FreeStyle Precision Neo Blood Glucose Monitoring System

K142928 · Abbott Diabetes Care, Inc. · NBW · Sep 30, 2015 · Clinical Chemistry

Device Facts

Record IDK142928
Device NameFreeStyle Precision Neo Blood Glucose Monitoring System
ApplicantAbbott Diabetes Care, Inc.
Product CodeNBW · Clinical Chemistry
Decision DateSep 30, 2015
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Indications for Use

The FreeStyle Precision Neo Blood Glucose Monitoring System is intended for single-patient use only.

Device Story

Device is a blood glucose monitoring system for home use by patients with diabetes. Input: fresh capillary whole blood sample applied to a test strip. Principle: amperometric technology; glucose dehydrogenase (GDH) enzyme on strip reacts with glucose; electrons transferred via mediator to meter; current generated is proportional to glucose concentration. Output: quantitative blood glucose result displayed on meter screen. Includes hypoglycemia/hyperglycemia trend indicators; alerts user via red/yellow arrow lights if results fall outside customizable target ranges. Data used by patients to monitor glucose levels and manage diabetes. System is for single-person use only.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on technological characteristics and performance specifications identical to the predicate device.

Technological Characteristics

Blood glucose meter; user-settable software-based hypoglycemic/hyperglycemic indicator feature. No changes to materials or fundamental sensing technology from predicate (K140371).

Indications for Use

Indicated for single-patient use for blood glucose monitoring.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} SPECIAL 510(k): Device Modification OIR Review Summary To: THE FILE RE: DOCUMENT NUMBER K 142928 This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable: 1. The name and 510(k) number of the SUBMITTER'S previously cleared device: FreeStyle Precision Neo Blood Glucose Monitoring System (k140371). 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. This change was for the addition of a user settable hypoglycemic (hypo) and hyperglycemic (hyper) indicator feature for the FreeStyle Precision Neo Blood Glucose Monitoring System. The Sponsor claims that the above changes in new device will not influence substantial equivalence (SE) to its predicate device. 4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, and physical characteristics. 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device. The FreeStyle Precision Neo Blood Glucose Monitoring System is intended for single-patient use only. Disinfection efficacy studies were previously performed in the 510(k) k140371. Robustness studies were also performed previously (k140371) by the sponsor demonstrating that there was no change in performance or external materials for each of the meters after 522 cleaning and 522 disinfection steps designed to simulate 2 cleaning and disinfection cycles each week for 5 years of single-patient use of using the Clorox Healthcare Bleach Germicidal wipes. The modification (addition of a user settable hypoglycemic (hypo) and hyperglycemic (hyper) indicator) does not affect the exterior {1} Page 2 of 2 features of the meter, therefore new cleaning/disinfection and robustness studies were not necessary. The labeling adequately notifies the user to the limitations of the device and proper interpretation of the test results.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...