LIQUID ASSAYED MULTIQUAL CONTROL

K043208 · Bio-Rad · JJY · Jan 7, 2005 · Clinical Chemistry

Device Facts

Record IDK043208
Device NameLIQUID ASSAYED MULTIQUAL CONTROL
ApplicantBio-Rad
Product CodeJJY · Clinical Chemistry
Decision DateJan 7, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Indications for Use

Liquid Assayed Multiqual is intended for use as an assayed quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in the package insert.

Device Story

Liquid Assayed Multiqual is a quality control serum derived from human serum supplemented with purified biochemical materials, chemicals, drugs, preservatives, and stabilizers. It is used by laboratory technicians to verify the performance of clinical chemistry instruments. The device serves as a control material to monitor the accuracy and precision of laboratory testing procedures. The current submission introduces Lactic acid as an additional analyte to the existing formulation. The product is provided in liquid form and is stored at 2 °C to 8 °C. Stability is verified through open-vial aging and accelerated stability testing to ensure analyte concentration remains within specified limits over time.

Clinical Evidence

Bench testing only. Stability studies were performed to determine open vial stability (14 days at 2-8°C, with exceptions) and shelf life (3 years at -20°C or colder).

Technological Characteristics

Quality control serum prepared from human serum, purified biochemical materials (human/animal tissue extracts), chemicals, drugs, preservatives, and stabilizers. Liquid form. Stability verified via open-vial aging and accelerated temperature testing.

Indications for Use

Indicated for use as an assayed quality control serum to monitor the precision of laboratory testing procedures for various analytes in human serum samples.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: K043208 B. Purpose for Submission: Reason for submission: formulation change – addition of one analyte. C. Measurand: Lactic acid has been added to this pre-existing quality control serum (see below). D. Type of Test: Quality Control Material E. Applicant: Bio-Rad Laboratories 9500 Jeronimo Road Irvine, California 92618 F. Proprietary and Established Names: Liquid Assayed Multiqual 1 Liquid Assayed Multiqual 2 Liquid Assayed Multiqual 3 Liquid Assayed Multiqual MiniPak G. Regulatory Information: 1. Regulation section: 21CFR862.1660 - Quality control material (assayed and unassayed). 2. Classification: Class I (reserved) 3. Product code: JJY 4. Panel: (75)Chemistry {1} H. Intended Use: 1. Intended use(s): Liquid Assayed Multiqual is intended for use as an assayed quality control serum to monitor the performance of laboratory testing procedures for the analytes listed in the package insert. 2. Indication(s) for use: Liquid Assayed Multiqual is intended for use as an assayed quality control serum to monitor the performance of laboratory testing procedures for the analytes listed in this package insert. 3. Special conditions for use statement(s): For Prescription Use Only 4. Special instrument requirements: Varies with test instrument. I. Device Description: Liquid Assayed Multiqual is prepared from human serum to which purified biochemical material (tissue extracts of human and animal origin), chemicals, drugs, preservatives, and stabilizers have been added. The product is provided as a liquid. J. Substantial Equivalence Information: 1. Predicate device name(s): Liquid Assayed Multiqual 2. Predicate 510(k) number(s): K011867 3. Comparison with predicate: The currently marketed device is similar in analyte composition and concentrations to the predicate device with the exception of the addition of Lactic acid. {2} K. Standard/Guidance Document Referenced (if applicable): None L. Test Principle: The product under submission is used by technicians to confirm the performance of lab or clinical instruments. The specific test varies with the instrument verified. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Not Applicable b. Linearity/assay reportable range: Not Applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): Expected Values Liquid Assayed Multiqual is prepared from human serum to which purified biochemical material (tissue extracts of human and animal origin), chemicals, drugs, preservatives, and stabilizers have been added. Analysis by the manufacturer of the test kit or instrument used: In this case the manufacturer develops the target values. Whenever possible, Bio-Rad prefers to have the values assign in the way Analysis by reference laboratory: In this case, Bio-Rad contracts with reference laboratories for data generation. Three laboratories are used for each method listed in order to generate representative values. Stability The stability of the component analytes are the primary criteria for acceptance. Stability is measured by: - Open vial aging which mimics handling by the users of the product. These studies involve verifying concentration of the analytes when the product is stored capped but unsealed at 2 °C to 8 °C. - Accelerated Stability Testing which involves storing tested samples at elevated temperatures in an effort to predict their long-term performance. Test data provided by the manufacturer indicates that the predicate and premarket device show similar, limited degradation of analytes as a function of time. The change in analyte concentrations is within the limits specified by the product inserts. {3} d. Detection limit: Not Applicable e. Analytical specificity: Not Applicable f. Assay cut-off: Not Applicable 2. Comparison studies: a. Method comparison with predicate device: Not Applicable b. Matrix comparison: Not Applicable 3. Clinical studies: a. Clinical Sensitivity: Not Applicable b. Clinical specificity: Not Applicable c. Other clinical supportive data (when a. and b. are not applicable): 4. Clinical cut-off: Not Applicable 5. Expected values/Reference range: Not Applicable N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...