K951014 · Isolab, Inc. · KLI · Oct 18, 1996 · Clinical Chemistry
Device Facts
Record ID
K951014
Device Name
ISOLAB'S T4 TEST KIT
Applicant
Isolab, Inc.
Product Code
KLI · Clinical Chemistry
Decision Date
Oct 18, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1700
Device Class
Class 2
Attributes
Pediatric
Intended Use
Isolab’s T4 Test Kit is intended for use in determining the total thyroxine (T4) in blood specimens dried on filter paper.
Device Story
In vitro diagnostic test kit for total thyroxine (T4) quantification in dried blood spots. Principle: sequential, competitive enzyme immunoassay. T4 eluted from blood disks; forms complex with biotin-labeled monoclonal antibody on streptavidin-coated microstrips. T4-horseradish peroxidase conjugate competes for binding sites. Enzymatic reaction with fluorogenic substrate (HPPA) produces fluorescence measured at 405 nm via Fluoroscan II Neonate reader. Used in clinical laboratory settings by trained technicians. Output provides T4 concentration in µg/dL, used to screen for congenital hypothyroidism. Benefits include early detection of neonatal thyroid dysfunction.
Clinical Evidence
Bench testing only. Linearity established from 0 to 25 µg/dL. Sensitivity (detection limit) is 1.0 µg/dL. Reproducibility study (20 days, duplicate runs) showed total CVs of 10.4% (at 5.5 µg/dL) and 15.6% (at 12.1 µg/dL). Recovery rates were 88% at low levels and 97% at normal levels. Interference testing performed per NCCLS EP-7; slight interference noted with thyroxine analogs and 5-propyl-2-thiouracil. Method comparison performed against Neometric and Wallac Oy assays.
Technological Characteristics
Competitive enzyme immunoassay. Components: streptavidin-coated microstrips, biotinylated monoclonal antibody, T4-HRP conjugate, HPPA fluorogenic substrate, H2O2, glycine stopping buffer. Detection: fluorometric (405 nm emission, 320 nm excitation). Form factor: microstrip plates. Reagents provided in kit sizes for 800 or 4000 samples.
Indications for Use
Indicated for screening neonates for congenital hypothyroidism using dried blood spot specimens. Results are reported in µg/dL; values at or below 8.5 µg/dL are considered presumptive positives requiring confirmatory diagnostic testing.
Regulatory Classification
Identification
A total thyroxine test system is a device intended to measure total (free and protein bound) thyroxine (thyroid hormone) in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Reference Devices
Neometric’s Accuscreen T4
Wallac Oy’s Delphia Neonatal T4
Related Devices
K103484 — GSP NEONATAL THYROXINE (T4) · Wallac Oy, A Subsidiary of Perkinelmer, Inc. · Apr 22, 2011
K030860 — ACCUBIND NEONATAL T4 MICROPLATE EIA, MODEL 2625-300 · Monobind · Aug 22, 2003
K973276 — AUTODELFIA NEONATAL HTSH L KIT (B077-112) · Eg &G Wallac · Nov 20, 1997
Submission Summary (Full Text)
{0}
K951014
ISOLAB INC.
510(K)
T4 TEST KIT
EXHIBIT C
# Summary of Safety and Effectiveness
OCT 18 1996
Isolab’s T4 Test Kit is intended for use in determining the total thyroxine (T4) in blood specimens dried on filter paper.
The Isolab’s T4 Test Kit is a sequential, competitive enzyme immunoassay with fluorometric detection in which T4 is eluted from dried blood disks and simultaneously forms a complex with the biotin-labeled monoclonal antibody which binds to the streptavidin coated solid phase. After sufficient incubation with biotinylated antibody, a conjugate of T4 with horseradish peroxidase is added to the wells and incubated with the reaction mixture. This conjugate competes with the analyte for the free binding sites on the bound monoclonal antibody. Excess of unbound reagent is washed away after a sequential incubation with biotinylated antibody and conjugate, and the enzymatic reaction with a fluorogenic substrate 3-p-hydroxyphenylpropionic acid (HPPA) [9] takes place. The reaction is stopped by addition of glycine buffer and the fluorescence in each well is measured with Fluoroscan II Neonate at 405 nm (excitation being 320 nm).
Isolab has two kit sizes available for the T4 Test Kit. Test Kit NT-1000 contains enough reagents and Microstrips to run 800 patient samples. Additional wells are available for Calibrators and controls. Test Kit NT-4000 contains enough reagents and coated Microstrips to run 4000 patient samples. Additional wells are available for Calibrators and Controls.
| Code | Code | Description |
| --- | --- | --- |
| NT-1000 | NT-4000 | Item |
| 10 plates | 50 plates | Streptavidin coated Microstrips® |
| 2 x 90 ml | 10 x 90 ml | Biotinylated Antibody (Bottle 1) |
| 2 X 30 ml | 10 X 30 ml | T4-HRP Conjugate (Bottle 2) |
| 210 ml | 5 X 210 ml | HPPA Substrate (Bottle 3) |
| 45 ml | 5 X 45 ml | H_{2}O_{2} Solution (Bottle 4) |
| 200 ml | 2 X 500 ml | Stopping Solution (Bottle 5) |
| 220 ml | 2 X 500 ml | Wash Solution (Bottle 6) |
| 1 card | 5 cards | T4 Calibrators |
| 10 | 50 | Microstrip Covers
(Plastic sheets to cover the Microstrips® during incubation.) |
The T4 values given for the Controls, Calibrators and thus patient results are given in µg/dL. This unit of measure reflects the estimated amount that would be found in serum. These values should be considered accurate for the purposes of screening. Due to differences in hematocrit values, these values should not be assumed as accurate as those from serum-based assays. The discrimination between normal and presumptive positives for CH is based on a predetermined fixed cut-off value generally regarded to be based on the lower 10%. The cut-off value that represents the lower 10% for this assay is 8.5 µg/dL. Any result at this value or below should be confirmed by another diagnostic test or procedure to confirm that the patient is positive for congenital hypothyroidism.
## Linearity
Results of the linearity study show that the assay is linear from concentration of 0 to 25 µg/dL of T4. Concentrations of T4 in the amounts of 6.25, 12.5, and 25 µg/dL were added to a whole blood sample and spotted onto filter Schleicher & Schuell No. 903 paper. Each spiked sample along with the unspiked sample was run five times.
{1}
# Recovery
The recovery was determined from the same data as shown under Reproducibility. The recovery can be calculated by the following formula:
$$
\text{mean} \times 100 = \% \text{recovery}
$$
amount added
$$
\begin{array}{l}
5.5 \times 100 = 88\% \\
6.25 \\
12.1 \times 100 = 97\% \\
12.5 \\
\end{array}
$$
The recovery is shown at the low T4 levels to be around 88% and at normal levels around 97%.
# Sensitivity
The detection limit for the T4 assay 1.0 µg/dL.
# Reproducibility
Two sample pools were run in duplicate twice a day for 20 days. The summary of the results are shown:
| | amount added | mean | total SD | total CV | within-run SD | within-run CV |
| --- | --- | --- | --- | --- | --- | --- |
| T4 depleted neonate | 6.25 µg/dL | 5.5 µg/dL | 0.57 | 10.4% | 0.43 | 7.8% |
| normal neonate | 12.5 µg/dL | 12.1 µg/dL | 1.89 | 15.6% | 1.52 | 12.6% |
# Comparison of Methods
Isolab’s T4 Test Kit was compared to Neometric’s Accuscreen T4 and Wallac Oy’s Delphia Neonatal T4. The values for the Neometrics and Wallac Oy assays were performed at two state laboratory facilities. the lower 10% of the values for each of the assays are shown.
| Test | State 1 Cutoff (µg/dL) | State 2 Cutoff (µg/dL) |
| --- | --- | --- |
| Isolab | 8.5 | 8.1 |
| Neometrics | - | 11.7 |
| Wallac Oy | 10.2 | - |
The distribution of values as well as the mean values, standard deviations and median values for the Isolab methods and the two comparative test kit methods are shown on the next page.
{2}


Interference testing was designed and executed using NCCLS document EP-7 as guidance [14]. The following substances were added to a control sample and tested as possible interferents to the T4 Test Kit. The control sample was assayed seven times and each of the whole blood samples spiked with the following substances were also assayed seven times. The two thyroxines (D-T4 and DL-T4) added about 100% to the calculated values as expected. Two other thyroxine analogs (DL-T₀ and TA₃) also produced a slight interference. The drug, 5-propyl-2-thiouracil (PTU) also caused a slight interference at the concentration tested. This level would probably be higher than in a neonate on this drug.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.