VITALAB AMYLASE REAGENT

K040534 · Clinical Data, Inc. · JFJ · Mar 11, 2004 · Clinical Chemistry

Device Facts

Record IDK040534
Device NameVITALAB AMYLASE REAGENT
ApplicantClinical Data, Inc.
Product CodeJFJ · Clinical Chemistry
Decision DateMar 11, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1070
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The Vitalab ot-Amylase Reagent Kit is intended for use with the Vitalab Selectra Analyzer for the quantitative determination of a-amylase in serum and plasma. Amylase results may be used for the diagnosis and treatment of pancreatitis (inflammation of the pancreas).

Device Story

Vitalab a-Amylase Reagent is a single-part reagent used with the Vitalab Selectra Analyzer. It measures amylase levels in serum and plasma samples via cleavage of 2-chloro-4-nitrophenyl-a-D-maltotrioside (CNP-a-G3) to produce 2-chloro-4-nitrophenol. The analyzer performs quantitative analysis; results are provided to clinicians to assist in diagnosing and treating pancreatitis. Used in clinical laboratory settings by trained personnel. The device provides quantitative amylase concentration values, aiding clinical decision-making regarding pancreatic inflammation.

Clinical Evidence

Bench testing only. Precision studies (n=60 per sample) showed total CVs of 0.9-1.7%. Linearity was confirmed up to 1,645 U/L (default) and 3,311 U/L (rerun). Method comparison (n=120) against the predicate device yielded a slope of 1.086 and intercept of -6.1 U/L. Interference testing evaluated bilirubin, hemoglobin, and triglycerides, with specific suppression thresholds identified. No clinical patient outcome studies were performed.

Technological Characteristics

Single-part liquid reagent; catalytic enzymatic assay using CNP-a-G3 substrate. Designed for use on Vitalab Selectra Analyzer. Quantitative photometric measurement. No specific materials of construction or connectivity standards provided.

Indications for Use

Indicated for the diagnosis and treatment of pancreatitis in adult patients.

Regulatory Classification

Identification

An amylase test system is a device intended to measure the activity of the enzyme amylase in serum and urine. Amylase measurements are used primarily for the diagnosis and treatment of pancreatitis (inflammation of the pancreas).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: k040534 B. Analyte: alpha-Amylase C. Type of Test: Quantitative D. Applicant: Clinical Data, Inc E. Proprietary and Established Names: Vitalab Alpha-Amylase Reagent Kit F. Regulatory Information: 1. Regulation section: 21 CFR 862.1070 2. Classification: Class II 3. Product Code: JFJ 4. Panel: Chemistry (75) G. Intended Use: 1. Intended use(s): The Vitalab Alpha-Amylase Reagent Kit is a device for the quantitative measurement of alpha-amylase activity in serum and plasma. 2. Indication(s) for use: Amylase measurements may be used for the diagnosis and treatment of pancreatitis (inflammation of the pancreas). 3. Special condition for use statement(s): Prescription use 4. Special instrument Requirements: The Vitalab Amylase Reagent is intended to be used with the Vitalab Selectra E Chemistry Analyzer. H. Device Description: The Vitalab Amylase Reagent and the Vitalab Selectra E Analyzer are used as a system for the quantitative analysis of alpha-amylase in serum and plasma. The Vitalab Amylase Reagent is Vitalab Amylase Reagent Kit {1} supplied as a single part liquid-stable reagent. The reagent components are: 2.25 mmol/L 2-chloro-4-nitrophenyl-α-D-maltotrioside, 300 mmol/L sodium chloride, 6.00 mmol/L calcium chloride, 900 mmol/L potassium thiocyanate, 50 mmol/L MES buffer. Vitalab Serum Control Kit is required but not provided with the reagent. There are no calibrators with this assay. I. Substantial Equivalence Information: 1. Predicate device name(s): Roche alpha-Amylase EPS Ver.2 Reagent, product 03183742 2. Predicate K number(s): k972250 (reagent was cleared with Cobas Integra analyzer) 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | Quantitative determination of alpha-amylase in serum and plasma | Quantitative determination of alpha-amylase in serum, plasma and urine | | Measurement method | Kinetic, measured at 405 nm | Kinetic, measured at 409 nm | | Analytical Range | 5 to 1,600 U/L (default range) 1,200 to 3,200 U/L (rerun range) | 0 to 2,000 U/L 0 to 10,000 U/L (with post dilution) | | Differences | | | | Item | Device | Predicate | | Chemical Reaction | Cleavage of 2-chloro-4-nitrophenyl-α-D-maltotrioside to produce 2-chloro-4-nitrophenol | Cleavage of 5-ethylidene-G7PNP to produce GnPNP, which is detected by an indicator reaction to produce 4-nitrophenol | | Reagent components | 2-chloro-4-nitrophenyl-α-D-maltotrioside, sodium chloride, calcium chloride, potassium thiocyanate, MES buffer | 5-ethylidene-G7PNP sodium chloride, calcium chloride magnesium chloride α-glucosidase HEPES buffer | | Calibration | Assay requires no external calibration material | Single point serum calibrator | J. Standard/Guidance Document Referenced (if applicable): The sponsor cites NCCLS publications EP3-T, Tentative Guidelines for Manufacturers for Establishing Performance Claims for Clinical Chemistry Methods, Replication Experiment, and NCCLS Document EP7-P, Volume 6 No.13. (Osmotic Shock Procedure) K. Test Principle: The Vitalab Amylase Reagent determines alpha-amylase through the cleavage of 2-chloro-4-nitrophenyl-α-D-maltotrioside (CNP-α-G₃) to produce 2-chloro-4-nitrophenol (CNP), a yellow dye. Results are calculated from rate of increase in absorbance of CNP at 405 nm. {2} Page 3 of 6 # L. Performance Characteristics (if/when applicable): ## 1. Analytical performance: ### a. Precision/Reproducibility: Precision using both the default and rerun sample volumes is demonstrated by the replicate assay of commercially available control serum. These controls are spiked with higher concentrations of alpha-amylase (from porcine pancreas) for the rerun precision study. Each sample is assayed in triplicate in twenty runs over at least 10 days using the Vitalab Amylase Reagent on a Selectra E Analyzer. Precision statistics, calculated analogous to the method described in NCCLS Guideline EP3-T, are shown below. Precision of Amylase Recoveries in U/L | Instrument Range | | Within Run | | | Total | | | --- | --- | --- | --- | --- | --- | --- | | Sample | n | mean | 1SD | %CV | 1SD | %CV | | Default | | | | | | | | Serum 1 | 60 | 98 | 1.0 | 1.0% | 1.5 | 1.5% | | Serum 2 | 60 | 565 | 3.6 | 0.6% | 5.0 | 0.9% | | Serum 3 | 60 | 1,034 | 7.0 | 0.7% | 9.7 | 0.9% | | Rerun | | | | | | | | Serum 1 | 60 | 1,961 | 17 | 0.9% | 25 | 1.3% | | Serum 2 | 60 | 2,517 | 26 | 1.0% | 41 | 1.6% | | Serum 3 | 60 | 3,002 | 40 | 1.3% | 50 | 1.7% | ### b. Linearity/assay reportable range: Linearity is shown for the default and rerun ranges in separate studies. The claimed linearity/reportable range for the assay is $5 - 1600\mathrm{U / L}$ (default range) and $5 - 3200\mathrm{U / L}$ (rerun range). The extended range on rerun is accomplished by reducing the sample size by $50\%$. To demonstrate linearity for the default range, seven amylase linearity standards are prepared by quantitatively diluting a stabilized stock solution of alpha-amylase (from porcine pancreas) with a buffered solution. These standards are assayed on a single Selectra Analyzer in ascending order over four analytical runs. Standard recoveries are compared to their dilution factors by least squares linear regression which is forced through the origin. A residual statistic is calculated for each standard as the difference between the mean recovery and its predicted value from the regression statistics. The maximum residual is less than 5 U/L of amylase indicating linearity to at least $1645\mathrm{U / L}$, which is the level of the highest standard. To demonstrate linearity for the rerun range, twelve amylase linearity standards are prepared as described above. These standards are assayed on a single Selectra Analyzer in ascending order over five analytical runs. Standard recoveries are compared to their dilution factors by least squares linear regression which is forced through the origin, and a residual statistic is calculated for each standard as described above. The maximum residual is less than $12\mathrm{U / L}$ of amylase indicating linearity to at least $3311\mathrm{U / L}$, which is the level of the highest standard. ### c. Traceability (controls, calibrators, or method): The sponsor states that the method is traceable to NIST SRM 938. ### d. Detection limit: {3} Page 4 of 6 Normal saline is assayed thirty times in a single analytical run. The detection limit is calculated as the mean plus two standard deviations of the results. The observed mean and standard deviation are 0.03 and 0.18 U/L respectively. The calculated detection limit of 0.40 U/L amylase is rounded up to 1 U/L, which is the round-off error of the assay. The claimed detection limit is 5 U/L. e. Analytical specificity: Potential interference from icterus (bilirubin), hemolysis (hemoglobin) and lipemia (triglycerides) is determined in three separate studies. In each study, a serum pool with approximately normal amylase levels is prepared from individual patient specimens and is divided into two aliquots. One aliquot is spiked with the potential interfering substance. The other aliquot is diluted with normal saline, if necessary, to mimic the dilution the spiked pool. These aliquots are then blended to prepare test pools with the interferant concentrations listed below. The red blood cell (RBC) hemolysate, which is used to spike the high pool for the hemolysis test, is prepared from at least five patient specimens according to the Osmotic Shock Procedure described in NCCLS Document EP7-P, Volume 6 No.13. | Interfering Substance | Levels tested | | --- | --- | | Ditaurobilirubin | 8, 16, 24, 32, 40 mg/dL (as bilirubin) | | RBC hemolysate | 40, 80, 120, 160, 200 mg/dL (as hemoglobin) | | Intralipid, 20% | 400, 800, 1,200, 1,600 and 2,000 mg/dL (as triglycerides) | Each set of original and spiked pools is assayed in an alternating order 9 and 6 times respectively in a single analytical run. Differences in recoveries between the original and spiked pools are reported with t-statistics. Statistically significant differences greater than 5 U/L are reported in the package insert. Bilirubin above 8 mg/dL suppresses results by approximately 8 U/L. Red blood cell hemolysate added to hemoglobin concentrations of 120 and 200 mg/dL decreases recoveries by 7 and 10 U/L respectively. The addition of Intralipid to 2000 mg/dL triglycerides suppresses recoveries by 5 U/L. f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: Fifty eight fresh serum and 55 fresh heparinized plasma specimens, selected randomly from individual adult patients, were supplemented with seven abnormal specimens, some of which were purchased frozen, to yield a total of 60 serum and 60 plasma specimens. These specimens were randomly assorted into groups of 15 serum and 15 plasma specimens each. These groups are assayed over four runs using the Vitalab α-Amylase Reagent on the Selectra E and the Roche α-Amylase EPS Ver.2 Reagent on the Cobas Integra 400. Seven specimens were re-assayed in a fifth run because the Cobas results were suppressed due to inadequate sample volume. The serum results, plasma results and the combined results for both specimen types are each compared by Deming regression assuming equal variances between methods. Regression statistics are given below. {4} Page 5 of 6 Serum Correlation | | Value | 95% Confidence Interval | | --- | --- | --- | | Intercept | -7.7 U/L | -9.5 to -6.0 U/L | | Slope | 1.097 | 1.087 to 1.108 | | s_{y.x} | 3.7 U/L | | | n | 60 | | | range | 31 to 1049 U/L | | Plasma Correlation | | Value | 95% Confidence Interval | | --- | --- | --- | | Intercept | -1.7 U/L | -3.7 to 0.4 U/L | | Slope | 1.034 | 1.014 to 1.055 | | s_{y.x} | 3.6 U/L | | | n | 60 | | | range | 26 to 394 U/L | | Combined Correlation | | Value | 95% Confidence Interval | | --- | --- | --- | | Intercept | -6.1 U/L | -7.4 to -4.7 U/L | | Slope | 1.086 | 1.076 to 1.096 | | s_{y.x} | 4.0 U/L | | | n | 120 | | | range | 26 to 1049 U/L | | Where $x =$ Predicate Results and $y =$ Selectra Results b. Matrix comparison: Serum and plasma specimens are individually compared to the predicate method. Deming regression statistics suggest that the serum results are approximately $6\%$ higher than plasma results. Although, this difference is small relative to the clinical requirements of the test, further analysis suggests that it is caused by differences in the distribution of the abnormal serum and plasma samples. This is supported by Passing & Bablok regression, which is less affected by extreme values. These statistics are tabulated below. | Specimen | y-intercept (95% CI) | slope (95% CI) | | --- | --- | --- | | Serum | 2.7 U/L (0.0 to 5.4 U/L) | 0.958 (0.917 to 1.000) | | Plasma | 0.0 U/L (0.0 to 2.1 U/L) | 1.000 (0.963 to 1.000) | 3. Clinical studies: a. Clinical sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when $a$ and $b$ are not applicable): Not applicable {5} Page 6 of 6 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: The published reference range for $\alpha$-amylase, measured at 37C is 22 – 80 U/L. Tietz Textbook of Clinical Chemistry, Third Edition, W.B. Saunders Company, Philadelphia, PA, (1999) **M. Conclusion:** Based upon the information provided for the file, I recommend that the Vitalab Alpha-Amylase Reagent Kit be found substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...