K960195 · Franklin Diagnostics, Inc. · JLW · Apr 15, 1996 · Clinical Chemistry
Device Facts
Record ID
K960195
Device Name
THYROCHEK ONE-STEP RAPID TSH ASSAY
Applicant
Franklin Diagnostics, Inc.
Product Code
JLW · Clinical Chemistry
Decision Date
Apr 15, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1690
Device Class
Class 2
Indications for Use
This new immunometric assay device is a single-use Class II medical device for professional use that is designed to indicate whether a serum sample contains TSH in a concentration > 5 µIU/ml.
Device Story
ThyroChek is a single-use immunometric assay for professional use; detects TSH concentrations > 5 µIU/ml in serum samples. Device provides qualitative indication of TSH levels; assists clinicians in thyroid function assessment. Operation involves testing serum samples; results compared against predicate device performance. Benefits include rapid identification of elevated TSH levels in clinical settings.
Clinical Evidence
Bench testing and comparative study performed. Parallel testing of 328 patient samples against predicate device: 65/66 samples with TSH > 5 µIU/ml correctly identified; 262/262 samples with TSH < 5 µIU/ml correctly identified. Intra-assay coefficient of variation < 3%. Cross-reactivity studies showed no interference from hCG (up to 150,000 mIU/ml), LH, or FSH (> 40 mIU/ml). Interference testing confirmed no impact from glucose, uremic metabolites, lipids, or hemolyzed red blood cells.
Technological Characteristics
Immunometric assay; single-use; professional use. Designed for serum sample analysis. No specific materials, connectivity, or software algorithms described.
Indications for Use
Indicated for professional use to detect TSH concentrations > 5 µIU/ml in serum samples.
Regulatory Classification
Identification
A thyroid stimulating hormone test system is a device intended to measure thyroid stimulating hormone, also known as thyrotrophin and thyrotrophic hormone, in serum and plasma. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders.
K023093 — ACCESS HYPERSENSITIVE HTSH ASSAY · Beckman Coulter, Inc. · Nov 25, 2002
Submission Summary (Full Text)
{0}
J. Ehrenkranz, M.D. p.3
K960195
# 14. SMDA Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. This new immunometric assay device is a single-use Class II medical device for professional use that is designed to indicate whether a serum sample contains TSH in a concentration > 5 µIU/ml. The intended use, technical characteristics, efficacy, and safety of this new device are substantially equivalent at detecting TSH > 5 µIU/ml to a number of other TSH immunoassays that have been commercially available both before and after 1978, and Franklin believes that demonstration of substantial equivalence to a legally marketed predicate device in accordance with section 510(k) has been demonstrated.
This document contains the results of a number of validation studies which compares the performance of ThyroChek at detecting TSH > > 5 µIU/ml to that of a predicate medical device, Immunolite Third Generation TSH Assay, 510(k) number K930007. The validating studies performed and the results obtained are as follows:
1. Quantitative Recovery. This demonstrated the ability of this assay to discriminate serum samples containing TSH at concentrations of 5.2, 5.7, and 6.5 µIU/ml, as determined by the predicated device, from samples containing TSH at a concentration of 4.1 and 4.5 µIU/ml.
2. Serial Dilution. This demonstrated the ability of this assay to distinguish serum samples containing TSH > 5 µIU/ml from those with a TSH concentration diluted below 5 µIU/ml.
3. Coefficient of Variation. 5 sets of 30 replicates showed uniform results, that is 30 identical results, using ThyroChek in 4 of 5 sets and 29 of 30 replicates agreed in 1 of 5 sets. Additionally 25 samples measured in duplicate showed 100% agreement between both replicates. This represents an intra-assay coefficient of variation of less than 3%.
4. Parallelism. 328 patient samples were assayed in parallel by the predicate device and by ThyroChek. 65 of 66 patients with TSH > 5 µIU/ml, as determined by the predicate device, were correctly identified using ThyroChek. 262 of 262 patients with TSH < 5 µIU/ml, as determined by the predicate device, were correctly identified by ThyroChek.
5. Cross-Reactivity. Because of the structural similarity between TSH and hCG, LH, and FSH, studies were performed to determine if maximal clinical concentrations of these hormones would interfere with ThyroChek performance. These studies showed that hCG, in concentrations as high as 150,000 mIU/ml, and LH / FSH > 40 mIU/ml, as encountered in 17 postmenopausal women, did not produce false negative or positive test results or affect ThyroChek's ability to determine whether a serum sample contained > 5 µIU/ml.
{1}
J. Ehrenkranz, M.D. p.4
6. Interference. Elevated concentrations of glucose, uremic metabolites, lipids, or hemolyzed red blood cells as encountered in patient samples did not interfere with device performance.
This 510(k) Summary Statement was revised by J. Ehrenkranz, M.D. on 20 March, 1996.
J. Ehrenkranz
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.