DIMENSION VISTA PROGESTERONE CALIBRATOR (PROG CAL)
Device Facts
| Record ID | K130698 |
|---|---|
| Device Name | DIMENSION VISTA PROGESTERONE CALIBRATOR (PROG CAL) |
| Applicant | Siemens Healthcare Diagnostics |
| Product Code | JIT · Clinical Chemistry |
| Decision Date | Apr 5, 2013 |
| Decision | SESE |
| Submission Type | Abbreviated |
| Regulation | 21 CFR 862.1150 |
| Device Class | Class 2 |
Indications for Use
The PROG CAL is an in vitro diagnostic product for the calibration of the Progesterone method on the Dimension Vista® System.
Device Story
Dimension Vista® Progesterone Calibrator (PROG CAL) is a liquid, frozen, human serum-based product containing progesterone and preservatives. Used in clinical laboratory settings to calibrate the Progesterone assay on the Dimension Vista® System. The calibrator is provided in ten vials across five concentration levels (0 to 44.0 ng/mL). Values are assigned using multiple Dimension Vista instruments and reagent lots, traceable to the Isotope Dilution Gas Chromatography Mass Spectrometry (ID/GC/MS) reference method. The healthcare provider uses the calibrator to establish the calibration curve on the instrument, ensuring accurate quantification of progesterone levels in patient samples. This process supports clinical decision-making regarding patient hormone status.
Clinical Evidence
Bench testing only. Stability studies confirmed 12-month shelf life at -15°C to -25°C and 15-day stability for opened vials at 2-8°C. Value assignment was validated using 45 replicates per level across multiple instruments and reagent lots.
Technological Characteristics
Human serum-based liquid calibrator; frozen storage (-15°C to -25°C); five concentration levels (0, 1.00, 8.00, 20.0, 44.0 ng/mL); traceable to ID/GC/MS reference method; used on Dimension Vista® 1500 Clinical Chemistry System.
Indications for Use
Indicated for the calibration of the Progesterone (PROG) assay on the Dimension Vista® System. For prescription use only.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Related Devices
- K113373 — DIMENSION VISTA LOCI 9 CALIBRATOR- LOCI 9 CAL · Siemens Healthcare Diagnostics, Inc. · Dec 30, 2011
- K163546 — Lumipulse G Progesterone-N Calibrators · Fujirebio Diagnostics,Inc. · Jan 13, 2017
- K103683 — IMMULITE/IMMULITE 1000 PROGESTERONE CALIBRATION VERIFICATION (CVM) · Siemens Healthcare Diagnostics · Apr 28, 2011