DADE INDIRECT IMT SYSTEM WITH TCO2 ELECTRODE
Device Facts
| Record ID | K970330 |
|---|---|
| Device Name | DADE INDIRECT IMT SYSTEM WITH TCO2 ELECTRODE |
| Applicant | Dade Intl., Inc. |
| Product Code | CEM · Clinical Chemistry |
| Decision Date | Apr 14, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1600 |
| Device Class | Class 2 |
Intended Use
The Dade Indirect IMT system with TCO₂ electrode performs four in vitro diagnostic tests: sodium (Na), potassium (K), chloride (Cl) and total carbon dioxide (TCO₂).
Device Story
System performs in vitro diagnostic testing for Na, K, Cl, and TCO₂; utilizes Dimension XL or RxL clinical chemistry system. Sample diluted automatically on-board; processed in multisensor cartridge using Integrated Multisensor Technology (IMT) to measure electrical potential proportional to ion activity. TCO₂ measured via Severinghaus electrode following sample acidification and CO₂ liberation. Used in clinical laboratory settings by trained personnel. Output provides quantitative electrolyte concentrations to healthcare providers for diagnostic assessment of patient metabolic and electrolyte status.
Clinical Evidence
Bench testing via split-sample comparison against Beckman SYNCHRON CX®3. Serum/plasma study (n=270) showed correlation coefficients: Na 0.983, K 0.996, Cl 0.968, TCO₂ 0.945. Urine study (n=95-105) showed correlation coefficients: Na 0.997, K 0.998, Cl 0.979. Slopes and intercepts provided for all analytes.
Technological Characteristics
Integrated Multisensor Technology (IMT) for ion-selective potentiometry; Severinghaus electrode for TCO₂. Automated dilution system. Operates on Dimension XL or RxL clinical chemistry platforms. Analyte range: Na (serum 98-160, urine 20-178), K (serum 2.3-9.5, urine 9-98), Cl (serum 71-119, urine 38-323), TCO₂ (serum 10.6-41.5).
Indications for Use
Indicated for the determination of sodium, potassium, chloride, and TCO₂ in human serum or plasma, and for the determination of sodium, potassium, and chloride in urine.
Regulatory Classification
Identification
A potassium test system is a device intended to measure potassium in serum, plasma, and urine. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.
Predicate Devices
- Dade direct IMT system (ISE sensor cartridge with revised TCO₂ electrode) (K970330)
Reference Devices
- Beckman SYNCHRON CX®3 clinical system
Related Devices
- K052724 — TECO ISE REAGENT SET FOR CX SYSTEM · Teco Diagnostics · Dec 14, 2005
- K970921 — STANDARD A MULTIPLY (A540-0), STANDARD B MULTIPLY (A550-0), STANDARD C MULTIPLY (A560-0) · Alko Diagnostic Corp. · Apr 8, 1997
- K100829 — EASYRA SODIUM ASSAY, EASYRA POTASSIUM ASSAY, EASYRA CHLORIDE ASSAY, EASYRA CARBON DIOXIDE ASSAY MODEL CAT NO. 5423 & 520 · Medica Corp. · Jan 14, 2011