The Axis-Shield Active-B12 (Holotranscobalamin) assay is an enzyme-immunoassay (EIA) for the quantitative determination of holotranscobalamin (HoloTC) in human serum. HoloTC (vitamin B12 bound to transcobalamin) is used as an aid in the diagnosis and treatment of vitamin B12 deficiency.
Device Story
The Axis-Shield Active-B12 is a manual enzyme-immunoassay (EIA) used in clinical laboratories to quantify holotranscobalamin (HoloTC) in human serum. The device utilizes a microtitre plate coated with anti-holotranscobalamin murine monoclonal antibodies. During the assay, HoloTC in the patient serum binds to the plate; a conjugate is added to bind the captured HoloTC, followed by a substrate to generate a colorimetric signal. The intensity of the color, measured via interpolation from a 6-point calibration curve, is directly proportional to the HoloTC concentration. Healthcare providers use these results to assist in diagnosing and managing vitamin B12 deficiency. The manual process provides a quantitative measurement in pmol/L, offering an alternative to automated chemiluminescent immunoassay systems.
Clinical Evidence
No clinical trials were performed. Substantial equivalence was established through non-clinical bench testing, including precision (total CV ≤ 12%), linearity (5.3 to 156.0 pmol/L), and interference studies. A method comparison study (n=111) against a commercially available assay showed a Passing-Bablok regression slope of 0.95 (95% CI: 0.89-1.01) and a Pearson correlation coefficient of 0.93 (95% CI: 0.90-0.95).
Technological Characteristics
Manual enzyme-linked immunosorbent assay (ELISA) using microtitre plates coated with murine monoclonal antibodies. Detection via alkaline phosphatase conjugate and pNPP substrate. Requires manual processing. Calibration uses a 6-point linear regression curve. Storage at 2-8°C. Compatible with serum and serum separator tubes.
Indications for Use
Indicated for the quantitative determination of holotranscobalamin (HoloTC) in human serum as an aid in the diagnosis and treatment of vitamin B12 deficiency in patients.
Regulatory Classification
Identification
A vitamin B12 test system is a device intended to measure vitamin B12 in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of anemias of gastrointestinal malabsorption.
K112443 — ARCHITECT ACTIVE B-12 (HOLOTRANSCOBALAMIN) · Axis-Shield Diagnostics, Ltd. · Dec 19, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K121946
#### Owners Name:
Axis-Shield Diagnostics Ltd. The Technology Park, Dundee DD2 1XA, UK Tel. +44(0) 1382 422 000 Fax. +44(0) 1382 422 088
# Name of contact person:
Dr Simon Richards QA/RA Director Axis-Shield Diagnostics Ltd.
#### Date summary prepared:
March 8th, 2013
Device Name: Axis-Shield Active-B12 (Holotranscobalamin)
Classification Name: Vitamin B12 test system Trade Name: Axis-Shield Active-B12 (Holotranscobalamin) Common Name: Holotranscobalamin test Governing Regulation: 862.1810 Device Classification: Class II Classification Panel: Clinical Chemistry Product Code: CDD
Legally marketed device to which equivalency is claimed: ARCHITECT Active-B12 (Holotranscobalamin) (K112443)
Axis-Shield Active-B12 (Holotranscobalamin), K121946 510(k) Summary Final
ਾ ਪ
{1}------------------------------------------------
#### Intended Use of Device:
The Axis-Shield Active-B12 (Holotranscobalamin) assay is an enzyme-immunoassay (EIA) for the quantitative determination of holotranscobalamin (HoloTC) in human serum. HoloTC (vitamin B12 bound to transcobalamin) is used as an aid in the diagnosis and treatment of vitamin B12 deficiency.
## Description of Device:
The Axis-Shield Active-B12 (Holotranscobalamin) device contains the following components: a microtitre plate with 8 x 12-well breakapart strips coated with a anti-holotranscobalamin murine monoclonal antibody, in a resealable foil pack with desiccant; ready-to-use calibrators, low and high controls (phosphate buffer containing protein (bovine) stabiliser and sodium azide preservative with or without recombinant HoloTC); ready-to-use pre-treatment solution; murine anti-human transcobalamin alkaline phosphatase conjugate; para-NitroPhenyl Phosphate (pNPP) substrate; wash buffer (8x); ready-to-use stop solution.
# Principle of the Assay:
The microtitre wells are coated with a highly specific monoclonal antibody for Active-B12 (Holotranscobalamin). During the first incubation holotranscobalamin in serum specifically binds to the antibody-coated surface. In the second incubation the Conjugate binds to any captured holotranscobalamin. The wells are then washed to remove unbound components. Bound holotranscobalamin is detected by incubation with the Substrate. Addition of Stop Solution terminates the reaction, resulting in a coloured end-product. The concentration of holotranscobalamin in pmol/L is directly related to the colour generated and can be estimated by interpolation from a dose-response curve based on Calibrators.
# Comparison of Technological Characteristics:
Axis-Shield Active-B12 (Holotranscobalamin) and ARCHITECT Active-B12 (Holotranscobalamin) are both immunoassays for the quantitative determination of Holotranscobalamin (HoloTC) in human serum.
The predicate device is an automated Chemiluminescent mircoparticle immunoassay (CMIA) whereas the submission device is a manual enzyme-linked immunosorbent assays also demonstrated substantial equivalence in terms antibodies employed and units of measure.
{2}------------------------------------------------
| | | | | | | | Comparison of the subject device with the predicate device: | |
|--|--|--|--|--|--|--|-------------------------------------------------------------|--|
|--|--|--|--|--|--|--|-------------------------------------------------------------|--|
| | Submission device | Predicate device |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Parameter | Axis-Shield Active-B12 (Holotranscobalamin) | |
| Intended use | The Axis-Shield Active-B12<br>(Holotranscobalamin) assay is an enzyme-<br>immunoassay (EIA) for the quantitative<br>determination of holotranscobalamin<br>(HoloTC) in human serum. HoloTC (vitamin<br>B12 bound to transcobalamin) is used as an<br>aid in the diagnosis and treatment of vitamin<br>B12 deficiency. | Chemiluminescent microparticle<br>immunoassay (CMIA) for the quantitative<br>determination of Holotranscobalamin in<br>human serum on the ARCHITECT i System.<br>Active-B12 (Holotranscobalamin) is used as<br>an aid in the diagnosis and treatment of<br>vitamin B12 deficiency. |
| Antibodies employed | Murine monoclonal antibody 3C4<br>Murine monoclonal antibody 3-11 | Murine monoclonal antibody 3C4<br>Murine monoclonal antibody 3-11 |
| Substrate / Signal<br>Generation | Alkaline phosphatase | Acridinium Tracer |
| Specimen type | Serum and Serum Separator | Serum and Serum Separator |
| Storage conditions | 2-8°C. | 2-8°C. |
| Calibration | 6-point calibration curve.<br>Linear regression curve-fit | 6-point calibration curve.<br>4PLC Y-weighted |
| Measuring Interval | 10 to 128 pmol/L | 5.0 to 128.0 pmol/L |
| Detection Limits | Limit of Quantitation of 8.3 pmol/L | Limit of Quantitation of ≤ 5.0 pmol/L |
| Linearity on Dilution | LOQ to Calibrator F | LOQ to Calibrator F |
| Expected Values in<br>Asymptomatic<br>Population | The mean Holotranscobalamin concentration<br>(derived from log-transformed data to<br>normalize the population) was 72 pmol/L<br>with a range from 15 to 147 pmol/L.<br>The central 95% of the population defined<br>the expected range of<br>21 to 123pmol/L (n= 135) | The mean Holotranscobalamin<br>concentration (derived from log-<br>transformed data to normalize the<br>population) was 71.9pmol/L with a range<br>from 20.6 to 196.7 pmol/L.<br>The central 95% of the population defined<br>the expected range of<br>25.1 to 165.0 pmol/L (n=181) |
| Cross-reactivity | No detectable carryover with;<br>Apotranscobalamin at 500 pmol/L<br>Haptocorrin at 5000 pmol/L | No detectable carryover with;<br>Apotranscobalamin at 500 pmol/L<br>Haptocorrin at 5000 pmol/L |
| Interference | ≤ 10% with;<br>Bilirubin at 0.3 mg/mL<br>Haemoglobin at 5 mg/mL<br>Triglycerides at 30 mg/mL<br>Rheumatoid Factor at 75 IU/mL<br>Total protein at 90 mg/mL | ≤ 10% with;<br>Bilirubin at 20 mg/dL<br>Haemoglobin at 200 mg/dL<br>Triglycerides at 850 mg/dL<br>Rheumatoid Factor at 70 IU/mL<br>Total protein at 10 g/dL |
| Imprecision | Total %CV ≤ 12%<br>Within-run %CV ≤ 9% | Total %CV ≤ 5.8%<br>Within-run %CV < 4.4% |
・
:
·
.
. .
・ .
.
'
. ..
.
{3}------------------------------------------------
#### Summary of Non-Clinical Performance:
Axis-Shield Active-B12 (Holotranscobalamin) The ARCHITECT Active-B12 (Holotranscobalamin) are both immunoassays for the quantitative determination of Holotranscobalamin (HoloTC) in human serum.
The Axis-Shield Active-B12 (Holotranscobalamin) assay demonstrated substantially equivalent performance to the ARCHITECT Active-B12 (Holotranscobalamin). A summary of the nonclinical performance data included in this 510(k) submission has been presented.
## Dilution Linearity
The Axis-Shield Active-B12 (Holotranscobalamin) EIA demonstrated linearity across the measuring range of the assay from 5.3 to 156.0 pmol/L.
#### Reportable Range
The Axis-Shield Active-B12 (Holotranscobalamin) ElA reportable range is 10 to 128 pmol/L.
#### Analytical limits at low levels
a representative study, the limit of blank of the Axis-Shield ln Active-B12 (Holotranscobalamin) ElA was 4.9 pmol/L, the limit of detection was 8.1 pmol/L and the limit of quantification was 8.3 pmol/L.
#### High Dose Hook
No high dose hook effect was detected up to a concentration of 2236pmol/L Holotranscobalamin.
## Cross-reactivity
The Axis-Shield Active-B12 (Holotranscobalamin) EIA is designed to have a maximum deviation in holotranscobalamin concentration of ≤10% in the presence of 500 pmol/L apotranscobalamin or 5000 pmol/L haptocorrin. The maximum deviation in holotranscobalamin concentration ranged from -5% to 1%.
# Interference
The Axis-Shield Active-B12 (Holotranscobalamin) EIA is designed to have a maximum deviation in holotranscobalamin concentration of ≤10% in the presence of potentially interfering compounds. The maximum deviation in holotranscobalamin concentration ranged from -10% to 8% for the potentially interfering compounds presented.
| Potential Interfering Substance | No interference found up to the<br>following concentration |
|------------------------------------|------------------------------------------------------------|
| Haemoglobin | 500 mg/dL |
| Bilirubin | 30 mg/dL |
| Triglyceride (Intralipid Solution) | 3000 mg/dL |
| Rheumatoid Factor | 7500 IU/dL |
| Total Protein | 9000 mg/dL |
Axis-Shield Active-B12 (Holotranscobalamin), K121946 510(k) Summary Final
{4}------------------------------------------------
# Precision-
8 human serum samples were assayed using 3 lots of reagents. Samples were assayed by 2 operators in replicates of 8, once a day for 5 days (total n=80). Data from this study are summarised in the following table: :
| Sample | n | Lot | Operator | Mean (pmol/l) | Intra-assay %CV | Total %CV |
|-----------------|----|-----|----------|---------------|-----------------|-----------|
| 1 A | 80 | 1 | 1 | 17.8 | 7.5% | 8.2% |
| | | 1 | 2 | 17.5 | 3.1% | 9.3% |
| | | 2 | 1 | 20.1 | 6.0% | 6.6% |
| | | 2 | 2 | 20.3 | 6.9% | 9.2% |
| | | 3 | 1 | 19.1 | 5.5% | 8.0% |
| | | 3 | 2 | 18.9 | 8.5% | 11.0% |
| 2 A | 80 | 1 | 1 | 21.8 | 5.5% | 9.9% |
| | | 1 | 2 | 21.8 | 3.9% | 7.5% |
| | | 2 | 1 | 22.6 | 5.6% | 8.7% |
| | | 2 | 2 | 23.5 | 9.0% | 10.3% |
| | | 3 | 1 | 23.9 | 7.0% | 10.2% |
| | | 3 | 2 | 23.2 | 5.8% | 8.9% |
| 3 A | 80 | 1 | 1 | 28.8 | 3.8% | 7.8% |
| | | 1 | 2 | 30.7 | 4.3% | 9.6% |
| | | 2 | 1 | 31.0 | 6.8% | 8.0% |
| | | 2 | 2 | 31.4 | 4.3% | 6.1% |
| | | 3 | 1 | 31.5 | 4.5% | 6.4% |
| | | 3 | 2 | 32.2 | 4.0% | 9.2% |
| 4 A | 80 | 1 | 1 | 49.3 | 3.9% | 7.4% |
| | | 1 | 2 | 52.6 | 4.1% | 6.7% |
| | | 2 | 1 | 50.8 | 5.6% | 10.0% |
| | | 2 | 2 | 51.7 | 4.7% | 5.9% |
| | | 3 | 1 | 52.6 | 4.6% | 4.8% |
| | | 3 | 2 | 55.0 | 5.5% | 6.1% |
| 5 A | 80 | 1 | 1 | 68.4 | 4.0% | 7.6% |
| | | 1 | 2 | 73.2 | 3.7% | 7.5% |
| | | 2 | 1 | 74.8 | 4.3% | 8.2% |
| | | 2 | 2 | 75.9 | 4.6% | 6.4% |
| | | 3 | 1 | 75.1 | 4.4% | 7.9% |
| | | 3 | 2 | 76.3 | 4.9% | 6.2% |
| 7 A | 80 | 1 | 1 | 115.9 | 4.2% | 5.9% |
| | | 1 | 2 | 121.1 | 3.6% | 7.0% |
| | | 2 | 1 | 123.2 | 4.3% | 10.2% |
| | | 2 | 2 | 124.0 | 4.2% | 6.4% |
| | | 3 | 1 | 127.0 | 4.8% | 10.1% |
| | | 3 | 2 | 129.5 | 3.2% | 5.6% |
| Low<br>Control | 80 | 1 | 1 | 23.7 | 9.4% | 10.9% |
| | | 1 | 2 | 23.8 | 5.1% | 11.5% |
| | | 2 | 1 | 20.0 | 6.0% | 7.5% |
| | | 2 | 2 | 18.6 | 5.8% | 8.5% |
| | | 3 | 1 | 20.3 | 8.3% | 9.7% |
| | | 3 | 2 | 20.1 | 8.3% | 10.0% |
| High<br>Control | 80 | 1 | 1 | 61.2 | 6.3% | 6.4% |
| | | 1 | 2 | 58.8 | 4.5% | 8.9% |
| | | 2 | 1 | 50.3 | 6.3% | 8.1% |
| | | 2 | 2 | 50.2 | 5.9% | 8.4% |
| | | 3 | 1 | 52.2 | 7.7% | 9.2% |
| | | 3 | 2 | 50.8 | 5.8% | 8.5% |
Axis-Shield Active-B12 (Holotranscobalamin), K121946
510(k) Summary
Final
{5}------------------------------------------------
# Matrix Comparison
Specimens collected in serum (clot) and serum separator (SST) tubes are compatible in the Axis-Shield Active-B12 (Holotranscobalamin) assay shown by a correlation study. The data obtained gave the following statistical values:
| Axis-Shield Active-B12 (Holotranscobalamin) EIA - Serum clot versus Serum Separator | |
|-------------------------------------------------------------------------------------|--------|
| Number of Specimens (n) | > 36 |
| Slope of regression line (Passing-Bablok regression) | > 0.97 |
| Correlation coefficient (r) (Pearson) | 0.98 |
| Overall percent bias (%) | < 2.5 |
# Summary of Clinical Performance:
The Axis-Shield Active-B12 (Holotranscobalamin) assay demonstrated substantially equivalent clinical performance to the ARCHITECT Active-B12 (Holotranscobalamin) as indicated by a method comparison study.
#### Method Comparison
A correlation study was performed with serum specimens from apparently healthy adults. All specimens were analysed using the Axis-Shield Active-B12 (Holotranscobalamin) EIA and a commercially available assay for Holotranscobalamin according to the CLSI document EP9-A2. Specimen concentrations ranged from 13.8 to 112.8 pmol/L in the assay. The data obtained gave the following statistical values:
| Axis-Shield Active-B12 (Holotranscobalamin) ElA versus a commercially available assay | | | | | |
|---------------------------------------------------------------------------------------|----------------------|--|--|--|--|
| Number of specimens | 111 | | | | |
| Slope of regression line (Passing-Bablok regression) (95% Cl) | 0.95 (0.89 to 1.01) | | | | |
| Y-intercept (Passing-Bablok regression) (95% CI) | 8.39 (5.73 to 11.77) | | | | |
| Correlation coefficient (r) (Pearson) (95% Cl) | 0.93 (0.90 to 0.95) | | | | |
#### Conclusion:
Based on the performance characteristics the Axis-Shield Active-B12 (Holotranscobalamin) assay (K121946) is substantially equivalent to the predicate device.
Axis-Shield Active-B12 (Holotranscobalamin), K121946 510(k) Summary Final
{6}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or other bird.
11
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 22, 2013
Axis-Shield Diagnostics, Ltd. c/o Simon J. Richards, PhD The Technology Park, Luna Place Dundee. Scotland United Kingdom, DD2 1XA
Re: k121946
Trade/Device Name: Axis-Shield Active-B12 (Holotranscobalamin) Regulation Number: 21 CFR 862.1810 Regulation Name: Vitamin B12 test system Regulatory Class: Class II Product Code: CDD Dated: February 08, 2013 Received: February 11, 2013
Dear Dr. Richards:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{7}------------------------------------------------
Page 2 Dr. Richards
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Carol G. Benson -S for
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indication for Use
510(k) Number (if known): k121946
Device Name: Axis-Shield Active-B12 (Holotranscobalamin)
Indication For Use:
The Axis-Shield Active-B12 (Holotranscobalamin) assay is an enzyme-immunoassay (EIA) for the quantitative determination of holotranscobalamin (HoloTC) in human serum. HoloTC (vitamin B12 bound to transcobalamin) is used as an aid in the diagnosis and treatment of vitamin B12 deficiency.
For in vitro diagnostic use.
Prescription Use _ X (21 CFR Part 801 Subpart D)
And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Image /page/8/Picture/11 description: The image shows the name "Yung W. Chan-S" in a stylized font. The letters "W" and "Chan" are designed with a geometric pattern, giving them a unique and modern look. The name is written in black against a white background, making it stand out. The overall design is clean and simple, with a focus on the typography.
Division Sign-Off Office of In Vitro Diagnostics and Radiological Health
510(k) k121946
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.