SYSTEM CALIBRATER
Device Facts
| Record ID | K141388 |
|---|---|
| Device Name | SYSTEM CALIBRATER |
| Applicant | Beckman Coulter Ireland, Inc. |
| Product Code | JIX · Clinical Chemistry |
| Decision Date | Jul 11, 2014 |
| Decision | SESE |
| Submission Type | Abbreviated |
| Regulation | 21 CFR 862.1150 |
| Device Class | Class 2 |
Indications for Use
The System Calibrator is a calibration serum intended to be used for the calibration of ALP, ALT, AST, Amylase, CK-NAC, GGT and LDH enzymes on Beckman Coulter AU480, AU680 and AU5800 analysers. For In Vitro Diagnostic use only.
Device Story
System Calibrator is a lyophilized human serum-based product containing chemical additives, preservatives, and enzymes of human, animal, and plant origin. It serves as a calibration reference for specific enzyme assays (ALP, ALT, AST, Amylase, CK-NAC, GGT, LDH) on Beckman Coulter AU480, AU680, and AU5800 clinical chemistry analyzers. The device is used in clinical laboratory settings by trained laboratory personnel. The calibrator is reconstituted and loaded onto the analyzer; the system uses the assigned calibrator values to adjust the instrument's response to reagent-analyte reactions, ensuring accurate quantitative determination of patient sample enzyme levels. This calibration process is essential for the clinical utility of the enzyme assays, enabling healthcare providers to make informed diagnostic and monitoring decisions based on reliable patient test results.
Clinical Evidence
Bench testing only. Stability studies (shelf life and open-vial) were conducted per CLSI EP25A on representative AU analyzer platforms. Value assignment was performed using 80 data points per reagent (10 replicates across 8 runs) to determine set points. No clinical data was required or provided.
Technological Characteristics
Lyophilized human serum matrix; contains chemical additives, preservatives, and enzymes. Designed for use on Beckman Coulter AU480, AU680, and AU5800 analyzers. Traceability to theoretical extinction coefficients. Stability: 36 months at 2-8°C (shelf life); 8 hours at 2-8°C (open vial).
Indications for Use
Indicated for the calibration of ALP, ALT, AST, Amylase, CK-NAC, GGT, and LDH enzyme assays on Beckman Coulter AU480, AU680, and AU5800 clinical chemistry analyzers. For professional in vitro diagnostic use.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Related Devices
- K161140 — VITROS Chemistry Products Calibrator Kit 3 · Ortho-Clinical Diagnostics, Inc. · May 20, 2016
- K040535 — LINICAL ENZYME CALIBRATION VERIFIERS LEVELS A-E FOR BECKMAN COULTER SYNCHRON SYSTEMS · Cliniqa Corporation · Mar 29, 2004
- K032287 — MODIFICATION TO SYNCHRON SYSTEMS ENZYME VALIDATOR SET LEVELS 1 AND 2 · Beckman Coulter, Inc. · Aug 18, 2003
- K103612 — DIMENSION VISA ENZ 2 CAL · Siemens Healthcare Diagnostics, Inc. · Jan 25, 2011