Dexcom G6 Continuous Glucose Monitoring (CGM) System
Applicant
Dexcom, Inc.
Product Code
QBJ · Clinical Chemistry
Decision Date
Jul 15, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1355
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Dexcom G6 Continuous Glucose Monitoring System (Dexcom G6 System) is a real time, continuous glucose monitoring device indicated for the management of diabetes in persons age 2 years and older. The Dexcom G6 System is intended to replace fingerstick blood glucose testing for diabetes treatment decisions. Interpretation of the Dexcom G6 System results should be based on the glucose trends and several sequential readings over time. The Dexcom G6 System also aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. The Dexcom G6 System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The Dexcom G6 System can be used alone or in conjunction with these digitally connected medical devices for the purpose of managing diabetes.
Device Story
Subcutaneous glucose sensor measures interstitial fluid glucose via glucose oxidase enzymatic reaction; generates electrical current proportional to glucose concentration. Transmitter processes signals using internal algorithm; transmits data via Bluetooth Low Energy (BLE) to receiver or mobile app. New Partner Web APIs enable cloud-based data sharing with authorized client software for non-medical applications, data analysis, secondary alarms, and patient monitoring. System provides real-time glucose readings, trend arrows, and alerts for hypo/hyperglycemia. Used by patients for diabetes management; replaces fingerstick testing for treatment decisions. Not intended for AID system integration via Partner Web APIs. Healthcare providers use output for therapy adjustments. Benefits include continuous monitoring, reduced fingersticks, and improved glycemic control.
Clinical Evidence
No new clinical data provided. Performance characteristics established in previous submissions (DEN170088, K200876) remain applicable. Bench testing and validation performed for software, interoperability, and cybersecurity.
Technological Characteristics
Subcutaneous enzymatic electrochemical sensor (glucose oxidase). BLE transmitter; receiver/mobile app display. Reportable range 40-400 mg/dL. Sterilization via radiation (ISO 11137). Biocompatibility per ISO 10993. Software lifecycle per IEC 62304. Cloud-based API connectivity added.
Indications for Use
Indicated for management of diabetes in persons age 2 years and older. Intended to replace fingerstick blood glucose testing for treatment decisions and aid in detection of hyperglycemia/hypoglycemia episodes.
Regulatory Classification
Identification
An integrated continuous glucose monitoring system (iCGM) is intended to automatically measure glucose in bodily fluids continuously or frequently for a specified period of time. iCGM systems are designed to reliably and securely transmit glucose measurement data to digitally connected devices, including automated insulin dosing systems, and are intended to be used alone or in conjunction with these digitally connected medical devices for the purpose of managing a disease or condition related to glycemic control.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following:
(i) Robust clinical data demonstrating the accuracy of the device in the intended use population.
(ii) The clinical data must include a comparison between iCGM values and blood glucose values in specimens collected in parallel that are measured on an FDA-accepted laboratory-based glucose measurement method that is precise and accurate, and that is traceable to a higher order (
*e.g.,* an internationally recognized reference material and/or method).(iii) The clinical data must be obtained from a clinical study designed to fully represent the performance of the device throughout the intended use population and throughout the measuring range of the device.
(iv) Clinical study results must demonstrate consistent analytical and clinical performance throughout the sensor wear period.
(v) Clinical study results in the adult population must meet the following performance requirements:
(A) For all iCGM measurements less than 70 milligrams/deciliter (mg/dL), the percentage of iCGM measurements within ±15 mg/dL of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 85 percent.
(B) For all iCGM measurements from 70 mg/dL to 180 mg/dL, the percentage of iCGM measurements within ±15 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 70 percent.
(C) For all iCGM measurements greater than 180 mg/dL, the percentage of iCGM measurements within ±15 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 80 percent.
(D) For all iCGM measurements less than 70 mg/dL, the percentage of iCGM measurements within ±40 mg/dL of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 98 percent.
(E) For all iCGM measurements from 70 mg/dL to 180 mg/dL, the percentage of iCGM measurements within ±40 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 99 percent.
(F) For all iCGM measurements greater than180 mg/dL, the percentage of iCGM measurements within ±40 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 99 percent.
(G) Throughout the device measuring range, the percentage of iCGM measurements within ±20 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 87 percent.
(H) When iCGM values are less than 70 mg/dL, no corresponding blood glucose value shall read above 180 mg/dL.
(I) When iCGM values are greater than 180 mg/dL, no corresponding blood glucose value shall read less than 70 mg/dL.
(J) There shall be no more than 1 percent of iCGM measurements that indicate a positive glucose rate of change greater than 1 mg/dL per minute (/min) when the corresponding true negative glucose rate of change is less than −2 mg/dL/min as determined by the corresponding blood glucose measurements.
(K) There shall be no more than 1 percent of iCGM measurements that indicate a negative glucose rate of change less than −1 mg/dL/min when the corresponding true positive glucose rate of change is greater than 2 mg/dL/min as determined by the corresponding blood glucose measurements.
(vi) Data demonstrating similar accuracy and rate of change performance of the iCGM in the pediatric population as compared to that in the adult population, or alternatively a clinical and/or technical justification for why pediatric data are not needed, must be provided and determined by FDA to be acceptable and appropriate.
(vii) Data must demonstrate that throughout the claimed sensor life, the device does not allow clinically significant gaps in sensor data availability that would prevent any digitally connected devices from achieving their intended use.
(2) Design verification and validation must include a detailed strategy to ensure secure and reliable means of iCGM data transmission to provide real-time glucose readings at clinically meaningful time intervals to devices intended to receive the iCGM glucose data.
(3) Design verification and validation must include adequate controls established during manufacturing and at product release to ensure the released product meets the performance specifications as defined in paragraphs (b)(1) and (b)(2) of this section.
(4) The device must demonstrate clinically acceptable performance in the presence of clinically relevant levels of potential interfering substances that are reasonably present in the intended use population, including but not limited to endogenous substances and metabolites, foods, dietary supplements, and medications.
(5) The device must include appropriate measures to ensure that disposable sensors cannot be used beyond its claimed sensor wear period.
(6) Design verification and validation must include results obtained through a usability study that demonstrates that the intended user can use the device safely and obtain the expected glucose measurement accuracy.
(7) The labeling required under § 809.10(b) of this chapter must include a separate description of the following sensor performance data observed in the clinical study performed in conformance with paragraph (b)(1) of this section for each intended use population, in addition to separate sensor performance data for each different iCGM insertion or use sites (
*e.g.,* abdomen, arm, buttock):(i) A description of the accuracy in the following blood glucose concentration ranges: less than 54 mg/dL, 54 mg/dL to less than 70 mg/dL, 70 to 180 mg/dL, greater than 180 to 250 mg/dL, and greater than 250 mg/dL.
(ii) A description of the accuracy of positive and negative rate of change data.
(iii) A description of the frequency and duration of gaps in sensor data.
(iv) A description of the true, false, missed, and correct alert rates and a description of the available glucose concentration alert settings, if applicable.
(v) A description of the observed duration of iCGM life for the device.
Predicate Devices
Dexcom G6 Continuous Glucose Monitoring System (K200876)
Related Devices
K231081 — Dexcom G7 Continuous Glucose Monitoring (CGM) System · Dexcom, Inc. · May 15, 2023
K183206 — Dexcom G6 Continuous Glucose Monitoring System · Dexcom, Inc. · Feb 14, 2019
K221259 — Dexcom G6 Continuous Glucose Monitoring (CGM) System, Dexcom G6 Glucose Program Continuous Glucose Monitoring (CGM) System, Dexcom G6 Professional Continuous Glucose Monitoring (CGM) System · Dexcom, Inc. · Jul 29, 2022
K223931 — Dexcom G6 Continuous Glucose Monitoring (CGM) System · Dexcom, Inc. · Jan 11, 2023
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
INSTRUMENT ONLY
## I Background Information:
A 510(k) Number
K201328
B Applicant
Dexcom, Inc.
C Proprietary and Established Names
Dexcom G6 Continuous Glucose Monitoring (CGM) System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QBJ | Class II | 21 CFR 862.1355 | CH - Clinical Chemistry |
## II Submission/Device Overview:
A Purpose for Submission:
This submission adds application programming interfaces (APIs) to the Dexcom G6 System to facilitate users sharing glucose data with authorized client software for specific and permitted use cases.
B Measurand:
Glucose in Interstitial Fluid
C Type of Test:
Quantitative, amperometric assay (Glucose Oxidase)
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The Dexcom G6 Continuous Glucose Monitoring System (Dexcom G6 System) is a real time, continuous glucose monitoring device indicated for the management of diabetes in persons age 2 years and older.
The Dexcom G6 System is intended to replace fingerstick blood glucose testing for diabetes treatment decisions. Interpretation of the Dexcom G6 System results should be based on the glucose trends and several sequential readings over time. The Dexcom G6 System also aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments.
The Dexcom G6 System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The Dexcom G6 System can be used alone or in conjunction with these digitally connected medical devices for the purpose of managing diabetes.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
- Remove the Dexcom G6 sensor, transmitter, and receiver before Magnetic Resonance Imaging (MRI), Computed Tomography (CT) scan, or high-frequency electrical heat (diathermy) treatment. The magnetic fields and heat could damage the components of the Dexcom G6 System, which may cause it to display inaccurate blood glucose readings or may prevent alerts.
- When wearing the device, ask for hand-wanding or full-body pat-down and visual inspection instead of going through the Advanced Imaging Technology (AIT) body scanner. Also avoid putting any part of the device through baggage x-ray machine.
- This device is not intended for pregnant women, people on dialysis, or critically ill patients.
- The device should not be used to make diabetes treatment decisions when:
- The user has not used the iCGM before or is unfamiliar with the Dexcom G6 System. (It may take days, weeks or months for a user to gain confidence in using the iCGM to make treatment decisions.)
- The user’s symptoms do not match the glucose values displayed by the device.
- The device does not show a glucose value or a trend arrow.
- During the first two hours of sensor warm-up period, the user should use a blood glucose meter to make treatment decisions.
- The user’s glucose is rising or falling rapidly.
- Although standard dosing of acetaminophen (1000 mg per every 6 hours) does not appear to cause significant bias, higher supra-therapeutic levels of acetaminophen have shown significant positive bias.
- Sensor glucose readings will be falsely higher if the user is taking hydroxyurea. Do not use the device for diabetes treatment decisions when taking hydroxyurea.
K201328 - Page 2 of 10
{2}
- Adult users should only use the abdomen and pediatric users should only use the buttock or abdomen. Sensor performance has not been evaluated in other insertion sites and may differ from expected iCGM performance.
- If a sensor wire breaks or detaches from the sensor, it could remain under the user’s skin. The user should contact their healthcare practitioner if this occurs.
- The transmitter should not be shared to avoid transmission of bloodborne illnesses.
- When using Bluetooth headphones, speakers, etc., the user's alarm/alerts may sound on the primary smart device or on the accessory. Each accessory is different. The user should test their device so that they know where they will hear the alarm/alerts.
- Before updating the smart device hardware or operating system, verify the compatibility of the updated hardware/software with the device system.
- The software application may not be used in environments not currently cleared for Dexcom G6 CGM System (e.g. hospital for inpatient care).
- The Partner Web APIs is not intended to be used by automated insulin delivery systems (AID).
IV Device/System Characteristics:
A Device Description:
The Dexcom G6 Continuous Glucose Monitoring (CGM) System uses a subcutaneous glucose sensor that measures glucose concentration in interstitial fluid and consists of three main components: a sensor, a Bluetooth Low Energy (BLE) transmitter and a BLE enabled display device (receiver and/or mobile application). The user can view glucose data on the receiver or on the G6 CGM App (i.e., a mobile medical application) running on a compatible mobile device, or on both simultaneously. An adhesive patch sits on the surface of the skin to position and hold the sensor, electrical contact elements, and an attachment point for a transmitter, which receives sensor signals, processes these, and sends glucose information to a dedicated receiver or mobile app.
G6 CGM SENSOR
The sensor component is a sterile device that consists of the sensor applicator, plastic base (“transmitter holder”), and sensor probe. The applicator is a single use, disposable unit that contains an introducer needle holding the sensor probe. The applicator deploys the needle and inserts the sensor under the skin. The needle is retracted back into the applicator after insertion. The sensor probe continuously measures glucose concentration in interstitial fluid and can be worn for up to 10 days.
G6 CGM TRANSMITTER
The G6 CGM Transmitter is a miniature radio transmitter that incorporates data processing functionality. The transmitter contains a Bluetooth radio transceiver for communication with a compatible display device (i.e., receiver and/or smart device). The transmitter attaches to the sensor and can be re-used for multiple sensing sessions up to three months.
G6 CGM RECEIVER
The G6 CGM Receiver is small hand-held device that wirelessly receives glucose information from the transmitter every five minutes and includes a touchscreen display. The 5 receiver displays the current glucose reading and glucose trends to the user. It alerts the user when glucose levels are outside of a target zone and when other important system conditions occur.
K201328 - Page 3 of 10
{3}
DEXCOM G6 System MOBILE APP
The G6 CGM App for iOS and G6 CGM App for Android provides an alternative display device to the receiver for users with a compatible, BLE-enabled smart device and behaves similarly to the receiver. The G6 CGM App is compatible with certain iOS, Android and Smart Device watches. A link to a list of compatible devices is included in the instructions for use.
The Dexcom G6 System is an interoperable connected device that can communicate glucose readings and other information wirelessly and securely to and from interoperable electronic interfaces; including compatible AID systems. The G6 CGM system is designed to communicate with interoperable devices in several ways, such as described below:
- Wireless communication from the transmitter directly to an interoperable device communicating through the same protocol.
- The app communicates to another app on a single mobile platform.
- The app communicates through the cloud to another software device
DEXCOM Partner Web APIs
The newly added software component, which consists of cloud-based application programming interfaces (APIs) enables communication of iCGM data to client software intended to receive that data through the cloud. The transmitted data can be used by authorized client software for specific and permitted use cases including non-medical device application, medical device data analysis, CGM secondary display alarm, active patient monitoring, and treatment decisions. The Partner Web APIs is not intended to be used by automated insulin delivery systems (AID). Dexcom display devices (receiver and mobile app) continue to serve as a primary display device for the (iCGM) data, which directly receives the data from the transmitter. The receiver and/or mobile app also continues to alert the user when glucose levels are outside of a target zone.
B Principle of Operation:
The Dexcom G6 CGM System uses a subcutaneous sensor to detect glucose levels from the fluid just beneath the skin (interstitial fluid). The sensor probe continuously measures the glucose concentration in the interstitial fluid via an enzymatic electrochemical reaction using glucose oxidase. The enzyme, glucose oxidase, catalyzes the oxidation of glucose and produces hydrogen peroxide. The production of hydrogen peroxide generates an electrical current that is proportionate to the interstitial glucose concentration. The transmitter converts the signal using an algorithm to a glucose value read in $\mathrm{mg/dL}$, the data is transmitted to a compatible display device for display to a user.
C Instrument Description Information:
1. Instrument Name: Dexcom G6 Continuous Glucose Monitoring System
2. Specimen Identification: Not applicable.
3. Specimen Sampling and Handling: Not applicable.
K201328 - Page 4 of 10
{4}
4. Calibration:
The Dexcom G6 CGM System does not require user calibration; however, users of this device have the option to calibrate the device manually (e.g., in situations where users do not use the calibration code, or in addition to code-based calibration).
5. Quality Control:
Not applicable.
V Substantial Equivalence Information:
A Predicate Device Name(s):
Dexcom G6 Continuous Glucose Monitoring System
B Predicate 510(k) Number(s):
K200876
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K201328 | K200876 |
| --- | --- | --- |
| Device Trade Name | Dexcom G6 Continuous Glucose Monitoring System | Dexcom G6 Continuous Glucose Monitoring System |
| General Device Characteristic Similarities | | |
| Intended Use | Same | An integrated continuous glucose monitoring system (iCGM) is intended to automatically measure glucose in bodily fluids continuously or frequently for a specified period of time. iCGM systems are designed to reliably and securely transmit glucose measurement data to digitally connected devices, including automated insulin dosing systems, and are intended to be used alone or in conjunction with these digitally connected medical devices for the purpose of managing a disease or condition related to glycemic control. |
| Indications For Use | Same | The Dexcom G6 Continuous Glucose Monitoring System (Dexcom G6 System) is a real |
K201328 - Page 5 of 10
{5}
K201328 - Page 6 of 10
| Device & Predicate Device(s): | K201328 | K200876 |
| --- | --- | --- |
| | | time, continuous glucose monitoring device indicated for the management of diabetes in persons age 2 years and older. The Dexcom G6 System is intended to replace fingerstick blood glucose testing for diabetes treatment decisions. Interpretation of the Dexcom G6 System results should be based on the glucose trends and several sequential readings over time. The Dexcom G6 System also aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. The Dexcom G6 System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The Dexcom G6 System can be used alone or in conjunction with these digitally connected medical devices for the purpose of managing diabetes. |
| General Device Characteristic Differences | | |
| Partner Web Application Programming Interfaces (APIs) | Addition of the Application Programming Interface (API) to enable communication of iCGM data to client software intended to receive that data through the cloud. The Partner Web APIs is not intended to be used by | |
{6}
| Device & Predicate Device(s): | K201328 | K200876 |
| --- | --- | --- |
| | automated insulin delivery (AID) systems. | |
VI Standards/Guidance Documents Referenced:
ISO 11137 - 1:2006/ Al :2013, Sterilization of health care products - Radiation, Part 1- Requirements for development, validation and routine control of a sterilization process for medical devices
ISO 11137-2:2 013, Sterilization of health care products - Radiation, Part 2 – Establishing the sterilization dose
ISO 11137-3:2006, Sterilization of health care products - Radiation, Part 3 - Guidance on dosimetric aspects
ISO 11737 -1:2006, Sterilization of Medical Devices - Microbiological Methods, Part 1 - Determination of a population of microorganisms on products
ISO 11737-2:2 00 9, Sterilization of Medical Devices - Microbiological Methods, Part 2 Tests of sterility performed in the definition, validation and maintenance of a sterilization process
ISO/TS 13004:2013, Sterilization of health care products - Radiation - Substantiation of selected sterilization dose: Method VDmaxSD
ISO 10993-1:2009/AC:2010, Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process
ISO 10993-2:2006, Biological Evaluation of Medical Devices - Part 2: Animal welfare requirements
ISO 10993-3: 2014 Biological Evaluation of Medical Devices - Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity
ISO 10993 -5:2009, Biological Evaluation of Medical Devices - Part 5: Tests for in vitro toxicity.
ISO 10993-6 :2016, Biological evaluation of Medical Devices - Part 6: Tests for local effects after implantation
ISO 10993-10:2010, Biological Evaluation of Medical Devices - Part 10: Tests for irritation and delayed-type hyper sensitivity
ISO 10993-11:2009, Biological Evaluation of Medical Devices - Part 11: Tests for systemic toxicity
K201328 - Page 7 of 10
{7}
ISO 10993-12:2012, Biological Evaluation of Medical Devices - Part 12: Sample preparation and reference materials
ISO 13485:2016 Medical devices - Quality management systems - Requirements for regulatory purposes
ISO 14971:2012 Medical devices - Application of risk management to medical devices
IEC 62304, Ed 1.1:2015 Medical Device Software - Software Lifecycle Processes (2006+ AMD I:2015)
## VII Performance Characteristics (if/when applicable):
### A Analytical Performance:
iCGM performance was previously evaluated in clinical studies described in DEN170088. The subject device uses the same updated glucose algorithm as cleared in K200876. Therefore, the same performance data applies for the proposed device.
1. Precision/Reproducibility: See k200876.
2. Linearity: The reportable range for the System is 40 to 400 mg/dL.
3. Analytical Specificity/Interference: Interference was previously assessed in DEN170088.
4. Assay Reportable Range: See Linearity section above.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): The sensor has a storage shelf-life of 12 months. Shelf life was evaluated at 32°-86° F and 10-90% relative humidity. Sensors should be stored at 32°-86° F.
The Dexcom G6 CGM System transmitter has sufficient battery life to function for 3 months as intended following its maximum storage time of 8 months. Shelf life was evaluated at 32°-113° F and 10-95% relative humidity.
6. Detection Limit: If a glucose measurement is less than 40 mg/dL, the result is displayed by the system as 'Lo'. If a glucose measurement exceeds 400 mg/dL, result is displayed as 'Hi'.
7. Assay Cut-Off: Not applicable.
K201328 - Page 8 of 10
{8}
8. Accuracy (Instrument): Not applicable.
9. Carry-Over: Not applicable.
## B Comparison Studies:
1. Method Comparison with Predicate Device: See k200876.
2. Matrix Comparison: Not applicable. Interstitial fluid is the only indicated matrix.
## C Clinical Studies:
1. Clinical Sensitivity: Not applicable.
2. Clinical Specificity: Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): See k200876.
## D Clinical Cut-Off:
Not applicable.
## E Expected Values/Reference Range:
Not applicable.
## F Other Supportive Instrument Performance Characteristics Data:
The following supportive instrument performance characteristics were established in the predicate for the Dexcom G6 System (k200876), and are not affected by the modifications in glucose algorithm in the current 510(k):
- Sterility
- Biocompatibility
- Mechanical Engineering
- Electromagnetic Compatibility
- Wireless
- Electrical Safety
- Environmental Testing
- Shelf-Life Stability
- Packaging Integrity/Shipping Integrity
- Contact Resistance
K201328 - Page 9 of 10
{9}
The following performance characteristics were verified or validated through studies of the current design:
- Human Factors
- Interoperability
- Cybersecurity
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K201328 - Page 10 of 10
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.