TD-4257 BLOOD GLUCOSE MONITORING SYSTEMS,TD 4257 MULTI BLOOD GLUCOSE MONITORING SYSTEMS
Applicant
Taidoc Technology Corporation
Product Code
NBW · Clinical Chemistry
Decision Date
Nov 22, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The TD-4257 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood samples from the finger. It is intended to be used by a single person and should not be shared. The TD-4257 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus or be used on neonates. The TD-4257 Blood Glucose Test Strips are for use with the TD-4257 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples. The TD-4257 Multi Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary, venous and neonatal whole blood samples. The TD-4257 Multi Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple patient use in professional healthcare settings as an aid in monitoring the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus. Professionals may test with capillary, venous and neonatal whole blood. Capillary samples may be drawn from the fingertip, and in the case of neonates, from the heel. The system is only used with single-use, auto-disabling lancing devices. The TD-4257 Multi Blood Glucose Test Strips are for use with the TD-4257 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary, venous and neonatal whole blood samples.
Device Story
System measures glucose concentration in fresh capillary, venous, or neonatal whole blood; utilizes amperometric assay with FAD-dependent glucose dehydrogenase (GDH) enzyme on test strips. User applies blood sample to strip; meter detects electrical current proportional to glucose concentration; displays result in mg/dL. Modified device features updated physical housing, increased memory (1000 tests), and Bluetooth data transmission to PC software. Used in home (single-patient) or professional clinical settings (multi-patient). Healthcare providers or patients use output to monitor diabetes management effectiveness. Benefits include rapid, quantitative glucose monitoring to aid in glycemic control.
Clinical Evidence
Bench testing only. Software verification and validation confirmed performance, safety, and effectiveness equivalent to predicate device. System accuracy performance compared to predicate.
Technological Characteristics
Electrochemical biosensor using glucose dehydrogenase. Connectivity: Bluetooth. Meter form factor modified from predicate. Software: minor modifications to glucose meter firmware.
Indications for Use
Indicated for quantitative blood glucose measurement in people with diabetes (home use) or professional healthcare settings. Supports capillary, venous, and neonatal whole blood samples. Not for diabetes diagnosis or screening.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
Taidoc, TD-4239 and TD-4239 Multi Blood Glucose Monitoring Systems (k101635)
Related Devices
K120042 — TD-4268 BLOOD GLUCOSE MONITORING SYSTEM TD-4268 MULTI BLOOD GLUCOSE MONITORING SYSTEM · Taidoc Technology Corporation · May 22, 2012
K101635 — TD-4239 BLOOD GLUCOSE MONITORING SYSTEM AND TD-4239 MULTI BLOOD GLUCOSE MONITORING SYSTEM · Taidoc Technology Corporation · Aug 31, 2011
K182286 — Oh'Care Lite Smart Blood Glucose Monitoring System · Osang Healthcare Co. , Ltd. · Sep 19, 2018
K103044 — AP-1010 AND AP-1020 BLOOD GLUCOSE MONITORING SYSTEM, MAJOR GLUCOSE CONTROL SOLUTION · Bestgen Biotech Corp. · Feb 10, 2012
K080283 — IDEAL LIFE GLUCO-MANAGER GMM 0001 BLOOD GLUCOSE MONITORING SYSTEM · Taidoc Technology Corporation · Jun 27, 2008
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k112963
B. Purpose for Submission:
Modified device (The manufacturer made the following modifications to their predicates: physical appearance, data transmission from RS-232 to Bluetooth and minor software modifications).
C. Measurand:
Whole blood glucose
D. Type of Test:
Whole blood glucose concentration through a quantitative amperometric assay (GDH-FAD)
E. Applicant:
Taidoc Technology Corporation
F. Proprietary and Established Names:
TD-4257 Blood Glucose Monitoring System
TD-4257 Multi Blood Glucose Monitoring System
G. Regulatory Information:
1. Regulation section:
21 CFR: 862.1345, Blood Glucose Test System
2. Classification:
Class II
3. Product code:
NBW, LFR
{1}
4. Panel:
75 (clinical chemistry)
H. Intended Use:
1. Intended use(s):
Same as indications for use below.
2. Indication(s) for use:
TD-4257 Blood Glucose Monitoring System
The TD-4257 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood samples from the finger. It is intended to be used by a single person and should not be shared.
The TD-4257 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus or be used on neonates.
The TD-4257 Blood Glucose Test Strips are for use with the TD-4257 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples.
TD-4257 Multi Blood Glucose Monitoring System
The TD-4257 Multi Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary, venous and neonatal whole blood samples. The TD-4257 Multi Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple patient use in professional healthcare settings as an aid in monitoring the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus.
Professionals may test with capillary, venous and neonatal whole blood. Capillary samples may be drawn from the fingertip, and in the case of neonates, from the heel.
The system is only used with single-use, auto-disabling lancing devices
The TD-4257 Multi Blood Glucose Test Strips are for use with the TD-4257 Multi Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary, venous and neonatal whole blood samples.
3. Special conditions for use statement(s):
- Not intended for diagnosis of or screening for diabetes mellitus
- For in vitro diagnostic use only
2
{2}
- Not for use on critically ill patients or dehydrated patients
- Not for use on severely hypotensive individuals or patients in shock.
- Not for use on individuals experiencing a hyperglycemic-hyperosmolar state, with or without ketosis.
- For TD-4257 Multi Blood Glucose Monitoring System; only Heparin should be used as an anticoagulant
- The TD-4257 Multi Blood Glucose Monitoring System must be disinfected between users following labeling recommendations.
- Only single use, auto-disabling lancing devices can be used with the TD-4257 Multi Blood Glucose Monitoring System
- Single-patient use devices are for use on single-patients only and should not be shared.
- The TD-4257 Blood Glucose Monitoring System is not intended for use on neonates.
4. Special instrument requirements:
TD-4257 Blood Glucose meter
TD-4257 Multi Blood Glucose meter
I. Device Description:
The TD-4257 Blood Glucose Monitoring Systems/TD-4257 Multi Blood Glucose Monitoring System is supplied as a kit which consists of:
- A glucose meter
- Test strips
- A storage case
Control solutions, and test strips may be supplied separately to the kit. The test strips are identical to the test strips cleared under k101635. The control solutions are cleared under k093724.
The modified device has the same intended use and method of glucose measurement comparing to the predicate (TD-4239 and the TD-4239 Multi Blood Glucose Monitoring System, cleared under k101635). The current devices differ from the predicates in the modification of meter physical appearance and data transmission method from data port to blue tooth.
Users have the option of uploading the stored test results into a personal computer (PC) through a Bluetooth connection with program pre-installed in their PC, Health Care System Software that has received clearance under K070941.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Taidoc, TD-4239 and TD-4239 Multi Blood Glucose Monitoring Systems
{3}
2 Predicate 510(k) number(s):
k101635
3. Comparison with predicate:
| Item | Predicate devices | Proposed devices |
| --- | --- | --- |
| Brand name | TD-4239/ TD-4239 Multi Blood Glucose Monitoring Systems | TD-4257/ TD4257 Multi Blood Glucose Monitoring Systems |
| Model no | TD-4239 | TD-4257 |
| Similarities | | |
| Indications for use (single patient use) | The TD-4239 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood samples from the finger. It is intended to be used by a single person and should not be shared.The TD-4239 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus or be used on neonates. | Same |
| Indications for use (Multi-patient use) | The TD-4239 Multi Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary, venous and neonatal whole blood samples. The TD-4239 Multi Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple patient use in professional healthcare settings as an aid in monitoring the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus.Professionals may test with capillary, venous and neonatal whole blood. | Same |
{4}
| | Capillary samples may be drawn from the fingertip, and in the case of neonates, from the heel.
The system is only used with single-use, auto-disabling lancing devices. | |
| --- | --- | --- |
| Detection mechanism | | |
| Detection method | Amperometry: measuring a current produced by a chemical reaction | Same |
| Enzyme | Glucose dehydrogenase (FAD-GDH) | Same |
| Specifications | | |
| Sample volume (μL) | 1.1 μL | Same |
| Reaction time (sec) | 5 seconds | Same |
| Measurement unit | mg/dL | Same |
| Measurement range | 20-600 mg/dL (single patient use)
10-600 mg/dL (Multi-patient use) | Same |
| Meter Storage/Transportation condition | -4°F – 140°F (-20°C - 60°C),
below 95% R.H. | Same |
| Functions | | |
| Power saving | Auto turn-off after 3 minutes without action | Same |
| Code Display | Code number is displayed on screen. | Same |
| Measuring mode | General, AC, PC and QC (quality control) | Same |
| Day average | 7-, 14-, 21-, 28-, 60- and 90-day average glucose result | Same |
| Alarm function | 4 alarms | Same |
| Strip ejection function | Yes | Same |
| Special message | Lo/Hi/Ketone? | Same |
| Test Strip Calibration | Automatic calibration by code strip | Same |
| Test Strip | | |
| Detection method | Amperometry: measuring a current produced by a chemical reaction | Same |
5
{5}
| Enzyme | Glucose dehydrogenase | Same |
| --- | --- | --- |
| Blood Volume | 1.1 μL | Same |
| Reaction time | 5 seconds | Same |
| Strip Storage/Transportation condition | 36°F- 90°F (2°C-32°C) below 85% R.H. | Same |
| Differences | | |
| Device | TD-4239/ TD-4239 Multi Blood Glucose Monitoring System | TD-4257/ TD4257 Multi Blood Glucose Monitoring System |
| Size (mm) Length X width X height | 95 mm (L) x 49 mm (W) x 14 mm (H) | 102mm(L) x 64 mm (W) x 29.5 mm (H) |
| Weight (g) | 42 g (without batteries) | 81g (without batteries) |
| Power source | One 3V CR2032 lithium battery | 2 x 1.5V AAA batteries |
| Backlight | No | Yes |
| Memory feature | 400 measurements | 1000 measurements |
| Transmission Function | Uses RS232 cable to transmit data to computer | Uses Bluetooth to transmit data to computer |
| User interface | M button, C button, SET button, test strip ejector, data port | M button, SET button, test strip ejector |
# K. Standard/Guidance Document Referenced (if applicable):
- ISO 14971:2007. Medical devices-Application of risk management to medical devices.
- ISO 15197. In vitro diagnostic test systems. Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus.
# L. Test Principle:
Same as predicate devices cleared under k101635.
The glucose biosensors recognize the glucose present in whole blood or control solutions by virtue of the specificity of the enzyme FAD dependent glucose dehydrogenase (GDH) present on the test strip. The electrons liberated by this reaction are transferred via a cofactor and mediator to the meter where they are read as a small electrical current. The
{6}
magnitude of the resultant current is proportional to the concentration of glucose in the specimen and the signal is converted into a readout displayed on the meter.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
The current devices differ from the predicates in the modification of meter physical appearance and data transmission method from data port to blue tooth. The modifications do not affect the analytical function of the meter and the current devices use the same test strips as cleared in the predicate devices in k101635.
#### a. Precision/Reproducibility:
Same as predicate devices cleared under k101635.
#### b. Linearity/assay reportable range:
The sponsor claims the same measuring range as established for the predicate devices. The sponsor claims 20-600 mg/dL for the TD-4257 Blood Glucose Monitoring System and 10-600 mg/dL for the TD-4257 Multi Blood Glucose Monitoring System.
#### c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability: Traceable to NIST SRM 91, dry D-glucose.
The control solutions were cleared under k093724.
Test strips were cleared under k101635. Data provided by the sponsor supports closed vial stability of TD-4239 test strips for 18 months and open vial stability of 3 months when stored at the recommended conditions of 2° - 32°C / 35.6° - 89.6°F, between 10%-85% R.H.
Calibration: calibration of the test strip is done by inserting the code strip into the meter when opening a new vial of test strips.
#### d. Detection limit:
The claimed measuring ranges are the same as the predicate devices; 20-600 mg/dL for the TD-4257 Blood Glucose Monitoring System and 10-600 mg/dL for the TD-4257 Multi Blood Glucose Monitoring System.
#### e. Analytical specificity:
Interference Study:
{7}
Same as predicate devices cleared under k101635.
The sponsor has the following limitations in their labeling:
Xylose: Do not test blood glucose during or soon after an absorption test. Xylose can give falsely elevated results.
There is no significant interference $(\leq 10\%)$ in the presence of galactose, maltose, fructose or mannitol observed in blood glucose tests as demonstrated in studies up to $1,000~\mathrm{mg / dL}$
Lipemic Effects: Blood triglycerides up to $2000\mathrm{mg / dL}$ (22.8 mmol/L) do not affect the results significantly $(\leq 10\%)$ , but may affect results at higher levels.
f. Assay cut-off:
Not Applicable
# 2. Comparison studies:
a. System accuracy as compared to YSI
The system accuracy was evaluated using 126 whole blood samples with glucose concentrations across the measuring range. The distribution of the glucose concentrations in these samples are depicted in the below table:
| Glucose Concentration by YSI ng/mL | Number of samples |
| --- | --- |
| 20-50 | 21 |
| 50-80 | 21 |
| 80-120 | 21 |
| 120-200 | 21 |
| 201-400 | 21 |
| 401-600 | 21 |
Each sample was tested in duplicate on three meters and the results are compared to the YSI method.
Result Summary:
Glucose $\leq 75\mathrm{mg / dL}$ (42 samples)
{8}
| Number of test result | Within ± 5 mg/dL | Within ± 10 mg/dL | Within ± 15 mg/dL |
| --- | --- | --- | --- |
| 252 | 100% (252/252) | 100% (252/252) | 100% (252/252) |
Glucose > 75 mg/dL (84 Samples)
| Number of test result | Within ± 5 mg/dL | Within ± 10 mg/dL | Within ± 15 mg/dL |
| --- | --- | --- | --- |
| 504 | 71% (358/504) | 100% (504/504) | 100% (504/504) |
Summary Assessment of Accuracy (126 Samples)
| Number of test result | Readings within ±15 mg/dL if <75 mg/dL, ±20% if ≥75 mg/dL |
| --- | --- |
| 756 | 100% (756/756) |
Conclusion:
The performance of the TD-4257 and the TD-4257 Multi Blood Glucose Monitoring System meets ISO 15197 accuracy requirement of more than 95% results within the acceptance limits of ±15 mg/dL if <75 mg/dL or ±20% if ≥75 mg/dL.
b. Matrix comparison:
The data was presented and reviewed in predicate devices cleared in k101635.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
The sponsor provides the following information in the labeling:
Reference values:
{9}
| Time of day | Normal plasma glucose range for people without diabetes |
| --- | --- |
| Fasting and before meal | Less than 100 mg/dL (5.6 mmol/L) |
| 2 hours after meals | Less than 140 mg/dL (7.8 mmol/L) |
Source: American Diabetes Association (2010). Clinical Practice Recommendations. Diabetes Care, 33 (Supplement 1): S1-S100.
N. Instrument Name:
TD-4257 Blood Glucose Meter,
TD-4257 Multi Blood Glucose Meter
O. System Descriptions:
1. Modes of Operation:
Each test strip is single use and must be replaced with a new strip for additional readings.
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
☐ Yes ☑ X or No
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
☐ Yes ☑ X or No
2. Software:
FDA reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
☐ Yes ☑ X or No
3. Specimen Identification:
There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected.
4. Specimen Sampling and Handling:
Professionals may test fingertip capillary and venous, and neonatal blood, while home-use is limited to fingertip capillary whole blood testing only.
5. Calibration:
{10}
These systems need to be calibrated with the code strip from every new vial of test strips
6. Quality Control:
Glucose control solutions are cleared under k 093724. The labeling provides instructions on when to test control solutions.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
1. An accuracy test using the Quality Control current board (QC board) was conducted to evaluate the accuracy of 10 TD-4257 meters in glucose range of 10-700 mg/dL.
The QC board is designed to simulate 12 levels of glucose concentration: 10, 20, 40, 60, 80, 100, 120, 160, 200, 280, 400, 550 and 700 mg/dL when connected with the meter, it will send an electrical current to the meter equivalent to the total amount of charge generated by the glucose oxidation reaction at a certain glucose concentration. The meter then process the current and display it in mg/dL unit.
The same procedure was also performed on 10 predicate meters of TD-4239. For the current device TD4257, the CV of the mean of 10 meter measurements was between 0.00%-0.40% at the 13 glucose levels. For the predicate meter TD-4239, the CV of the mean of 10 meter measurements was between 0.00%-0.32% at the 13 glucose levels. The sponsor concluded that the accuracy of the modified TD-4257 meter was consistent through out the 13 glucose intervals with CV significantly below the set acceptance criteria of 5%.
2. Altitude:
The sponsor claims that the TD 4257 BMGS can be used at altitude up to 15,000 feet based on study performed on the predicate device TD4239 in k101635.
3. Hematocrit:
The sponsor claims a hematocrit range of 20-70% based on study performed on the predicate device TD4239 in k101635.
4. Temperature and humidity:
The sponsor claims that the TD-4257 meters operates at 50 -104 °F (10°C - 40°C), and relative humidity between 10%-85% based on study performed on the predicate meters k101635.
5. Drop tests and vibration tests were conducted. The study protocol and test report were reviewed and deemed acceptable.
6. EMC testing was performed for TD4257 and TD4257 multi Blood Glucose Monitoring Systems by an accredited EMC Testing Laboratory ETC. A test certificate was issued to Taidoc Technology Corporation on December 28, 2009.
11
{11}
7. Disinfection and Robustness studies:
Due to the casing changes in the new devices as compared to the predicates, new Disinfection and Robustness studies were performed for TD4257 and TD4257 multi Blood Glucose Monitoring Systems.
The disinfection studies were performed on these meters by an outside commercial testing service to determine the robustness of the meter to the recommended cleaning and disinfection protocol, and its effectiveness in preventing the spread of blood borne pathogens, particularly Duck hepatitis B virus (HBV). Micro-Kill Plus™ disposable wipes (EPA Reg. No: 59894-10-37549) were validated, demonstrating complete inactivation of live virus for use with the meters. The sponsor also demonstrated that there was no change in performance or in the external materials of the meters and lancing device after 10,000 cleaning and disinfection cycles designed to simulate 3 years of healthcare professional use and 5 years of use by lay users. Labeling has been reviewed for adequate instructions in validated cleaning and disinfection procedures.
8. The software changes related to the modified functions (Bluetooth data transmission and memory storage of up to 1000 tests) were tested and the intended functions were validated.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
12
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.