CARESTART CARDIAC 3-IN-1 TROPONIN 1/CK-MB/MYOGLOBIN

K051925 · Access Bio Incorporate · MMI · Nov 29, 2005 · Clinical Chemistry

Device Facts

Record IDK051925
Device NameCARESTART CARDIAC 3-IN-1 TROPONIN 1/CK-MB/MYOGLOBIN
ApplicantAccess Bio Incorporate
Product CodeMMI · Clinical Chemistry
Decision DateNov 29, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1215
Device ClassClass 2

Intended Use

CareStart™ Cardiac Troponin I Test is intended for the qualitative detection of cardiac troponin I in human blood, serum, or plasma. Measurement of cardiac troponin I aids in rapid diagnosis of acute myocardial infarction. The performance characteristics of the test have not been evaluated in a point-of-care (POC) setting. CareStart™ Cardiac CKMB Test is intended for the qualitative detection of cardiac CKMB in human blood, serum, or plasma. Measurement of cardiac CKMB aids in rapid diagnosis of acute myocardial infarction. The performance characteristics of the test have not been evaluated in a point-of-care (POC) setting. CareStart™ Cardiac 2-in-1 Troponin I/Myoglobin Test is intended for the qualitative detection of cardiac troponin I, and myoglobin in human blood, serum, or plasma. Measurement of cardiac troponin I, and myoglobin aids in rapid diagnosis of acute myocardial infarction. The performance characteristics of the test have not been evaluated in a point-of-care (POC) setting. CareStart™ Cardiac 3-in-1 Troponin I/CKMB/Myoglobin Test is intended for the qualitative detection of cardiac troponin I, CKMB and myoglobin in human blood, serum, or plasma. Measurement of cardiac troponin I, CKMB and myoglobin aids in rapid diagnosis of acute myocardial infarction. The performance characteristics of the test have not been evaluated in a point-of-care (POC) setting.

Device Story

Lateral flow immunochromatographic assay; detects cardiac troponin I, CK-MB, and myoglobin. Input: human blood, serum, or plasma sample applied to well. Mechanism: red blood cell separation membrane; analyte complexes with dye-labeled antibody and biotinylated antibody; complex migrates to immobilized streptavidin on membrane. Output: visual pinkish-purple bands in test area if analyte concentration exceeds cutoff (Troponin I 0.6 ng/mL, Myoglobin 70 ng/mL, CK-MB 5.0 ng/mL); control band confirms validity. Used in clinical settings; operated by healthcare professionals. Results aid rapid diagnosis of acute myocardial infarction. Benefits: provides qualitative diagnostic information for cardiac events.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Solid-phase chromatographic immunoassay. Components: plastic housing, chromatographic membrane, red blood cell separation membrane, dye pad with monoclonal antibodies. Detection principle: streptavidin-biotin binding. Qualitative visual readout. No energy source required. Shelf stability: 8 months at room temperature.

Indications for Use

Indicated for qualitative detection of cardiac troponin I, CKMB, and/or myoglobin in human blood, serum, or plasma to aid in the diagnosis of acute myocardial infarction. Prescription use only.

Regulatory Classification

Identification

A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 NOV 2 9 2005 Young Ho Choi Access Bio, Inc. 2033 Route 130, Unit H Monmouth Junction, NJ 08852 Re: k051925 Trade/Device Name: CareStart Cardiac 3-in-1 Troponin I/CK-MB/Myoglobin Test CareStart Cardiac 2-in-1 Troponin I/Myoglobin Test CareStart Cardiac Troponin I CareStart Cardiac CK-MB Regulation Number: 21 CFR 862.1215 Regulation Name: Creatine phosphokinase/creatine kinase or isoenzymes test system Regulatory Class: Class II Product Code: MMI, JHT, DDR Dated: October 24, 2005 Received: October 26, 2005 Dear Mr. Choi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {1}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Alberto Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ 510(k) Number (if known): k051925 Device Name: CareStart™ Cardiac 3-in-1 Troponin I/CK-MB/Myoglobin Test Indications For Use: CareStart™ Cardiac 3-in-1 Troponin I/CK-MB/Myoglobin Test is intended for the qualitative detection of cardiac troponin I, CK-MB and myoglobin in human blood, serum qualitative detection of cardiac troponin I, CK-MB and myoglobin aids in rapid diagnosis of acute myocardial infarction. The performance characteristics of the test have not been evaluated in a point-of-care (POC) setting. Prescription Use _ × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Lucy Phillips Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety (k) K051925 Page 1 of {3}------------------------------------------------ 510(k) Number (if known): k051925 Device Name: CareStart™ Cardiac 2-in-1 Troponin I/Myoglobin Test Indications For Use: CareStart™ Cardiac 2-in-1 Troponin I/Myoglobin Test is intended for the qualitative detection of cardiac troponin I, and myoglobin in human blood, serum or plasma. detooking of cardiae troponin I, and myoglobin aids in rapid diagnosis of acute myocardial infarction. The performance characteristics of the test have not been evaluated in a point-of-care (POC) setting. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Page 1 of ____________________________________________________________________________________________________________________________________________________________________ Tracy Phillips Division Sign-Off ice of In Vitro Diagnostic Device valuation and Safety K051925 {4}------------------------------------------------ 510(k) Number (if known): k051925 Device Name:_CareStart™ Cardiac CKMB Test Indications For Use: CareStart™ Cardiac CKMB Test is intended for the qualitative detection of CK-MB in human blood, serum or plasma. Measurement of CK-MB aids in rapid diagnosis of numan blood, corant of on. The performance characteristics of the test have not been evaluated in a point-of-care (POC) setting. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Page 1 of Lucy Phillips Vision Sign-Off fice of In Vitro Diagnostic Device valuation and Safety K051925 {5}------------------------------------------------ 510(k) Number (if known): k051925 Device Name:_CareStart™ Cardiac Troponin 1 Test Indications For Use: CareStart™ Cardiac Troponin I Test is intended for the qualitative detection of cardiac oarcotar troponiir in haman broomyocardial infarction. The performance characteristics of the test have not been evaluated in a point-of-care (POC) setting. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Page 1 of Tracy Phillips Sign-Off Office of In Vitro Diagnostic Device Valuation and Safety K051925
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...