RIGHTEST BLOOD GLUCOSE MONITORING SYSTEM, MODEL GM110
Device Facts
| Record ID | K081450 |
|---|---|
| Device Name | RIGHTEST BLOOD GLUCOSE MONITORING SYSTEM, MODEL GM110 |
| Applicant | Bionime Corporation |
| Product Code | NBW · Clinical Chemistry |
| Decision Date | Sep 11, 2008 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 862.1345 |
| Device Class | Class 2 |
| Attributes | Pediatric |
Indications for Use
The Rightest Blood Glucose Monitoring System is intended for in vitro diagnostic use (outside of body). It is indicated to be used by professional healthcare personnel or diabetics at home to measure the glucose concentration for aiding diabetes management. The glucose concentration is measured with quantitative capillary whole blood from the fingertip, palm, and forearm by using Rightest Blood Glucose Monitoring System. This device is not intended for testing neonate blood samples.
Device Story
Rightest Blood Glucose Monitoring System (GM110) measures glucose concentration in fresh capillary whole blood; utilizes meter, test strips, control solutions, and lancing device. Principle of operation: amperometry using glucose oxidase (Aspergillus niger) and potassium ferricyanide mediator on noble metal electrodes. User applies blood sample (1.4 uL) to test strip; meter provides quantitative result in 8 seconds. Used by patients at home or healthcare professionals in clinical settings. Output displayed on meter; aids diabetes management decisions. System provides plasma-equivalent results.
Clinical Evidence
Clinical study evaluated alternative site testing (fingertip, palm, forearm) against Olympus AU2700 plasma reference. Sample size: 164 tests per site. Results showed high correlation (r=0.990 to 0.991) and similar slopes (0.95-0.98) and intercepts (2.13-9.80) across all sites. Bench testing confirmed precision, linearity, interference, and hematocrit range (30-55%).
Technological Characteristics
Electrochemical blood glucose meter; manual coding mechanism; mechanical housing; internal memory storage. Fundamental scientific technology consistent with predicate device.
Indications for Use
Indicated for professional healthcare personnel or diabetics at home to measure glucose concentration in capillary whole blood from fingertip, palm, or forearm for diabetes management. Contraindicated for neonate blood samples.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
- Rightest Blood Glucose Monitoring System (K042678)
- Rightest Blood Glucose Monitoring System (K053635)
- Rightest Blood Glucose Monitoring System (K062567)
Related Devices
- K072054 — RIGHTEST BLOOD GLUCOSE MONITORING SYSTEM, MODEL GM310 · Bionime Corporation · Feb 28, 2008
- K081451 — RIGHTEST BLOOD GLUCOSE MONITORING SYSTEM, MODEL GM100 · Bionime Corporation · Sep 18, 2008
- K042678 — RIGHTEST BLOOD GLUCOSE MONITORING SYSTEM · Bionime Corporation · Jan 3, 2005
- K062829 — DIACHEX BLOOD GLUCOSE MONITORING SYSTEM · Tyson Bioresearch, Inc. · Oct 10, 2006
- K062567 — RIGHTEST BLOOD GLUCOSE MONITORING SYSTEM · Bionime Corporation · Feb 16, 2007