ADVIA Centaur Cortisol (COR)

K192788 · Siemens Healthcare Diagnostics, Inc. · JFT · Nov 25, 2019 · Clinical Chemistry

Device Facts

Record IDK192788
Device NameADVIA Centaur Cortisol (COR)
ApplicantSiemens Healthcare Diagnostics, Inc.
Product CodeJFT · Clinical Chemistry
Decision DateNov 25, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1205
Device ClassClass 2

Indications for Use

For in vitro diagnostic use in the quantitative determination of cortisol in serum, plasma (EDTA and lithium heparin), and urine using the ADVIA Centaur® XP system. Measurements of cortisol are used in the diagnosis and treatment of disorders of the adrenal gland.

Device Story

The ADVIA Centaur Cortisol (COR) assay is a competitive immunoassay using direct chemiluminescent technology. It processes patient samples (serum, plasma, or urine) on the ADVIA Centaur XP system. During the assay, cortisol in the sample competes with acridinium ester-labeled cortisol for binding to polyclonal rabbit anti-cortisol antibodies attached to paramagnetic particles. The system measures the resulting chemiluminescent signal to quantify cortisol levels. Results are calculated using a 2-point calibration curve and a master curve provided via reagent barcode. The device is used in clinical laboratory settings by trained personnel. Healthcare providers use the quantitative cortisol results to diagnose and manage adrenal gland disorders, such as adrenal insufficiency or Cushing's syndrome. This submission specifically adds EDTA and lithium heparin plasma as validated sample types.

Clinical Evidence

No clinical studies were performed. Evidence is based on analytical bench testing. Specimen equivalence was demonstrated by comparing serum to EDTA plasma (N=83) and lithium-heparin plasma (N=99), showing high correlation (r=1.00) and slopes of 0.95 and 0.96, respectively. Interference testing per CLSI EP07-ed3 confirmed that EDTA (9.0 mg/mL) and heparin (75 U/mL) do not significantly bias results (bias ≤ 2.9%).

Technological Characteristics

Competitive chemiluminescent immunoassay. Reagents: acridinium ester-labeled cortisol, rabbit anti-cortisol antibody, monoclonal mouse anti-rabbit IgG antibody, paramagnetic particles. Energy source: instrument-based light detection. Connectivity: ADVIA Centaur XP system. Standards: CLSI EP07-Ed3 for interference testing.

Indications for Use

Indicated for the quantitative determination of cortisol in serum, plasma (EDTA and lithium heparin), and urine to aid in the diagnosis and treatment of adrenal gland disorders.

Regulatory Classification

Identification

A cortisol (hydrocortisone and hydroxycorticosterone) test system is a device intended to measure the cortisol hormones secreted by the adrenal gland in plasma and urine. Measurements of cortisol are used in the diagnosis and treatment of disorders of the adrenal gland.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY ## I Background Information: A 510(k) Number K192788 B Applicant Siemens Healthcare Diagnostics, Inc. C Proprietary and Established Names ADVIA Centaur Cortisol (COR) D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | JFT | Class II | 21 CFR 862.1205 - Cortisol (Hydrocortisone And Hydroxycorticosterone) Test System | CH - Clinical Chemistry | ## II Submission/Device Overview: A Purpose for Submission: Addition of K2-EDTA and lithium heparin plasma sample types. B Measurand: Cortisol C Type of Test: Chemiluminescent Immunoassay K192788 - Page 1 of 5 {1} III Intended Use/Indications for Use: A Intended Use(s): See Indication(s) for Use below. B Indication(s) for Use: For in vitro diagnostic use in the quantitative determination of cortisol in serum, plasma (EDTA and lithium heparin), and urine using the ADVIA Centaur XP system. Measurements of cortisol are used in the diagnosis and treatment of disorders of the adrenal gland. C Special Conditions for Use Statement(s): Prescription use only D Special Instrument Requirements: ADVIA Centaur XP IV Device/System Characteristics: A Device Description: The ADVIA Centaur Cortisol (COR) assay is a competitive immunoassay using direct chemiluminescent technology. Results are determined using a calibration curve that is generated specifically on each instrument by a 2-point calibration and a master curve with the reagent bar code. The ADVIA Centaur COR reagent kit contains the following: ADVIA Centaur ReadyPack® primary reagent pack contains Lite Reagent and Solid Phase Reagent. Lite reagent consists of 2.5 mL reagent pack with cortisol labeled with acridinium ester in buffered saline with sodium salicylate (~50 mg/mL), sodium azide (0.1%) and preservatives. Solid phase reagent consists of 12.5 mL reagent pack with rabbit anti-cortisol antibody (~1.1 µg/mL) bound to monoclonal mouse anti-rabbit IgG antibody (~56 µg/mL) covalently coupled to paramagnetic particles in buffered saline with sodium azide (0.1%) and preservatives. B Principle of Operation: The ADVIA Centaur Cortisol (COR) assay is a competitive immunoassay using direct chemiluminescent technology. Cortisol in the patient sample competes with acridinium ester labeled cortisol in the Lite Reagent for binding to polyclonal rabbit anti-cortisol antibody in the Solid Phase. The polyclonal rabbit anti-cortisol antibody is bound to monoclonal mouse anti-rabbit antibody, which is covalently coupled to paramagnetic particles in the Solid Phase. An inverse relationship exists between the amount of cortisol present in the patient sample and the amount of relative light units (RLUs) detected by the instrument. K192788 - Page 2 of 5 {2} K192788 - Page 3 of 5 V Substantial Equivalence Information: A Predicate Device Name(s): ADVIA Centaur Cortisol (COR) Assay B Predicate 510(k) Number(s): K142723 C Comparison with Predicate(s): | Device & Predicate Device(s): | K192788 | K142723 | | --- | --- | --- | | Device Trade Name | ADVIA Centaur Cortisol (COR) Assay | Same | | General Device Characteristic Similarities and differences | | | | Intended Use/Indications For Use | For in vitro diagnostic use in the quantitative determination of cortisol in serum, plasma (EDTA and lithium heparin), and urine using the ADVIA Centaur Systems. Measurements of cortisol are used in the diagnosis and treatment of disorders of the adrenal gland. | Same | | Traceability | Internal Standards traceable to GCMS | Same | | Assay Range | Serum and Plasma: 0.50–75 μg/dL Urine: 0.50–53 μg/dL | Serum: 0.50–75 μg/dL Urine: 0.50–53 μg/dL | | Sample Type | Serum, Plasma (EDTA and lithium heparin), Urine. | Serum and Urine | VI Standards/Guidance Documents Referenced: CLSI EP07-Ed3: Interference Testing in Clinical Chemistry {3} VII Performance Characteristics (if/when applicable): A Analytical Performance: 1. Precision/Reproducibility: Previously established in k142723. 2. Linearity: Previously established in k142723. 3. Analytical Specificity/Interference: Interference testing was conducted according to CLSI EP07-ed3: Interference Testing in Clinical Chemistry. Interfering substances were tested using human K2-EDTA and heparin pools. Various concentration of interferents were spiked into one pool at low and one at high cortisol concentrations for each matrix. The same sample pools with no interferents were used as controls. All samples were run in triplicate using ADVIA Centaur XP instrument and one reagent lot. The sponsor defined non-significant interference as ≤10% difference between the spiked samples and control. Results are summarized below. | Potential Interferent | Highest concentration tested that showed no interference | | --- | --- | | K2-EDTA | 9.0 mg/mL | | Lithium heparin | 75 U/mL | 4. Assay Reportable Range: Previously established in k142723. 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): Previously established in k142723. 6. Detection Limit: Previously established in k142723. 7. Assay Cut-Off: Not applicable. B Comparison Studies: 1. Method Comparison with Predicate Device: Previously established in k142723. K192788 - Page 4 of 5 {4} K192788 - Page 5 of 5 2. Matrix Comparison: A matrix comparison study was performed to demonstrate sample equivalence between serum and plasma (K2-EDTA and lithium heparin). Matched sample sets were tested using one ADVIA Centaur (COR) assay reagent lot and one ADVIA Centaur XP instrument. One replicate was tested for each sample. A total of 83 dipotassium EDTA plasma/serum and 99 lithium heparin plasma/serum sample pairs were tested. Deming linear regression analysis was used to analyze measured values using serum as the comparator. The results are shown in the table below. | Comparison | N* | Correlation Coefficient (r) | Regression Equation | Sample Range | | --- | --- | --- | --- | --- | | K2EDTA Plasma vs. Serum | 83 | 1.00 | y = 0.95x + 0.24 | 0.29-67.06 μg/dL | | Lithium Heparin Plasma vs. Serum | 99 | 1.00 | y = 0.96x + 0.54 | 0.29-67.06 μg/dL | *N=Number of samples tested C Clinical Studies: 1. Clinical Sensitivity: Not applicable. 2. Clinical Specificity: Not applicable. 3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): Not applicable. D Clinical Cut-Off: Not applicable. E Expected Values/Reference Range: Previously established in k142723. VIII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...