ATAC DIRECT LDL REAGENT AND ATAC DIRECT LDL CALIBRATOR KITS
Applicant
Elan Diagnostics
Product Code
MRR · Clinical Chemistry
Decision Date
Apr 12, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1475
Device Class
Class 1
Indications for Use
The ATAC Direct LDL Reagent Kit, the ATAC Direct LDL Calibrator and the ATAC 8000 Random Access Chemistry System are intended for the quantitative determination of LDL-cholesterol in serum and plasma. These are intended for the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases, and for the assessment of the risk of developing cardiovascular disease. This reagent is intended to be used by trained personnel in a professional setting and is not intended for home use.
Device Story
ATAC Direct LDL Reagent Kit and Calibrator used on ATAC 8000 Random Access Chemistry System; performs quantitative in vitro diagnostic measurement of LDL-cholesterol in serum/plasma. Two-step enzymatic reaction: first step consumes non-LDL cholesterol without color production; second step uses detergent to solubilize LDL, reacting with cholesterol oxidase, cholesterol esterase, and peroxidase to produce color proportional to LDL concentration. Used in clinical laboratories by trained personnel; output is numerical LDL-cholesterol concentration. Healthcare providers use results to diagnose lipid disorders, monitor treatment, and assess cardiovascular risk. Benefits include standardized, automated quantification of LDL-cholesterol to guide clinical management of metabolic and cardiovascular conditions.
Clinical Evidence
Bench testing only. Linearity established from 5 to 700 mg/dL. Precision studies (n=59-60) showed within-run CVs of 1.5-2.2% and total CVs of 2.7-3.5%. Method comparison against a competitive reagent (n=159) yielded r=0.996. Comparison against beta-quantification reference method (n=42) yielded r=0.986. Detection limit documented at 0.73 mg/dL. Stability studies confirmed 14-day on-board and calibration stability with total imprecision <3%.
Technological Characteristics
In vitro diagnostic reagent kit and calibrator. Two-step enzymatic colorimetric assay (cholesterol oxidase/esterase/peroxidase). Designed for use on ATAC 8000 Random Access Chemistry System. Automated spectrophotometric detection. No specific material standards or software algorithm class specified.
Indications for Use
Indicated for quantitative determination of LDL-cholesterol in serum and plasma for adult patients to diagnose/treat lipid disorders (e.g., diabetes mellitus, atherosclerosis, liver/renal diseases) and assess cardiovascular disease risk. Not for home use.
Regulatory Classification
Identification
A lipoprotein test system is a device intended to measure lipoprotein in serum and plasma. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
K161691 — Direct LDL Cholesterol (LDL) · Randox Laboratories Limited · Mar 20, 2017
K032936 — DIRECT LDL CHOLESTEROL LIQUID COLOR AND STANBIO DIRECT HDL/LDL CHOLESTEROL CALIBRATOR · Stanbio Laboratory · Mar 12, 2004
K022772 — RAICHEM LDL-CHOLESTEROL DIRECT REAGENT · Hemagen Diagnostics, Inc. · Sep 17, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
élan diagnostics
Image /page/0/Picture/2 description: The image shows a logo for "élan". The logo features a stylized letter "e" formed by multiple parallel lines, creating a sense of movement or flow. Below the symbol, the word "élan" is written in a bold, sans-serif font, with an acute accent over the "e".
020575
## SUMMARY OF 510(K) SAFETY AND EFFECTIVENESS INFORMATION
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The ATAC Direct LDL Reagent Kit, the ATAC Direct LDL Calibrator and the ATAC 8000 Random Access Chemistry THE ATAC DICCLEDE Reagent Not the quantitative determination of LDL-cholesterol in serum and planna. System ate intended for the diagnosis and treatment of lipid disorders (such as diabetes mellius), atherosclerosis, LDL-choices liver and renal diseases, and for the assessment of the risk of developing cardiovascular disease.
The ATAC Direct LDL Reagent determines LDL-cholesterol through a two step reaction, which depecial The ATAC Direct LDL Neagult determines in checked are consumed in a series of reactions without producing color. In the second reaction step, the chromogen is added, and a second detergent solubilizes the LDLproducing color. In ano socond receit with cholesterol oxidase, cholesterol esterase and peroxidase in the reagent to produce a color that is proportional to the amount of LDL-cholesterol in the sample.
The ATAC Direct LDL Reagent Kit, calibrated with the ATAC Direct LDL Calibrator, is substantially equivalent of the collibrary The ATAC Ditch LDL Reagent Tit, callerined with 1.80-4598-00, calibrated with N-geneous LDL Cholesterol Calibrator, N-gencous DDI Choicerer reagent ready marketed by Genzyme Corporation of Cambridge, MA.
The effectiveness of ATAC Direct LDL Reagent Kit and the ATAC Direct LDL Calibrator used on the ATAC 8000 Random Access Chemistry System is shown by the following studies.
The recovery of LDL-cholesterol using the ATAC Direct LDL Reagent is linear from 5 to 700 mg/dL as shown by the I ne recovery of LDL-notestere used insable range. Linear regression statistics, with the regression line forced through the origin, compare standard recoveries to standard factors.
(ATAC Recoveries) = 0 mg/dL + 1.023 x (Standard Factors), n = 35
Precision is demonstrated by the replicate assay of commercially available control serum and a serum pool. Precision statistics, calculated analogous to the method described in NCCLS Guideline EP3-T, are shown below.
| | Precision of LDL-Cholesterol Recoveries in mg/dL | | | | | |
|---------|--------------------------------------------------|------|------------|------|-------|------|
| Sample | n | mean | Within Run | | Total | |
| | | | 1SD | %CV | 1SD | %CV |
| Serum 1 | 60 | 93 | 1.4 | 1.5% | 2.7 | 2.9% |
| Serum 2 | 60 | 184 | 3.3 | 1.8% | 4.9 | 2.7% |
| Serum 3 | 59 | 555 | 11.9 | 2.2% | 19.2 | 3.5% |
Mixed serum and plasma specimens, collected from adult patients, were assayed for LDL cholesterol using the ATAC 8000 Random Access Chemistry System and another commercially available method. Results were compared by least squares linear regression and the following statistics were obtained.
Serum/Plasma Comparison ATAC 8000 = - 1.7 mg/dL + 1.078 x Competitive Reagent range = 19 - 265 mg/dL r = 0.996 n = 159
510(k) Notification, ATAC Direct LDL Reagent Kit, 20 Sept., 2002, p 57
Elan Diagnostics 1075 W. Lambert Road · Brea California · 92821 Phone: 714.672.3553 · Fax: 714.672.3554
Elan Diagnostics is a division of Elan Pharmaceuticals
{1}------------------------------------------------
To verify accuracy, 42 serum and plasma specimens, collected from 22 adults were assayed for LDL cholesterol using the ATAC 8000 Random Access Chemistry System and the beta-quantification reference method. Results were compared by least squares linear regression and the following statistics were obtained.
Serum/Plasma Comparison ATAC 8000 = - 4.2 mg/dL + 1.035 x reference method r = 0.986 n = 42 range = 59 - 178 mg/dL
The detection limit claim of 5 mg/dL is documented through the repetitive assay of a diluted serum pool. The observed detection limit, calculated as two standard deviations of a 30 replicate within run precision study, is 0.73 mg/dL and is below the claimed limit.
The 14 day on board reagent stability and 14 day calibration stability claims are documented through the assay of serum controls over the claimed periods. In all cases, the total imprecision of LDL-cholesterol recoveries over the test periods are less than 3%.
Wyman Stocking
Wynn Stocking Manager of Regulatory Affairs Elan Diagnostics
510(k) Notification, ATAC Direct LDL Reagent Kit, 20 Sept., 2002, p 57
Elan Diagnostics 1075 W. Lambert Road · Brea California 92821 Phone: 714.672.3553 · Fax: 714.672.3554
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of a caduceus, a symbol often associated with medicine and healthcare. The caduceus is depicted with a staff entwined by two snakes and topped with wings.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
PR 1 2 2002
Mr. Wynn Stocking Manager, Regulatory Affairs Elan Diagnostics 1075 W. Lambert Road, Building D Brea, California 92821
Re: k020575
Trade/Device Name: ATAC Direct LDL Reagent and ATAC Direct LDL Calibrator Regulation Number and Name: 21 CFR § 862.1475-Low density lipoprotein test system Regulation Number and Name: 21 CFR § 862.1150-Calibrator, secondary Product Code: MRR Regulatory Class: I Regulatory Class: II Product Code: JIT Dated: February 20, 2002 Received: February 21, 2002
Dear Mr. Stocking:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket I this lotter will and hype and al equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and If you dosire opening ad in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, (Jorg Sy Potter of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small mornation on 90 m 100 Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known):
02-0575
ATAC Direct LDL Reagent and ATAC Direct LDL Calibrator
Device Name:
Indications For Use:
The ATAC Direct LDL Reagent Kit, the ATAC Direct LDL Calibrator and the ATAC 8000 Random Access Chemistry THE ATAC DIFOLLEDE Notes and INT, the Quantitative determination of LDL-cholesterol in serum and plasma. LDL-System are included to the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atheroscherosis, and various liver and renal diseases, and for the assessment of the risk of developing cardiovascular disease.
This reagent is intended to be used by trained personnel in a professional setting and is not intended for home use.
Respectfully,
Wynn Stocking
Regulatory Affairs Manager Elan Diagnostics
20 February, 2002
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use __
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.