EASYRA TBIL REAGENT, EASYRA DBII REAGENT, EASYRA CK REAGENT
Applicant
Medica Corp.
Product Code
CIG · Clinical Chemistry
Decision Date
Mar 1, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1110
Device Class
Class 2
Intended Use
The EasyRA TBIL reagent is intended for the quantitative measurement of Total Bilirubin in human serum and plasma of adults on the Medica EasyRA analyzer in clinical laboratories. Bilirubin measurements are used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gall bladder block. For in vitro diagnostic use only. The EasyRA DBIL reagent is intended for the quantitative measurement of Direct Bilirubin in human serum and plasma of adults on the Medica EasyRA analyzer in clinical laboratories. Bilirubin measurements are used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gall bladder block. For in vitro diagnostic use only. The EasyRA CK reagent is intended for the quantitative determination of Creatine Kinase (CK) in human serum and plasma, using the MEDICA "EasyRA Chemistry analyzer" in clinical laboratories. Measurements of CK are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy. For in vitro diagnostic use only.
Device Story
Reagent kits for quantitative colorimetric chemistry analysis on Medica EasyRA Chemistry Analyzer; inputs are human serum or plasma samples. TBIL/DBIL assays utilize non-enzymatic endpoint reactions with 2,4-dichlorophenyldiazonium salt to form azobilirubin; absorbance measured at 550 nm. CK assay utilizes coupled enzymatic reactions; CK catalyzes phosphate transfer from creatine phosphate to ADP; subsequent reactions involving hexokinase and G6PDH produce NADPH; absorbance measured at 340 nm. Used in clinical laboratories by trained personnel. Output is analyte concentration; assists clinicians in diagnosing liver, metabolic, or muscle disorders. Device expands sample compatibility to include plasma (lithium heparin) in addition to previously cleared serum.
Indicated for quantitative measurement of Total Bilirubin, Direct Bilirubin, and Creatine Kinase in adult human serum and plasma. Used for diagnosis/treatment of liver, hemolytic, hematological, metabolic disorders (TBIL/DBIL), myocardial infarction, and muscle diseases (CK). For clinical laboratory use.
Regulatory Classification
Identification
A bilirubin (total or direct) test system is a device intended to measure the levels of bilirubin (total or direct) in plasma or serum. Measurements of the levels of bilirubin, an organic compound formed during the normal and abnormal distruction of red blood cells, if used in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block.
Related Devices
K080874 — EASYRA ALB REAGENT AND EASYRA ALP REAGENT, MODEL 10218/10214, EASYRA AST REAGENT, MODEL 10206, EASYRA CO2 REAGENT · Medica Corp. · Feb 20, 2009
K130080 — EASYRA CK-MB REAGENT, EASYRA CRP REAGENT, EASY CAL CRP CALIBRATOR KIT, EASYQC CRP QUALITY CONTROL MATERIAL · Medica Corporation · May 13, 2014
K091991 — EASY RA D BIL REAGENT MODEL 10212, EASY RA GLU-H REAGENT MODEL 10200 · Medica Corp. · Sep 4, 2009
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k123802
B. Purpose for Submission:
Total Bilirubin: Addition of plasma to the already cleared device (k080810)
Direct Bilirubin: Addition of plasma to the already cleared device (k080810)
Creatine Kinase: Addition of plasma to the already cleared device (k080874)
C. Measurand:
Total Bilirubin (TBIL), Direct Bilirubin (DBIL), Creatine Kinase (CK)
D. Type of Test:
Quantitative colorimetric chemistry tests
E. Applicant:
Medica Corporation
F. Proprietary and Established Names:
EasyRA Total Bilirubin Reagent
EasyRA Direct Bilirubin Reagent
EasyRA Creatine Kinase Reagent
G. Regulatory Information:
| Device Classification Name | Regulation section | Classification | Product Code | Panel |
| --- | --- | --- | --- | --- |
| Total Bilirubin | 21 CFR § 862.1110 | II | CIG | Chemistry (75) |
| Direct Bilirubin | 21 CFR § 862.1110 | II | CIG | Chemistry (75) |
| Creatine Kinase | 21 CFR § 862.1215 | II | CGS | Chemistry (75) |
H. Intended Use:
1. Intended use(s):
{1}
See indications for use below.
2. **Indication(s) for use:**
**EasyRA Total Bilirubin Reagent**
The EasyRA TBIL reagent is intended for the quantitative determination of Total Bilirubin in human serum and plasma of adults on the MEDICA “EasyRA Chemistry Analyzer” in clinical laboratories. Bilirubin measurements are used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gall bladder block. For *in vitro* diagnostic use only.
**EasyRA Direct Bilirubin**
The EasyRA DBIL reagent is intended for the quantitative determination of Direct Bilirubin (DBIL) in human serum and plasma of adults on the MEDICA “EasyRA Chemistry Analyzer” in clinical laboratories. Bilirubin measurements are used in the diagnosis and treatment of liver, hemolytic, hematological and metabolic disorders, including hepatitis and gall bladder block. For *in vitro* diagnostic use only.
**EasyRA Creatine Kinase**
The EasyRA CK reagent is a device intended for the quantitative determination of Creatine Kinase (CK) in human serum and plasma, using the MEDICA “EasyRA Chemistry analyzer” in clinical laboratories. Measurements of CK are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy. For *in vitro* diagnostic use only.
3. **Special conditions for use statement(s):**
Prescription use only
4. **Special instrument requirements:**
Medica EasyRA Analyzer
I. **Device Description:**
The EasyRA TBIL reagent kit is packaged as two ready to use reagents, R1 and R2. The formulation of each reagent is presented in the following table:
{2}
| R1 | Reactive Ingredients | Concentration |
| --- | --- | --- |
| | NaCl | 154 mmol/L |
| | HCl | 190 mmol/L |
| | Surfactants and Preservatives | |
| R2 | HCl | 417 mmol/L |
| | 2,4- dichlorophenyldiazonium salt | 5 mmol/L |
| | Surfactant | |
The EasyRA DBIL reagent kit is packaged as two ready to use reagents, R1 and R2. The formulation of each reagent is presented in the following table:
| R1 | Reactive Ingredients | Concentration |
| --- | --- | --- |
| | EDTA-Na2 | 0.1 mmol/L |
| | NaCl | 154 mmol/L |
| | Sulfamic acid | 100 mmol/L |
| R2 | 2,4 - dichlorophenyldiazonium salt | 5 mmol/L |
| | HCl | 0.9 mmol/L |
| | EDTA - Na2 | 0.13 mmol/L |
The EasyRA CK reagent kit is packaged as two ready to use reagents, R1 and R2. The formulation of each reagent is presented in the following table:
| R1 | Reactive Ingredients | Concentration |
| --- | --- | --- |
| | Imidazole Buffer (pH 6.7) | 100 mmol/L |
| | D-Glucose | 20 mmol/L |
| | N-Acetyl-L-Cysteine | 20 mmol/L |
| | Magnesium Acetate | 10.0 mmol/L |
| | NADP | 2.0 mmol/L |
{3}
4
| | EDTA | 2.0 mmol/L |
| --- | --- | --- |
| | Hexokinase (Baker’s yeast) | 2500 U/L |
| R2 | Imidazole Buffer (pH 6.7) | 100 mmol/L |
| | Creatine phosphate | 30 mmol/L |
| | ADP | 2.0 mmol/L |
| | AMP | 5.0 mmol/L |
| | Diadenosine pentaphosphate | 10.0 mmol/L |
| | Glucose-6-PDH (Baker’s yeast) | 1500 U/L |
| | EDTA | 2.0 mmol/L |
## J. Substantial Equivalence Information:
| New Device | Predicate Device Name | Predicate K Number(s) |
| --- | --- | --- |
| EasyRA Total Bilirubin Reagent | EasyRA Total Bilirubin Reagent | k080810 |
| EasyRA Direct Bilirubin | EasyRA Direct Bilirubin Reagent | k080810 |
| EasyRA Creatine Kinase Reagent | EasyRA Creatine Kinase Reagent | k080874 |
Comparison with predicate:
| Characteristic | Predicate device: EasyRA Total Bilirubin Reagent (k080810) | Candidate device: EasyRA Total Bilirubin Reagent |
| --- | --- | --- |
| Intended Use | Intended for the quantitative determination of Total Bilirubin in human serum, using the MEDICA “EasyRA Chemistry Analyzer” in clinical laboratories | Same |
| Sample Type | Serum | Serum, Plasma |
| Reagent | Liquid Ready to use | Same |
| Assay Principle | Non-enzymatic reaction based on the binding of the analyte to an organic/inorganic compound to produce a chromogen at endpoint. | Same |
{4}
5
| | The chromogen absorbs light of specific wavelength and the resulting absorbance is measured. | |
| --- | --- | --- |
| Analytical Range | 0.08 – 20 mg/dL | Same |
| Characteristic | Predicate device: EasyRA Direct Bilirubin Reagent (k080810) | Candidate device: Easy RA Direct Bilirubin Reagent |
| --- | --- | --- |
| Intended Use | Intended for the quantitative determination of Direct Bilirubin in human serum, using the MEDICA “EasyRA Chemistry Analyzer” in clinical laboratories | Same |
| Sample Type | Serum | Serum, Plasma |
| Reagent | Liquid ready to use | Same |
| Assay Principle | Endpoint | Same |
| Analytical Range | 0.06 – 10 mg/dL | Same |
| Characteristic | Predicate device: EasyRA Creatine Kinase Reagent (k080874) | Candidate device: EasyRA Creatine Kinase Reagent |
| --- | --- | --- |
| Intended Use | Intended for the quantitative determination of Creatine kinase in human serum, using the MEDICA “EasyRA Chemistry Analyzer” in clinical laboratories | Same |
| Sample Type | Serum | Serum, Plasma |
| Reagent | Liquid ready to use | Same |
| Assay Principle | Enzymatic reaction | Same |
| Linearity | 3-1200 U/L | Same |
K. Standard/Guidance Document Referenced (if applicable):
Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline- Second Edition (CLSI EP5-A2)
{5}
Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline-Second Edition (CLSI EP9-A2)
6
L. Test Principle
EasyRA Total Bilirubin
This endpoint reaction method measures total bilirubin (conjugated and unconjugated) binding to the 2,4-dichlorophenyldiazonium salt in the presence of surfactant to form azo-bilirubin.
Surfactant
Total bilirubin +2,4-dichlorophenyldiazonium → Azobilirubin salt
The increase in absorbance at 550 nm is directly proportional to the Total Bilirubin concentration in the sample.
EasyRA Direct Bilirubin
This endpoint reaction is based on using a 2,4-dichlorophenyldiazonium salt as the diazo reagent:
Direct bilirubin +2,4-dichlorophenyldiazonium → Azobilirubin salt
Direct bilirubin (conjugated) binds with the diazo reagent to form azobilirubin. The increase in absorbance measured at 550 nm is proportional to the direct bilirubin concentration.
EasyRA Creatine Kinase
CK catalyzes transfer of phosphate groups from creatine phosphate to ADP to form ATP. The rate of production of ATP is measured by two coupled reactions. Hexokinase (HK) catalyzes the reaction of Glucose and ATP to form glucose-6-phosphate. The glucose-6-phosphate with NAD in the presence of glucose-6-phosphate dehydrogenase (G6PDH) forms 6-phosphogluconate and NADPH. The increase in NADPH absorbance at 340 nm is directly related to CK activity in the sample.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
{6}
Serum and Plasma: Precision studies were performed on the EasyRA analyzer in conjunction with the matrix comparison study with lithium heparin tubes. Duplicate plasma and serum samples from the method studies were calculated for average SD for 3 partitioned bins covering the ranges tested. The average SD in each of the 3 bins for both plasma and serum are comparable.
TBIL
| | Low | Medium | High |
| --- | --- | --- | --- |
| Samples Range (mg/dL) | 0.08 – 0.5 mg/dL | 0.5 – 0.75 mg/dL | 0.75 – 20 mg/dL |
| Number of Samples | 23 | 23 | 16 |
| Number of Replicates | 2 | 2 | 2 |
| | | | |
| Serum Mean | 0.38 | 0.62 | 3.51 |
| Serum SD | 0.016 | 0.018 | 0.027 |
| Serum % CV | 4.2% | 2.9% | 0.77% |
| | | | |
| Plasma Mean | 0.37 | 0.60 | 3.48 |
| Plasma SD | 0.017 | 0.013 | 0.030 |
| Plasma %CV | 4.6% | 2.2% | 0.86% |
DBIL
| | Low | Medium | High |
| --- | --- | --- | --- |
| Samples Range (mg/dL) | 0.06 – 0.1 mg/dL | 0.1 – 0.16 mg/dL | 0.16 – 10 mg/dL |
| Number of Samples | 20 | 15 | 17 |
| Number of Replicates | 2 | 2 | 2 |
| | | | |
| Serum Mean | 0.09 | 0.14 | 1.64 |
| Serum SD | 0.006 | 0.013 | 0.010 |
| Serum % CV | 6.7% | 9.3% | 0.61% |
| | | | |
| Plasma Mean | 0.09 | 0.12 | 1.64 |
| Plasma SD | 0.007 | 0.008 | 0.010 |
| Plasma %CV | 7.8% | 6.7% | 0.61% |
{7}
CK
| | Low | Medium | High |
| --- | --- | --- | --- |
| Samples Range (U/L) | 3 – 95 U/L | 95 – 160 U/L | 160 – 1200 U/L |
| Number of Samples | 19 | 16 | 18 |
| Number of Replicates | 2 | 2 | 2 |
| | | | |
| Serum Mean | 68.0 | 121.5 | 370.4 |
| Serum SD | 0.63 | 1.06 | 1.69 |
| Serum % CV | 0.93% | 0.87% | 0.46% |
| | | | |
| Plasma Mean | 64.3 | 118.5 | 369.2 |
| Plasma SD | 0.78 | 0.84 | 2.75 |
| Plasma %CV | 1.2% | 0.71% | 0.75% |
In addition, a simplified within - run precision study was performed on the EasyRA analyzer by analyzing 3 plasma patient samples using lithium heparin tubes $(N = 20)$ . The within - run precision data is summarized in the table below:
| | Mean | SD | %CV |
| --- | --- | --- | --- |
| TBIL (mg/dL) | Level 1 –0.78 | Level 1 – 0.02 | Level 1 – 2.16 |
| | Level 2 – 4.82 | Level 2 – 0.09 | Level 2 – 1.82 |
| | Level 3 – 12.49 | Level 3 – 0.05 | Level 3 – 0.43 |
| DBIL (mg/dL) | Level 1 – 0.75 | Level 1 – 0.01 | Level 1 – 1.78 |
| | Level 2 – 2.08 | Level 2 – 0.03 | Level 2 – 1.63 |
| | Level 3 – 5.01 | Level 3 – 0.02 | Level 3 – 0.46 |
| CK (U/L) | Level 1 – 10.66 | Level 1 – 0.13 | Level 1 – 1.26 |
| | Level 2 – 75.45 | Level 2 – 1.05 | Level 2 – 1.39 |
| | Level 3 – 942.90 | Level 3 – 4.92 | Level 3 – 0.52 |
# b. Linearity/assay reportable range:
Plasma - Linearity studies were not conducted in plasma. See previously cleared linearity data in k080810 (TBIL and DBIL) and k080874 (CK) for serum samples. The linear reportable ranges for each assay are summarized below.
{8}
| | TBIL | DBIL | CK |
| --- | --- | --- | --- |
| Linear Reportable Range | 0.08 – 20 mg/dL | 0.06 – 10 mg/dL | 3 – 1200 U/L |
An extended linearity study was performed for all analytes with the EasyRA analyzer to evaluate accuracy and precision. The sponsor recommends a dilution of 1:2 when the patient TBIL, DBIL, and CK results in plasma fall outside the upper measuring range of 20 mg/dL, 10 mg/dL and 1200 U/L, respectively. A dilution study was performed for TBIL on 5 different patient plasma samples spiked with standard TBIL stock solution to increase the TBIL level in the range of 20 to 40 mg/dL, for DBIL 5 different patient plasma samples were spiked with standard DBIL stock solution to increase the DBIL level in the range of 10 to 20 mg/dL, and for CK 5 different patient plasma samples were spiked with standard CK stock solution to increase the CK level in the range of 1200 to 2400 U/L. Each sample was then diluted with saline at 1:2 dilution by the analyzer or manually. Each diluted sample was run in triplicate on two EasyRA analyzers. The % recovery range of the system for each analyte is provided in the table below:
| | EasyRA % Recovery |
| --- | --- |
| TBIL | 98.77 to 101.01 |
| DBIL | 98.02 to 100.82 |
| CK | 101.00 to 102.79 |
In addition a simplified within-run precision study was performed in the extended linearity range for all analytes by analyzing 3 plasma samples, 20 consecutive times. The within-run precision data is summarized in the table below:
| | Mean | SD | %CV |
| --- | --- | --- | --- |
| TBIL | Level 1 – 29.46 | Level 1 – 0.21 | Level 1 – 0.72 |
| | Level 2 – 33.06 | Level 2 – 0.34 | Level 2 – 1.04 |
| | Level 3 – 39.45 | Level 3 – 0.27 | Level 3 – 0.68 |
| DBIL | Level 1 – 13.23 | Level 1 – 0.07 | Level 1 – 0.55 |
| | Level 2 – 16.16 | Level 2 – 0.13 | Level 2 – 0.79 |
| | Level 3 – 18.83 | Level 3 – 0.12 | Level 3 – 0.66 |
| CK | Level 1 – 1398.95 | Level 1 – 9.92 | Level 1 – 0.71 |
| | Level 2 – 1706.65 | Level 2 – 12.52 | Level 2 – 0.73 |
| | Level 3 – 2050.40 | Level 3 – 11.78 | Level 3 – 0.57 |
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Calibrators were previously cleared under k080810 (TBIL and DBIL) and k080874 (CK)
{9}
d. Detection limit:
Limit of detection studies were not conducted for plasma. See previously cleared serum data under k080810 (TBIL and DBIL) and k080874 (CK)
e. Analytical specificity:
Plasma interference studies were not performed. See previously cleared interference data in k080810 (TBIL and DBIL) and k080874 (CK) for serum samples.
f. Assay cut-off: Not Applicable
2. Comparison studies:
a. Method comparison with predicate device:
Plasma - See previously cleared method comparison data in k080810 (TBIL and DBII) and k080874 (CK) for serum samples. Matrix comparison data is provided below in part (b) of this section.
b. Matrix comparison:
Plasma (Lithium Heparin) – A matrix comparison study was performed in conjunction with CLSI EP9-A2 guidelines using lithium heparin tubes. The study was conducted with human plasma and serum samples. 62 total samples (56 unaltered and 6 altered samples) were analyzed for TBIL, 52 total samples (47 unaltered and 5 altered samples) were analyzed for DBIL, and 51 total samples (46 unaltered and 5 altered samples) were analyzed for CK. Each sample was analyzed in duplicate using the EasyRA chemistry analyzer. One single set of plasma samples were used as test samples, while duplicate serum samples were used as references. Results of the linear regression correlations are as follows:
| Analyte | Slope | Intercept | R2 | Sample Range |
| --- | --- | --- | --- | --- |
| TBIL | 0.9952 | -0.0079 | 0.9998 | 0.17 to 16.78 mg/dL |
| DBIL | 1.0064 | -0.01 | 0.9992 | 0.07 to 9.95 mg/dL |
| CK | 1.0081 | -4.0906 | 0.9986 | 10 to 931 U/L |
3. Clinical studies:
a. Clinical Sensitivity:
{10}
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
EasyRA Total Bilirubin: Adult: 0.2 – 1.0 mg/dl¹
EasyRA Direct Bilirubin: 0.0 – 0.20 mg/dL¹
EasyRA Creatine Kinase: Male: 24-195 U/L, Female: 24-170 U/L²
The expected serum values for all the analytes were cited from the following references:
¹ Burtis, C.A. and Ashwood, E.R. (Eds), Tietz Textbook of Clinical Chemistry, 2nd edition, W.B. Saunders CO., Philadelphia (1994).
² Tietz NW. Editor, Clinical Guide to Laboratory Tests, 3rd ed., WB Saunders and Co., Philadelphia, PA, (1995) p. 180.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
11
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.