MODIFICATION TO HYPOGUARD ADVANCE BLOOD GLUCOSE MONITORING SYSTEM
K031388 · Hypoguard USA, Inc. · CGA · May 22, 2003 · Clinical Chemistry
Device Facts
Record ID
K031388
Device Name
MODIFICATION TO HYPOGUARD ADVANCE BLOOD GLUCOSE MONITORING SYSTEM
Applicant
Hypoguard USA, Inc.
Product Code
CGA · Clinical Chemistry
Decision Date
May 22, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1345
Device Class
Class 2
Indications for Use
The Hypoquard Advance Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood. Testing is done outside the body (in vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control.
Device Story
System comprises portable battery-operated meter, test strips, and control solution. Used for in vitro quantitative measurement of glucose in fresh capillary whole blood. Operated by patients at home or healthcare professionals in clinical settings. User applies blood sample to test strip inserted into meter; device performs electrochemical measurement of glucose concentration. Output displayed on meter screen; assists in monitoring diabetes control effectiveness. Control solution used to verify system accuracy.
Clinical Evidence
Clinical accuracy and method correlation study compared modified system against original system and YSI 2300 reference analyzer. Study population included men and women, ages 19 to 80s, with Type I and Type II diabetes. Tested glucose range: 25-80 mg/dL to >250 mg/dL. Results met acceptance criteria, including Clarke Error Grid Analysis.
Technological Characteristics
Portable, battery-operated blood glucose meter. Uses electrochemical test strips and control solution for in vitro diagnostic measurement. Quantitative measurement of glucose in fresh capillary whole blood.
Indications for Use
Indicated for quantitative glucose measurement in fresh capillary whole blood for home (OTC) use by persons with diabetes or in clinical settings by healthcare professionals to monitor diabetes control effectiveness.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Reference Devices
YSI 2300 analyzer
Related Devices
K092887 — EASYPLUS MINI MASTERDRIVER SELF MONITORING BLOOD GLUCOSE SYSTEM, MODEL EPS09009 · Eps Bio Technology Corp. · Oct 21, 2009
K122110 — ACON ON CALL VIVID PAL BLOOD GLUCOSE MONITORING SYSTEM · ACON Laboratories, Inc. · Nov 21, 2012
K092104 — ASSURE PLATINUM BLOOD GLUCOSE MONITORING SYSTEM · Arkray Factory USA, Inc. · Nov 24, 2009
K053079 — ASSURE PRO BLOOD GLUCOSE MONITORING SYSTEM, MODEL 460001 · Hypoguard USA, Inc. · Mar 28, 2006
K091102 — ASSURE PLATINUM AND GLUCOCARD VITAL BLOOD GLUCOSE MONITORING SYSTEMS · Arkray Factory USA, Inc. · Oct 23, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
K031388
MAY 22 2003
# 510 (k) SUMMARY
| 1. SUBMITTED BY: | Bruce A. MacFarlane, Ph.D.<br>Hypoguard USA, Inc.<br>5182 West 76 <sup>th</sup> Street<br>Minneapolis, MN 55439<br>952-646-3188 (phone)<br>952-646-3110 (fax)<br>Summary prepared: 29 April, 2003 |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. NAME OF DEVICE: | |
| Trade Name: | Hypoguard Advance Blood Glucose Monitoring<br>System |
| Common Names/Descriptions: | Blood glucose meter system |
| Classification Names: | Glucose test system, product code 75CGA, 21<br>CFR 862.1345 |
Hypoguard Advance Blood Glucose Monitoring System
## 4. DEVICE DESCRIPTION:
3. PREDICATE DEVICE:
The Hypoquard Advance Blood Glucose Monitoring System consists of a meter, test strips, and control solution. It is intended for over-the-counter, home use by diabetics to monitor their blood glucose levels, or for use in a clinical setting by health care professionals. The system tests fresh capillary whole blood. The meter is a portable, battery-operated instrument designed for use with Hypoguard Advance Blood Glucose Test Strips.
- 5. INTENDED USE:
This modification does not alter the original intended use:
The Hypoquard Advance Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood. Testing is done outside the body (in vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control.
{1}------------------------------------------------
# 6. SUMMARY OF TECHNOLOGICAL CHARACTERISTICS
The modification has not altered the Hypoquard Advance Blood Glucose Monitoring System technological characteristics.
The modified Hypoquard Advance System has the same technological characteristics as the original device.
## 7. NON-CLINICAL TESTING
Precision Study: Testing was performed identically to the original Advance 510(k). Within run, between run and total CV% were calculated identically to original 510(k). Acceptance criteria were met. Dynamic range claims and minimum sample volume claims were validated.
## 8. CLINICAL TESTING
Accuracy/method correlation testing was done comparing the modified Hypoquard Advance System against the original Hypoquard Advance System and the YSI 2300 analyzer (reference method). Testing included both men and women, both Type I and Type II diabetes, ages from nineteen to the eighties, and a wide range of educational levels. Tested blood glucose values encompassed the 25-80 mq/dL range on the low end to values over 250 mg/dL at the high level. Acceptance criteria were met, including Clarke Error Grid Analysis.
## 9. CONCLUSIONS FROM TESTING
Testing demonstrated that the results were within acceptance criteria. Therefore the modification did not adversely affect performance characteristics.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circle around the symbol. The logo is black and white.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
MAY 22 2003
Bruce A. MacFarlane, Ph.D. Quality Assurance Manager Hypoguard USA, Inc. 5182 West 76th Street Minneapolis, MN 55439
Re: k031388
Trade/Device Name: Hypoguard Advance Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: NBW, CGA Dated: April 29, 2003 Received: May 2, 2003
Dear Dr. MacFarlane:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
#### Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## INDICATIONS FOR USE
510(k) Number (if known):
Hypoquard Advance Blood Glucose Monitoring System Device Name:
#### Indications For Use:
Hypoquard Advance Blood Glucose Monitoring System:
The Hypoquard Advance Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood. Testing is done outside the body (in vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control.
Hypoquard Advance Blood Glucose Meter:
The Hypoquard Advance Blood Glucose Meter is intended for the quantitative measurement of glucose in fresh capillary whole blood. Testing is done outside the body (in vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control.
Hypoquard Advance Blood Glucose Test Strips:
The Hypoquard Advance Blood Glucose Meter is intended for the quantitative measurement of dlucose in fresh capillary whole blood. Testing is done outside the body (in vitro diagnostic use). It is indicated for use at home (over the counter (OTC1) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control.
Hypoquard Advance Control Solution:
Hypoquard Advance Control Solution is intended for use with the Hypoquard Advance Meter and Hypoquard Advance Test Strips as a quality control check to verify the accuracy of blood qlucose test results.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Over-The-Counter Use
Prescription Use (Per 21 CFR 801.109)
Dean Cooper
Division Sign-Off
OR
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K031388
(Optional Format 3-10-98)
5
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.