate, as determined by FDA, training plan must be established for users and healthcare providers to assure the safety and performance of the device when used. This may include, but not be limited to, training on device contraindications, situations in which the device should not be used, notable differences in device functionality or features compared to similar alternative therapies, and information to help prescribers identify suitable candidate patients, as applicable. (8) The labeling required under § 809.10(b) of this chapter must include: (i) A contraindication for use in pediatric populations except to the extent clinical performance data or other available information demonstrates that it can be safely used in pediatric populations in whole or in part. (ii) A prominent statement identifying any populations for which use of this device has been determined to be unsafe. (iii) A prominent statement identifying by name the therapeutic agents that are compatible with the controller, including their identity and concentration, as appropriate. (iv) The identity of those digitally connected devices with which the controller can be used, including descriptions of the specific system configurations that can be used, per the detailed strategy submitted under paragraph (b)(1)(iii) of this section. (v) A comprehensive description of representative clinical performance in the hands of the intended user, including information specific to use in the pediatric use population, as appropriate. (vi) A comprehensive description of safety of the device, including, for example, the incidence of severe hypoglycemia, diabetic ketoacidosis, and other relevant adverse events observed in a study conducted to satisfy paragraph (b)(1)(i) of this section. (vii) For wireless connection enabled devices, a description of the wireless quality of service required for proper use of the device. (viii) For any controller with hardware components intended for multiple patient reuse, instructions for safely reprocessing the hardware components between uses.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ## I Background Information: A 510(k) Number K234055 B Applicant Deka Research and Development C Proprietary and Established Names DEKA Loop D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | QJI | Class II | 21 CFR 862.1356 - Interoperable Automated Glycemic Controller | CH - Clinical Chemistry | E Purpose for Submission New device ## II Intended Use/Indications for Use: A Intended Use(s): See Indications for Use below. B Indication(s) for Use: DEKA Loop is intended for use with compatible integrated continuous glucose monitors (iCGM) and the DEKA alternate controller enabled (ACE) insulin infusion pump to automatically increase, decrease, and suspend delivery of basal insulin based on iCGM readings and predicted glucose values. It can also recommend, and with the user's confirmation, control the delivery of correction boluses when glucose values are predicted to exceed user configurable thresholds. Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {1} DEKA Loop is intended for the management of Type 1 diabetes mellitus in persons six years of age and greater. DEKA Loop is intended for single patient use and requires a prescription. ## C Special Conditions for Use Statement(s): Rx - For Prescription Use Only DEKA Loop is not indicated for use by anyone who is pregnant, on dialysis, or critically ill. It is not known how different conditions or medications used in these scenarios may affect performance of the DEKA ACE Pump System with DEKA Loop. CGM readings being used by the system may be inaccurate in these cases, meaning the DEKA Loop will not have accurate information to use for dosage recommendations. This could result in over-delivery or under-delivery of insulin, which can lead to low and high blood glucose. Do not use DEKA Loop when using hydroxyurea, a medication used in the treatment of diseases including cancer and sickle cell anemia. CGM readings being used by the system may be inaccurate while taking hydroxyurea, meaning the DEKA Loop will not have accurate information to use for dosage recommendations. This could result in over-delivery of insulin, which can lead to hypoglycemia. Talk to your healthcare provider about using your BG meter for treatment decisions instead. Do not use DEKA Loop if you are taking GLP-1 agonists or SGLT-2 inhibitors, metformin, or DPP-4. DEKA Loop has not been tested in populations using additional diabetes medications and the performance of the system in conjunction with these medications is unknown. Do not take external insulin, such as manual insulin injections or inhaled insulins, while the DEKA Loop automation is enabled on the DEKA ACE Pump System. DEKA Loop does not receive information about insulin taken outside the system. If you choose to take additional insulin with another method while the DEKA Loop automation is enabled on the DEKA ACE Pump System, over-delivery or under-delivery of insulin may occur, which can lead to low and high blood glucose. Consult with your healthcare provider about how long to wait after manually taking insulin before enabling DEKA Loop automated insulin delivery. ## III Device/System Characteristics: ### A Device Description: The DEKA Loop is a software-only device and resides on the DEKA ACE pump (K233952). The DEKA Loop is an interoperable Alternate Glycemic Controller (iAGC) and works to control an ACE (Alternate Controller Enabled) insulin pump to automatically increase, decrease, and suspend delivery of basal insulin based on readings from an iCGM (integrated continuous glucose monitor) and glucose values predicted by the DEKA Loop. The DEKA Loop can also recommend, and with the user's confirmation, control the delivery of correction boluses when glucose values are predicted to exceed user configurable thresholds. The DEKA Loop is a ported version of the Tidepool Loop iAGC cleared in K203689 and is embedded within the DEKA ACE Pump System. As with the cleared Tidepool Loop device, the K234055 - Page 2 of 7 {2} DEKA Loop predicts glucose levels up to 6 hours in the future based on prior iCGM readings, insulin delivery history, and user input (e.g., carbohydrate intake and exercise), and uses that prediction to adjust insulin delivery. DEKA Loop requires that users input their specific therapy settings, which are established with the help of their health care provider before using the device, as part of the prescription process. These settings include: - Target Correction Ranges for normal operation, pre-meal, and workout presets - Settings can be programmed as low as 87 mg/dL and as high as 180 mg/dL and different ranges can be added for different times of day. - Glucose Safety Limit - Settings can be programmed as low as 67 mg/dL, or as high as 110 mg/dL or the Correction Range minimum, whichever is lower. - Carb to Insulin Ratios - Insulin sensitivity factors - Basal rates - Max basal rate - Max bolus ## IV Substantial Equivalence Information: ### A Predicate Device Name(s): Tidepool Loop ### B Predicate 510(k) Number(s): K203689 ### C Comparison with Predicate(s): | Device & Predicate Device(s): | K234055 | K203689 | | --- | --- | --- | | Device Trade Name | DEKA Loop | Tidepool Loop | | General Device Characteristic Similarities | | | | Intended Use/Indications For Use | The device is intended for use with compatible integrated continuous glucose monitors (iCGM) and alternate controller enabled (ACE) pumps to automatically increase, decrease, and suspend delivery of basal insulin based on iCGM readings and predicted glucose values. It can also recommend and deliver correction boluses when glucose values are predicted to exceed user configurable thresholds. | Same | K234055 - Page 3 of 7 {3} | Intended use population | Persons with Type 1 Diabetes Mellitus ages 6 and above | Same | | --- | --- | --- | | General Device Characteristic Differences | | | | Device Hosting Controller | DEKA ACE Pump System | iPhone | | Compatible ACE Pump | DEKA ACE Pump System | ACE pumps that have the specifications and meet the pre-specified acceptance criteria. | V Standards/Guidance Documents Referenced: Special controls established under 21 CFR 862.1356 for interoperable automated glycemic controller. VI Performance Characteristics (if/when applicable): A Other Supportive Instrument Performance Characteristics Data: 1. In-silico simulation testing The DEKA Loop is a ported version of the Tidepool Loop device, cleared under k203689. The DEKA Loop is intended to function identically to the Tidepool Loop, with the only differences being the physical location of the algorithm (i.e., integrated into the DEKA ACE pump in the case of the DEKA Loop, versus residing on a separate smartphone app in the case of the Tidepool Loop), and the programming language the device software is written in. To demonstrate that the DEKA Loop functions identically to the Tidepool Loop, and to support leveraging the clinical validation used for the Tidepool Loop, DEKA provided the results of software equivalency tests. DEKA used a layered approach as the test strategy to verify algorithm equivalence between the Tidepool Loop and the DEKA Loop. In addition to unit testing and software requirement verification testing for supporting the porting strategy, DEKA has performed equivalence testing to demonstrate clinical equivalence between the DEKA Loop and the Tidepool Loop, including single input vector integration testing, targeted scenario in-silico testing, and massively distributed random in-silico testing. Integration-level in-silico testing was conducted using target hardware for the DEKA Loop and Tidepool Loop (e.g., the DEKA ACE pump hardware, and an Apple iPhone developer's emulator, respectively). DEKA conducted side-by-side unit testing on the DEKA Loop and Tidepool Loop devices with 100% code coverage. Results of the unit testing demonstrated that the two devices are logically equivalent at the unit level. The primary equivalence metric used by DEKA to compare the two iAGC devices during integration testing with time series data is the Raw Glucose Equivalence over the duration of testing, which is defined as the root mean squared (RMS) percent difference. K234055 - Page 4 of 7 {4} The single input vector integration testing challenged both the DEKA Loop and the Tidepool Loop with a series of pre-defined input vectors and verified the equivalence of their outputs, as well as equivalence of intermediate calculations used in the calculation of outputs. In the single input vector integration testing, 25 input vectors were used, and the results of multiple different computations within the iAGC algorithms were compared to evaluate the equivalence of the two iAGC devices, including device output (bolus and basal recommendations), as well as multiple intermediate calculations. The overall Raw Glucose Equivalence metric was 99.99%, and the range across individual scenarios was 99.62% to 100%. Comparing device outputs, bolus recommendations showed overall equivalence of 100%, and basal recommendations showed overall equivalence of 99.97% (range: 99.62 - 100%). All instances of discrepant results were analyzed and determined to be the result of differences in the floating-point hardware used on the two different hardware platforms and not the result of different iAGC functionality. Results supported a finding of substantial equivalence. Targeted scenario in-silico testing provided additional supporting evidence of clinical equivalence between the subject device and the predicate. This testing added an in-silico human body model to the test system to challenge the combined algorithm/body model systems with clinically-relevant scenario-based inputs. In the targeted scenario in-silico testing, 15 scenarios were chosen to characterize the relative responses of the two iAGC devices across the expected variety of inputs including meals, exercise, CGM bias, CGM signal loss, and boluses. These scenarios were a series of events that would be expected under normal operation of the system and were designed to last 24 hours. The Raw Glucose Equivalence metric range for these 15 scenarios was from 99.76% to 100%. Results supported a finding of substantial equivalence. Massively distributed random in-silico testing was used to create approximately one million random variants of a 24-hour sequence of input conditions to test the equivalence of the DEKA Loop and the Tidepool Loop for 24 million simulated-use hours. In the massively distributed random in-silico testing, the test system used software to generate 24-hour sequences of random inputs from a set of defined variables. The same set of input scenarios were applied to both the Tidepool Loop and DEKA Loop algorithms to permit assessment of overall equivalency between the algorithms to be evaluated. Results showed numerically and/or clinically equivalent performance in 99.76% of scenarios. In 0.24% of scenarios, non-equivalent performance was observed. All non-equivalent results were reviewed and analyzed individually. The differences in simulation results were determined to be the result of numerical differences in floating-point calculations that, on rare occasions, could result in a transition above or below a hard threshold in the simulation framework or in the Loop algorithm, leading to different simulation results. In all instances it was determined that the DEKA Loop algorithm behaved as expected. Taken together, the results of all software equivalency tests support a finding of substantial equivalence. ## 2. Human Factors Human factors validation tests were conducted with the DEKA Loop in its finished configuration (i.e., integrated into the DEKA ACE Pump System). All study participants K234055 - Page 5 of 7 {5} received training that was consistent with the training that healthcare professionals and patients would receive with the commercial product. Usability evaluations assessed comprehension and usability of the device for critical tasks. Results of the study demonstrated that the DEKA Loop within the DEKA ACE Pump System is validated for its intended use. ## 3. User Training The following user training modes were validated as part of the human factors validation: - In-person training of health care professionals - Certified Diabetes Care and Education Specialists (CDCES) - Registered Nurses (RNs) - In-person training, 1:1 from the trained trainers described above The commercial implementation of the device utilizes a “train the trainer” approach wherein CDCES and RNs receive a 5 hour training from the manufacturer of the device and become qualified trainers. The device manufacturer determined if the trainers had sufficient system knowledge to then train the patients that use the device. This method of training the trainers was validated in the human factors study. 1:1 in-person user training provided by trained healthcare professionals representative of those available in a clinical setting was validated as part of human factors that had a duration of 3 hours. As part of this evaluation, a trainer guide was provided for the trainer to follow to train the intended users of this device. The trainer guide covers how the device works, what is included in the box, setting up the device, and assessment of user training results at each step for use of the device. A user training guide was also provided for device users. The trainer is expected to determine that users are competent in the use of the system or if additional training is required. ## 4. Interoperability A plan and approach for interoperability were provided and determined to be adequate to support and clearly specify expectations, requirements, and interface specifications for potential interoperable devices. The sponsor has established plans for interfacing with ACE pumps and iCGMs. The DEKA ACE Pump System is designed to be interoperable with the Dexcom G6 using Bluetooth LE. Details were provided in the Software Requirements and Design for iCGM Interface Subsystem. Further, the sponsor has performed software and bench testing to verify and validate that these interface control specifications meet the special controls and their controller’s requirements for performance, communications, and data logging. The same approach will be used when integrating with other iCGMs in the future. The DEKA Loop uses the same requirements for compatible iCGM performance as established for Tidepool Loop. DEKA Loop will not allow use of an unvalidated ACE pump or iCGM. By design, DEKA Loop can only be used with compatible ACE Pumps (e.g., DEKA ACE Pump System, K233952) and iCGMs (e.g., Dexcom G6, K203089) with which it has been validated for use. DEKA Loop currently receives glucose information through the DEKA ACE Pump System that only contains the interface to the Dexcom G6 iCGM. K234055 - Page 6 of 7 {6} In addition, the plans provided by the sponsor covered their approach to working with connected device companies regarding contractual issues, interfaces for data communication and exchange, and post-market reporting procedures and responsibilities (e.g., who is responsible for investigating and reporting complaints, malfunctions, and adverse events). The approach was determined to acceptably align with FDA-recognized ANSI/AAMI/UL 2800-1:2019 Standard for Safety for Medical Device Interoperability. 5. Postmarket Surveillance Study: As described in the decision summary for the predicate device, at the time of clearance for the predicate Tidepool Loop (K203689) there was uncertainty remaining regarding the clinical performance and risk/benefit profile of that device when used in the broader intended use population as compared to those studied in the Tidepool premarket study, and a postmarket surveillance study would be ordered by FDA to confirm understanding of safety. As of the time of this clearance for the DEKA Loop device, the Tidepool Loop device has not begun commercial distribution. Therefore, the planned postmarket surveillance study for the Tidepool Loop has not begun. The premarket clinical study of the Tidepool Loop was leveraged to support clearance of the DEKA Loop. As a result, the same uncertainty regarding clinical performance and risk/benefit profile exists for the DEKA Loop device. Accordingly, a postmarket surveillance study will be ordered by FDA to confirm performance and understanding of safety of this device. VII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. VIII Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. K234055 - Page 7 of 7
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...