QUICKSTREAM ONE STEP PREGNANCY TEST MODEL NUMBER 9012

K980532 · Phamatech · LCX · Mar 11, 1998 · Clinical Chemistry

Device Facts

Record IDK980532
Device NameQUICKSTREAM ONE STEP PREGNANCY TEST MODEL NUMBER 9012
ApplicantPhamatech
Product CodeLCX · Clinical Chemistry
Decision DateMar 11, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1155
Device ClassClass 2

Indications for Use

Over-the-Counter HCG (pregnancy) tests are for those individuals who, for any reason, believe they may be pregnant. This test detects HCG in urine. HCG is a hormone produced by the placenta shortly after implantation. Since HCG is present in the urine of a pregnant woman, it is an excellent marker for confirming pregnancy. Early identification is important in the management of pregnancy. This condition must be identified so that fetal exposure to potentially harmful circumstance may be minimized or eliminated. Many individuals hesitate to seek clinical confirmation of this condition (pregnancy) due to medical examination costs and appointment scheduling difficulties. This device offers an alternative to the initial office visit and will allow for more rapid initiation of pre-natal care upon confirmation tests by trained clinical technicians.

Device Story

QuickStream Pregnancy Test is an over-the-counter diagnostic device for qualitative detection of human chorionic gonadotropin (HCG) in urine. Device utilizes lateral flow immunoassay technology; urine sample applied to test strip; HCG binds to specific antibodies; visual indicator confirms presence of hormone. Intended for home use by individuals to identify pregnancy early; facilitates rapid initiation of prenatal care. Results provide preliminary confirmation; users encouraged to seek clinical follow-up. Device serves as alternative to initial office visit for pregnancy testing.

Clinical Evidence

No clinical data provided in the document; bench testing only.

Technological Characteristics

Lateral flow immunoassay; qualitative detection of HCG in urine; visual readout; non-electronic; standalone diagnostic test.

Indications for Use

Indicated for individuals who believe they may be pregnant to detect HCG in urine as a marker for pregnancy.

Regulatory Classification

Identification

A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The logo is presented in black and white. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Carl Monqiovi Director of Operations Phamatech 9265 Activity Road #112 San Diego, California 92126 MAR | | 1998 Re : K980532 QuickStream One Step Pregnancy Test Regulatory Class: II Product Code: LCX Dated: February 9,1998 Received: February 11, 1998 Dear Mr. Mongiovi: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set -forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or ... - requlations. {1}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" . Sincerely yours, Steven Slutman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## INDICATIONS FOR USE Applicant: Phamatech 510(k) Number (if known):(k) 980532 Device Name: QuickStream Pregnancy Test Indications for Use: Over-the-Counter HCG (pregnancy) tests are for those individuals who, for any reason, believe they may be pregnant. This test detects HCG in urine. HCG is a hormone produced by the placenta shortly after implantation. Since HCG is present in the urine of a pregnant woman, it is an excellent marker for confirming pregnancy. Early identification is important in the management of pregnancy. This condition must be identified so that fetal exposure to potentially harmful circumstance may be minimized or eliminated. Many individuals hesitate to seek clinical confirmation of this condition (pregnancy) due to medical examination costs and appointment scheduling difficulties. This device offers an alternative to the initial office visit and will allow for more rapid initiation of pre-natal care upon confirmation tests by trained clinical technicians. ## PLEASE DO NOT WRITE BELOW THIS LINE Concurrence of CDRH Office of Device Evaluation (ODE) | (Division Sign-Off) | | |-----------------------------------------|---------| | Division of Clinical Laboratory Devices | | | 510(k) Number | K980532 | Division Sign-Off Division of Clinical Laboratory Devices | <b>510(k) Number:</b> | | |-----------------------|--| |-----------------------|--| | Prescription Use:<br>Per 21 CFR 801.109 | OR Over the Counter | |-----------------------------------------|---------------------| |-----------------------------------------|---------------------|
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...