BIOCIRCUITS IOS QUANTITATIVE HCG TEST CARTRIDGES/BIOCIRCUITS IOS IMMUNOASSAY CONTROLS
Applicant
Biocircuits Corp.
Product Code
JJY · Clinical Chemistry
Decision Date
Dec 5, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Indications for Use
The IOS™ Quantitative hCG Test Cartridges are to be used for the quantitative determination of human chorionic gonadotropin (hCG) in serum for the early detection of pregnancy. They are intended to be used with the IOS™ instrument in clinical laboratories, physicians' office laboratories, and other alternate sites of use close to the point of patient care. The IOS™ Immunoassay Controls Kit is to be used to assist in monitoring accuracy and precision in the IOS™ immunoassay test cartridges.
Device Story
Device is a two-site sandwich immunoassay for quantitative hCG detection in serum. Input: patient serum sample added to IOS™ test cartridge. Operation: sample rehydrates dried reagents (polyclonal capture antibody, monoclonal alkaline phosphatase-labeled detection antibody, substrate); hCG-conjugate complex binds to immobilized capture antibody; unbound material washed away; substrate reaction produces fluorescent signal. Instrument uses front-surface fluorometry to measure signal rate. Output: quantitative hCG concentration calculated via instrument-stored calibration curve. Used in clinical labs/physician offices by technologists or medical assistants. Instrument automates buffer addition, washing, signal reading, and calculation. If signal exceeds range, instrument automatically reads a second, user-diluted sample track to extend assay range up to 250,000 mIU/mL. Results displayed on LCD and printed. Benefits: rapid, quantitative pregnancy testing and monitoring of hCG-secreting tumors or ectopic pregnancy/spontaneous abortion.
Clinical Evidence
Bench testing and clinical comparison studies performed. Precision: total CVs ranged from 8.98% to 11.57% across control levels in both manufacturer and physician office labs. Accuracy: 165 samples tested in manufacturer lab (r=0.981 to 0.986); 121 samples tested in three physician office labs (OB/GYN, Internal Medicine, Reproductive Endocrinology) showed correlation coefficient of 0.996 compared to predicate device.
Technological Characteristics
Two-site sandwich fluorometric enzyme immunoassay. Materials: plastic cartridge. Energy: fluorometric signal detection. Connectivity: standalone instrument with LCD and printer. Reagents: dried polyclonal capture antibody, monoclonal detection antibody, and substrate. Calibration: factory-generated, 90-day stability. Operating environment: 15-30°C.
Indications for Use
Indicated for quantitative determination of hCG in serum for early pregnancy detection. Intended for use in clinical labs, physician offices, and point-of-care settings.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
Dade Stratus Fluorometric Enzyme hCG Immunoassay
Related Devices
K961792 — BIOCIRCUITS IOS SERUM PREGNANCY TEST CARTRIDGES/IMMUNOASSAY CONTROLS · Biocircuits Corp. · Aug 21, 1996
{0}
K963770
DEC - 5 1996
# Attachment B
Revised Summary of Information
Respecting Safety and Effectiveness
Biocircuits IOS™ Quantitative hCG Test Cartridges
and IOS™ Immunoassay Controls
{1}
510(k) Summary of Information Respecting Safety and Effectiveness
A. Submitter:
Biocircuits Corporation
1324 Chesapeake Terrace
Sunnyvale, CA 94089
(408) 745-1961
Contact: Sheila Ramerman
Date Prepared: November 1, 1996
B. Device Names:
Proprietary Name: Biocircuits IOS™ Quantitative hCG Test Cartridges
Biocircuits IOS™ Immunoassay Controls
(formerly IOS™ Thyroid Controls)
Common Names: Reagents for hCG (human chorionic gonadotropin)
Quality control materials (assayed and unassayed)
Classification Name: human chorionic gonadotropin (hCG) test system
Quality control materials (assayed and unassayed)
C. Legally Marketed Device:
The IOS Quantitative hCG Test Cartridges are substantially equivalent to the Dade Stratus Fluorometric Enzyme hCG Immunoassay currently manufactured and distributed by Dade International.
D. Device Description:
Quantitative hCG Test Cartridges: Human chorionic gonadotropin is a glycoprotein that is synthesized by the placenta during pregnancy and appears in both serum and urine relatively soon after implantation of the developing embryo. The presence of hCG, and its rapid rise following conception, is thus the basis of pregnancy testing. (1) Quantitation of hCG can aid in the determination of ectopic pregnancy and spontaneous abortion (2, 3). hCG is also secreted by a wide variety of tumors: gestational trophoblastic tumors, testicular and prostatic tumors, and some breast cancers. (4,5).
Human chorionic gonadotropin is composed of two noncovalently linked polypeptides: the alpha and beta subunits. The individual subunits lack biological activity, but become active when combined to form the intact hormone. The alpha subunit of hCG is structurally homologous to the alpha subunits of LH (luteinizing hormone), FSH (follicle stimulating hormone), and TSH (thyroid stimulating hormone). The beta subunit of each of the hormones is structurally unique and confers the biologic and immunologic specificity to each hormone. The beta subunit of hCG contains a unique chain of 30 carboxy-terminal amino acids that confers its specificity (1). Monoclonal antibodies directed against this portion of the beta subunit permit differentiation between hCG and the other pituitary glycoprotein hormones.
10
{2}
Principle of the Test
Quantitative hCG Test Cartridges: The IOS™ Quantitative hCG test is a two-site sandwich immunoassay which utilizes two antibodies directed against hCG: a polyclonal antibody immobilized on the plastic cartridge surface (capture antibody) and a monoclonal antibody labeled with alkaline phosphatase (detection antibody, 'conjugate'). hCG in the patient sample binds to conjugate in the test cartridge, and this hCG-conjugate complex then binds to the immobilized capture antibody. After a short incubation period, unbound hCG-conjugate complex is washed away. Substrate is added, which reacts with the alkaline phosphatase conjugate-hCG-capture antibody 'sandwich' and produces a fluorescent signal. The level of fluorescence is directly proportional to the amount of bound hCG-conjugate, which is directly proportional to the amount of hCG present in the patient sample. All reagents necessary to perform the test are dried in the IOS™ cartridge, and are rehydrated by the addition of patient sample by the operator, or by the addition of buffer by the instrument.
To perform a test, the operator inserts an IOS™ Quantitative hCG cartridge in the IOS™ instrument. When prompted, the operator adds sample to one sample well. A second portion of patient sample is diluted off-line, using the dilution vial supplied with the kit, and pipetted into the second well of the test cartridge. The operator starts the test sequence when sample addition is complete. Patient sample rehydrates conjugate dried in the sample well; the conjugate binds to hCG present in the patient serum sample. This hCG-conjugate complex then flows into the incubation/reaction chamber where it binds to the immobilized capture antibody. At the end of the incubation time, unbound sample-conjugate is washed away by buffer. Buffer is used to rehydrate the dried substrate necessary for signal generation and quantitation of hCG, and the rehydrated substrate is then delivered to the incubation/reaction chamber. The fluorescent signal produced is read as a rate by front-surface fluorometry, compared to the rates produced by a series of calibrators stored in the instrument memory, and the amount of hCG present in the patient sample is calculated from the stored calibration curve. If the signal from the first track (undiluted patient serum) is higher than that of the highest calibrator, the instrument will take a reading from the second track (user-diluted sample) to obtain a quantitative value for the patient sample. This increases the assay range 2500 mIU/mL (undiluted track) to 250,000 mIU/mL (diluted track).
Immunoassay Controls: The use of materials derived from human blood to monitor quality control of clinical chemistry testing in the clinical laboratory has been widely established over the past several years. The Biocircuits IOS™ Immunoassay Controls are two levels of blood-based material for use with Biocircuits IOS™ Test Cartridges.
To run a control, the operator inserts the Control Cartridge (packaged with the controls) into the IOS™ instrument. The instrument reads the lot number and ranges of acceptable values for the control solutions from the Control Cartridge barcode, and then ejects the Control Cartridge. The operator then inserts a test cartridge and follows the instrument prompts to identify the control level, apply control solutions, and begin the test sequence. The IOS™ instrument performs the required buffer additions to rehydrate assay reagents and perform wash steps as necessary, reads the fluorescence signal generated, and calculates and prints the control result just as it would if the cartridge were used to test a patient sample.
E. Intended Use:
The IOS™ Quantitative hCG Test Cartridges are to be used for the quantitative determination of human chorionic gonadotropin (hCG) in serum for the early detection of pregnancy. They are intended to be used with the IOS™ instrument in clinical laboratories, physicians' office laboratories, and other alternate sites of use close to the point of patient care.
14
{3}
NUU-20-20
The IOS™ Immunoassay Controls Kit is to be used to assist in monitoring accuracy and precision in the IOS™ immunoassay test cartridges.
## F. Comparison with the Predicate Device:
Table I summarizes the comparative features of the IOS and Stratus assays.
## G. Performance Data:
**Quantitative hCG Test Cartridges:**
Non-clinical testing performed in the manufacturer's laboratories gave the following results:
1. Precision:
| Control Level | 1 | 2 | 3 |
| --- | --- | --- | --- |
| Mean, overall (mIU/mL) | 36.8 | 169.2 | 339.3 |
| SD, overall (mIU/mL) | 4.26 | 16.39 | 34.46 |
| % CV, within-day (n ≥ 19) | 8.88 | 8.35 | 5.76 |
| % CV, between-day (n ≥ 60) | 10.89 | 9.06 | 10.75 |
| % CV, total | 11.57 | 9.69 | 10.16 |
2. Accuracy: A comparison of methods was performed by testing a total of 165 patient samples in the manufacturer's laboratories using the IOS™ Quantitative hCG cartridges and a commercially available fluorescent enzyme immunoassay. Of the range of hCG concentration from $2\mathrm{mIU/mL}$ to $2270\mathrm{mIU/mL}$ (neat sera), 107 samples tested gave a correlation coefficient (r) of 0.981 with the line of regression described by the equation $y = 5.63 + 0.89x$. Diluted samples with hCG concentrations up to $181,000\mathrm{mIU/mL}$ ($n = 58$) were also tested in the study, giving a correlation coefficient of 0.986 and a line of regression of $y = 0.002 + 0.955x$ for the entire range of samples tested ($2 - 181,000\mathrm{mIU/mL}$ hCG).
Clinical testing performed in typical physicians' office laboratories gave the following results:
1. Precision:
| Control Level | 1 | 2 |
| --- | --- | --- |
| Number of replicates | 39 | 31 |
| Mean (mIU/mL) | 35.34 | 152.79 |
| SD, overall (mIU/mL) | 4.08 | 13.72 |
| % CV, total | 11.54 | 8.98 |
2. Accuracy: A comparison of methods was performed by users in three typical physicians' office laboratories. Sites of testing included an OB/GYN practice, an Internal Medicine practice, and a Reproductive Endocrinology/Fertility practice. Users included medical technologists, medical assistants, and front-office personnel. A total of 121 patient samples were tested using the IOS™ Quantitative hCG cartridges in the office laboratories; the samples were split and sent to the manufacturer's laboratory for retesting on the predicate device. These studies gave a correlation coefficient (r) of 0.996 with the line of regression described by the equation $y = 0.002 + 0.880x$. Samples tested ranged from $2.0\mathrm{mIU/mL}$ to $202,598\mathrm{mIU/mL}$, and included diluted samples.
{4}
# Immunoassay Controls
The following ranges for the IOS™ Immunoassay Controls were determined in studies in the manufacturer's laboratories. To establish the ranges, the controls were tested in at least 40 cartridges each, over several days, using several IOS™ instruments. These values only apply to this lot of IOS™ Immunoassay Controls. Different lots of Immunoassay Controls will likely have slightly different ranges. Your laboratory should establish its own range for these controls over time.
| Analyte | Control Level 1 | | Control Level 2 | |
| --- | --- | --- | --- | --- |
| | Mean | Range | Mean | Range |
| T4 (ug/dL) | 8.1 | 6.48-9.72 | 12.9 | 10.32-15.4 |
| T-Uptake (%) | 32.4 | 29.25-35.55 | 39.7 | 36.85-42.55 |
| hCG (mIU/mL) | | | | |
It is self-evident from the data and information presented here that the Biocircuits IOS™ Quantitative hCG Test Cartridges are as safe, effective, and perform as well as the Dade Stratus hCG Fluorometric Immunoassay manufactured and distributed by Dade International.
Attachment: Table I: Assay Comparison
{5}
TABLE 1
Dade STRATUS vs. Biocircuits IOS™
Assay Comparison
| ATTRIBUTE | STRATUS hCG | IOS™ hCG |
| --- | --- | --- |
| Technology | Fluorometric enzyme immunoassay | Fluorometric Enzyme immunoassay |
| Assay format | Sandwich (two-site) | Sandwich (two-site) |
| Enzyme label | Alkaline phosphatase | Alkaline phosphatase |
| Substrate | Methylumbelliferyl phosphate | Methylumbelliferyl phosphate |
| Reagents | | |
| Immobilization Medium | Reaction tab | Plastic cartridge |
| Dry | Monoclonal antibody only | Polyclonal antibody, monoclonal antibody-
conjugate, substrate |
| Wet | 2 (assay-specific), loaded by operator at start
of each run | 1 (used for all assays), continuously on board |
| Delivery | Fully automated | Fully automated |
| Calibration | User-generated | Factory-generated |
| Calibration Stability | 14 days (minimum) | 90 days (minimum) |
| Storage | Refrigerated (2-8°C) | Room Temperature (15-30°C) |
| Sample | | |
| Type | Serum or plasma | Serum |
| Volume | 0.2 ml (minimum) | 0.1 mL |
| Measurement Needed | Non-precision | Non-precision |
| Dilution | Manual or programmable | Manual, run at same time as undiluted sera |
| Operating environment | 22°-32° C | 15°-30° C |
| Data analysis | Microprocessor-controlled
Stored standard curves | Microprocessor-controlled
Stored standard curves |
| Data output | LCD display
Printed alphanumeric hard copy | LCD display
Printed alphanumeric hard copy |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.