Dexcom G6 Pro Continuous Glucose Monitoring System
K191833 · Dexcom, Inc. · QII · Oct 7, 2019 · Clinical Chemistry
Device Facts
Record ID
K191833
Device Name
Dexcom G6 Pro Continuous Glucose Monitoring System
Applicant
Dexcom, Inc.
Product Code
QII · Clinical Chemistry
Decision Date
Oct 7, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1355
Device Class
Class 2
Attributes
Pediatric
Intended Use
The Dexcom G6 Pro Continuous Glucose Monitoring System (Dexcom G6 Pro System) is a real time continuous glucose monitoring device indicated for the management of diabetes in persons age 2 years and older in a home environment while under the supervision of a healthcare professional. The Dexcom G6 Pro System is intended to replace fingerstick blood glucose testing for diabetes treatment decisions. Interpretation of the real-time Dexcom G6 Pro System results should be based on the glucose trends and several sequential readings over time. The Dexcom G6 Pro System may also be used as a retrospective glucose recording device indicated for assessing glycemic variability in persons age 2 years and older in a home environment while under the supervision of a healthcare professional. Retrospective interpretation of data recorded by the Dexcom G6 Pro System should be conducted solely by a healthcare professional. The Dexcom G6 Pro System aids in detecting glucose excursions facilitating care plan adjustments. The Dexcom G6 Pro System is also intended to interface with digitally connected devices. The Dexcom G6 Pro System can be used alone or in conjunction with these digitally connected medical devices or assessing glycemic variability.
Device Story
System uses sterile sensor probe inserted into interstitial fluid; enzymatic electrochemical reaction (glucose oxidase) generates electrical current proportional to glucose concentration. Transmitter samples current; onboard algorithm converts measurements to glucose readings. Data transmitted via Bluetooth to mobile app/receiver; displays real-time glucose values (updated every 5 minutes) and trends (up to 12 hours). Used in home environment; physician-enabled for either real-time monitoring or retrospective data logging. Healthcare professional (HCP) manages care plan adjustments based on glucose excursions. System replaces fingerstick testing for treatment decisions; supports interoperability with digitally connected devices.
Clinical Evidence
No new clinical data provided. Analytical and clinical performance established in predicate K191450. Device utilizes identical sensor and algorithm as predicate.
Technological Characteristics
Amperometric sensing of interstitial fluid glucose via glucose oxidase reaction; single-use 10-day transmitter; mobile app interface; factory calibrated; no optional user calibration; real-time and retrospective data modes.
Indications for Use
Indicated for management of diabetes or assessment of glycemic variability in persons age 2 years and older in a home environment under healthcare professional supervision.
Regulatory Classification
Identification
An integrated continuous glucose monitoring system (iCGM) is intended to automatically measure glucose in bodily fluids continuously or frequently for a specified period of time. iCGM systems are designed to reliably and securely transmit glucose measurement data to digitally connected devices, including automated insulin dosing systems, and are intended to be used alone or in conjunction with these digitally connected medical devices for the purpose of managing a disease or condition related to glycemic control.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following:
(i) Robust clinical data demonstrating the accuracy of the device in the intended use population.
(ii) The clinical data must include a comparison between iCGM values and blood glucose values in specimens collected in parallel that are measured on an FDA-accepted laboratory-based glucose measurement method that is precise and accurate, and that is traceable to a higher order (
*e.g.,* an internationally recognized reference material and/or method).(iii) The clinical data must be obtained from a clinical study designed to fully represent the performance of the device throughout the intended use population and throughout the measuring range of the device.
(iv) Clinical study results must demonstrate consistent analytical and clinical performance throughout the sensor wear period.
(v) Clinical study results in the adult population must meet the following performance requirements:
(A) For all iCGM measurements less than 70 milligrams/deciliter (mg/dL), the percentage of iCGM measurements within ±15 mg/dL of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 85 percent.
(B) For all iCGM measurements from 70 mg/dL to 180 mg/dL, the percentage of iCGM measurements within ±15 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 70 percent.
(C) For all iCGM measurements greater than 180 mg/dL, the percentage of iCGM measurements within ±15 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 80 percent.
(D) For all iCGM measurements less than 70 mg/dL, the percentage of iCGM measurements within ±40 mg/dL of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 98 percent.
(E) For all iCGM measurements from 70 mg/dL to 180 mg/dL, the percentage of iCGM measurements within ±40 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 99 percent.
(F) For all iCGM measurements greater than180 mg/dL, the percentage of iCGM measurements within ±40 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 99 percent.
(G) Throughout the device measuring range, the percentage of iCGM measurements within ±20 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 87 percent.
(H) When iCGM values are less than 70 mg/dL, no corresponding blood glucose value shall read above 180 mg/dL.
(I) When iCGM values are greater than 180 mg/dL, no corresponding blood glucose value shall read less than 70 mg/dL.
(J) There shall be no more than 1 percent of iCGM measurements that indicate a positive glucose rate of change greater than 1 mg/dL per minute (/min) when the corresponding true negative glucose rate of change is less than −2 mg/dL/min as determined by the corresponding blood glucose measurements.
(K) There shall be no more than 1 percent of iCGM measurements that indicate a negative glucose rate of change less than −1 mg/dL/min when the corresponding true positive glucose rate of change is greater than 2 mg/dL/min as determined by the corresponding blood glucose measurements.
(vi) Data demonstrating similar accuracy and rate of change performance of the iCGM in the pediatric population as compared to that in the adult population, or alternatively a clinical and/or technical justification for why pediatric data are not needed, must be provided and determined by FDA to be acceptable and appropriate.
(vii) Data must demonstrate that throughout the claimed sensor life, the device does not allow clinically significant gaps in sensor data availability that would prevent any digitally connected devices from achieving their intended use.
(2) Design verification and validation must include a detailed strategy to ensure secure and reliable means of iCGM data transmission to provide real-time glucose readings at clinically meaningful time intervals to devices intended to receive the iCGM glucose data.
(3) Design verification and validation must include adequate controls established during manufacturing and at product release to ensure the released product meets the performance specifications as defined in paragraphs (b)(1) and (b)(2) of this section.
(4) The device must demonstrate clinically acceptable performance in the presence of clinically relevant levels of potential interfering substances that are reasonably present in the intended use population, including but not limited to endogenous substances and metabolites, foods, dietary supplements, and medications.
(5) The device must include appropriate measures to ensure that disposable sensors cannot be used beyond its claimed sensor wear period.
(6) Design verification and validation must include results obtained through a usability study that demonstrates that the intended user can use the device safely and obtain the expected glucose measurement accuracy.
(7) The labeling required under § 809.10(b) of this chapter must include a separate description of the following sensor performance data observed in the clinical study performed in conformance with paragraph (b)(1) of this section for each intended use population, in addition to separate sensor performance data for each different iCGM insertion or use sites (
*e.g.,* abdomen, arm, buttock):(i) A description of the accuracy in the following blood glucose concentration ranges: less than 54 mg/dL, 54 mg/dL to less than 70 mg/dL, 70 to 180 mg/dL, greater than 180 to 250 mg/dL, and greater than 250 mg/dL.
(ii) A description of the accuracy of positive and negative rate of change data.
(iii) A description of the frequency and duration of gaps in sensor data.
(iv) A description of the true, false, missed, and correct alert rates and a description of the available glucose concentration alert settings, if applicable.
(v) A description of the observed duration of iCGM life for the device.
Predicate Devices
Dexcom G6 Continuous Glucose Monitoring System (K191450)
Related Devices
K193642 — Dexcom G6 Glucose Program Continuous Glucose Monitoring (CGM) System · Dexcom, Inc. · Jan 29, 2020
K182041 — Dexcom G6 Glucose Program Continuous Glucose Monitoring System · Dexcom, Inc. · Oct 26, 2018
K191450 — Dexcom G6 Continuous Glucose Monitoring System, Dexcom G6 Glucose Program Continuous Glucose Monitoring System and Dexcom Pro Q Continuous Glucose Monitoring System · Dexcom, Inc. · Jun 28, 2019
P050012 — DEXCOM STS CONTINOUS MONITORS · Dexcom, Inc. · Mar 24, 2006
Submission Summary (Full Text)
{0}
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K191833
B Applicant
Dexcom, Inc.
C Proprietary and Established Names
Dexcom G6 Pro Continuous Glucose Monitoring System,
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QII | Class II | 21 CFR 862.1355 - Integrated Continuous Glucose Monitoring System | CH - Clinical Chemistry |
## II Submission/Device Overview:
A Purpose for Submission:
New Device
B Measurand:
Glucose in interstitial fluid
C Type of Test:
Quantitative, amperometric assay (Glucose Oxidase)
K191833 - Page 1 of 8
{1}
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The Dexcom G6 Pro Continuous Glucose Monitoring System (Dexcom G6 Pro System) is a real time continuous glucose monitoring device indicated for the management of diabetes in persons age 2 years and older in a home environment while under the supervision of a healthcare professional. The Dexcom G6 Pro System is intended to replace fingerstick blood glucose testing for diabetes treatment decisions. Interpretation of the real-time Dexcom G6 Pro System results should be based on the glucose trends and several sequential readings over time.
The Dexcom G6 Pro System may also be used as a retrospective glucose recording device indicated for assessing glycemic variability in persons age 2 years and older in a home environment while under the supervision of a healthcare professional. Retrospective interpretation of data recorded by the Dexcom G6 Pro System should be conducted solely by a healthcare professional.
The Dexcom G6 Pro System aids in detecting glucose excursions facilitating care plan adjustments. The Dexcom G6 Pro System is also intended to interface with digitally connected devices. The Dexcom G6 Pro System can be used alone or in conjunction with these digitally connected medical devices for managing diabetes or assessing glycemic variability.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
D Special Instrument Requirements:
Not Applicable
IV Device/System Characteristics:
A Device Description:
G6 PRO CGM SENSOR
The sensor component is a sterile device that consists of the sensor applicator, plastic base ("transmitter holder"), and sensor probe. The applicator is a single use, disposable unit that contains an introducer needle holding the sensor probe. The applicator deploys the needle and inserts the sensor under the skin. The needle is retracted back into the applicator after insertion. The sensor probe continuously measures glucose concentration in interstitial fluid and can be worn for up to 10 days.
K191833 - Page 2 of 8
{2}
The sensor may be worn in the abdomen for adults, and both the abdomen and buttock for children ages 2-17 years old.
## G6 PRO TRANSMITTER
The transmitter component is a miniature radio transmitter that incorporates data processing functionality. The transmitter contains a Bluetooth radio transceiver for communication with a compatible display device (i.e., mobile device). The transmitter attaches to the sensor and can be re-used for multiple sensing sessions up to three months. The transmitter includes an auto-start feature which enables the transmitter to start a session immediately upon attachment of the transmitter to the on-body wearable. The G6 Pro Transmitter can be used for retrospective CGM data logging and it can also send real-time estimated glucose values to a mobile application.
## G6 PRO MOBILE APP
The G6 CGM App for iOS and G6 CGM App for Android is the primary display system for the physician enabled, real-time users with a compatible, BLE-enabled smart device. The G6 CGM App is compatible with certain iOS, Android and Smart Device watches. A link to a list of compatible devices is included in the instructions for use.
The Dexcom G6 Pro CGM System is an interoperable connected device that can communicate glucose readings and other information wirelessly and securely to and from interoperable electronic interfaces. The G6 Glucose Program CGM system is designed to communicate with interoperable devices in several ways, such as described below:
- Wireless communication from the transmitter directly to an interoperable device communicating through the same protocol.
- The app communicates to another app on a single mobile platform.
- The app communicates through the cloud to another software device.
## B Principle of Operation:
The G6 Pro system detects glucose levels from the fluid just beneath the skin (interstitial fluid). The sensor probe continuously measures glucose concentration in the interstitial fluid via an enzymatic electrochemical reaction using glucose oxidase. The enzyme, glucose oxidase, catalyzes the oxidation of glucose and produces hydrogen peroxide. The production of hydrogen peroxide generates an electrical current that is proportionate to the interstitial glucose concentration. The transmitter then samples the electrical current produced by the sensor and converts these measurements into glucose readings using an onboard algorithm. The transmitter sends glucose data to the receiver and/or mobile app which displays the current glucose reading (updated every 5 minutes) and glucose trends (up to 12 hours) from the transmitter. The receiver and/or mobile app displays the current glucose reading and glucose trends to the user. It alerts the user when glucose levels are outside of a target zone and when other important system conditions occur. The G6 Pro System does not require calibrations using self-monitoring of
K191833 - Page 3 of 8
{3}
blood glucose (SMBG), and the sensor life has an expected wear time of up to 10 days. The system can be used for either a real-time or retrospective CGM session. A users Healthcare Professional (HCP) elects which type of CGM session the patient uses (retrospective vs. real-time). The G6 Pro Transmitter Firmware supports a single-use 10-day sensor session per transmitter.
C Instrument Description Information:
| Modes of Operation | Yes | No |
| --- | --- | --- |
| Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device? | ☑ | ☐ |
| Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission? | ☑ | ☐ |
| Software | | |
| FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types. | ☑ | ☐ |
1. Instrument Name: Dexcom G6 Pro Continuous Glucose Monitoring System
2. Specimen Identification: Not applicable.
3. Specimen Sampling and Handling: Not applicable.
4. Calibration: The Dexcom G6 Pro Continuous Glucose Monitoring System does not require user calibration, user calibration is not possible for this device.
5. Quality Control: Not applicable.
V Substantial Equivalence Information:
A Predicate Device Name(s): Dexcom G6 Continuous Glucose Monitoring System
B Predicate 510(k) Number(s): K191450
K191833 - Page 4 of 8
{4}
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K191450 | K191833 |
| --- | --- | --- |
| Device Trade Name | Dexcom G6 Continuous Glucose Monitoring (CGM) System | Dexcom G6 Pro Continuous Glucose Monitoring System |
| General Device Characteristic Similarities | | |
| Intended Use | Intended to automatically measure glucose in bodily fluids continuously or frequently for a specified period of time. Designed to reliably and securely transmit glucose measurement data to digitally connected devices and intended to be used alone or in conjunction with these digitally connected medical devices for the purpose of managing a disease or condition related to glycemic control. | Same |
| Principle of Operation | Amperometric measurement of current proportional to glucose concentration in interstitial fluid via glucose oxidase chemical reaction | Same |
| Data Presented | Estimated Glucose Value (EGV): The EGV is the nominal glucose value presented to the user. Glucose Trend: Based off the glucose rate of change, users are shown their glucose trend with a corresponding arrow. Historical Glucose Data: Users can view their previous three, six, twelve, or twenty-four hours of glucose data. | Same |
| Features | Connect to Dexcom Share: Users can share their glucose data with followers. | Same |
| Human Factors | Easy to understand UI/UX. Commonly understood navigation tools and features. Color-coded graphics. | Same |
K191833 - Page 5 of 8
{5}
| General Device Characteristic Differences | | |
| --- | --- | --- |
| Data Availability | Real-time | Real-time or retrospective at the discretion of a Healthcare Professional |
| Clinical application | Management of diabetes mellitus | Management of diabetes mellitus or assessing glycemic control in other conditions |
| Clinical setting/sites of use | Home use | Sensor insertion, transmitter attachment and retrospective glucose data download occurs in a clinic with a healthcare professional. Used at home in between clinic visits. |
| Transmitter Use Life | Reusable (90 days) | Single-Use (10-day) |
VI Standards/Guidance Documents Referenced:
ISO 13485:2003; Medical Devices-Quality management systems-Requirements for regulatory purposes
ISO 14971:2012; Medical devices – Applications of risk management to medical devices
IEC 62304, Ed. 1.1:2015; Medical Device Software – Software Lifecycle Processes (2006+AMD1:2015)
IEC 62366 Ed. 1.1:2014; Medical Devices – Application of usability engineering to medical devices, Edition 1.1 (2014)
IEC 60601-1:2005; Medical Electrical Equipment – Part 1: Requirements for basic safety and essential performance
ISO 15223-1:2012; Medical devices – Symbols to be used with medical device labels, labeling and information to be supplied – Part 1: General requirements
Guidance for Industry and FDA Staff, Format for Traditional and Abbreviated 510(k)s, dated August 2005
Guidance for Industry and FDA Staff, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, dated May 11, 2005
General Principles of Software Validation; Final Guidance for Industry and FDA Staff, dated January 11, 2002
Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, Guidance for Industry and Food and Drug Administration Staff, dated October 2, 2014
Applying Human Factors and Usability Engineering to Medical Devices – Guidance for Industry and FDA Staff, dated February 3, 2016
K191833 - Page 6 of 8
{6}
Guidance for Industry and FDA Staff Use of Symbols on Labels and in Labeling of In Vitro Diagnostic Devices Intended for Professional Use, dated November 30, 2004
eCopy Program for Medical Device Submissions, Guidance for Industry and FDA Staff, dated December 31, 2015
Refuse to Accept Policy for 510(k)s, dated January 30, 2018
## VII Performance Characteristics (if/when applicable):
The analytical and clinical performance of this device was established in K191450. The sensor and algorithm remain unchanged. The candidate transmitter has the same overall form, fit, and function as the predicate transmitter, the difference being updated transmitter firmware.
Firmware was modified from K191450 to additionally support the storage of glucose data for the entire sensor wear period and enable retrospective analysis by a healthcare provider using an authorized extraction device. Transmitter firmware for the G6 pro is also updated to enable an auto-start feature, ensure only a single, ten-day use, and prevent calibration by the user. The predicate (K191450) mobile App (iOS and Android) has been updated to accommodate use by the candidate device.
## A Analytical Performance:
1. Precision/Reproducibility:
As established in K191450.
2. Linearity:
As established in K191450.
3. Analytical Specificity/Interference:
As established in K191450.
4. Assay Reportable Range:
As established in K191450.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
As established in K191450.
6. Detection Limit:
As established in K191450.
7. Assay Cut-Off:
Not applicable.
K191833 - Page 7 of 8
{7}
8. Accuracy (Instrument):
As established in K191450.
9. Carry-Over:
Not applicable.
B Comparison Studies:
1. Method Comparison with Predicate Device:
Not applicable
2. Matrix Comparison:
Not applicable. Interstitial fluid is the only indicated matrix.
C Clinical Studies:
1. Clinical Performance:
As established in K191450.
2. Clinical Specificity:
As established in K191450.
D Clinical Cut-Off:
As established in K191450.
E Expected Values/Reference Range:
Not applicable.
F Other Supportive Instrument Performance Characteristics Data:
As established in K191450.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K191833 - Page 8 of 8
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.