K212418 · Healstone Biotech, Inc. · LCX · Jun 30, 2022 · Clinical Chemistry
Device Facts
Record ID
K212418
Device Name
hCG One Step Pregnancy Test
Applicant
Healstone Biotech, Inc.
Product Code
LCX · Clinical Chemistry
Decision Date
Jun 30, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1155
Device Class
Class 2
Intended Use
The hCG One Step Pregnancy Test is an in vitro diagnostic test for the qualitative determination of human chorionic gonadotropin hormone in human urine samples. It is used as an aid in early detection of pregnancy, in some cases as early as five (5) days before the expected period, i.e., as early as six (6) days before the day of the missed period.
Device Story
Lateral flow sandwich immunochromatographic assay; detects human chorionic gonadotropin (hCG) in human urine. User applies urine via mid-stream or dip method to absorbent tip; sample migrates via capillary action along membrane. If hCG present, it reacts with monoclonal antibodies to form colored test line; control line confirms procedural validity. Used in home/point-of-care settings by lay users. Provides qualitative visual result (pregnant/not pregnant) within 3 minutes. Assists in early pregnancy detection; informs clinical decision-making regarding pregnancy status.
Clinical Evidence
Clinical specificity evaluated in 612 non-pregnant women (ages 18-66), showing 100% specificity. Early pregnancy detection study followed 59 pregnant women (ages 25-45) from day -9 to +1 of expected period; device detected hCG in 68% of samples 5 days before expected period and 100% of samples 2 days before. Lay-user study (n=202) demonstrated 100% agreement with professional users for both dip and midstream methods.
Technological Characteristics
Lateral flow sandwich immunochromatographic assay. Uses monoclonal antibodies for hCG detection. Form factor: mid-stream or dip test device. Qualitative visual readout via colored lines. Storage: 2-30°C. No electronic components, software, or external energy source.
Indications for Use
Indicated for qualitative detection of hCG in human urine to aid in early pregnancy detection. Intended for over-the-counter use by individuals seeking to determine pregnancy status.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
Predicate Devices
First Response Early Result Pregnancy Test (K123436)
Related Devices
K032661 — ACON QUICK-CHECK II HOME PREGNANCY TEST DEVICE (CASSETTE) · ACON Laboratories, Inc. · Oct 14, 2003
K081150 — RAPIDAN OPTIMA EARLY PREGNANCY TEST AND TOYO PREGNANCY TEST · Turklab Medical Devices, Inc. · Jun 16, 2009
K233174 — FaStep Pregnancy Rapid Test Strip; FaStep Pregnancy Rapid Test Midstream · Assure Tech. (Hangzhou) Co, Ltd. · Feb 5, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 30, 2022
Healstone Biotech Inc. % Guang Gao Principal Consultant Axteria Biomed Consulting 8040 Cobble Creek Circle Potomac, Maryland 20854
Re: K212418
Trade/Device Name: hCG One Step Pregnancy Test Regulation Number: 21 CFR 862.1155 Regulation Name: Human Chorionic Gonadotropin (HCG) Test System Regulatory Class: Class II Product Code: LCX Dated: May 5, 2022 Received: May 5, 2022
Dear Guang Gao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Marianela Perez-Torres, Ph.D. Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) k212418
Device Name hCG One Step Pregnancy Test
### Indications for Use (Describe)
The hCG One Step Pregnancy Test is an in vitro diagnostic test for the qualitative determination of human chorionic gonadotropin hormone in human urine samples. It is used as an aid in early detection of pregnancy, in some cases as early as five (5) days before the expected period, i.e., as early as six (6) days before the day of the missed period.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K212418
## 510(K) SUMMARY
Submitter's Identification: Healstone Biotech Inc. Unit 650, 655 West Kent Ave. N., Vancouver, BC, V6P 6T7, Canada www.orientgene.com
Date: June 29, 2022
Contact Person:
Guang Gao, Principal Consultant Axteria Biomed Consulting 8040 Cobble Creek Circle Potomac, MD 20854 (301) 814-4985 ggw101413@icloud.com
Proprietary Name of the Device: hCG One Step Pregnancy Test
Common Name: hCG One Step Pregnancy Test
Classification Name: 21 CFR § 862.1155 Human chorionic gonadotropin (HCG) test system
Product Code: LCX
#### 1.0 Device Description
The hCG One Step Pregnancy Test is a lateral flow sandwich immunochromatographic assay for the qualitative determination of human chorionic gonadotropin (hCG) in human urine samples.
#### 2.0 Intended Use
The hCG One Step Pregnancy Test is an in vitro diagnostic test for the qualitative determination of human chorionic gonadotropin hormone in human urine samples. It is used as an aid in early detection of pregnancy, in some cases as early as five (5) days before the expected period, i.e., as early as six (6) days before the day of the missed period.
{4}------------------------------------------------
#### 3.0 Test Principle
The hCG One Step Pregnancy Test is a lateral flow sandwich immunochromatographic assay that utilizes a combination of monoclonal antibodies for qualitative detection of hCG in human urine samples. The test uses two lines to show the results. When the urine sample is added into the absorbent tip, the sample migrates via capillary action along the membrane. If present in the sample, hCG will react with the specific monoclonal hCG antibody and form a colored line at the test line region of the membrane, indicating a positive result (pregnant). Absence of a colored line at the test line region suggests a negative result (not pregnant). To serve as a procedural control, a colored line will always appear in the control line region. If the control line does not appear (no color appears at the control line region 3 minutes after application of the sample), the test is invalid and the specimen should be retested with a new device.
| Similarities and Differences | | |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------|
| | Proposed device | Predicate device |
| Name | hCG One Step Pregnancy Test | First Response Early Result Pregnancy Test (K123436) |
| Device Type | In vitro diagnostic | Same |
| Classification | Class II | Same |
| Regulatory Number | 21 CFR §862.1155 Human Chorionic Gonadotropin | Same |
| Product Usage | OTC | Same |
| Principle of Operation | Lateral flow sandwich immunochromatographic assay | Same |
| Specimen Collection Method | Mid-stream or dip mode | Same |
| Intended Use | A chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine. The device is intended for use as an aid in early detection of pregnancy, in some cases as early as five (5) days before the expected period, i.e., as early as six (6) days before the day of the missed period. | Same |
#### 4.0 Comparison of the Predicate Device
{5}------------------------------------------------
| Type of Specimen | Human urine | Same |
|--------------------------|---------------------------------------------------------------|--------------------------------------------------------------------------------|
| Sensitivity | 10 mIU/mL | Same |
| Time to Result | 3 minutes | Same |
| Result | Qualitative | Same |
| Shelf Life | 24 months shelf life when stored in<br>a dry place at 2-30°C. | Same |
| hCG Isoforms<br>Detected | Intact hCG | Intact hCG,<br>hyperglycosylated hCG, hCG<br>ß-subunit, hCG ß-core<br>fragment |
#### Test Performance 5.0
#### Analytical Performance 5.1
- Limit of Detection 5.1.1
The Limit of Detection (LOD) was determined by testing serial dilutions of the hCG standard substance that is commercially available and traceable to the 5th WHO international Standard. Three internal sites used 3 lots of the proposed device with a simulated stream method and dip method. The panel was created by spiking 30 negative urine samples (from males and nonpregnant females) with the hCG standard substance at eight different hCG concentrations (below, at and above the initial assigned LOD). These panel members were randomized, blind coded and tested with the 3 lots of the proposed device by three different lab technicians on different days. Each operator performed the test with one lot of the proposed device and interpreted the results independently.
The results showed that the LOD (the lowest concentration that yields ≥ 95% positive results) is 10 mIU/mL for urine obtained by both the simulated stream method and the dip method.
#### Precision/Reproducibility 5.1.2
Panels were prepared from 100% pooled human urine containing no analyte. The hCG human urine samples were spiked with hCG positive specimen that is traceable to the 5th WHO international standard to obtain the hCG at concentrations 0, 5, 7.5, 8.75, 10, 12.5, and 25 mIU/mL.
Repeatability testing was conducted with twenty replicates of each urine sample in one day with one lot of the hCG One Step Pregnancy Test. For intermediate precision testing, each panel member was tested by three operators with 10 replicates and 3 lots of device for 5 nonconsecutive days. A total of 450 replicates of each urine sample was tested over multiple days by multiple operators. When testing hCG at 10 mIU/mL and above, a 100% positive rate was
{6}------------------------------------------------
obtained for all operators with all device lots, at both the simulated stream method and the dip method.
The results show that the proposed device can produce consistent results when detecting hCG in urine samples.
#### Analytical Specificity 5.1.3
# 5.1.3.1 Potential Interfering Substances Study
The interference substance study was performed with urine containing prescription/OTC drugs, elevated levels of chemical analytes (e.g., caffeine, ascorbic acid), and elevated levels of biological analytes (e.g., glucose, protein, albumin, bilirubin, hemoglobin).
For the potential interfering substances study, negative female urine samples containing 0 and 10 mIU/mL of hCG were individually spiked with the potential interfering substances. Five devices were used for the testing. Based on the five replicates results, no interference was observed from the potential interfering substances at the level evaluated.
# 5.1.3.2 Cross Reactivity Study
The hCG One Step Pregnancy Test was evaluated for cross-reactivity with other substances that are similar in structure to hCG. Human urine specimens with hCG concentrations of approximately 0 and 10 mIU/mL were supplemented with potentially cross-reacting compounds. The compounds were tested and found to have no cross-reactivity.
#### 5.2 Comparison Study
The performance the hCG One Step Pregnancy Test was evaluated by professional users at 2 sites with 2 lots of the proposed device. A total of 158 urine samples were collected from 158 subjects with ages ranging between 18 to 45 who were in need of determining their pregnancy status. The urine samples were collected at various times of day in a random fashion. The urine specimens were tested with the proposed device and with the predicate device (First Response Early Result Pregnancy Test) by independent operators. The urine samples were masked with a code and randomized by individuals who were not conducting the test. The test results of the hCG One Step Pregnancy Test and the First Response Early Result Pregnancy Test were compared to evaluate the clinical test performance.
The test performance of the hCG One Step Pregnancy Test is with 100% concordance when compared to the predicate.
#### 6.0 Conclusion
Based on the test principle and acceptable performance characteristics including interfering substances, cross reactivity, method comparison, clinical specificity, and lay-user studies of the devices, it is concluded that the hCG One Step Pregnancy Test is substantially equivalent to the predicate.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.