Afinion HbA1c Dx on Afinion 2
Device Facts
| Record ID | K182988 |
|---|---|
| Device Name | Afinion HbA1c Dx on Afinion 2 |
| Applicant | Alere Technologies AS |
| Product Code | PDJ · Clinical Chemistry |
| Decision Date | Nov 29, 2018 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 862.1373 |
| Device Class | Class 2 |
Intended Use
The AfinionTM A1c Dx is a multi-assay analyzer for in vitro diagnostic use. The AfinionTM A1c Dx is a compact, rapid, multi-assay analyzer that provides in vitro diagnostic tests. The AfinionTM A1c Dx is intended for professional use in clinical settings. The AfinionTM A1c Dx is intended for use by healthcare professionals.
Device Story
Afinion 2 is a compact, point-of-care, multi-assay analyzer for quantitative HbA1c measurement. It processes Afinion test cartridges containing human venous or capillary whole blood. The device performs automated boronate affinity assay processing, including pumping, heating, and image processing. It physically manipulates cartridges via locking, docking, splitting, foil penetration, and merging. The system provides HbA1c results to clinicians to assist in diabetes diagnosis and metabolic monitoring. It is designed for use in clinical settings by healthcare professionals. The Afinion 2 replaces the AS100 analyzer, featuring updated electronics, injection-molded plastic components, and integrated connectivity, while maintaining the same assay sequence timing and core functionality as the predicate.
Clinical Evidence
Bench testing only. Verification and validation studies were performed based on risk analysis to confirm that the modified device (Afinion 2) maintains the performance and safety of the Afinion system. All acceptance criteria were met.
Technological Characteristics
Compact multi-assay analyzer; boronate affinity assay principle; automated cartridge processing (pumping, heating, image processing); injection-molded plastic and metal components; 9 PCBs; integrated Ethernet connectivity; traceable to IFCC reference method; NGSP certified.
Indications for Use
Indicated for quantitative determination of HbA1c in human venous and capillary whole blood to aid in diabetes diagnosis, identify at-risk patients, and monitor long-term metabolic control in patients with diabetes mellitus.
Regulatory Classification
Identification
A hemoglobin A1c test system is a device used to measure the percentage concentration of hemoglobin A1c in blood. Measurement of hemoglobin A1c is used as an aid in the diagnosis of diabetes mellitus and as an aid in the identification of patients at risk for developing diabetes mellitus.
Special Controls
(b) Classification. Class II (special controls). Hemoglobin A1c test systems must comply with the following special controls: 1) The device must have initial and annual standardization verification by a certifying glycohemoglobin standardization organization deemed acceptable by FDA. 2) The premarket notification submission must include performance testing to evaluate precision, accuracy, linearity and interference, including the following: i) Performance testing of device precision must, at a minimum, use blood samples with concentrations near 5.0%, 6.5%, 8.0% and 12% hemoglobin A1c. This testing must evaluate precision over a minimum of 20 days using at least 3 lots of the device and 3 instruments, as applicable. ii) Performance testing of device accuracy must include a minimum of 120 blood samples that span the measuring interval of the new device and compare results of the new device to results of the standardized test method. Results must demonstrate little or no bias versus the standardized method. iii) Total error of the new device must be evaluated using single measurements by the new device compared to results of the standardized test method, and this evaluation must demonstrate a total error less than or equal to 6%. iv) Performance testing must demonstrate that there is little to no interference from common hemoglobin variants, including Hemoglobin C, Hemoglobin D, Hemoglobin E, Hemoglobin A2 and Hemoglobin S. 3) When assay interference from Hemoglobin F or interference with other hemoglobin variants with low frequency in the population is observed, a warning statement must be placed in a black box and must appear in all labeling material for these devices describing the interference and any affected populations.
*Classification.* Class II (special controls). The special controls for this device are:(1) The device must have initial and annual standardization verification by a certifying glycohemoglobin standardization organization deemed acceptable by FDA. (2) The premarket notification submission must include performance testing to evaluate precision, accuracy, linearity, and interference, including the following: (i) Performance testing of device precision must, at a minimum, use blood samples with concentrations near 5.0 percent, 6.5 percent, 8.0 percent, and 12 percent hemoglobin A1c. This testing must evaluate precision over a minimum of 20 days using at least three lots of the device and three instruments, as applicable. (ii) Performance testing of device accuracy must include a minimum of 120 blood samples that span the measuring interval of the device and compare results of the new device to results of a standardized test method. Results must demonstrate little or no bias versus the standardized method. (iii) Total error of the new device must be evaluated using single measurements by the new device compared to results of the standardized test method, and this evaluation must demonstrate a total error less than or equal to 6 percent. (iv) Performance testing must demonstrate that there is little to no interference from common hemoglobin variants, including Hemoglobin C, Hemoglobin D, Hemoglobin E, Hemoglobin A2, and Hemoglobin S. (3) When assay interference from Hemoglobin F or interference with other hemoglobin variants with low frequency in the population is observed, a warning statement must be placed in a black box and must appear in all labeling material for these devices describing the interference and any affected populations.
Predicate Devices
- Afinion™ HbA1c Dx on Alere Afinion™ AS100 Analyzer (k180296)
Related Devices
- K171650 — Alere Afinion HbA1c and ACR on Afinion 2 analyzer · Abbott Rapid Diagnostics · Nov 21, 2017
- K214117 — Afinion HbA1c, Afinion 2 and Alere Afinion AS100 Analyzer · Abbott Diagnostics Technologies AS · Sep 27, 2023
- K050574 — AFINION HBALC · Axis-Shield Poc AS · Jul 22, 2005