The Olympus Creatine Kinase reagent is intended for use in Olympus automated clinical chemistry analyzers for the quantitative determination of creatine kinase activity in human serum. Measurements of creatine kinase are used in the diagnosis and treatment of myocardial infarction and muscle disease, such as progressive Duchenne-type muscular dystrophy.
Device Story
Olympus Creatine Kinase Reagent is an in vitro diagnostic reagent used on Olympus automated clinical chemistry analyzers. It measures creatine kinase activity in human serum samples. The device is operated by laboratory professionals in clinical settings. The quantitative output provides clinicians with data to assist in the diagnosis and management of myocardial infarction and various muscle diseases, such as Duchenne-type muscular dystrophy.
Clinical Evidence
No clinical data provided; device is an in vitro diagnostic reagent for clinical chemistry analysis.
Technological Characteristics
In vitro diagnostic reagent for use on automated clinical chemistry analyzers. Quantitative enzymatic assay for creatine kinase activity in human serum.
Indications for Use
Indicated for quantitative determination of creatine kinase activity in human serum to aid in diagnosis and treatment of myocardial infarction and muscle diseases, including progressive Duchenne-type muscular dystrophy.
Regulatory Classification
Identification
A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
Related Devices
K043202 — OLYMPUS CREATINE KINASE REAGENT OSR6179/OSR6279 · Olympus America, Inc. · Dec 23, 2004
K980012 — CREATINE KINASE REAGENT · Ac Biochemicals, Inc. · Feb 25, 1998
K971817 — OLYMPUS CK-MB REAGENT · Olympus America, Inc. · Jun 11, 1997
K013912 — CREATINE KINASE (CK-NAC) · Jas Diagnostics, Inc. · Jan 18, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized symbol that resembles a human figure with three wavy lines extending from the head, possibly representing hair or movement.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JAN 2 0 2000
Olympus Diagnostics GMBH c/o Ms. Laura Storms-Tyler Olympus America Inc. Two Corporate Center Drive Melville, New York 11747
Re: K994189
> Trade Name: Olympus Creatine Kinase Reagent Regulatory Class: II Product Code: CGS Dated: December 9, 1999 Received: December 13, 1999
Dear Ms. Storms-Tyler:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{1}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| 510(k) Number (if known): | K994189 |
|----------------------------------------------|---------|
| Device Name: Olympus Creatine Kinase Reagent | |
Indications for Use:
The Olympus Creatine Kinase reagent is intended for use in Olympus automated clinical chemistry analyzers for the quantitative determination of creatine kinase activity in human serum.
Measurements of creatine kinase are used in the diagnosis and treatment of myocardial infarction and muscle disease, such as progressive Duchenne-type muscular dystrophy.
Lian Coggin
(Division Sign-Off) Division of Clinical Laboratory Laboratory Levices 510(k) Number ...
(Please does not write below this line. Continue on another page is needed.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-the-Counter Use
-
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.