SD GlucoNavii Mentor BT Blood Glucose Monitoring System, SD GlucoNavii Mentor BT Multi Blood Glucose Monitoring System
Applicant
Sd Biosensor
Product Code
NBW · Clinical Chemistry
Decision Date
Jun 24, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The SD GlucoNavii Mentor BT Blood Glucose Monitoring System is intended for single-patient use. The SD GlucoNavii Mentor BT Multi Blood Glucose Monitoring System is intended for multiple-patient use in a professional healthcare setting.
Device Story
Blood glucose monitoring system; measures glucose levels via test strips; provides digital readout. Modification adds Bluetooth for wireless data transmission to mobile devices; reduces internal memory capacity from 300 to 100 records. Used by patients for self-monitoring or professionals in clinical settings. Output enables glucose tracking and clinical decision-making regarding diabetes management. Robustness validated for 3-year multi-patient and 5-year single-patient use cycles.
Clinical Evidence
No clinical data. Bench testing only. Robustness studies confirmed no degradation in performance or materials after 10,950 cleaning/disinfection cycles using Discide Ultra Disinfecting towlettes, simulating 3-5 years of use.
Technological Characteristics
Glucose meter with integrated Bluetooth wireless module. Memory capacity: 100 records. Materials validated for repeated cleaning/disinfection with Discide Ultra (EPA Reg No. 10492-4).
Indications for Use
Indicated for quantitative blood glucose measurement in fresh capillary whole blood (fingertip, palm, forearm, upper arm) in patients with diabetes. Not for neonates. Not for diagnosis/screening of diabetes. Single-patient version for home use; multi-patient version for professional healthcare settings (requires auto-disabling lancing devices).
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
SD GlucoNavii® Mentor NFC Blood Glucose Monitoring System and SD GlucoNavii® Mentor NFC Multi Blood Glucose Monitoring System (k132929)
Related Devices
K142785 — PRODIGY iConnect Blood Glucose Monitoring System · Ok Biotech Co., Ltd. · Dec 17, 2015
K191657 — Bioland Blood Glucose Monitoring System · Bioland Technology, Ltd. · Mar 20, 2020
K152534 — Medisign MM1000 BT Blood Glucose Monitoring System, Medisign MM1100 BT Blood Glucose Monitoring System, Medisign MM1200 BT Blood Glucose Monitoring System, Smart Diabetes Bluetooth Blood Glucose Monitoring System, Medisign MM 1000 BT MULTI Blood Glucose Monitoring System, Medisign MM 1100 BT MULTI Blood Glucose Monitoring System, Medisign MM 1200 BT MULTI Blood Glucose Monitoring System, Smart Diabetes Bluetooth Pro Blood Glucose Monitoring System · Tianjin Empecs Medical Device Co., Ltd. · Mar 21, 2017
K170143 — GE Blood Glucose Monitoring System 180, GE Blood Glucose Monitoring System 182, Rightest Blood Glucose Monitoring System GM280, Rightest Blood Glucose Monitoring System GM280B, · Bionime Corporation · Jul 6, 2017
K180196 — On Call Express Mobile Blood Glucose Monitoring System · ACON Laboratories, Inc. · Mar 7, 2018
Submission Summary (Full Text)
{0}
SPECIAL 510(k): Device Modification OIR Decision Memorandum
To: THE FILE
RE: DOCUMENT NUMBER k153201
This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary):
1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.)
SD GlucoNavii® Mentor NFC Blood Glucose Monitoring System and SD GlucoNavii® Mentor NFC Multi Blood Glucose Monitoring System (k132929)
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use).
3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.
This change was for the following items:
a. The addition of Bluetooth for wireless transmission of data to a mobile device.
b. Memory capacity for stored glucose test results decreased from 300 to 100 records.
c. The trade name of the systems has changed from SD GlucoNavii Mentor NFC Blood Glucose Monitoring System and SD GlucoNavii Mentor NFC Multi Blood Glucose Monitoring System to SD GlucoNavii Mentor BT Blood Glucose Monitoring System and SD GlucoNavii Mentor BT Multi Blood Glucose Monitoring System.
4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, and specifications.
5. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
The SD GlucoNavii Mentor BT Blood Glucose Monitoring System is intended for single-patient use. The SD GlucoNavii Mentor BT Multi Blood Glucose Monitoring System is intended for multiple-patient use in a professional healthcare setting. There were no physical changes to the subject device relative to the predicate that would warrant new disinfection efficacy studies. Disinfection efficacy studies were performed on the materials comprising the meter for the predicate device (k132929) by an outside
{1}
Page 2 of 2
commercial testing laboratory and demonstrated complete inactivation of Hepatitis B Virus (HBV) with the chosen disinfectant, Discide Ultra Disinfecting towlettes (EPA Registration No. 10492-4). Robustness studies were performed on the subject device to demonstrate that there were no changes in performance or external materials of the meter after 10,950 cleanings and 10,950 disinfection steps with Discide Ultra Disinfecting towlettes. The robustness studies were designed to simulate 3 years of multi-patient use and 5 years of single-patient use. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.