OLYMPUS UNSATURATED IRON BINDING CAPCITY REAGENT, OSR6124/0SR6524
Applicant
Olympus America, Inc.
Product Code
JMO · Clinical Chemistry
Decision Date
Jul 16, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1415
Device Class
Class 1
Indications for Use
An iron-binding capacity test system is a device intended to measure iron-binding capacity in serum and plasma. Iron-binding capacity measurements are used in the diagnosis and treatment of anemia.
Device Story
The Olympus Unsaturated Iron Binding Capacity test system is an in vitro diagnostic device used to measure iron-binding capacity in serum and plasma samples. It is intended for use by healthcare professionals in clinical laboratory settings to assist in the diagnosis and treatment of anemia. The device functions as a chemical test system; it provides quantitative results that clinicians use to assess iron status and guide therapeutic decisions for patients suspected of having anemia.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Reagent R2a modified from powder to liquid state; linearity range 400-450 µg/dL.
Indications for Use
Indicated for the measurement of iron-binding capacity in serum and plasma for the diagnosis and treatment of anemia in patients.
Regulatory Classification
Identification
An iron-binding capacity test system is a device intended to measure iron-binding capacity in serum. Iron-binding capacity measurements are used in the diagnosis and treatment of anemia.
Related Devices
K990216 — MAGNETIC TIBC KIT · Reference Diagnostics, Inc. · Feb 16, 1999
K012340 — BAYER ADVIA SYSTEM; UNSATURATED IRON BINDING CAPACITY (UIBC) · Bayer Corp. · Oct 1, 2001
K030124 — OLYMPUS FERRITIN REAGENT · Olympus America, Inc. · Feb 6, 2003
K972363 — IL TEST IRON · Instrumentation Laboratory CO · Aug 15, 1997
K020816 — IRON-PC-SL ASSAY MODEL # 151-10, 151-26 · Diagnostic Chemicals , Ltd. · May 14, 2002
Submission Summary (Full Text)
{0}
SPECIAL 510(k): Device Modification ODE Review Memorandum
To: THE FILE
RE: DOCUMENT NUMBER K041793
This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary):
1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.)
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials.
3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.
This change was for reagent R2a going from a powder to a liquid. The linearity extended from 400 to 450 µg/dL.
4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, and reagent components remain the same.
5. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied
c) A declaration of conformity with design controls. The declaration of conformity should include:
i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and
ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review.
6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices).
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.