K032992 · Biotech Atlantic, Inc. · LCX · Nov 21, 2003 · Clinical Chemistry
Device Facts
Record ID
K032992
Device Name
UNIMARK HOME PREGNANCY TEST DEVICE
Applicant
Biotech Atlantic, Inc.
Product Code
LCX · Clinical Chemistry
Decision Date
Nov 21, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1155
Device Class
Class 2
Intended Use
The UniMark® Home Pregnancy Test Device is intended for the qualitative measurement of human Chorionic Gonadotropin (hCG) in urine as a home use screening test for pregnancy.
Device Story
UniMark® Home Pregnancy Test Device is a rapid, visual, one-step, solid-phase chromatographic immunoassay for qualitative hCG detection in urine. Device consists of a test strip within a plastic cassette. Principle of operation: urine specimen migrates via capillary action through a conjugate pad containing mouse anti-beta hCG antibodies conjugated to colloidal gold particles; hCG in specimen binds to conjugate; complex moves to test zone where immobilized goat anti-hCG antibodies capture the complex, forming a visible purple band. Positive result (purple band) indicates presence of hCG; negative result (absence of band) indicates no detectable hCG. Control window confirms proper reagent performance. Used by consumers at home for early pregnancy screening. Results are visually interpreted by the user to inform pregnancy status.
Clinical Evidence
Bench testing: Sensitivity 25 mIU/ml (WHO 3rd IS 75/537). Accuracy: 364 urine specimens (189 positive, 175 negative) showed 100% agreement with a quantitative visual hCG test. Specificity: No cross-reactivity with hLH (200 mIU/ml), hFSH (1000 mIU/ml), or hTSH (1000 uIU/ml). Interference testing: No interference from common substances (acetaminophen, caffeine, glucose, etc.) at specified concentrations. Consumer study: 100 home users tested coded urine samples (0, 20, 30, 1000 mIU/ml hCG). 100% correlation between home user results and professional laboratory testing.
Technological Characteristics
Solid-phase chromatographic immunoassay. Qualitative test. Assay cut-off: 25 mIU/ml. Form factor: test strip in plastic cassette with dropper. Traceability: WHO 3rd IS 75/357.
Indications for Use
Indicated for the early detection of pregnancy in women via qualitative determination of hCG in urine. Intended for consumer home use.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
K960174 — UNIMARK MIDSTREAM PREGNANCY TEST STICK · Biotech Atlantic, Inc. · Apr 1, 1996
K960117 — PLATFORM ONESTEP PREGNANCY HOME TEST · Syntron Bioresearch, Inc. · Mar 19, 1996
K963341 — MIDSTREAM ONESTEP PREGNANCY HOME TEST · Syntron Bioresearch, Inc. · Dec 2, 1996
K071930 — ONE STEP HCG URINE PREGNANCY TEST (STRIP), MODEL B01-S, ONE STEP HCG URINE PREGNANCY TEST (CASSETTE) , MODEL B01C · Blue Cross Bio-Medical Co., Ltd. · Dec 18, 2007
K203246 — Atlas One Step hCG Urine Pregnancy Test (Strip), Atlas One Step hCG Urine Pregnancy Test (Cassette), Atlas One Step hCG Urine Pregnancy Test (Midstream) · Atlas Link Technology Co., Ltd. · Aug 4, 2021
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE
A. 510(k) Number: K032992
B. Analyte: human chorionic gonadotropin
C. Type of Test: qualitative solid phase chromatographic immunoassay
D. Applicant: Biotech Atlantic Inc.
E. Proprietary and Established Names: UniMark® Home Pregnancy Test Device
F. Regulatory Information:
1. Regulation section: 21CFR862.1155, Human Chorionic Gonadotropin Test System
2. Classification: Class II
3. Product Code: LCX
4. Panel: Chemistry (75)
G. Intended Use:
1. Indication(s) for use: The UniMark® Home Pregnancy Test Device (provided as a test strip contained within a plastic test strip holding cassette) is for the rapid and qualitative determination of human Chorionic Gonadotropin (hCG) in urine. It is intended for consumer use at home. It is indicated for use in the early detection of pregnancy.
2. Special condition for use statement(s):
3. Special instrument Requirements:
H. Device Description:
The device includes a test strip within a cassette holder and a dropper for application.
I. Substantial Equivalence Information:
1. Predicate device name(s): UniMark® hCG Pregnancy Test device
2. Predicate K number(s): K941090
3. Comparison with predicate: The device is the same device as the predicate. The difference is that the indications for use is expanded to include over-the counter use. It is indicated for use in the early detection of pregnancy
{1}
Page 2 of 3
J. Standard/Guidance Document Referenced (if applicable):
None referenced
K. Test Principle: The test is a solid phase chromatographic immunoassay.
L. Performance Characteristics (if/when applicable):
1. Analytical performance:
All analytical and method comparison studies were performed previously on the predicate device. Therefore, only the over-the counter lay user study was included in this 510(k).
a. Precision/Reproducibility: Qualitative assay:
Reproducibility of the device in the hands of lay users was evaluated in a consumer study of 100 users selected on a random basis as they presented themselves at study sites. Participants ranged in age from 16 to over 40 and ranged in education level from high school to graduate levels. Participants were given a set of coded samples containing spiked specimens at concentrations of 0, 20, 30 and 1000 mIU/ml. The only instructions provided were draft copies of the home use package insert. All samples at 30 mIU and 1000 mIU/ml tested positive. Results were compared to those obtained by laboratory professionals. In this study there was 100% agreement between results obtained by lay users and those obtained by professionals.
b. Linearity/assay reportable range: Not applicable - qualitative test
c. Traceability (controls, calibrators, or method): Reviewed in the 510(k) for the predicate device. WHO 3rd IS 75/357
d. Detection limit: Reviewed in the 510(k) for the predicate device.
e. Analytical specificity: Reviewed in the 510(k) for the predicate device.
f. Assay cut-off: 25 mIU/ml. Reviewed in the 510(k) for the predicate device.
2. Comparison studies:
a. Method comparison with predicate device:
Comparison to professional use device is described in the reproducibility section above.
3. Clinical studies:
a. Clinical sensitivity: Not applicable. Not typically provided for this device type.
{2}
Page 3 of 3
b. Clinical specificity: Not applicable. Not typically provided for this device type.
c. Other clinical supportive data (when a and b are not applicable): Not applicable
4. Clinical cut-off: Not applicable. Not typically provided for this device type.
5. Expected values/Reference range: Not applicable. Qualitative assay.
**M. Conclusion:**
I recommend that the Biotech Atlantic Inc. UniMark® Home Pregnancy Test Device is substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.