For the IN VITRO diagnostic use as calibrators for the DCL CRP-ADVANCE Assay for the quantitation of C-reactive protein in serum. The CRP-ADVANCE Multi-Calibrator Set, Cat. No SE-250 are C-reactive protein standards containing known quantities of human c-reactive protein. These calibrators are to be used with the DCL CRP-ADVANCE Assay.
Device Story
CRP-ADVANCE Multi-Calibrator Set (Model SE-250) consists of 5 levels of human C-reactive protein standards (3.0–362 mg/L). Used as calibrators for DCL CRP-ADVANCE immuno-agglutination assays on clinical chemistry analyzers (e.g., Hitachi 911). Liquid, ready-to-use serum matrix. Donor units tested non-reactive for HBsAg, HCV, and HIV-1/2. Used by laboratory professionals to establish calibration curves for quantitative CRP measurement, ensuring accuracy of patient test results.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Human C-reactive protein standards in serum matrix. 5 levels (3.0-362 mg/L). Packaged in 2 mL vials. Ready-to-use liquid. Traceable to ERM 470. Compatible with Hitachi 911 or equivalent analyzers. Storage at 2-8°C.
Indications for Use
Indicated for in vitro diagnostic use as calibrators for the DCL CRP-ADVANCE Assay to quantify C-reactive protein in human serum.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
K021607 — RDI HS CRP CALIBRATOR SET · Reference Diagnostics, Inc. · Jun 27, 2002
K991210 — K-ASSAY CRP CALIBRATOR SET · Kamiya Biomedical Co. · May 11, 1999
K030546 — CRP (II) CALIBRATORS · Denka Seiken Co., Ltd. · Jun 2, 2003
Submission Summary (Full Text)
{0}
1 of 4
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k060140
B. Purpose for Submission:
New Submission
C. Measurand:
Calibrator materials for C-Reactive Protein
D. Type of Test:
Not applicable
E. Applicant:
Diagnostic Chemicals LTD.
F. Proprietary and Established Names:
CRP-ADVANCE MULTI CALIBRATOR SET, MODEL SE-250
G. Regulatory Information:
1. Regulation section:
21CFR Sec.- 862.1150-Calibrator
2. Classification:
Class 2
3. Product code:
JIT - Calibrator, Secondary
4. Panel:
Chemistry (75)
H. Intended Use:
1. Intended use(s):
See indications for use below.
2. Indication(s) for use:
For the IN VITRO diagnostic use as calibrators for the DCL CRP-ADVANCE Assay for the quantitation of C-reactive protein in serum.
The CRP-ADVANCE Multi-Calibrator Set, Cat. No SE-250 are C-reactive protein standards containing known quantities of human c-reactive protein. These calibrators are to be used with the DCL CRP-ADVANCE Assay.
{1}
3. Special conditions for use statement(s):
For prescription use
4. Special instrument requirements:
The Hitachi 911 or any analyzer with assay compatible specifications
I. Device Description:
The calibrators included are C-reactive protein standards containing known quantities of human C-reactive protein. These calibrators are to be used with the DCL CRP-ADVANCE Assay. There are 5 levels of calibrators ranging from 3.0 mg/L to 362 mg/L. They are packaged ready to use in 2 mL vials.
Each serum donor unit used in the preparation of this product has been tested by an FDA approved method and found non-reactive for the presence of HBsAg, HCV, and antibody to HIV-1/2.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Equal Diagnostics', CRP Ultra Wide Range Calibrator Set B manufactured by Denka Seiken Co., Ltd.
2. Predicate 510(k) number(s):
k030546
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended use | Intended to calibrate
Quantitative C-Reactive
Protein Assay | Intended to calibrate
Quantitative C-Reactive
Protein Assay |
| Sample | Serum | Serum |
| Matrix | Liquid ready to use | Liquid ready to use |
K. Standard/Guidance Document Referenced (if applicable):
CLSI - Evaluation of the Linearity of Quantitative Analytical Methods; Approved Guideline - EP06
CLSI - Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline - Second Edition - EP09
L. Test Principle:
Calibrator for Quantitative immuno-agglutination assay
2 of 4
{2}
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Not Applicable
b. Linearity/assay reportable range:
Not Applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The calibrator is traceable to European Reference Materials (ERM) 470 and the assigned values of 3.0, 6.1, 30.0, 181.0 and 362.0 mg/L have been validated by calibrator comparison studies on the Hitachi 911 yielding results of 2.98, 6.12, 30.18, 184.71 and 358.75 mg/L respectively.
Stability studies include accelerated and ongoing real time studies for unopened shelf life when stored at 2-8°C and real time studies for opened calibrators when stored at 2-8°C at 30 days.
d. Detection limit:
Not Applicable
e. Analytical specificity:
Not Applicable
f. Assay cut-off:
Not Applicable
2. Comparison studies:
a. Method comparison with predicate device:
Not Applicable
b. Matrix comparison:
Not Applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not Applicable
4. Clinical cut-off:
{3}
Not Applicable
5. Expected values/Reference range: Not Applicable
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
4 of 4
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.