ELECSYS PROBNP II STAT IMMUNOASSAY AND ELECSYS PROBNP II CALSET, MODELS 05390109-160, 04842472-190
Applicant
Roche Diagnostics Corp.
Product Code
NBC · Clinical Chemistry
Decision Date
Feb 4, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1117
Device Class
Class 2
Intended Use
Immunoassay for the in vitro quantitative determination of N-terminal pro-brain natriuretic peptide in human serum and plasma. The Elecsys proBNP II STAT assay is used as an aid in the diagnosis of individuals suspected of having congestive heart failure. The test is further indicated for the risk stratification of patients with acute coronary syndrome and congestive heart failure. The test may also serve as an aid in the assessment of increased risk of cardiovascular events and mortality in patients at risk for heart failure who have stable coronary artery disease. The electrochemiluminescence immunoassay "ECLIA" is intended for use on the cobas e 601 immunoassay analyzer.
Device Story
In vitro diagnostic immunoassay; measures N-terminal pro-brain natriuretic peptide (NT-proBNP) in human serum/plasma. Uses electrochemiluminescence immunoassay (ECLIA) technology; sandwich assay format with two monoclonal antibodies directed at NT-proBNP epitopes; capture antibody biotinylated for streptavidin-coated microparticles; signal antibody tagged with ruthenium. Operated on cobas e 601 analyzer by laboratory professionals. Provides quantitative results in 9 minutes (STAT application). Results used by clinicians alongside medical history and clinical examination to aid diagnosis of congestive heart failure, risk stratification for acute coronary syndrome, and assessment of cardiovascular event risk. Benefits include rapid turnaround time for critical cardiac assessment.
Clinical Evidence
Bench testing only. Performance characteristics evaluated include method comparison, recovery, limit of detection, limit of quantitation, precision, hook effect, and stability.
Technological Characteristics
In vitro diagnostic assay; modified from two-step to one-step incubation. Software updated to accommodate assay duration change. Reagent concentrations unchanged.
Indications for Use
Indicated for in vitro quantitative determination of NT-proBNP in human serum/plasma to aid in diagnosis of suspected congestive heart failure, risk stratification of patients with acute coronary syndrome and congestive heart failure, and assessment of increased risk of cardiovascular events/mortality in patients at risk for heart failure with stable coronary artery disease.
Regulatory Classification
Identification
The B-type natriuretic peptide (BNP) test system is an in vitro diagnostic device intended to measure BNP in whole blood and plasma. Measurements of BNP are used as an aid in the diagnosis of patients with congestive heart failure.
Special Controls
*Classification.* Class II (special controls). The special control is “Class II Special Control Guidance Document for B-Type Natriuretic Peptide Premarket Notifications; Final Guidance for Industry and FDA Reviewers.”
Predicate Devices
k072437
Related Devices
K210546 — Elecsys proBNP II, Elecsys proBNP II STAT · Roche Diagnostics · Mar 31, 2022
K072437 — ELECSYS PROBNP II, PRECICONTROL CARDIAC II, PROBNP II CALSET · Roche Diagnostics · Feb 5, 2008
{0}------------------------------------------------
K0926649
### Appendix | Updated 510(k) Summary
ستر،
FEB - 4 2010
| Immunoassay Comparison | | |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | Predicate Device: Elecsys proBNP II Assay (K072437) (18 Minute) | Elecsys proBNP II STAT Assay |
| | General Assay Features | |
| Intended<br>Use/<br>Indications<br>for Use | Immunoassay for the in vitro<br>quantitative determination of N-<br>terminal pro-brain natriuretic peptide<br>in human serum and plasma. The<br>Elecsys proBNP II assay is used as an<br>aid in the diagnosis of individuals<br>suspected of having congestive heart<br>failure. The test is further indicated<br>for the risk stratification of patients<br>with acute coronary syndrome and<br>congestive heart failure. The test may<br>also serve as an aid in the assessment<br>of increased risk of cardiovascular<br>events and mortality in patients at risk<br>for heart failure who have stable<br>coronary artery disease.<br><br>The electrochemiluminescence<br>immunoassay "ECLIA" is intended<br>for use on Elecsys and cobas e<br>immunoassay analyzers. | Same, except:<br>The assay name is changed:<br>... The Elecsys proBNP II STAT assa<br>is used as an aid in the diagnosis...<br><br>Elecsys instruments are removed:<br>The electrochemiluminescence<br>immunoassay "ECLIA" is intended<br>for use on the cobas e 601<br>immunoassay analyzer. |
| Assay<br>Protocol | Sandwich assay | Same |
| Detection<br>Protocol | Electrochemiluminescent<br>Immunoassay | Same |
| Application | 18 Minute | STAT (9 Minute) |
| Instrument<br>Platform | Roche Elecsys 2010/cobas e 411 and<br>MODULAR ANALYTICS<br>E170/cobas e 601 | cobas e 601 |
| Sample<br>Volume | 15 µL | Same |
| Immunoassay Comparison | | |
| Feature | Predicate Device: Elecsys proBNP<br>II Assay (K072437) (18 Minute) | Elecsys proBNP II STAT Assay |
| General Assay Features | | |
| Sample Type | Human serum and plasma treated with<br>K2-EDTA, K3-EDTA, lithium heparin<br>and Na-heparin plasma. | Same |
| Reagents | The proBNP II Assay is a sandwich<br>immunoassay with two antibodies<br>directed towards epitopes within the<br>N-terminal portion of the proBNP<br>molecule. The capture antibody is<br>biotinylated to react with streptavidin-<br>coated microparticles. The signal<br>antibody is tagged with ruthenium.<br>Both antibodies are monoclonal | Same antibodies and same epitopes |
| Traceability | The Elecsys proBNP II assay has been<br>standardized against the Elecsys<br>proBNP assay (k022516). This in turn<br>was standardized against reference<br>standards by weighing pure synthetic<br>NT-proBNP (1-76 amino acids) into<br>equine serum matrix. | Same |
| Calibrator | Elecsys proBNP II CalSet (k072437) | Elecsys proBNP II STAT CalSet will<br>have a different material number to<br>avoid confusion on instrument use.<br>The stability, value assignment and<br>matrix will be identical. |
{1}------------------------------------------------
{2}------------------------------------------------
| | Immunoassay Comparison | |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | Predicate Device: Elecsys proBNP<br>II Assay (K072437) (18 Minute) | Elecsys proBNP II STAT Assay |
| | General Assay Features | |
| Calibration<br>Interval | Calibration must be performed once<br>per reagent lot using fresh reagent<br>(i.e. not more than 24 hours since the<br>reagent kit was registered on the<br>analyzer). Renewed calibration is<br>recommended as follows:<br><br>MODULAR ANALYTICS E170,<br>Elecsys 2010 and cobas e analyzers:<br>After 1 month (28 days) when<br>using the same reagent lotAfter 7 days (when using the<br>same reagent kit on the<br>analyzer)<br>Elecsys 1010 analyzer:<br>With every reagent kitAfter 7 days (ambient<br>temperature 20-25 °C)After 3 days (ambient<br>temperature 25-32 °C) | Calibration must be performed once<br>per reagent lot using fresh reagent (i.e.<br>not more than 24 hours since the<br>reagent kit was registered on the<br>analyzer). Renewed calibration is<br>recommended as follows:<br><br>cobas e 601:<br>After 1 month (28 days) when<br>using the same reagent lotAfter 7 days (when using the<br>same reagent kit on the<br>analyzer) |
| Controls | Elecsys PreciControl Cardiac II<br>(k072437) | Same |
{3}------------------------------------------------
# Premarket Notification, 510(k) for Elecsys proBNP II
#### STAT Assay, Continued
| Immunoassay Comparison | | |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| Feature | Predicate Device: Elecsys proBNP<br>II Assay (K072437) (18 Minute) | Elecsys proBNP II STAT Assay |
| General assay features | | |
| Traceability /<br>Standardization | This method has been standardized<br>against the Elecsys proBNP assay.<br>This in turn was standardized against<br>reference standards by weighing pure<br>synthetic NT-proBNP (1-76) into an<br>equine serum matrix. | Same |
| Reagent<br>Stability/<br>Storage | Store at 2-8 °C. Store the Elecsys<br>proBNP II reagent kit upright in order<br>to ensure complete availability of the<br>microparticles during automatic<br>mixing prior to use.<br><br>Stability:<br>Unopened at 2-8 °C—up to the stated<br>expiration date<br>After opening at 2-8 °C—12 weeks<br>On MODULAR ANALYTICS E170<br>and cobas e 601—8 weeks<br>On Elecsys2010 and cobas e 411—8<br>weeks | Same, but only reported on the cobas<br>e 601. |
{4}------------------------------------------------
# Premarket Notification, 510(k) for Elecsys proBNP II
#### STAT Assay, Continued
| | Immunoassay Comparison | |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | Predicate Device: Elecsys proBNP<br>II Assay (K072437) (18 Minute) | Elecsys proBNP II STAT Assay |
| | Labeled Performance Characteristics | |
| Measuring<br>Range | 5-35,000 pg/mL | Same |
| Precision | Elecsys 1010/2010 and cobas e411 analyzers<br>Within-run (Repeatability)<br>4.2% CV @ 44.0 pg/mL<br>2.4% CV @ 126.0 pg/mL<br>1.3% CV @ 2410.0pg/mL<br>2.7% CV @ 33606.0 pg/mL<br>2.6% CV @ 82.0 pg/mL<br>1.2% CV @ 2318.0 pg/mL<br>Total (Intermediate)<br>4.6% CV @ 44.0 pg/mL<br>2.6% CV @ 126.0 pg/mL<br>1.8% CV @ 2410.0pg/mL<br>3.8% CV @ 33606.0 pg/mL<br>2.8% CV @ 82.0 pg/mL<br>1.6% CV @ 2318.0 pg/mL<br>MODULAR ANALYTICS E170 and cobas<br>e601:<br>Within-run (Repeatability)<br>1.9% CV @ 64.0 pg/mL<br>1.5% CV @ 124.0 pg/mL<br>1.3% CV @ 14142.0 pg/mL<br>1.8% CV @ 77.0 pg/mL<br>1.2% CV @ 2105.0 pg/mL<br>Total (Intermediate)<br>3.1% CV @ 46.0 pg/mL<br>2.7% CV @ 125.0 pg/mL<br>2.7% CV @ 32805.0 pg/mL<br>2.7% CV @ 77.0 pg/mL<br>2.7% CV @ 2170.0 pg/mL | cobas e 601:<br>Within-run (Repeatability)<br>2.4% CV @ 130.0 pg/mL<br>2.2% CV @ 4942.0 pg/mL<br>3.5% CV @ 59.0 pg/mL<br>2.0% CV @ 142.0 pg/mL<br>1.8% CV @ 522.0 pg/mL<br>1.9% CV @ 934.5 pg/mL<br>2.0% CV @ 6552.0 pg/mL<br>2.9% CV @ 30,870.0 pg/mL<br>Total (Intermediate)<br>2.5% CV @ 130.0 pg/mL<br>2.6% CV @ 4942.0 pg/mL<br>3.5% CV @ 59.0 pg/mL<br>2.5% CV @ 142.0 pg/mL<br>2.0% CV @ 522.0 pg/mL<br>2.5% CV @ 934.5 pg/mL<br>2.3% CV @ 6552.0 pg/mL<br>5.4% CV @ 30,870.0 pg/mL |
| Immunoassay Comparison | | |
| | Predicate Device: Elecsys proBNP<br>II Assay (K072437) (18 Minute) | Elecsys proBNP II STAT Assay |
| Feature | Labeled Performance Characteristics | |
| Analytical<br>Sensitivity | Limit of Blank (LoB) not reported<br>Limit of Detection (LoD) 5.00 pg/mL<br>Limit of Quantitation (LoQ)<br>Functional Sensitivity: 50.00 pg/mL | Limit of Blank (LoB): 5.00 pg/mL<br>Limit of Detection (LoD): 5.00 pg/mL<br>Limit of Quantitation (LoQ)<br>Functional Sensitivity: 50.00 pg/mL |
| Hook Effect | There is no high-dose hook effect at<br>proBNP concentrations up to 300,000<br>pg/mL (300 ng/mL). | Same |
| Limitations | The assay is unaffected by:<br>• Hemoglobin ≤ 0.1 g/dL<br>• Bilirubin up ≤ 25 mg/dL<br>• Triglycerides ≤ 1,500 mg/dL<br>• Biotin ≤ 30 ng/mL<br>• Rheumatoid factors ≤ 1,500 IU/mL<br>In vitro tests were performed on 51<br>commonly used pharmaceuticals. No<br>interference with the assay was found.<br>• The risk of interference from potential<br>immunological interactions between<br>test components and rare sera has been<br>minimized by the inclusion of suitable<br>additives.<br>• As with all tests containing monoclonal<br>mouse antibodies, erroneous findings<br>may be obtained from samples taken<br>from patients who have been treated<br>with monoclonal mouse antibodies or<br>have received them for diagnostic<br>purposes.<br>• In rare cases, interference due to<br>extremely high titers of antibodies to<br>streptavidin and ruthenium can occur.<br>The test contains additives which<br>minimize these effects.<br>• For diagnostic purposes, the results<br>should always be assessed in<br>conjunction with the patient's medical<br>history, clinical examination and other<br>findings | Same |
/
Continued on next page
।
{5}------------------------------------------------
## Premarket Notification, 510(k) for Elecsys proBNP II
#### STAT Assay, Continued
{6}------------------------------------------------
| Immunoassay Comparison | | | | |
|-------------------------------------|-----------------------------------------------------------------|------------------------------|-------------------|-------|
| Feature | Predicate Device: Elecsys proBNP II Assay (K072437) (18 Minute) | Elecsys proBNP II-STAT Assay | | |
| Labeled Performance Characteristics | | | | |
| Method Comparison | Both assays run on the cobas e 601 | | | |
| | n = 120<br>Min = 5.10 pg/mL<br>Max = 30,583 pg/mL | | | |
| | Passing/Bablok | Linear Regression | Deming Regression | |
| | Slope | 0.992 | 0.986 | 0.988 |
| | Intercept | -2.141 | 13.964 | 8.83 |
| | Tau/r/rho | 0.987 | 0.998 | 1.00 |
{7}------------------------------------------------
| CalSet Comparison | | |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | Predicate Device: Elecsys proBNP II CalSet (K072437) | Elecsys proBNP II STAT CalSet |
| Labeled Performance Characteristics | | |
| Intended Use | Elecsys proBNP II CalSet is used for calibrating the quantitative Elecsys proBNP II assay on Elecsys and cobas e immunoassay analyzers. | Elecsys proBNP II STAT CalSet is used for calibrating the quantitative Elecsys proBNP II STAT assay on the cobas e 601 immunoassay analyzer. |
| Levels | Two | Same |
| Format | Lyophilized equine serum | Same |
| Analyte<br>Concentration | Cal 1: ~140 pg/mL<br>Cal 2: ~2,700 pg/mL | Same |
| Handling | Dissolve carefully the contents of one bottle by adding exactly 1.0 mL of distilled water and allow to stand closed for 15 minutes to reconstitute. Mix carefully avoiding the formation of foam. Transfer the reconstituted calibrator into the empty, labeled snap-cap bottles supplied. | Same |
| Stability | Unopened:<br>• Store at 2 – 8°C until expiration date.<br>Reconstituted:<br>• 2 – 8°C: 2 weeks<br>• -20°C: 3 months (freeze only once)<br>• On Elecsys 1010/2010 and cobas e411 analyzers at 20 – 25°C: up to 5 hours<br>• On MODULAR ANALYTICS E170 and cobas e601 analyzers: use only once | Unopened:<br>• Store at 2 – 8°C until expiration date.<br>Reconstituted:<br>• 2 – 8°C: 2 weeks<br>• -20°C: 3 months (freeze only once)<br>• On the analyzer: use only once |
{8}------------------------------------------------
Roche Diagnostics Corporation requests that the FDA not disclose the nature Confidentiality or existence of the premarket notification until the substantial equivalence decision has been reached.
Closing
We trust that the information provided in this Premarket Notification [Special 510(k)] will support a determination of substantial equivalence for the Elecsys proBNP II STAT Immunoassay.
If you should have questions or require further information, please do not hesitate to contact me.
- Phone: (317) 521-3338
- · FAX: (317) 521-2324
- · jane.phillips(@roche.com
Sincerely,
Jane Phillips, PhD Regulatory Affairs Principal US Regulatory Submissions Roche Diagnostics Corporation
{9}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/9/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
FEB - 4 2010
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002
Roche Diagnostics Corp. c/o Dr. Jane Phillips, Regulatory Affairs Principal 9115 Hague Road Indianapolis, IN 46250
Re: k092649 Trade Name: Elecsys proBNP II STAT Immunoassay, Elecsys proBNP II STAT CalSet Regulation Number: 21 CFR §862.1117 Regulation Name: B-type Natriuretic Peptide Test System Regulatory Class: Class II Product Codes: NBC, JIT Dated: January 4, 2010 Received: January 5, 2010
Dear Dr. Phillips:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21
{10}------------------------------------------------
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Signature
Courtney C. Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
# Indications for Use Form
510(k) Number (if known):
Device Name: Elecsys proBNP II STAT Immunoassay
Immunoassay for the in vitro quantitative determination of N-terminal pro-brain natriuretic peptide in human serum and plasma. The Elecsys proBNP II STAT assay is used as an aid in the diagnosis of individuals suspected of having congestive heart failure. The test is further indicated for the risk stratification of patients with acute coronary syndrome and congestive heart failure. The test may also serve as an aid in the assessment of increased risk of cardiovascular events and mortality in patients at risk for heart failure who have stable coronary artery disease.
The electrochemiluminescence immunoassay "ECLIA" is intended for use on the cobas e 601 immunoassay analyzer.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Carol Benson
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K092649
Page 1 of 2
{12}------------------------------------------------
### Indications for Use Form
510(k) Number (if known):
Device Name: Elecsys proBNP II STAT CalSet
Elecsys proBNP II STAT CalSet is used for calibrating the quantitative Elecsys proBNP II STAT assay on the cobas e 601 immunoassay analyzer.
Prescription Use __ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Carol Benson
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K092649
· Page 2 of 2
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.