FACT PLUS ONE STEP PREGNANCY TEST

K990373 · Abbott Laboratories · LCX · Feb 22, 1999 · Clinical Chemistry

Device Facts

Record IDK990373
Device NameFACT PLUS ONE STEP PREGNANCY TEST
ApplicantAbbott Laboratories
Product CodeLCX · Clinical Chemistry
Decision DateFeb 22, 1999
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1155
Device ClassClass 2

Indications for Use

The Fact plus® One Step Pregnancy Test is a qualitative, consumer home-use, in vitro diagnostic immunoassay for the detection of human chorionic gonadotropin (hCG) in human urine for the early detection of pregnancy.

Device Story

Fact plus One Step Pregnancy Test is a qualitative in vitro diagnostic immunoassay; detects human chorionic gonadotropin (hCG) in human urine. Designed for consumer home-use; provides early pregnancy detection. Device utilizes immunoassay technology to identify hCG presence in urine samples. User performs test at home; interprets visual results directly from device. Output informs user of pregnancy status. Benefits include convenient, rapid, private pregnancy confirmation without clinical visit.

Clinical Evidence

No clinical data provided; substantial equivalence based on design control documentation and performance characteristic comparisons to the predicate device.

Technological Characteristics

Qualitative in vitro diagnostic immunoassay; lateral flow or similar test strip format for hCG detection in urine. Standalone device; no electronic components or software.

Indications for Use

Indicated for qualitative detection of human chorionic gonadotropin (hCG) in human urine for early pregnancy detection in a consumer home-use setting.

Regulatory Classification

Identification

A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 2/22/99 Image /page/0/Picture/1 description: The image shows a handwritten string of characters, "K990373". The characters are written in a dark ink, and the handwriting is somewhat stylized. The string appears to be a code or serial number. ## Special 510(k) Summary Fact plus® One Step Pregnancy Test ## Information Supporting a Substantially Equivalent Determination The conformance to Design Control as presented in the Special 510(k) for the Fact plus One Step Pregnancy Test documents that the modifications made to the predicate device to prepare the modified device have not changed the original intended use or performance characteristics of the Fact plus One Step Pregnancy Test. The intended use of both pregnancy tests is as a qualitative, consumer home-use, in vitro diagnostic immunoassay for the detection of human chorionic gonadotropin (hCG) in human urine. The modified device has demonstrated comparable performance characteristics to the predicate device. This information is documented in the Design History File for the assay, maintained at Abbott Laboratories. Prepared and Submitted February 5, 1999 by: Abbott Laboratories Katherine M. Wortley, Ph.D. Regulatory Specialist ADD Regulatory Affairs 1-847-938-7707 Bldg. AP31, Dept. 9V6 200 Abbott Park Road Abbott Park, IL 60064-6200 {1}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Image /page/1/Picture/2 description: The image shows the date February 2, 1999. The month is represented by the abbreviation FEB. The day is the number 2, and the year is 1999. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Katherine M. Wortley, Ph.D. Regulatory Specialist ADD Regulatory Affairs Abbott Laboratories Diagnostic Division Dept. 9V6 Building AP31 200 Abbott Park Road Abbott Park, Illinois 60064-6200 Re: K990373 > Trade Name: Fact Plus One Step Pregnancy Test Regulatory Class: II Product Code: LCX Dated: February 5, 1999 Received: February 8, 1999 Dear Dr. Wortley: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2}------------------------------------------------ ## Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Steven Sutman Steven I. Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ 510(k) Number (if known): __ Device Name: Fact plus® One Step Pregnancy Test Indications For Use: The Fact plus® One Step Pregnancy Test is a qualitative, consumer home-use, in vitro diagnostic immunoassay for the detection of human chorionic gonadotropin (hCG) in human urine for the early detection of pregnancy. K990373 Sean Cooper (Division Sign-Off) vision of Clinical I aboratory Devices 510(k) Number 190372 (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The Over-The-Counter Use (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...