K161856 · Nova Biomedical Corporation · PZI · Nov 15, 2016 · Clinical Chemistry
Device Facts
Record ID
K161856
Device Name
StatStrip Xpress Glucose Hospital Meter System
Applicant
Nova Biomedical Corporation
Product Code
PZI · Clinical Chemistry
Decision Date
Nov 15, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The StatStrip Xpress Glucose Hospital Meter System is intended for point-of-care, in vitro diagnostic, multiple-patient use for the quantitative determination of glucose in capillary finger stick, venous whole blood, arterial whole blood, neonate arterial whole blood and neonate heel stick specimens. The StatStrip Xpress Glucose Hospital Meter System is also intended for use in the quantitative determination of glucose in venous whole blood, arterial whole blood, neonatal heel stick and neonatal arterial whole blood samples throughout all hospital and all professional healthcare settings. The system should only be used with single-use, auto-disabling lancing devices when performing a capillary finger stick or neonate heel stick. It is not intended for use with neonate cord blood specimens. It is not intended for the screening or diagnosis of diabetes mellitus but is indicated for use in determining dysglycemia.
Device Story
StatStrip Xpress Glucose Hospital Meter System; point-of-care blood glucose monitoring device for hospital use. Modification: expansion of operating temperature range from 15°C-40°C to 5°C-40°C. Input: blood sample via test strip; electrochemical measurement. Output: glucose concentration displayed on meter. Operation: healthcare professionals perform testing; device provides rapid results for clinical decision-making. Modification involves software update to temperature error threshold; hardware remains unchanged. Benefit: increased utility in colder clinical environments.
Clinical Evidence
Bench testing only. Verification activities performed to confirm device performance within the new operating temperature range (5°C to 15°C). Disinfection efficacy and durability testing (10,950 cycles) previously established in K150461.
Indicated for point-of-care, in vitro diagnostic, multiple-patient use for quantitative glucose determination in capillary finger stick, venous whole blood, arterial whole blood, and neonate heel stick/arterial specimens in hospital and professional healthcare settings. Not for neonate cord blood or diabetes screening; indicated for dysglycemia management.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
StatStrip Xpress Glucose Hospital Meter System (K150461)
Related Devices
K152986 — StatStrip Xpress 2 Glucose Hospital Meter System · Nova Biomedical Corporation · Jan 27, 2016
K150461 — StatStrip Xpress Glucose Hospital Meter System · Nova Biomedical Corporation · May 20, 2015
Submission Summary (Full Text)
{0}
SPECIAL 510(k): Device Modification OIR Decision Memorandum
To: THE FILE
RE: DOCUMENT NUMBER K161856
This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary):
1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.) StatStrip Xpress Glucose Hospital Meter System - k150461
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use).
3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.
This change was for:
- Changing the lower limit of the operating temperature range from 59°F (15°C) to 41°F (5°C).
- Modifying the software for the temperature error code from displaying the error when the operating temperature is less than 15°C to displaying the error when the operating temperature is less than 5°C.
Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, and specificaitons.
4. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
Infection Control Studies:
The StatStrip Xpress Glucose Hospital Meter System is intended for multiple-patient use. Disinfection efficacy studies were previously performed (k150461) on the external materials of the meter by an outside commercial testing service and demonstrated complete inactivation of
{1}
live Hepatitis B virus with Clorox Germicidal Wipes, EPA registration # 67619-12). The sponsor also demonstrated that there was no change in performance or in the external materials of the meter after 10,950 cleaning and disinfection cycles (one cycle includes one cleaning wipe plus one disinfecting wipe) to simulate 3 years of device use. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.
This device was cleared after the FDA issued final guidance documents for prescription use blood glucose monitoring systems (BGMS) and over-the-counter use blood glucose monitoring systems (SMBG). However, the recommendations in the guidance documents were not followed for this device since the submission was received prior to the finalization of the guidance documents.
2
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.