K033983 · Clinical Data, Inc. · JIY · May 21, 2004 · Clinical Chemistry
Device Facts
Record ID
K033983
Device Name
VITALAB IRON REAGENT
Applicant
Clinical Data, Inc.
Product Code
JIY · Clinical Chemistry
Decision Date
May 21, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1410
Device Class
Class 1
Attributes
3rd-Party Reviewed
Indications for Use
Vitalab Iron Reagent is for the quantitative determination of total iron in serum and plasma using the Vitalab Selectra Analyzer.
Device Story
Vitalab Iron Reagent Kit and Vitalab Selectra Analyzer function as a system for quantitative iron measurement in serum and plasma. Principle of operation: photometric measurement of iron concentration via increase in absorbance at 578 nm. Used in clinical laboratory settings by trained personnel. Output: iron concentration values (µg/dL). Healthcare providers use results to diagnose and monitor iron deficiency anemia, hemochromatosis, and chronic renal disease. Benefits include accurate assessment of iron status to guide clinical management.
Clinical Evidence
Bench testing only. Precision evaluated via NCCLS EP3-T (n=66 per sample, CVs 0.4-2.7%). Linearity confirmed 0-1,000 μg/dL. Detection limit 2.4 μg/dL. Interference testing showed minimal bias for bilirubin (≤4.1%) and Intralipid (≤-2.3%). Method comparison (n=118) against predicate device yielded slope 0.988 and intercept 1.1 μg/dL, demonstrating equivalence.
Technological Characteristics
Photometric iron test system; measures absorbance at 578 nm. Includes reagent and calibrator. Designed for use with Vitalab Selectra Analyzer. Analytical method based on iron dissociation and indicator reaction.
Indications for Use
Indicated for the quantitative determination of total iron in serum and plasma for patients suspected of iron deficiency anemia, hemochromatosis, or chronic renal disease.
Regulatory Classification
Identification
An iron (non-heme) test system is a device intended to measure iron (non-heme) in serum and plasma. Iron (non-heme) measurements are used in the diagnosis and treatment of diseases such as iron deficiency anemia, hemochromatosis (a disease associated with widespread deposit in the tissues of two iron-containing pigments, hemosiderin and hemofuscin, and characterized by pigmentation of the skin), and chronic renal disease.
Predicate Devices
Beckman Synchron Systems Iron (Fe) and Total Iron Binding Capacity (IBCT) Reagents (K960485)
Related Devices
K972363 — IL TEST IRON · Instrumentation Laboratory CO · Aug 15, 1997
K051115 — SENTINEL IRON LIQUID · Sentinel Ch. Srl · Aug 25, 2005
K013097 — WIENER LAB.FER-COLOR AA, MODEL 5X 20 ML CAT.N 1492003 · Wiener Laboratories Saic · Nov 13, 2001
K974604 — RANDOX IRON · Randox Laboratories, Ltd. · Jan 6, 1998
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE
A. 510(k) Number:
K033983
B. Purpose for Submission:
New product
C. Analyte:
Total iron
D. Type of Test:
Quantitative
E. Applicant:
Clinical Data, Inc.
F. Proprietary and Established Names:
Vitalab Iron Reagent and Calibrator
G. Regulatory Information:
1. Regulation section:
21 CFR §862.1410 Iron (non-heme) test system
2. Classification:
Class I
3. Product Code:
JIY
4. Panel:
Chemistry (75)
H. Intended Use:
1. Intended use(s):
Vitalab Iron Reagent is for the quantitative determination of total iron in serum and plasma using the Vitalab Selectra Analyzer.
2. Indication(s) for use:
The Vitalab Iron Reagent Kit, which contains both reagent and calibrator, is intended for use with the Vitalab Selectra E Analyzer as a system for the quantitative determination of total iron in serum and plasma. Iron results may be used for the diagnosis and treatment of diseases such as iron deficiency anemia, hemochromatosis (a disease associated with widespread deposit in the tissues of two iron-containing pigments, hemosiderin and hemofuscin, and characterized by pigmentation of the skin), and chronic renal disease.
{1}
Page 2 of 5
3. Special condition for use statement(s):
None
4. Special instrument Requirements:
The Vitalab Iron Reagent Kit is intended for use with the Vitalab Selectra Analyzer.
I. Device Description:
The device consists of vials of Vitalab Iron Reagent 1, Vitalab Reagent 2, and Vitalab Iron Calibrator. This liquid product is used to quantitatively determine total iron in serum and plasma using the Vitalab Selectra Analyzer. The sample is added to the first reagent component allowing for sample blank reading before the addition of the second reagent. Total iron concentrations are calculated from the change in absorbance at 578 nm after the reaction is finished.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Beckman Synchron Systems Iron (Fe) and Total Iron Binding Capacity (IBCT) Reagents
2. Predicate K number(s):
K960485
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Quantitative determination of total iron in serum and plasma | same |
| Measurement method | Endpoint with sample blank measured at approximately 570 nm (578 nm) | same (560 nm) |
| Calibration | Single point with reagent blank | same |
| Matrix | Serum or heparinized plasma | same |
| Differences | | |
| Item | Device | Predicate |
| Product type | Calibrated endpoint reagent with sample blank | Calibrated endpoint reagent |
| Chemical Reaction | Chelation of Fe^{2+} cation by ferene after reduction in acetate buffer | Chelation of Fe^{2+} cation by ferrozine after reduction in acetate buffer |
| Reagent components | ferene (iron chelator) thiourea (sulfhydryl reducing agent) acetic acid buffer other ingredients | ferrozine (iron chelator) thioglycolic acid (sulfhydryl reducing agent) acetic acid and hydroxylamine hydrochloride other ingredients |
| Analytical Range | 10 to 1,000 μg/dL | 5 to 500 μg/dL |
{2}
Page 3 of 5
## K. Standard/Guidance Document Referenced (if applicable):
NCCLS Publication: Vol. 2 No. 20, Tentative Guidelines for Manufacturers for Establishing Performance Claims for Clinical Chemical Methods, Replication Experiment. NCCLS, Villanova, PA, (1982). Precision is calculated using NCCLS EP3-T, “Tentative Guidelines for Manufacturers for Establishing Performance Claims for Clinical Chemical Methods, Replication Experiment.”
## L. Test Principle:
Iron in the specimen is released from the transferrin using acetic acid. This iron is reduced to the ferrous cation and is bound by ferene, which is a sensitive iron indicator. The increase in absorbance at 578 nm after subtracting the sample blank is proportional to the iron concentration of the sample.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Serum controls are assayed in triplicate twice per day over eleven days on a single Selectra Analyzer. Estimates of within run and total imprecision are calculated as described in NCCLS publication EP3-T.
**Precision of Vitalab Iron Reagent**
| Sample | n | mean | Within Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | 1SD | %CV | 1SD | %CV |
| Serum 1 | 66 | 57 | 0.9 | 1.6% | 1.5 | 2.7% |
| Serum 2 | 66 | 158 | 0.9 | 0.6% | 3.0 | 1.9% |
| Serum 3 | 66 | 260 | 1.0 | 0.4% | 5.2 | 2.0% |
#### b. Linearity/assay reportable range:
Eleven iron standards are prepared to span the linear range of the application by dissolving NIST iron in a dilute hydrochloric acid matrix. These standards are assayed on a single Selectra E Analyzer in ascending order over four independently calibrated analytical runs. Standard recoveries are compared to their reference values by least squares linear regression. A residual statistic is calculated for each standard as the difference between the mean recovery and its value predicted from the regression statistics. The assay is linear from 0 ug/dL to 1,000 µg/dL as shown by residuals of less than 5 µg/dL.
#### c. Traceability (controls, calibrators, or method):
Calibrator set points are traceable to NIST SRM 937. The calibrator was approved under K935712 as a kit component. The shelf-life of the iron reagent is 15 months. Data supporting this claim were provided by the OEM.
{3}
Page 4 of 5
d. Detection limit:
Normal saline is assayed thirty times in a single analytical run. The detection limit is calculated as the mean plus two standard deviations of the results. The observed mean and standard deviation are 0.00 and 1.20 µg/dL respectively. Therefore, the detection limit of the assay is 2.4 µg/dL iron.
e. Analytical specificity:
Potential interference from icterus (bilirubin) and lipemia (triglycerides) was determined in separate studies. In each study, a serum pool with approximately normal iron levels is prepared from individual patient specimens and is divided into two aliquots. One aliquot was spiked with the potential interfering substance. The other aliquot was diluted with normal saline, as necessary, to mimic the dilution the spiked pool. Ditaurobilrubin (for bilirubin) was tested at concentrations up to 40 mg/dL with ≤4.1% bias while Intralipid was tested at concentrations up to 2000 mg/dL with ≤ -2.3% change in recovery.
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
Random specimens from individual anonymous adult patients were collected from local clinical labs. These unaltered samples were supplemented with one additional frozen serum specimen with an elevated iron concentration to yield a total of 60 serum and 60 heparinized plasma specimens. These specimens were randomly assorted into groups of 15 serum and 15 plasma specimens each. One group of serum and plasma specimens was assayed in each of four runs using the Vitalab Selectra Iron Application and the Beckman Iron Reagent on the Synchron CX5; each instrument was calibrated for the iron reagent with its required calibrator before each run.
The serum results, plasma results and the combined results for both specimen types are each compared by Deming regression assuming equal variances between methods. Regression statistics are given below:
{4}
Page 5 of 5
Regression Statistics: Comparison of Candidate and Predicate
| | Values in ug/dL (95% confidence intervals) | | |
| --- | --- | --- | --- |
| Value | Serum | Plasma | Combined |
| Intercept | -1.6
(-3.0 to -0.1) | 2.2
(0.4 to 4.0) | 1.1
(0 to 2.3) |
| Slope | 1.005 | 0.992 | 0.988 |
| s_{v,x} | 1.8 | 2.4 | 2.3 |
| N | 60 | 58 | 118 |
| Range | 13 to 287 | 9 to 161 | 9 to 287 |
b. Matrix comparison:
Serum and plasma specimens are individually compared to the predicate method by Deming regression. Regression slope statistics for these two specimen types agree to with 1.3% and the substantial overlap in their 95% confidence intervals indicates equivalency. The y-intercept statistics agree to within 4 ug/dL, which is an insignificant error compared to the clinical requirements of the test.
3. Clinical studies:
a. Clinical sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a and b are not applicable):
Not applicable.
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
Reference ranges are established in the literature and quoted from Tietz Textbook of Clinical Chemistry, Third Edition, Burtis and Ashwood, editors, W. B. Saunders Company (1999).
N. Conclusion:
The submitted information in this pre-market notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.