Diazyme Potassium Enzymatic Assay Kit in conjunction with Diazyme Potassium Low and High Calibrators, are intended for the quantitative determination of sodium (K) in serum. Diazyme Potassium Enzymatic Assay Kit contains a low level standard and a high level standard. The standards are used to generate a linear graph that will be used in the calculation of potassium concentrations in unknown serum samples. Diazyme Potassium Enzymatic Assay has controls for normal serum potassium level and abnormal serum potassium level. The controls are used as reference samples for checking the functionality of the Diazyme Potassium Enzymatic Assay.
Device Story
Spectrophotometric enzymatic assay for quantitative determination of potassium ions in human serum. Reagents (diluent 1/enzyme 1 and diluent 2/substrate 2) mixed with serum sample; incubated at 37 °C. Absorbance measured at 450 nm at 3 and 5 minutes. Uses bi-level serum calibrators (3 mM and 7 mM) to generate linear calibration curve. Controls (normal 4.4 mM, abnormal 6.4 mM) verify assay functionality. Performed manually or on automated chemistry analyzers. Output is potassium concentration (mmol/L) used by clinicians to assess patient electrolyte status.
Clinical Evidence
Bench testing only. Performance evaluated via correlation study with the predicate device (r=0.96). Analytical recovery averaged 104% and 97% for spiked serum samples. Precision and linearity were assessed. Interference studies conducted for various substances (e.g., NH4Cl, NaPi, CaCl2, NaCl, CuCl2, ZnCl2, FeCl3, Ascorbic Acid, Glucose, Bilirubin) showed minimal interference at tested concentrations.
Technological Characteristics
Two-reagent (R1 and R2) kinetic assay system. Principle: Spectrophotometric measurement at 450 nm. Measuring range: 2 to 8 mmol/L. Compatible with automated clinical analyzers. No software algorithm class specified; relies on standard spectrophotometric absorbance quantification.
Indications for Use
Indicated for quantitative determination of potassium in human serum for prescription use.
Regulatory Classification
Identification
A potassium test system is a device intended to measure potassium in serum, plasma, and urine. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.
K053566 — MODIFICATION TO: DIAZYME POTASSIUM ENZYMATIC ASSAY, MODEL DZ113A · Diazyme Laboratories · Feb 15, 2006
K091455 — DIAZYME LIQUID STABLE ENZYMATIC POTASSIUM ASSAY KIT, MODEL DZ113C-X, DIAZYME LIQUID STABLE ENZYMATIC POTASSIUM ASSAY CAL · Diazyme Laboratories · Aug 14, 2009
K043267 — POTASSIUM TEST KIT AND ELECTROLYTE CAL 1 AND 2 · Randox Laboratories, Ltd. · Feb 10, 2005
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE
A. 510(k) Number:
k042191
B. Purpose for Submission:
New Device
C. Analyte:
Potassium
D. Type of Test:
Quantitative
E. Applicant:
Diazyme Laboratories
F. Proprietary and Established Names:
Diazyme Potassium Enzymatic Assay Kit
Diazyme Enzymatic Potassium Serum Controls
G. Regulatory Information:
1. Regulation section:
21 CFR 862.1600 Potassium Test System
21 CFR 862.1660 Quality Controls
2. Classification:
Class II - 21 CFR 862.1600 Potassium Test System
Class I (reserved) - 21 CFR 862.1660 Quality Controls
3. Product Code:
Diazyme Potassium Enzymatic Assay Kit and Calibrators- MZV, JIT
Diazyme Enzymatic Potassium Serum Controls- JJX
4. Panel:
Chemistry (75)
H. Intended Use:
1. Intended use(s):
Diazyme Potassium Enzymatic Assay Kit in conjunction with Diazyme Potassium Low and High Calibrators, are intended for the quantitative determination of sodium (K) in serum.
{1}
Page 2 of 6
Diazyme Potassium Enzymatic Assay Kit contains a low level standard and a high level standard. The standards are used to generate a linear graph that will be used in the calculation of potassium concentrations in unknown serum samples.
Diazyme Potassium Enzymatic Assay has controls for normal serum potassium level and abnormal serum potassium level. The controls are used as reference samples for checking the functionality of the Diazyme Potassium Enzymatic Assay.
2. Indication(s) for use:
Diazyme Potassium Enzymatic Assay Kit in conjunction with Diazyme Potassium Low and High Calibrators, are intended for the quantitative determination of sodium (K) in serum.
Diazyme Potassium Enzymatic Assay Kit contains a low level standard and a high level standard. The standards are used to generate a linear graph that will be used in the calculation of potassium concentrations in unknown serum samples.
Diazyme Potassium Enzymatic Assay has controls for normal serum potassium level and abnormal serum potassium level. The controls are used as reference samples for checking the functionality of the Diazyme Potassium Enzymatic Assay.
3. Special condition for use statement(s):
For Prescription Use Only
4. Special instrument Requirements:
This assay can be performed manually or in an automated chemistry analyzer capable of maintaining 37 °C and reading absorbance at 450 nm.
I. Device Description:
The Diazyme Potassium Enzymatic Assay is a spectrophotometric assay intended for the quantitative determination of potassium ions in human serum. The assay includes high and low calibrators. The assay controls are sold separately. The Diazyme Potassium Enzymatic Assay consists of 4 components that are used in 2 stages (diluent 1 and enzyme 1 forms reagent 1 and diluent 2 and substrate 2 forms reagent 2). Reagent 1 is mixed with a serum sample and incubated at 37 °C and followed by reagent 2. The absorbance is read at 450 nm after 3 minutes and then again after 2 additional minutes.
The Diazyme Potassium Enzyme Calibrator set is a bi-level serum calibrator with high (7 mM) and low (3 mM) potassium levels. The calibrator is prepared with Tris, pH 7.5 and potassium chloride equivalent to the high or low level calibrator. The Diazyme Enzymatic Potassium Serum Control kit is sold as a lyophilized bi-level serum control with normal (4.4 mM) and abnormal (6.4 mM) potassium levels. The controls are pooled sodium free human sera that were tested and found negative for HIV, HBV and HCV using FDA approved methods.
{2}
Page 3 of 6
# J. Substantial Equivalence Information:
1. Predicate device name(s): Sychron LX I 725 Clinical System
2. Predicate K number(s): k023049
3. Comparison with predicate:
Diazyme Potassium Enzyme Assay Kit
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Indications for Use | Diazyme Potassium Enzymatic Assay Kit is for quantitative in vitro determination of Potassium in human serum. | Synchron LX Systems is used for quantitative in vitro determination of Potassium in human serum, plasma and urine |
| Sample | Serum | Serum, Plasma and Urine |
| Analyte | Potassium | Potassium |
| Differences | | |
| Item | Device | Predicate |
| Test Principle | Potassium is determined enzymatically via potassium dependent urea amidolyase activity with 1-methoxy-5-methyl-phenazinium methyl sulfate (PMS) as a substrate. After a series of reactions, the absorbance at 450 nm of the product Formazan is proportional to the potassium concentration. | Potassium is determined through a microprocessor-based instrument using ion-selective electrodes. |
| Assay Range | 2-8 mmol/L | 1-15 mmol/L |
{3}
Page 4 of 6
Diazyme Enzymatic Potassium Serum Controls
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Indications for Use | Diazyme Potassium Enzymatic Assay has controls for normal serum Potassium level and abnormal serum Potassium level. The controls are used as reference samples for checking the functionality of the Diazyme Potassium Enzymatic Assay. | Synchron LX Systems is has controls for 2 serum Potassium levels. The Controls are used as a reference for checking the functionality of the Synchron LX System. |
| Analyte | Potassium | Potassium |
K. Standard/Guidance Document Referenced (if applicable):
None Referenced
L. Test Principle:
Potassium is determined enzymatically via potassium dependent urea amidolyase activity with 1-methoxy-5-methyl-phenazinium methyl sulfate (PMS) as a substrate. After a series of reactions, the absorbance at 450 nm of the product Formazan is proportional to the potassium concentration.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Two levels of potassium specimens, 4.4 mM and 6.4 mM, were tested within run 80 times time in 4 replicates over 20 days to obtain a CV of 3.2% and 3.0% respectively. The same levels were tested for precision to obtain a CV of 5.3% and 3.3% respectively.
The Diazyme Enzymatic Potassium Serum Controls were tested for reproducibility at two laboratories with the Hitachi 917 on serum control samples containing 3.9 mM (low) and 6.1 mM (high) potassium. Site 1 obtained with an average of 3.9 mM and 6.1 mM potassium for the low and high controls respectively. Site 2 obtained an average of 4.0 mM and 6.2 mM potassium for the low and high controls respectively.
b. Linearity/assay reportable range:
Linearity was tested on serum samples spiked with potassium that ranged from 2 mM to 10 mM. Samples were diluted with distilled water or potassium chloride to achieve targeted concentrations. The linearity equation was Y = 0.0777x + 0.3784 with a R² of .9952 using the Cobas Mira.
{4}
Page 5 of 6
c. Traceability (controls, calibrators, or method):
The Diazyme Potassium controls are sold separately and are prepared with pooled human sera that were tested for HIV, HBV and HCV using FDA approved methods.
The Diazyme Potassium calibrators are included with the Diazyme Potassium Assay and are prepared gravimetrically using commercially available solutions.
Both a real time and an accelerated stability study were conducted for the Diazyme Enzymatic Potassium Assay kit and they supported a stability of 8 months at 2-8 °C for the lyophilized form and 14 days at 2-8 °C for the reconstituted form.
The Diazyme Potassium Enzymatic Calibrators were studied in real time at 4 °C and the results supported a stability of 6 months at 4 °C. There is an ongoing real time study.
The Diazyme Potassium Serum Controls were studied using an accelerated rate equation and a shelf life of 3 months at 2-8 °C in the lyophilized form and 14 days at 2-8 °C.
d. Detection limit:
See Linearity Above
e. Analytical specificity:
Interference was tested on cations and substances normally present in serum with the Diazyme Enzymatic Potassium Assay by spiking 4.4 mM potassium serum samples. The study found less than 10% deviation when tested at the following levels and substances: 1 mM NH₄Cl, 1.5 mM NaPi, 5 mM CaCl₂, 200 mM NaCl, 0.25 mM CuCl₂, 0.25 ZnCl₂, 0.025 mM FeCl₃, 5 mM Ascorbic Acid, 5 mM Glucose, 10 mg/dL Bilirubin and 500 mg/dL Triglyceride.
f. Assay cut-off:
See Linearity Above.
2. Comparison studies:
a. Method comparison with predicate device:
Thirty four serum samples, 26 diluted samples and eight calibrators and/or controls were tested with both the Diazyme Enzymatic Potassium Assay and the Synchron LX I 725 Clinical System. An equation of Y = 1.001X + 0.18 was obtained with an R² of .9665.
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical sensitivity:
Not applicable
{5}
Page 6 of 6
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a and b are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
A normal potassium serum values range from 3.5 to 5.5 mM (13.7- 19.9 mg/dL) was provided as obtained from the literature.
N. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.