EZ SMART

K040848 · Vip International Wholesalers, Corp. · NBW · Aug 11, 2004 · Clinical Chemistry

Device Facts

Record IDK040848
Device NameEZ SMART
ApplicantVip International Wholesalers, Corp.
Product CodeNBW · Clinical Chemistry
Decision DateAug 11, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Indications for Use

The EZ Smart Blood Glucose Test Strips are used with the EZ Smart Blood Glucose Meter to measure glucose (sugar) in whole blood. The EZ Smart Test Strips are for testing outside the body (in vitro diagnostic use). The EZ Smart Blood Glucose Monitoring System is intended for use in the home and in professional settings to monitor blood glucose levels. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.

Device Story

System measures glucose in whole blood samples using test strips and a meter. Principle of operation: glucose oxidase chemical reaction produces current measured by meter. Device displays glucose concentration directly without user calculation. Used in home and professional settings by patients or clinicians. Output provides blood glucose levels to assist in glucose control. Benefits include convenient, rapid monitoring of glucose levels.

Clinical Evidence

OTC clinical study with 203 patients; patients performed fingerstick tests unassisted. Compared EZ Smart results against Bayer Elite and YSI 2300 reference method. 91.1% of results were within 20% of YSI values. Precision studies (within-run and between-run) conducted across 5 glucose concentrations (30-400 mg/dL) showed CVs ranging from 1.17% to 5.35%.

Technological Characteristics

Glucose oxidase electrochemical biosensor. Measuring range: 20-600 mg/dL. Measurement time: 10 seconds. Dimensions: 94 x 49 x 17 mm. Standalone meter.

Indications for Use

Indicated for monitoring blood glucose levels in diabetics in home and professional settings. Not for neonates, diabetes diagnosis, or screening.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} Page 1 of 8 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: k040848 B. Purpose For Submission: Premarket Notification 510(k) of intention to manufacture and market the VIP International Wholesalers, Corp., EZ Smart Blood Glucose Monitoring System. C. Analyte: Glucose D. Type of Test: Quantitative electrochemical biosensor. E. Applicant: VIP International Wholesalers. Corp. F. Proprietary and Established Names: EZ Smart Blood Glucose Monitoring System G. Regulatory Information: 1. Regulation section: 21 CFR §862.1345, Glucose test system. 2. Classification: Class II 3. Product Code: NBW, CGA 4. Panel: 75 H. Intended Use: 1. Intended use(s): The EZ Smart Blood Glucose Test Strips are used with the EZ Smart Blood Glucose Meter to measure glucose (sugar) in whole blood. The EZ Smart Test Strips are for testing outside the body (in vitro diagnostic use). The EZ Smart Blood Glucose Monitoring System is intended for use in the home and in professional settings to monitor blood glucose levels for better glucose level control among diabetics. 2. Indication(s) for use: The EZ Smart Blood Glucose Test Strips are used with the EZ Smart Blood Glucose Meter to measure glucose (sugar) in whole blood. The EZ Smart Test Strips are for testing outside the body (in vitro diagnostic use). The EZ Smart Blood Glucose Monitoring System is intended for use in the home and in {1} Page 2 of 8 professional settings to monitor blood glucose levels for better glucose level control among diabetics. 3. Special condition for use statement(s): EZ Smart Blood Glucose Test Strips and EZ Smart Control Solutions are to be used only with the EZ Smart Blood Glucose Meter to test glucose in fresh capillary whole blood only. This meter is not to be used for Alternate Site Testing or Neonatal Testing. 4. Special instrument Requirements: EZ Smart Blood Glucose Meter. I. Device Description: The EZ Smart Blood Glucose Test Strips are used with the EZ Smart Blood Glucose Meter to quantitatively measure glucose in capillary whole blood. When the edge of the EZ Smart test strip is touched to a drop of blood, the test strip draws the blood into the sample chamber and your glucose reading is displayed on the meter after 10 seconds. The test measures glucose from 20 mg/dL (1.1mmol/L) to 600 mg/dL (33.3 mmol/L). The EZ Smart Test Strip is calibrated to display the equivalent of plasma glucose values to allow easy comparison of results with laboratory methods. J. Substantial Equivalence Information: 1. Predicate device name(s): Bayer Elite with the Elite Test Strips 2. Predicate K number(s): k964630 (Bayer Elite) k992342 (Bayer Elite Test Strips) 3. Comparison with Predicate: The VIP International Wholesalers, Corp., EZ Smart Blood Glucose Monitoring System is equivalent to the Bayer Elite Blood Glucose Monitoring System previously cleared under (K964630). The table below lists the similarities and differences between the Predicate and Proposed device. Substantial Equivalence Comparison Similarities | Item | Predicate Device Bayer ELITE (K964630) | Proposed Device EZ Smart | | --- | --- | --- | | Similarities | 1. Monitors glucose using whole blood. | 1. Monitors glucose using whole blood. | {2} Page 3 of 8 | | 2. Directly displays results without requiring calculation. 3. Test principle includes measuring a current produced by a chemical reaction. 4. Test principle: Uses glucose oxidase reaction. 5. Measuring range: 20 to 600 ng/dL. | 2. Directly displays results without requiring calculation. 3. Test principle includes measuring a current produced by a chemical reaction. 4. Test principle: Uses glucose oxidase reaction. 5. Measuring range: 20 to 600 mg/dL. | | --- | --- | --- | ## Differences | Item | Predicate Device Bayer ELITE (K964630) | Proposed Device EZ Smart | | --- | --- | --- | | Differences | 1. Size: 97.8 x 56 x 14.5 mm. 2. Measuring time: 30 seconds. | 1. Size: 94 x 49 x 17 mm. 2. Measuring time: 10 seconds. | ## K. Standard/Guidance Document Referenced (if applicable): The EZ Smart Blood Glucose Monitoring System has been tested with the listed standards and found in compliance with the council EMC directive 89/336/EEC. It is possible to use CE markings to demonstrate the compliance with this EMC Directive. | Test Standards | | | | --- | --- | --- | | EN 60601-1-2/1993 | Medical electrical equipment – Electromagnetic compatibility | | | | EN 55011/1998 + A1/1999 | Emissions, Limits and methods of measurement of radio disturbance characteristics of industrial, scientific and media (ISM) radio-frequency equipment. | | | IEC 801-3/1984 | Immunity, Electrostatic discharge | | | IEC 801-3/1984 ENV 50204/1995 | Immunity, Radio-frequency electromagnetic field. | CDRH document, Write it Right, Recommendations for Developing User Instruction Manuals for Medical Devices Used in House Health Care. {3} NCCLS Guideline Gp-14P, Labeling for Home Use In Vitro Testing Products, (SMOG) NCCLS Guideline EP6-A. ## L. Test Principle: The test principle is based on electrochemical biosensor technology using glucose oxidase. The strip uses the enzyme glucose oxidase to produce a current that will stimulate a chemical reaction. This reaction is measured by the meter and displayed as the blood glucose result. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: #### a. Precision/Reproducibility: The following precision studies were obtained from spiked venous blood specimens and control solution in the R&D laboratories at the manufacturer's site. This precision study is a measurement of strip-to-strip (within-run) reproducibility by calculating the mean, standard deviation, and coefficient of variation for test conducted on strips from three different lots (between-run). Whole blood samples from volunteers were collected with EDTA Vacutainer® tubes and spiked with β-D-Glucose to desired glucose concentrations. The spikes samples were gently mixed by inverting 20 times. The desired glucose concentrations used ranges at 5 different levels; low concentration (~30-50 mg/dL) normal concentration (~51-110 mg/dL), high normal (~111-150 mg/dL), mid high (~151-250 mg/dL), and high concentration (~251-400 mg/dL). A YSI 2300 glucose analyzer (Yellow Springs Instrument, Yellow Springs, Ohio, USA) was used to measure the glucose concentration in venous samples. This YSI 2300 glucose analyzer was used throughout the study as the "gold standard". All measurements were done in "Normal Mode" and a calibration mode of "5/15" (calibrating every 5 measurements or every 15 minutes). Results presented in the table below are the results of the precision study conducted using the EZ Smart meters. | Glucose Level (mg/dL) | 44.9 | 89.8 | 131 | 224 | 361 | | --- | --- | --- | --- | --- | --- | | Number of Tests | 30 | 30 | 30 | 30 | 30 | | Average (mg/dL) | 49.7 | 85.0 | 121.4 | 221.0 | 362.1 | | S.D. (mg/dL) | 2.1 | 1.6 | 4.0 | 6.8 | 9.0 | | C.V. (%) | 4.2% | 1.8% | 3.3% | 3.1% | 2.5% | Regression (y) = 1.003x-2.46 $$ r_2 = 0.9981 $$ {4} The results obtained from test conducted on strips from three different lots (between-run) are presented below: | Lot No. | YSI (mg/dL) | Mean (mg/dL) | SD (mg/dL) | CV% | | --- | --- | --- | --- | --- | | Meter Serial # 1 M020242 | 44.9 | 49.20 | 1.32 | 2.68 | | | 89.8 | 80.60 | 2.67 | 3.32 | | | 131 | 122.30 | 5.19 | 4.24 | | | 224 | 215.70 | 7.02 | 3.26 | | | 361 | 357.80 | 4.18 | 1.17 | | Meter Serial # 2 M0202426 | 44.9 | 49.70 | 2.11 | 4.25 | | | 89.8 | 85.00 | 1.56 | 1.84 | | | 131 | 121.40 | 3.68 | 3.28 | | | 224 | 221.00 | 6.80 | 3.08 | | | 361 | 362.10 | 9.02 | 2.49 | | Meter Serial # 3 M0202426 | 44.9 | 49.90 | 1.73 | 3.46 | | | 89.8 | 82.60 | 2.41 | 2.92 | | | 131 | 118.20 | 6.32 | 5.35 | | | 224 | 227.60 | 7.00 | 3.07 | | | 361 | 357.30 | 6.33 | 1.77 | # b. Linearity/assay reportable range: Linearity (analytical range) studies were designed in accordance with NCCLS Guideline EP6-A. Venous blood was drawn from healthy volunteers and collected into lithium heparin vacutainer® tubes. The blood was then placed in a room temperature environment overnight, until glycolysis took place to reduce the glucose concentration to nearly zero. The blood was then pooled and allocated to lithium heparin tubes. A small amount of high concentration glucose $(10,000\mathrm{mg / dL})$ was added to each tube to obtain the desired blood glucose levels needed to perform the test. Recovery tests were performed by confirming the blood glucose concentration with the YSI 2300. Blood samples with glucose concentrations ranging from 20 to $600\mathrm{mg / dL}$ were then tested with the EZ Smart Blood Glucose Monitoring System. Three lots of EZ Smart Test Strips, Chosen at random, were used during the test. The test results were evaluated to describe accuracy over the entire range of blood glucose values. A linear regression analysis was performed by the method of least squares $(Y = 0.9717X + 3.66$ , $R^2 = 0.9987)$ . The sensitivity of the system was determined by the slope of the regression line whereas the linearity was determined by the correlation coefficient of the regression line. All measurements determined by the EZ Smart System are within a $15\%$ bias of the reference (YSI 2300) results (glucose concentration $>100~\mathrm{mg / dL}$ ) and a $15~\mathrm{mg / dL}$ bias of the reference (YSI 2300) results (glucose concentration $< 100~\mathrm{mg / dL}$ ) see table below: {5} Page 6 of 8 The sensitivity of EZ Smart Blood Glucose Monitoring System | YSI (mg/dL) | 20 | 75 | 152 | 258 | 359 | 482 | 600 | | --- | --- | --- | --- | --- | --- | --- | --- | | EZ Smart Results Lot R030212 | 25 | 77 | 148 | 269 | 358 | 478 | 595 | | | 23 | 82 | 139 | 258 | 354 | 459 | 589 | | | 24 | 77 | 158 | 254 | 368 | 486 | 570 | | | 20 | 71 | 156 | 245 | 348 | 469 | 591 | | | 21 | 74 | 147 | 249 | 347 | 465 | 600 | | | 28 | 76 | 155 | 243 | 345 | 462 | 597 | | | 30 | 83 | 148 | 255 | 350 | 471 | 582 | | | 26 | 76 | 155 | 248 | 351 | 467 | 591 | | MEAN | 24.63 | 77.00 | 150.75 | 252.63 | 352.63 | 469.63 | 589.38 | | SD | 3.38 | 3.93 | 6.36 | 8.37 | 7.44 | 8.78 | 9.55 | | CV% | 13.72% | 5.10% | 4.22% | 3.31% | 2.11% | 1.87% | 1.62% | | Bias% | 23.13% | 2.67% | -0.82% | -2.08% | -1.78% | -2.57% | -1.77% | c. Traceability (controls, calibrators, or method): Traceability has been reference by the manufacturer to ISO 17511. d. Detection limit: The detection range is from 20 – 600 mg/dL (1.1 to 33.3 mmol/L). See linearity/assay detection limit above. e. Analytical specificity: Interference testing was conducted to determine the effect of select endogenous and exogenous substances. The following study was used to elucidate whether the EZ Smart blood glucose Monitoring System is capable of a precise reading in blood samples containing the interferences. Interference studies were conducted according to NCCLS EP7-P. The Interfering Effect of Drugs and Chemical Compounds on EZ Smart Test Strips. | Interfering Compounds | Physiological Levels Test Conc. | Test Concentration | No Interference Level of EZ Smart Test Strip | | --- | --- | --- | --- | | Acetone | | 6 mg/ml | 6 mg/ml | | Acetaminophen | 20 μg/ml | 40 μg/ml | 30μg/ml | | Ascorbic Acid | 12 μg/ml | 150 μg/ml | 75 μg/ml | | Alcohol | | 35 μg/ml | 35 mg/ml | | Barbital | | 1 mg/ml | 1 mg/ml | | Benzoic Acid | | 14.4 mg/ml | 14.4 mg/ml | | Bilirubin | | 200 μg/ml | 200 μg/ml | {6} # f. Assay cut-off: NA # 2. Comparison studies: a. Method comparison with predicate device: 203 patients were enrolled in an OTC study that used the Bayer Elite as the predicate device and the YSI model 2300 as the reference method. VIP International instituted the clinical trials at three different locations in the continental United States. The patients performed their own fingerstick and ran the test unassisted with the EZ Smart meter. The results were recorded on the questionnaire form provided. The patient then performed their own test with the Bayer Elite meter using blood from the original fingerstick and recorded the results. A technician collected a blood sample from the patient when possible for analysis on the YSI Model 2300 comparative laboratory method. The results obtained by the patients on the EZ Smart Glucometer, Bayer Elite Glucometer, and by the technical staff on the YSI using the capillary blood are presented in the chart below. Using capillary blood, 184/202 or $91.1\%$ of the results were within $20\%$ of the YSI values. | Site # | N | X-Method Capillary | Y-Method Capillary | R | Regular Slope | Regular Intercept | Deming Slope | Deming Intercept | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | #1 | 38 | YSI-Ref | EZ Smart | 0.9574 | 1.023 | 2.7 | 1.072 | -5.1 | | #2 | 87 | YSI | EZ Smart | 0.9768 | 0.9338 | 1.7 | 0.9550 | -1.2 | | #3 | 77 | YSI | EZ Smart | 0.9654 | 0.9446 | 0.6 | 0.9776 | -4.7 | | #4 | 100 | YSI | EZ Smart | 0.9776 | 0.9446 | 0.6 | 0.9776 | -4.7 | | #5 | 100 | YSI | EZ Smart | 0.9776 | 0.9446 | 0.6 | 0.9776 | -4.7 | | #6 | 100 | YSI | EZ Smart | 0.9776 | 0.9446 | 0.6 | 0.9776 | -4.7 | | #7 | 100 | YSI | EZ Smart | 0.9776 | 0.9446 | 0.6 | 0.9776 | -4.7 | | #8 | 100 | YSI | EZ Smart | 0.9776 | 0.9446 | 0.6 | 0.9776 | -4.7 | | #9 | 100 | YSI | EZ Smart | 0.9776 | 0.9446 | 0.6 | 0.9776 | -4.7 | {7} | #1 | 38 | Bayer Elite | EZ Smart | 0.9676 | 0.8651 | 15.6 | 0.8907 | 11.1 | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | #2 | 87 | Bayer Elite | EZ Smart | 0.9835 | 0.8890 | 11.4 | 0.9023 | 9.7 | | #3 | 77 | Bayer Elite | EZ Smart | 0.9789 | 0.8557 | 10.7 | 0.8716 | 8.0 | | All | 202 | YSI | EZ Smart | 0.9644 | 0.9678 | -0.1 | 1.004 | -5.5 | | All | 202 | Bayer Elite | EZ Smart | 0.9815 | 0.8778 | 10.6 | 0.8925 | 8.4 | | All | 202 | YSI | Bayer Elite | 0.9637 | | | | | b. Matrix comparison: NA 3. Clinical studies: a. Clinical sensitivity: NA b. Clinical specificity: NA c. Other clinical supportive data (when $a$ and $b$ is not applicable): NA 4. Clinical cut-off: NA 5. Expected values/Reference range: The normal fasting glucose range for a non-diabetic adult is 70 to $110\mathrm{mg / dL}$ . (3.9 to 6.1 mmol/L) $^{1}$ . One to two hours after meals, normal glucose values should be less than 120 mg/dL $(6.7\mathrm{mmol / L})^2$ . 1. Burtis CA Ashwood ER, eds: Tietz Textbook of Clinical Chemistry. $2^{\mathrm{nd}}$ Edition. W.B. Saunders. Philadelphia. 1994. p. 2190. 2. Krall LP and Beaser RS: Joslin Diabetes Manual. Lea and Febiger. Philadelphia 1989. p. 138. # N. Conclusion: The submitted material in this premarket notification is complete and supports a substantial equivalence decision
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...