ELITECH CLINICAL SYSTEMS ELICAL 2, ELITECH CLINICAL SYSTEMS ELITROL I, ELITECH CLINICAL SYSTEMS ELITROL II MODEL CALI-05
Applicant
Elitechgroup
Product Code
JIX · Clinical Chemistry
Decision Date
Aug 30, 2011
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 862.1150
Device Class
Class 2
Indications for Use
ELITech Clinical Systems ELICAL 2 is a multi-parametric calibrator for in vitro diagnostic use in the calibration of quantitative ELITech Clinical Systems methods on ELITech Clinical Systems Selectra analyzers. ELITech Clinical Systems ELITROL I & ELITROL II are multi-parametric control sera for in vitro diagnostic use in quality control of quantitative ELITech Clinical Systems methods on ELITech Clinical Systems Selectra analyzers.
Device Story
ELITech Clinical Systems ELICAL 2 (calibrator) and ELITROL I/II (quality control sera) are lyophilized human serum-based products used to calibrate and monitor quantitative clinical chemistry assays on ELITech Selectra analyzers. These products are sourced from a commercial vendor and re-labeled. The calibrator ensures accurate system calibration, while the controls verify assay performance. Healthcare providers, such as laboratory technicians, reconstitute the lyophilized material with distilled or deionized water and run the samples on the Selectra analyzer. The analyzer compares the measured values against assigned target ranges provided in lot-specific value sheets. If results fall outside specified ranges, the provider performs recalibration or maintenance. These products benefit patients by ensuring the accuracy and reliability of clinical chemistry test results used for diagnosis and monitoring.
Clinical Evidence
Bench testing only. No clinical data provided. Verification of calibrator and control values was performed by running samples in quadruplicate on two Selectra analyzers using two reagent lots. Results were compared against assigned bottle values, with acceptance criteria set at +/- 9% of the previously assigned values.
Technological Characteristics
Lyophilized human serum-based calibrator and control materials. Reconstitution required with distilled or deionized water. Stability: 2-8°C unopened; specific post-reconstitution stability windows provided for 15-25°C, 2-8°C, and -25 to -15°C. Traceability to IFCC formulations and ERM-AD454. Designed for use on ELITech Clinical Systems Selectra analyzers.
Indications for Use
Indicated for in vitro diagnostic use as a multi-parametric calibrator (ELICAL 2) and multi-parametric control sera (ELITROL I & II) for quantitative clinical chemistry assays on ELITech Clinical Systems Selectra analyzers.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
Roche Diagnostics Calibrator for Automated Systems (C.f.a.s) (K033501)
K130765 — ELITECH CLINICAL SYSTEMS ELICAL 2; ELITECH CLINICAL SYSTEMS ELITROL I, ELITECH CLINICAL SYSTEMS ELITROL II · Elitechgroup · Apr 4, 2013
K151552 — ELITech Clinical Systems ELICAL 2, ELITech Clinical Systems ELITROL I, ELITech Clinical Systems ELITROL II · Elitechgroup · Jul 8, 2015
K102016 — PRECICONTROL CLINCHEM MULTI 1 AND 2 · Roche Diagnostics · Sep 1, 2010
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k112029
B. Purpose for Submission:
Claim of previously cleared analytes (ALP, ALT/GPT, Gamma-GT and LDH) for use on Elitech's test system under Elitech's label using previously cleared multi-analyte controls and calibrators
C. Measurand:
Calibrators and assayed controls for ALP, ALT/GPT, Gamma-GT and LDH
D. Type of Test:
Calibrators and Quality Control Material
E. Applicant:
ELITech Clinical Systems, SEPPIM S.A.S.
F. Proprietary and Established Names:
ELITech Clinical Systems ELITROL I
ELITech Clinical Systems ELITROL II
ELITech Clinical Systems ELICAL 2
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JIX
Calibrator, Multi-analyte mixture | Class II | 21 CFR 862.1150
Calibrator | Clinical Chemistry (75) |
| JJY
Multi-analyte controls, all kinds (assayed) | Class I, reserved | 21 CFR 862.1660
Quality control material (assayed and unassayed) | Clinical Chemistry (75) |
{1}
H. Intended Use:
1. Intended use(s):
Refer to indication for use below
2. Indication(s) for use:
ELITech Clinical Systems ELICAL 2 is a multi-parametric calibrator for in vitro diagnostic use in the calibration of quantitative ELITech Clinical Systems methods on ELITech Clinical Systems Selectra analyzers.
ELITech Clinical Systems ELITROL I & ELITROL II are multi-parametric control sera for in vitro diagnostic use in quality control of quantitative ELITech Clinical Systems methods on ELITech Clinical Systems Selectra analyzers.
3. Special conditions for use statement(s):
For prescription use
4. Special instrument requirements:
ELITech Clinical Systems Selectra analyzers
I. Device Description:
ELITech Clinical Systems purchases the below described products from a commercial vendor and re-label it for sale for use with the Elitech test systems. These products were cleared under k033501 and k041227 sponsored by the supplier. Elitech did not change the content and the packaging except the labeling change.
ELITech Clinical Systems ELICAL 2 is a lyophilized calibrator based on human serum containing constituents to ensure optimal calibration.
ELITech Clinical Systems ELITROL I and ELITROL II are two level quality control products consisting of lyophilized human serum containing constituents at desired levels.
ELICAL 2, Elitrol I and Elitrol II are prepared exclusively from the blood of donors tested individually and found to be negative for HbsAg and to antibodies to HCV and HIV according to FDA-approved methods or methods in compliance with the European Directive 98/79/EC, Annex II, List A.
J. Substantial Equivalence Information:
1. Predicate device name(s):
{2}
Roche Calibrator f.a.s.
Roche Precinorm U/Precipath U
2. Predicate 510(k) number(s):
Roche Calibrator f.a.s. (k033501)
Roche Precinorm U/Precipath U (k041227)
3. Comparison with predicate:
| Characteristic | ELICAL 2 (New Device) | Roche Calibrator f.a.s. (Predicate Device, k033501) |
| --- | --- | --- |
| Similarities | | |
| Intended Use | Same | Is a multi-parametric calibrator used for calibration of quantitative assays |
| Format | Same | Lyophilized |
| Levels | Same | 1 |
| Handling | Same | Reconstitute with exactly 3.0 mL distilled or deionized water. |
| Matrix | Same | Human Serum |
| Stability | Same | Unopened: • Store at 2-8°C until expiration date Reconstituted: Between 15-25 °C : 12 hours Between 2-8 °C : 5 days Between (-25)-(-15) °C : 4 weeks (when frozen once) |
| Differences | | |
| Labeling | Labeled under ELITech | Labeled under Roche |
| Claim | 9 of the 16 analytes | 16 analytes |
| Characteristic | Elitrol I/Elitrol II (New Device) | Roche Precinorm U/Precipath U (Predicate Device, k041227) |
| --- | --- | --- |
| Similarities | | |
| Intended Use | Same | Are multi-parametric control sera used in quality control of quantitative assays |
| Format | Same | Lyophilized |
| Matrix | Same | Human serum |
| Levels | Same | 2 |
| Handling | Same | Reconstitute with exactly 5.0 mL distilled or deionized water |
| Stability | Same | Unopened: • Store at 2-8°C until expiration date Reconstituted: Between 15-25 °C : 12 hours Between 2-8 °C : 5 days |
{3}
4
| | | Between (-25)-(-15) °C : 4 weeks
(when frozen once) |
| --- | --- | --- |
| Differences | | |
| Labeling | Labeled under ELITech | Labeled under Roche |
| Claim | 9 of the 16 analytes | 16 analytes |
K. Standard/Guidance Document Referenced (if applicable):
None were referenced
L. Test Principle:
Not applicable
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Not applicable
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability
| Components | Traceability |
| --- | --- |
| ALP | IFCC formulation (Tietz, 1983), manual measurement |
| ALT-GPT | ERM-AD454/IFCC |
| Gamma-GT | Original formulation (Persijn/v.d. Slik, 1976), manual measurement |
| LDH | IFCC formulation (Schumann, 2002), manual measurement |
Abbreviations:
ERM: European Reference Material
IFCC = International Federation of Clinical Chemistry and Laboratory Medicine
Value Assignment:
{4}
EliTech performs verification on the provided calibrator values using Elitech's test systems on each calibrator lot. If the value provided by the manufacturer verifies, then it is used on the value sheet provided to the end user. When each new lot is received, the new lot is used to calibrate the system using the provided value. Then already-validated lots of control and an earlier lot of ELICAL are run as unknowns in quadruplicate on 2 Selectra analyzers, using 2 reagent lots, and 3 test vials per lot. If the result (mean of 48 tests) obtained is within the specification, +/-9% of the previously assigned bottle values, then the supplier's value is accepted.
Value assignment for the ELITROL I and ELITRIOL II control solutions was performed by quadruplicate runs on 2 Selectra analyzers, using 2 reagent lots, and 3 test vials per lot. The mean analyte value is calculated and a target value is assigned. The labeling states that obtained control solution values should fall within the specified range provided on lot-specific value sheets and that laboratories should establish appropriate quality control procedures when using this product for its intended use. The labeling also states that laboratories should recalibrate when reagents lots change, when quality control results fall out of range, and after a maintenance operation.
**Stability:**
Calibrator material is purchased from a commercial vendor (previously cleared under k033501) with no change to the content and the packaging except the labeling change. The sponsor claimed the following stability determined by the provider: ELICAL 2 is stable until the expiration date printed on the label when stored at 2-8°C prior to reconstitution. After reconstitution the stability is 8 hours when stored at 15-25°C, 2 days at 2-8°C or 4 weeks (when frozen once) at -25°C and -15°C.
Control material is purchased from a commercial vendor (previously cleared under k041227) with no change to the content and the packaging except the labeling change. The sponsor claimed the following stability determined by the provider: Before reconstitution, the shelf-life of the ELITech Clinical Systems Elitrol I and Elitrol II is 30 months at 2-8°C. After reconstitution the stability is 12 hours when stored at 15-25°C, 5 days when stored at 2-8°C or 4 weeks (when frozen once) at -25°C and -15°C.
**d. Detection limit:**
Not applicable
**e. Analytical specificity:**
Not applicable
**f. Assay cut-off:**
Not applicable
{5}
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Not applicable
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
6
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.