The Roche Diagnostics OMNI S Analyzer is a fully automated critical care analyzer intended to be used for the measurement of pH, PO2, PCO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, bilirubin, total hemoglobin, oxygen saturation, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin and methemoglobin in samples of whole blood, serum, plasma and aqueous solutions as appropriate.
Device Story
The OMNI S Analyzer is a fully automated critical care analyzer used in clinical settings. It incorporates an oximeter module, which is an optical sensor module, to measure bilirubin levels in patient samples. The device utilizes spectrophotometric analysis based on Lambert-Beer's law to determine bilirubin concentration. By measuring various hemoglobin derivatives, the instrument provides a bilirubin measurement that is corrected for hemoglobin interference. The output is a quantitative bilirubin value, which assists clinicians in the management of neonatal jaundice and related conditions. The device is intended for prescription use.
Clinical Evidence
Bench testing only. Precision evaluated using Autotrol Plus levels and whole blood (CVs ≤1.67% and ≤7.56% respectively). Linearity demonstrated across 6.86-41.52 mg/dL range (R=0.9996). Method comparison against four commercial methods (Hitachi, Beckman, Kodak, Radiometer) showed strong correlation (R=0.974-0.986). Interference studies conducted per NCCLS EP7-P; methylene blue identified as an interferent.
Technological Characteristics
Fully automated critical care analyzer. Measures bilirubin via spectrophotometric or electrochemical sensing principles (implied by critical care analyzer class). System supports whole blood, serum, plasma, and aqueous solutions. Connectivity and software details not specified. Sterilization and material specifications not provided.
Indications for Use
Indicated for the quantitative measurement of bilirubin in whole blood, serum, plasma, and aqueous solutions in neonates.
Regulatory Classification
Identification
A bilirubin (total and unbound) in the neonate test system is a device intended to measure the levels of bilirubin (total and unbound) in the blood (serum) of newborn infants to aid in indicating the risk of bilirubin encephalopathy (kernicterus).
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE
A. 510(k) Number:
K040174
B. Purpose for Submission:
Addition of bilirubin parameter on the Omni S Analyzer
C. Analyte:
Bilirubin (total & unbound) in neonate test system
D. Type of Test:
Quantitative
E. Applicant:
Roche Diagnostics Corporation
F. Proprietary and Established Names:
Bilirubin Assay on the OMNI S Analyzer
G. Regulatory Information:
1. Regulation section:
21 CFR 862.1113
2. Classification:
I
3. Product Code:
MQM
4. Panel:
75
H. Intended Use:
1. Intended use(s):
The Roche Diagnostics OMNI S Analyzer is a fully automated critical care analyzer intended to be used for the measurement of pH, PO2, PCO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, bilirubin, total hemoglobin, oxygen saturation, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin and methemoglobin in samples of whole blood, serum, plasma and aqueous solutions as appropriate.
2. Indication(s) for use:
The Roche Diagnostics OMNI S Analyzer is a fully automated critical care analyzer intended to be used for the measurement of pH, PO2, PCO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN,
{1}
Page 2 of 4
bilirubin, total hemoglobin, oxygen saturation, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin and methemoglobin in samples of whole blood, serum, plasma and aqueous solutions as appropriate.
3. Special condition for use statement(s):
Prescription use
4. Special instrument Requirements:
OMNI S Analyzer (K032311)
I. Device Description:
Refer to K032311 for details on the fully automated OMNI S instrument. Contained within the instrument is the oximeter module, an optical sensor module that is used for determining bilirubin.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Radiometer ABL735
2. Predicate K number(s):
K991417
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Quantitative determination of neonatal bilirubin in newborns using whole blood samples | Quantitative determination of neonatal bilirubin in newborns using whole blood samples |
| Technology | Automated co-oximetry | Automated co-oximetry |
| Differences | | |
| Item | Device | Predicate |
| Instrument | OMNI S | ABL735 |
K. Standard/Guidance Document Referenced (if applicable):
None referenced
L. Test Principle:
Bilirubin is measured spectrophotometrically on the basis of Lambert-Beer's law.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Precision was evaluated using Autotrol Plus Levels 1-3, 4B and 5B, and three levels of whole blood, forty (40) runs each. The Autotrol levels ranged between 4 and 24 mg/dL. The whole blood levels
{2}
Page 3 of 4
ranged between 8 and 44 mg/dL. The CVs for the Autotrol materials were ≤1.67%. The CVs for the human whole blood were ≤7.56%.
b. Linearity/assay reportable range:
The assay reportable range is 3-50 mg/dL. Linearity was evaluated using human whole blood (3 measurements per level) and two OMNI S instruments. The mean values ranged from 6.86 to 41.52 mg/dL, and the recovery ranged from 93.85 to 114.33%. The correlation results were as follows: slope (0.9038-1.1064), intercept (+/-1.514), and correlation coefficient (0.9996).
c. Traceability (controls, calibrators, or method):
Refer to K032311.
d. Detection limit:
The detection limit is 3 mg/dL. This is supported in various method comparison and verification studies.
e. Analytical specificity:
Interference studies were performed on methylene blue, Intralipid, Lipofundin (triglyceride), Propofol, Indocyanine, Beta-carotin, and Evans Blue. Hemoglobin interference was deemed unnecessary due to the design of the instrument. The measured concentration of hemoglobin derivatives is used to provide a bilirubin measurement virtually free of interference from hemoglobin.
The interference studies were performed according to NCCLS Document EP7-P "Interference Testing in Clinical Chemistry." The analysis performed on all potential interfering substances was the control group bias versus the reference value. Methylene blue was found to be an interferent.
f. Assay cut-off:
See Detection limit above.
2. Comparison studies:
a. Method comparison with predicate device:
The bilirubin assay on the OMNI S analyzer was compared to four commercially available methods and yielded the following results:
| Comparison method | N | Slope and Intercept | Bias | R value |
| --- | --- | --- | --- | --- |
| Hitachi TBil | 85 | y =0.968x - 0.127 | -3.7% | 0.986 |
| Beckman LX 20 tBil | 76 | y = 1.060x - 0.537 | +1.4% | 0.980 |
| Kodak Vitros tBil | 73 | y = 0.988x - 0.119 | -2.4% | 0.984 |
| Radiometer | 82 | y = 1.044x + 0.327 | +10.5% | 0.974 |
{3}
Page 4 of 4
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a and b are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
The expected values/reference ranges were established based on the literature. Specific reference was made to Tietz “Textbook of Clinical Chemistry” 3rd Edition 1999.
N. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.