STRATUS CS ACUTE CARE NT-PROBNP TESTPAK, PBNP CALPAK CALIBRATOR, PBNP DILPAK DILUENT

K043476 · Dade Behring, Inc. · NBC · Feb 15, 2005 · Clinical Chemistry

Device Facts

Record IDK043476
Device NameSTRATUS CS ACUTE CARE NT-PROBNP TESTPAK, PBNP CALPAK CALIBRATOR, PBNP DILPAK DILUENT
ApplicantDade Behring, Inc.
Product CodeNBC · Clinical Chemistry
Decision DateFeb 15, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1117
Device ClassClass 2

Intended Use

The Stratus® CS Acute Care™ NT-proBNP method (pBNP) is an in vitro diagnostic assay for the quantitative determination of N-terminal pro-brain natriuretic peptide (NT-proBNP) in buman plasma. In individuals suspected of having congestive heart failure (CHF), measurements of NT-proBNP are used as an aid in the diagnosis and assessment of severity. The test is further indicated for the risk stratification of patients with acute coronary syndrome and heart failure. The NT-proBNP Calibrator (pBNP CalPak) is an in vitro diagnostic product intended to be used for calibration of the NT-proBNP method on the Stratus® CS analyzer. The NT-proBNP Dilution Pak (DilPak) is an in vitro diagnostic product intended to be used for dilution of the NT-proBNP Method on the Stratus® CS analyzer.

Device Story

Stratus CS Acute Care NT-proBNP TestPak is an in vitro diagnostic assay for quantitative measurement of NT-proBNP in heparinized plasma. Device uses Radial Partition Immunoassay (RPIA) technology; dendrimer-linked polyclonal sheep antibodies immobilized on glass fiber paper capture NT-proBNP from sample; enzyme-labeled secondary antibody forms sandwich; substrate wash initiates enzymatic reaction; optical system monitors front-surface fluorescence. Analyzer microprocessor performs all data analysis. Used in clinical settings; operated by laboratory personnel. Output is quantitative NT-proBNP concentration (pg/mL). Clinicians use results to aid diagnosis/severity assessment of CHF and risk stratification for acute coronary syndrome and heart failure.

Clinical Evidence

Clinical study included 308 individuals without CHF (reference group) and 234 patients with CHF (disease group). Performance compared to Roche Elecsys® proBNP. Sensitivity for males <75 yrs: 90% (Stratus) vs 91% (predicate); males ≥75 yrs: 92% (both). Specificity for males <75 yrs: 92% (Stratus) vs 93% (predicate); males ≥75 yrs: 73% (both). Females <75 yrs sensitivity: 84% (both); females ≥75 yrs: 95% (both). Specificity for females <75 yrs: 92% (Stratus) vs 94% (predicate); females ≥75 yrs: 85% (Stratus) vs 87% (predicate).

Technological Characteristics

Solid-phase Radial Partition Immunoassay (RPIA) on glass fiber paper. Uses Roche polyclonal (sheep) antibodies. Detection via front surface fluorescence. Reportable range 15-20,000 pg/mL. Analyte: synthetic NT-proBNP (1-76). Calibration updated per lot and every 30 days. Sample volume 50 uL. Analyzed on Stratus® CS analyzer.

Indications for Use

Indicated for quantitative measurement of NT-proBNP in human plasma to aid in diagnosis and severity assessment of congestive heart failure (CHF) and for risk stratification of patients with acute coronary syndrome and heart failure.

Regulatory Classification

Identification

The B-type natriuretic peptide (BNP) test system is an in vitro diagnostic device intended to measure BNP in whole blood and plasma. Measurements of BNP are used as an aid in the diagnosis of patients with congestive heart failure.

Special Controls

*Classification.* Class II (special controls). The special control is “Class II Special Control Guidance Document for B-Type Natriuretic Peptide Premarket Notifications; Final Guidance for Industry and FDA Reviewers.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ FEB 1 5 2005 K093976 #### SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. | Submitter's Name: | George M. Plummer<br>Dade Behring Inc.<br>P.O. Box 6101<br>Newark, DE 19714-6101 | |-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Date of Preparation: | December 14, 2004 | | Name of Product(s): | Stratus® CS Acute Care™ NT-proBNP (pBNP) TestPak assay<br>Stratus® CS pBNP CalPak (the assay calibrator)<br>Stratus® CS pBNP DilPak (the assay diluent) | | FDA Classification Name(s): | B-type natriuretic peptide test system and calibrator (all Class II) | | FDA Guidance Documents: | "Class II Special Control Guidance Document for B-Type Natriuretic<br>Peptide Premarket Notifications; Final Guidance for Industry and<br>FDA Reviewers" - 11/30/2000 | | Predicate Device(s): | Roche Diagnostics Elecsys® proBNP immunoassay and CalSet<br>calibrator (K032646/K022516) | #### Device Description(s): #### Method The Stratus® CS Acute Care™ NT-proBNP method is a two-site sandwich assay based upon solid phase Radial Partition Immunoassav (RPIA) technology. In this procedure, dendrimer linked polyclonal antibody is added to the center portion of a square piece of glass fiber paper in the pBNP TestPak. This antibody recognizes a distinct antigenic site on the NT-proBNP molecule. Sample is then added onto the paper where it reacts with the immobilized antibody. After a short incubation, a conjugate consisting of enzyme-labeled polvclonal antibody directed against a second distinct antigenic site on the NT-proBNP molecule is pipetted onto the reaction zone of the paper. During this second incubation period, enzyme-labeled antibody reacts with the bound NT-proBNP , forming an antibody-antigen-labeled antibody sandwich. The unbound labeled antibody is later eluted from the field of view of the Stratus® CS analyzer by applying a substrate wash solution to the center of the reaction zone. By including substrate for the enzyme within the wash solution, initiation of enzyme activity occurs simultaneously with the wash. The enzymatic rate of the bound fraction increases directly with the concentration of NT-proBNP in the sample. The reaction rate can then be measured by an optical system that monitors the {1}------------------------------------------------ reaction rate via front surface fluorescence. All data analysis functions are performed by the microprocessor within the analyzer. ### Calibrator The Dade Behring pBNP Calibrator is a frozen liquid product containing synthetic human NT-proBNP in a bovine albumin matrix with stabilizers and preservative. The kit consists of five CalPaks at a single calibrator level. Each CalPak contains calibrator reagent in three wells. ### Diluent The Dade Behring pBNP Diluent is a refrigerated product containing a buffered bovine protein matrix with stabilizers and preservative. The kit consists of 5 DilPaks with diluent in one well. #### Intended Use: #### Method The Stratus® CS Acute Care™ NT-proBNP method (pBNP) is an in vitro diagnostic test for the quantitative measurement of N-terminal pro-brain natriuretic peptide (NT-proBNP) in heparanized plasma. In individuals suspected of having congestive heart failure (CHF), measurements of NT-proBNP are used as an aid in the diagnosis and assessment of severity. The test is further indicated for the risk stratification of patients with acute coronary syndrome and heart failure. #### Calibrator The NT-proBNP (pBNP) Calibrator (CalPak), REF CPBNP-C, is an in vitro diagnostic product intended to be used for calibration of the Stratus® CS Acute Care™ NT-proBNP method (pBNP). #### Diluent The NT-proBNP Dilution Pak (DilPak), REF CPBNP-D, is an in vitro diagnostic product intended to be used in conjunction with the pBNP TestPak, REF CPBNP, on the Stratus® CS analyzer for the measurement of samples with elevated levels of NT-proBNP. ### Comparison to Predicate Device: ### Method A summary of the features of the Dade Behring Stratus® CS Acute Care™ pBNP TestPak and the predicate Roche Diagnostics Elecsys® proBNP immunoassay (K032646/K022516) is provided in the following charts. The Dade Behring pBNP TestPak utilizes the Roche polyclonal (sheep) antibody/antigen set. {2}------------------------------------------------ # Method: : · | Feature | Stratus® CS Acute Care™ pBNP | Roche Elecsys® proBNP | |---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | For the <i>in vitro</i> quantitative<br>determination of N-terminal<br>pro-brain natriuretic peptide in<br>human plasma as an aid in the<br>diagnosis and assessment of<br>severity of individuals suspected of<br>having congestive heart failure.<br>The test is further indicated for the<br>risk stratification of patients with<br>acute coronary syndrome and heart<br>failure. | For the <i>in vitro</i> quantitative<br>determination of N-terminal pro-brain<br>natriuretic peptide in human serum<br>and plasma as an aid in the diagnosis<br>of individuals suspected of having<br>congestive heart failure.<br>The test is further indicated for the<br>risk stratification of patients with<br>acute coronary syndrome and<br>congestive heart failure. | | Assay Type<br>(detection) | immunoassay<br>(fluorometric) | immunoassay<br>(electrochemiluminescent) | | Reportable Range | 15- 20,000 pg/mL | 5 - 35,000 pg/mL | | Antibody | Roche Diagnostics'<br>polyclonal (sheep) antibody | polyclonal (sheep) antibody | | Cut-off | 125 pg/mL for patients less than 75<br>years and 450 pg/mL for patients 75<br>years and older | 125 pg/mL for patients less than 75<br>years and 450 pg/mL for patients 75<br>years and older | | Analytical<br>Sensitivity | 15 pg/mL | 5 pg/mL | | Functional<br>Sensitivity | < 50 pg/mL | < 50 pg/mL | | Analytical<br>Specificity | The pharmaceutical Natrecor®<br>shows no significant cross reactivity<br>at 0 and 125 pg/mL NT-proBNP;<br>sixteen other substances also show<br>no significant cross reactivity | The pharmaceutical Natrecor® shows<br>no significant cross reactivity at 300<br>pg/mL and 3000 pg/mL NT-proBNP;<br>sixteen other substances also show no<br>significant cross reactivity | | Interferences | No significant interference from: bilirubin, conj. up to 60 mg/dL<br>bilirubin, unconjugated up to 60 mg/dL hemoglobin up to 1000 mg/dL triglycerides up to 3000 mg/dL rheumatoid factors up to 750 IU/mL | No significant interference from: bilirubin up to 35 mg/dL<br>hemoglobin up to 1.4 g/dL triglycerides up to 4000 mg/dL rheumatoid factors up to 1500 IU/mL | | Reference | Roche purified synthetic NT-proBNP (1-76) | Roche purified synthetic NT-proBNP (1-76) | | Hook Effect | No high dose effect ( up to 1,400,000 pg/mL) | No high dose effect ( up to 300,000 pg/mL) | | Calibration Interval | Calibration curve updated for each lot, using one level and every 30 days, thereafter with the same reagent lot. After calibration update at completion of each test, recovered values are calculated from stored calibration coefficients. | Calibration curve updated for each lot, using two levels every 30 days with the same reagent lot. | | Sample Volume | 50 uL | 20 uL | : {3}------------------------------------------------ #### Calibrator: | Feature | Stratus® CS Acute Care™ pBNP | Roche Elecsys® proBNP | |--------------|-------------------------------|-------------------------------| | Intended Use | pBNP method calibration | proBNP method calibration | | Analyte | Synthetic NT-proBNP | Synthetic NT-proBNP | | Matrix | Bovine albumin | Horse serum | | Form | Liquid | Lyophilyzed | | Volume | 150 uL in each of three wells | 1 mL for each level | | Levels | 1 level | 2 levels (140 and 2700 pg/mL) | ### Method performance Summary: #### Clinical Results For the Reference Study Group, NT-proBNP concentrations were determined in 308 individuals without congestive heart failure (163 Women and 145 men); this population included apparently healthy individuals and individuals with diabetes, hypertension, and pulmonary disease. For the Disease Study Group, blood samples were obtained from 234 patients diagnosed with congestive heart failure (CHF); this population included 70 women and 164 men. {4}------------------------------------------------ The high level of equivalence between the two assays on clinical performance measures justifies using the same cutoffs for the Stratus® CS Acute Care™ pBNP assay as the Elecsys® proBNP predicate device, as shown below: | Patients < 75 years: | 125 pg/mL [14.8 pmol/L] | |----------------------|-------------------------| | Patients ≥ 75 years: | 450 pg/mL [53.2 pmol/L] | The clinical performance of the Dade Behring assay presented in the tables below was substantially equivalent to that of the predicate device. Data used to calculate the values are from the method comparison and reference interval data sets generated at the University of Maryland Medical Center. | Clinical Performance of Dade Behring NT-DroBNP assay versus predicate device on Male patients. | | | |------------------------------------------------------------------------------------------------|--|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | | LE SALES CONSULTION CONSULTION COLLECTION COLLECTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION | | Males | | <75 yrs (95% CI) | ≥ 75 yrs (95% CI) | |-----------------|------------------|------------------|-------------------| | Sensitivity (%) | Stratus® CS pBNP | 90% (84 – 95) | 92% (84 – 99) | | | Elecsys® proBNP | 91% (86 – 96) | 92% (84 – 99) | | Specificity (%) | Stratus® CS pBNP | 92% (86 – 98) | 73% (61 – 84) | | | Elecsys® proBNP | 93% (87 – 98) | 73% (61 – 84) | | NPV | Stratus® CS pBNP | 86% (79 – 94) | 92% (84 – 100) | | | Elecsys® proBNP | 89% (82 – 95) | 92% (84 – 100) | Clinical Performance of Dade Behring NT-proBNP assay versus predicate device on Female Patients | Females | | <75 yrs (95% CI) | ≥75 yrs (95% CI) | |-----------------|------------------|------------------|------------------| | Sensitivity (%) | Stratus® CS pBNP | 84% (74-94) | 95% (85-100) | | | Elecsys® proBNP | 84% (74-94) | 95% (85-100) | | Specificity (%) | Stratus® CS pBNP | 92% (87-97) | 85% (76-94) | | | Elecsys® proBNP | 94% (90-99) | 87% (78-95) | | NPV | Stratus® CS pBNP | 92% (87-97) | 98% (94-100) | | | Elecsys® proBNP | 92% (87-97) | 98% (94-100) | #### Analytical Results ### Method Comparison A split sample method comparison demonstrated good agreement between the Dade Behring Stratus® CS Acute Care™ NT-proBNP TestPak method and the predicate Roche Elecsys® proBNP immunoassay with heparinized plasma patient samples. | Comparative<br>Method | Slope | Intercept<br>(pg/mL) | Correlation<br>Coefficient | n | |-----------------------|-------|----------------------|----------------------------|-----| | Roche Elecsys® proBNP | 0.96 | 5.5 | 0.99 | 481 | The model equation for Passing-Bablok linear regression statistics is: [results for Stratus = slope x [comparative method results] + intercept. The range of NT-proBNP values in the correlation study was: 16.1 - 17691.9 pg/mL. {5}------------------------------------------------ ### Lithium Heparin versus Sodium Heparin Comparison of lithium heparin versus sodium heparin samples on the Stratus® CS system showed very good agreement. Lithium heparin samples (n=19) ranging from < 15 to 15,810 pg/mL when compared to sodium heparin samples gave a slope of 1.00, correlation coefficient of 0.999, and intercept of 28 pg/mL using Passing-Bablok regression statistics. ### Reproducibility Typical precision observed for the Stratus® CS NT-proBNP TestPak method is summarized below: | | | Within-Run Precision | | Total Precision | | | |-----------------------------------|-------------------|----------------------|-----|-----------------|-----|--| | Sample | Mean (pg/mL) | SD (pg/mL) | %CV | SD (pg/mL) | %CV | | | | Human Plasma Pool | | | | | | | Pool 1 | 153 | 6.5 | 4.2 | 7.8 | 5.1 | | | Pool 2 | 463 | 15.0 | 3.2 | 15.0 | 3.2 | | | Pool 3 | 6726 | 177.3 | 2.6 | 177.3 | 2.6 | | | MAS® CardioImmune® proBNP Control | | | | | | | | Level 1 | 156 | 7.7 | 4.9 | 7.7 | 4.9 | | | Level 2 | 5371 | 115.5 | 2.1 | 158.6 | 3.0 | | * MAS® and CardioImmune® are registered trademarks of Medical Analysis Systems, Inc., Camarillo, CA The reproducibility testing was conducted in accordance with the NCCLS Approved Guideline for User Evaluation of Precision Performance of Clinical Chemistry Devices EP5-A, 1999. Specimens at each level were analyzed in duplicate once per day for 20 days. The within-run and total standard deviations were calculated by the analysis of variance method. ### Calibrator The Stratus® CS pBNP CalPak is similar to other calibrator products associated with their assays, such as the Roche Elecsys® proBNP CalSet calibrator. ### Diluent The Stratus® CS pBNP DilPak is similar to other diluent products associated with their assays, such as the Roche Elecys® proBNP CalSet diluent. {6}------------------------------------------------ #### Comments on Substantial Equivalence: Both the Stratus® CS Acute Care™ NT-proBNP (pBNP) TestPak and the Roche Elecsys® proBNP immunoassays are intended for the quantitative determination of NT-proBNP. Comparative data for human plasma samples demonstrate good analytical and clinical agreement between the methods. #### Conclusion: The Dade Behring ® CS Acute Care™ NT-proBNP (pBNP) TestPak and the predicate Roche Elecsys® proBNP immunoassay (K032646/K022516) are substantially equivalent based on their intended use and performance characteristics as described above. The calibrator and diluent products are also equivalent in their design and intended use with their respective assay systems. > George M. Plummer Regulatory Affairs and Compliance Manager December 14, 2004 {7}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with three wave-like lines extending from its head. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle symbol. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 FEB 1 5 2005 Mr. George M. Plummer Regulatory Affairs and Compliance Manager Dade Behring Inc. Glasgow Business Community Building 500, MS 514 PO Box 6101 Newark, DE 19714 Re: k043476 K043470 Trade/Device Name: Stratus® CS Acute Care™ NT-proBNP (pBNP) TestPak assay Stratus® CS pBNP CalPak Stratus® CS pBNP DilPak Regulation Number: 21 CFR 862.1117 Regulation Name: B-type natriuretic peptide test system Regulatory Class: Class II Product Code: NBC; JIT Dated: December 14, 2004 Received: December 16, 2004 Dear Mr. Plummer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, do floos that have boon require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may attress, provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device n may of caojoer of the 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean r houre of action and a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any 1 with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {8}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html Sincerely yours, Sean M. Cooper MS, DUM Jean M. Cooper, MS, D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {9}------------------------------------------------ #### INDICATIONS FOR USE STATEMENT #### Ko43476 510(k) Number (If Known): Device(s) Name(s): Stratus® CS Acute Care™ NT-proBNP (pBNP) TestPak assay Stratus® CS pBNP CalPak Stratus® CS pBNP DilPak #### Indications for Use: The Stratus® CS Acute Care™ NT-proBNP method (pBNP) is an in vitro diagnostic assay for the quantitative determination of N-terminal pro-brain natriuretic peptide (NT-proBNP) in buman plasma. In individuals suspected of having congestive heart failure (CHF), measurements of NT-proBNP are used as an aid in the diagnosis and assessment of severity. The test is further indicated for the risk stratification of patients with acute coronary syndrome and heart failure. The NT-proBNP Calibrator (pBNP CalPak) is an in vitro diagnostic product intended to be used for calibration of the NT-proBNP method on the Stratus® CS analyzer. The NT-proBNP Dilution Pak (DilPak) is an in vitro diagnostic product intended to be used for dilution of the NT-proBNP Method on the Stratus® CS analyzer. ✓ 510(k) Numbe Prescription Use (Part 21 CFR 801 Subpart D) ------------------------- and/or Over-the-counter Use (21 CFR 801 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) (Division Sign-Off) Division of Clinical Laboratory Devices
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...