K101344 · Glytec, LLC (Formally Known AS Glucotec, Inc.) · NDC · Jun 28, 2010 · Anesthesiology
Device Facts
Record ID
K101344
Device Name
G+ SYSTEM
Applicant
Glytec, LLC (Formally Known AS Glucotec, Inc.)
Product Code
NDC · Anesthesiology
Decision Date
Jun 28, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.1890
Device Class
Class 2
Attributes
Software as a Medical Device, Therapeutic
Indications for Use
The G+ System is intended to evaluate the current as well as cumulative patient blood glucose values, and, based on the aggregate of those measurements, whether one or many, regulate the infusion of I.V. fluids, through an I.V. infusion pump, and drive the blood glucose level towards a predetermined target range. Once that target blood glucose range has been reached, the system's function is to recommend a titration of insulin, glucose, and saline for the purpose of maintaining the patient's blood glucose level in that target range. The system is provide intravenous dosing of glucose, saline, and insulin, as well as subcutaneous dosing of glucose and insulin. The G+ System logic is not a substitute for, but rather an assist to clinical reasoning. The measurements and calculations generated are intended to be used by qualified and trained medical personnel in evaluating patient conditions in conjunction with clinical history, symptoms, and other diagnostic measurements, as well as the medical professional's clinical judgment. No medical decision should be based solely on the recommended guidance provided by this software program.
Device Story
Software device (G+ System) evaluates current and cumulative patient blood glucose values; processes aggregate data to calculate and recommend titration of insulin, glucose, and saline; regulates I.V. infusion pumps to drive blood glucose toward target range; maintains glucose levels once target is reached. Used by qualified medical personnel in clinical settings; provides decision support for dosing; does not replace clinical judgment. Available in Standard or Enterprise (server-based) editions. Benefits include automated, precise glycemic control; reduces manual calculation errors; assists clinicians in maintaining patient safety.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and intended use comparison to predicate.
Technological Characteristics
Software-based drug dose calculator; operates as standalone or server-based (Enterprise) system; interfaces with I.V. infusion pumps; utilizes programmed logic for glycemic control; no specific hardware materials or energy sources defined beyond software execution platform.
Indications for Use
Indicated for patients aged 18 and older requiring management of blood glucose levels via intravenous or subcutaneous dosing of insulin, glucose, and saline. Contraindicated for patients with known insulin allergies.
Regulatory Classification
Identification
A predictive pulmonary-function value calculator is a device used to calculate normal pulmonary-function values based on empirical equations.
Predicate Devices
CollaborativeMed, LLC Glucommander Plus (G+) System (K061110)
Submission Summary (Full Text)
{0}------------------------------------------------
## JUN 2 8 2010
K10134Y
Glytec, LLC G+ System (Glucommander Plus System)
Special 510(k) Premarket Notification
# 510(k) Summary of Safety and Effectiveness
| Proprietary Name: | Glytec LLC, G+ System |
|------------------------------------|---------------------------------------------|
| Common Name: | G+ System™ |
| Product Code/Classification Panel: | NDC – General and Plastic Surgery |
| Classification Name: | Drug Dose Calculator Class II per §868.1890 |
#### Submitter Information
#### Submitter's Name and Address:
Glytec, LLC (a wholly owned subsidiary of Glucotec, Inc., formally known as CollaborativeMed, LLC) 665 N. Academy Street Greenville, SC 29601 FDA Establishment Registration Number: 3005853093
#### Contact Information:
.
William Matthews, Managing Partner BioDevice Solutions, LLC 2607 Woodruff Rd. Simpsonville, SC 29681 Telephone (760) 574-9476 Fax (864) 297-5270
#### Performance Standards
No applicable performance standards have been issued under section 514 of the Food, Drug and Cosmetic Act for Automatic Peritoneal Delivery Systems.
#### Predicate Device
The predicate device for the GlyTec, LLC G+ System is the following:
- CollaborativeMed, LLC Glucommander Plus (G+) System #K061110 (6/7/2006); .
{1}------------------------------------------------
#### Glytec, LLC G+ System (Glucommander Plus System) Special 510(k) Premarket Notification
## 510(k) Summary of Safety and Effectiveness
#### Device Description
The Glytec, LLC G+ System is a software device intended to evaluate the current as well as cumulative patient blood glucose values, and, based on the aggregate of those measurements, whether one or many, regulate the infusion of I.V. fluids, through an I.V. infusion pump, and direct the blood glucose level towards a predetermined target range. Once that target blood glucose range has been reached, the system's function is to recommend a titration of insulin, glucose, and saline for the purpose of maintaining the patient's blood glucose level in that target range. The system is provide intravenous dosing of glucose, saline, and insulin, as well as subcutaneous dosing of glucose and insulin.
The device is available in a Standard Edition or server based Enterprise Edition.
#### Indications for Use/Intended Use
The G+ System is intended to evaluate the current as well as cumulative patient blood glucose values, and, based on the aggregate of those measurements, whether one or many, regulate the infusion of I.V. fluids, through and I.V. infusion pump, and drive the blood glucose level towards a predetermined target range. Once that target blood glucose range has been reached, the system's function is to recommend a titration of insulin, glucose, and saline for the purpose of maintaining the patient's blood glucose level in that target range. The system is provide intravenous dosing of glucose, saline, and insulin, as well as subcutaneous dosing of glucose and insulin.
The G+ System logic is not a substitute for, but rather an assist to clinical reasoning. The measurements and calculations generated are intended to be used by qualified and trained medical personnel in evaluating patient conditions in conjunction with clinical history, symptoms, and other diagnostic measurements, as well as the medical professional's clinical judgment. No medical decision should be based solely on the recommended guidance provided by this software program.
### Statement of Substantial Equivalence
Glytec, LLC believes that, within the meaning of the Medical Device Amendments of 1976, the Glytec, LLC G+ System is substantially equivalent to the following medical device in commercial distribution:
- CollaborativeMed, LLC, Glucommander Plus #K061110 (6/7/2006) .
{2}------------------------------------------------
## Glytec, LLC G+ System (Glucommander Plus System) Special 510(k) Premarket Notification
# 510(k) Summary of Safety and Effectiveness
## Conclusion
Based upon an analysis of the overall performance characteristics for the Glytec, LLC, G+ System, Glytec, Inc. believes that no significant differences exist between this system and the predicate systems quoted therefore, the Glytec, LLC G+ System does not impose any new safety or effectiveness concerns.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
JUN 2 8 2010
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Glytec, LLC (Formally known as GlucoTec, Incorporated) C/O Mr. William Matthews BioDevice Solutions, LLC 2607 Woodruff Road Simpsonville, South Carolina 29681
Re: K101344
Trade/Device Name: G+ System Regulation Number: 21 CFR 868.1890 Regulation Name: Predictive Pulmonary - Function Value Calculator Regulatory Class: II Product Code: NDC Dated: June 17, 2010 Received: June 18, 2010
Dear Mr. Matthews:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2- Mr. Matthews
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting < (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Swor Ranns
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): K011344
Device Name: G+™ System
#### Indications For Use:
The Glucommander is intended to evaluate the current as well as cumulative patient blood glucose values, and, based on the aggregate of those measurements, whether one or many, calculate and recommend a dose of saline, glucose, and insulin to drive the blood glucose level, either up or down, towards a predetermined target range. Once that target blood glucose range has been reached, the system's function is to recommend dosing of insulin, glucose, and saline for the purpose of maintaining the patient's blood glucose level in that target range. The system is provide intravenous dosing of glucose, saline, and insulin; as well as subcutaneous dosing of glucose and insulin. The device is not intended for use with patients with known insulin allergies or patients under the age of 18.
The Glucommander's programmed logic is not a substitute for, but rather an assist to clinical reasoning. The measurements and calculations generated by the GBGDS are intended to be used by qualified and trained medical personnel in evaluating patient conditions in conjunction with clinical history, symptoms, and other diagnostic measurements, as well as the medical professional's clinical judgment. No medical decision should be based solely on he recommended guidance provided by this software program.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ris C. Cla
Division of Anesthesiology, Genera Infection Control, Dental Devices
510(k) Number: -
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.