QUICKSTICK ONE STEP PREGNANCY TEST(9010)
Device Facts
| Record ID | K962219 |
|---|---|
| Device Name | QUICKSTICK ONE STEP PREGNANCY TEST(9010) |
| Applicant | Phamatech |
| Product Code | LCX · Clinical Chemistry |
| Decision Date | Jul 23, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1155 |
| Device Class | Class 2 |
Indications for Use
The Quickstick™ One Step HCG Pregnancy Test is intended to detect the presence of Human Chorionic Gonadotropin (hCG) in Urine for the women who suspect they may be pregnant. HCG is a well known and established analyte used to confirm pregnancy because of its early appearance in urine following conception followed by a dramatic increase in concentration.
Device Story
Lateral flow immunoassay for qualitative hCG detection in urine; utilizes visual color sandwich immunoassay technology; antibody-hCG-antibody complex formation indicates presence of analyte. Used by lay users for home pregnancy testing. Provides visual qualitative result; assists in early pregnancy confirmation. Benefits patient through rapid, accessible pregnancy status determination.
Clinical Evidence
Clinical sample correlation study and blind labeled spiked HCG study performed. Correlation study (>99% agreement) compared against Quidel Rapidvue and Syntron Bioresearch Be Sure. Blind consumer study (n=158) demonstrated 100% sensitivity (79/79), 100% specificity (79/79), and 100% accuracy (158/158).
Technological Characteristics
Visual color sandwich immunoassay; lateral flow technology; antibody-based detection of hCG; qualitative output.
Indications for Use
Indicated for women suspecting pregnancy to qualitatively detect human chorionic gonadotropin (hCG) in urine.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
Predicate Devices
- Unipath Clearview HCG
- Abbott Laboratories’ Fact Plus
- Syntron Bioresearch Be Sure Pregnancy Test
Reference Devices
- Quidel Rapidvue
- Syntron Bioresearch Be Sure Pregnancy Test
Related Devices
- K981722 — QUIDEL HOME PREGNANCY TEST · Quidel Corp. · Jun 3, 1998
- K131236 — UCP PREGNANCY CASSETTE TEST, DIP/CARD STRIP, HOME PREGNANCY CASSETTE TEST, MIDSTREAM TEST AND DIP/CARD STRIP · Ucp Biosciences, Inc. · Nov 7, 2013
- K193132 — CLUNGENE HCG Pregnancy Rapid Test Cassette, CLUNGENE HCG Pregnancy Rapid Test Strip, CLUNGENE HCG Pregnancy Rapid Test Midstream · Hangzhou Clongene Biotech Co., Ltd. · Dec 11, 2019
- K032661 — ACON QUICK-CHECK II HOME PREGNANCY TEST DEVICE (CASSETTE) · ACON Laboratories, Inc. · Oct 14, 2003
- K212418 — hCG One Step Pregnancy Test · Healstone Biotech, Inc. · Jun 30, 2022