Lumipulse G FSH-N Calibrators
Device Facts
| Record ID | K163534 |
|---|---|
| Device Name | Lumipulse G FSH-N Calibrators |
| Applicant | Fujirebio Diagnostics,Inc. |
| Product Code | JIT · Clinical Chemistry |
| Decision Date | Jan 13, 2017 |
| Decision | SESE |
| Submission Type | Abbreviated |
| Regulation | 21 CFR 862.1150 |
| Device Class | Class 2 |
Indications for Use
Lumipulse G FSH-N Calibrators are for in vitro diagnostic use in the calibration of Lumipulse G FSH-N on the LUMIPULSE G System.
Device Story
Lumipulse G FSH-N Calibrators are liquid, ready-to-use reagents used to calibrate the Lumipulse G FSH-N assay on the LUMIPULSE G System. The calibrator kit consists of two levels (0 mIU/mL and 250 mIU/mL) containing follicle-stimulating hormone (FSH) in a Tris buffer with bovine protein stabilizer and sodium azide preservative. The device is used in clinical laboratory settings by trained laboratory personnel. Master calibration data are encoded in a 2D barcode on the immunoreaction cartridge case. The system uses this data and the calibrator measurements to generate a calibration curve, which is then used to automatically calculate the FSH concentration of patient specimens in mIU/mL. This quantitative output assists clinicians in assessing patient FSH levels, which is relevant for reproductive health and endocrine function diagnostics.
Clinical Evidence
Bench testing only. No clinical data. Performance was validated through analytical studies including stability (real-time and open-vial), transport simulation, and value assignment traceability to the 1st International Standard (NIBSC 92/510). Stability studies confirmed a 12-month shelf life at 2-10°C. Accuracy and reproducibility met pre-defined acceptance criteria (e.g., CV ≤10%, accuracy within ±20%).
Technological Characteristics
Liquid, ready-to-use calibrator set. Contains FSH antigen in 0.15 M sodium chloride, Tris buffer, bovine protein stabilizer, and sodium azide preservative. Traceable to NIBSC 92/510. Designed for use with the LUMIPULSE G System. Stability validated per CLSI EP25-A. Storage at 2-10°C.
Indications for Use
Indicated for in vitro diagnostic use in the calibration of the Lumipulse G FSH-N assay on the LUMIPULSE G System. No specific patient population is described.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Related Devices
- K163521 — Lumipulse G LH Calibrators · Fujirebio Diagnostics,Inc. · Jan 12, 2017
- K012399 — MODIFICATION TO ELECSYS FSH II CALSET · Roche Diagnostics Corp. · Aug 24, 2001
- K111023 — ABBOTT ARCHITECT LH CALIBRATORS · Abbott Laboratories · Jun 10, 2011
- K163546 — Lumipulse G Progesterone-N Calibrators · Fujirebio Diagnostics,Inc. · Jan 13, 2017
- K101460 — LOCI 8 CALIBRATOR, MODEL KC 646 · Siemens Healthcare Diagnostics · Jul 9, 2010